K101133 · Dpcom AS · GWM · Nov 18, 2010 · Neurology
Device Facts
Record ID
K101133
Device Name
SENSOMETRICS SOFTWARE
Applicant
Dpcom AS
Product Code
GWM · Neurology
Decision Date
Nov 18, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1620
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
Sensometrics® Software is intended for use by a qualified nurse or physician to display variables from existing legally marketed primary measuring devices used to monitor neurosurgical and neurological patients. The parameters displayed by the Sensometrics® Software include intracranial pressure, arterial blood pressure, oxygen, core/cerebral temperature, and these are continuously trended in charts. The presentation of both measured parameters and the relationships between them is intended as an adjunct to the information provided by existing primary monitors. The Sensometrics Software should not be used alone as the sole basis for decisions as to diagnosis or therapy.
Device Story
Sensometrics® Software is a secondary display system for neurosurgical intensive care units; collects digital inputs from primary monitors measuring intracranial pressure, arterial blood pressure, oxygen, and core/cerebral temperature. Software runs on a PC connected via RJ45/RS232 cables to primary monitors; aggregates data onto a single large color display screen. Displays measured and derived variables in digital trace form; stores and trends data for up to 240 hours. Used by physicians and nurses to visualize parameter relationships; serves as an adjunct to primary monitor information. Does not replace primary monitors; intended to assist clinical decision-making by providing consolidated, trended views of patient status.
Clinical Evidence
Bench testing only. Performance comparison test conducted against Phillips and Integra devices. Evaluated mean ICP, mean ABP, systolic pressure, diastolic pressure, and cerebral temperature. Results demonstrated that parameters computed by Sensometrics software are equivalent to those of the predicate devices.
Technological Characteristics
Software-based secondary display system; runs on a PC. Connectivity via RJ45/RS232 cables to primary monitors. Displays digital traces of intracranial pressure, arterial blood pressure, oxygen, and core/cerebral temperature. Supports data storage and trending up to 240 hours. No specific materials or sterilization methods applicable as it is a software-only secondary display system.
Indications for Use
Indicated for neurosurgical and neurological patients requiring monitoring of intracranial pressure, arterial blood pressure, oxygen, and core/cerebral temperature. Intended for use by qualified nurses or physicians as an adjunct to primary monitoring devices; not for use as the sole basis for diagnosis or therapy.
Regulatory Classification
Identification
An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.
K171666 — MPR2 logO DATALOGGER · Raumedic AG · Jul 6, 2017
K192572 — CNS Envision · Moberg Research, Inc. · Dec 17, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
K101133, Sensometrics® Software - Response November 16, 2010
#### Attachment C
### NDV 1 9 2010
#### 510(k) Summary
Submitter: dPCom AS
Address: Solveien 29 C, N-1177 Oslo, Norway Postal: Pb 14, Bekkelagshøgda, N-1109 Oslo, Norway Phone: +47 22 95 82 77
Contact Information: Constance Bundy C. G. Bundy Associates, Inc. 435 Rice Creek Terrace Fridley, MN 55432
Submission Date: April 16, 2010, Revised November 12, 2010
Device Name and Classification: Sensometrics® Software, Intracranial Pressure Monitoring Device, Class II, 21 CFR 882.1620 Product Code: GWM
Equivalent Device Identification: NeuroSystems 1™ Monitor, K050702, Integra MPM-1 Monitor, K962928, Phillips IntelliVue Mp40, K032858
Device Description: Sensometrics® Software is a secondary display system that collects together on a single large color display screen the measured pressure variables relevant to the multimodality monitoring of the patient in a neurosurgical intensive care unit. It receives digital inputs from primary monitors used in the measurement of intracranial pressure, arterial blood pressure and cerebral temperature. The Sensometrics® Software displays the measured and derived variables in digital trace form and can store and display trends over periods of up to 240 hours.
Intended Use: Sensometrics® Software is intended for use by a qualified nurse or physician to display variables from existing legally marketed primary measuring devices used to monitor neurosurgical and neurological patients. The parameters displayed by the Sensometrics® Software include intracranial pressure, arterial blood pressure, oxygen, core/cerebral temperature, and these are continuously trended in charts. The presentation of both measured parameters and the relationships between them is intended as an adjunct to the information provided by existing primary monitors. The Sensometrics Software should not be used alone as the sole basis for decisions as to diagnosis or therapy.
Summary of Substantial Equivalence: Sensometrics® Software is similar in design, intended use and performance to the NeuroSystems device. It has the same intended use in that it is an adjunct to primary measuring devices. There are no major differences between the two devices. Sensometrics Software is also substantially equivalent to the Phillips and Integra device. All three devices have the same outputs. No new issues of safety or effectiveness are introduced by using this device.
{1}------------------------------------------------
## K101133, Sensometrics® Software -- Response November 16, 2010
・
### Comparison Table
:
| Parameter | Proposed device | Predicate Device 1 | Predicate Devices 2 & 3 |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Identification | Sensometrics<br>Software | NeuroSystems 1<br>Monitor, K050702 | Phillips IntelliVue<br>Mp40, Integra MPM-1 |
| 510(k) No. | K101133 | K050702 | K032858, K962928 |
| Technology | Intracranial pressure<br>monitoring device | Same | Same. |
| Intended Use | Sensometrics®<br>Software is intended<br>for use by a qualified<br>nurse or physician to<br>display variables<br>from existing legally<br>marketed primary<br>measuring devices<br>used to monitor<br>neurosurgical and<br>neurological patients.<br>The parameters<br>displayed by the<br>Sensometrics®<br>Software include<br>intracranial pressure,<br>arterial blood<br>pressure, oxygen,<br>core/cerebral<br>temperature, and<br>these are<br>continuously trended<br>in charts. The<br>presentation of both<br>measured parameters<br>and the relationships<br>between them is<br>intended as an<br>adjunct to the<br>information provided<br>by existing primary<br>monitors. The<br>Sensometrics®<br>Software should not<br>be used alone as the<br>sole basis for | Same | Same except these<br>devices are the primary<br>measuring devices |
| | diagnosis or therapy. | | |
| Variables<br>measured | The measured<br>parameters are<br>described above in<br>detail. All parameters<br>have their origin<br>from the measured<br>signal which is<br>equivalent to the<br>predicate device. | Same but might have<br>different presentation<br>policy with regard to<br>pressure parameters<br>presented. | Same |
| Design: | Software installed on<br>a PC connected via a<br>RJ45/RS232 cable to<br>primary monitors | Monitor connected to<br>primary monitors | Primary Monitor |
| Data Storage | Dynamic and limit<br>depends on<br>configuration | 1, 2, 8 or 24 hrs. | Unknown |
| Trace Form | Digital | Analog or digital | Analog or digital |
.
. .
{2}------------------------------------------------
#### Summary of Testing:
Sensometrics® Software has undergone testing on several levels:
Module/Unit testing
· Subsystem testing
Integration test
Verification and validation testing throughout the development life cycle
Final acceptance testing of the final product
The device performed according to specifications.
The Software was also compared with the Phillips and Integra devices in a performance comparison test. The test objective was to compare Mean ICP, Mean ABP, systolic pressure output, diastolic pressure output and cerebral temperature output between the three devices. The test results show that the parameters computed by Sensometrics software are equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
dPCom AS c/o Ms. Constance G. Bundy C.G. Bundy Associates, Inc. 435 Rice Ccreek Terrace NE Fridley, MN 55432
NOV 1 9 2010
Re: K101133
Trade/Device Name: Sensometrics® Software Regulation Number: 21 CFR 882.1620 Regulation Name: Intracranial Pressure Monitoring Device Regulatory Class: Class II Product Code: GWM Dated: August 25, 2010 Received: August 31, 2010
Dear Ms. Bundy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title-21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Constance G. Bundy
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Erich Fromm fr
Malvina B. Eydelman, M.D Director Division of Ophthalmic. Neurological. and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K101133
### dPCom AS Oslo, Norway
# Indications for Use
510(k) Number (if known):
Device Name: Sensometrics® Software
Indications For Use:
Sensometrics Software is intended for use by a qualified nurse or physician to display variables from existing legally marketed primary measuring devices used to monitor neurosurgical and neurological patients. The parameters displayed by the Sensometrics Software include intracranial pressure, arterial blood pressure, oxygen, core/cerebral temperature, and these are continuously trended in charts. The presentation of both measured parameters and the relationships between them is intended as an adjunct to the information provided by existing primary monitors. The Sensometrics® Software should not be used alone as the sole basis for decisions as to diagnosis or therapy.
Prescription Use X
AND/OR Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Michael Hoffmann
Page 1 of
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
101133
510(k) Number
3
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.