MAGSTIM DOUBLE 70-2 COIL

K130403 · The Magstim Company , Ltd. · GWF · Aug 23, 2013 · Neurology

Device Facts

Record IDK130403
Device NameMAGSTIM DOUBLE 70-2 COIL
ApplicantThe Magstim Company , Ltd.
Product CodeGWF · Neurology
Decision DateAug 23, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.1870
Device ClassClass 2

Intended Use

The Magstim D702 coil is a stimulating coil intended for use with the Magstim® 2002 Stimulating Unit or the Magstim® Rapid2 Stimulating Unit for the purpose of peripheral nerve stimulation for diagnostic purposes. The coil is an accessory to the Magstim ® 2002 Stimulating Unit, and the coil is an accessory to the Magstim® Rapid2 Stimulating Unit.

Device Story

The Magstim D702 coil is a non-invasive, reusable accessory for Magstim 2002 or Rapid2 stimulating units. It functions by delivering high-intensity magnetic pulses to peripheral nerves via electromagnetic induction, creating eddy currents in conductive body tissue. This allows for deep nerve stimulation without skin preparation or removal of clothing. The device is used in clinical or laboratory settings by trained professionals. The coil features overlapping windings, which improve efficiency in achieving motor thresholds with lower stimulator output compared to previous models. The healthcare provider positions the coil over the target nerve; the resulting stimulation assists in diagnostic procedures. The device is transportable and intended for short-term contact (up to 24 hours).

Clinical Evidence

Bench testing only. Testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), biocompatibility (ISO 10993-1), usability, and magnetic field measurement. Results demonstrated performance consistent with predicate specifications.

Technological Characteristics

Reusable, non-sterile stimulating coil. Features overlapping windings (11 turns, 2 windings), 12.5 mm2 wire cross-sectional area, and 1.8m cable. Operates via electromagnetic induction. Complies with IEC 60601-1, IEC 60601-1-2, and ISO 10993-1. No active cooling. Integral adapter included.

Indications for Use

Indicated for peripheral nerve stimulation for diagnostic purposes in patients requiring such assessment.

Regulatory Classification

Identification

An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KI304023 ### 510(k) Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 C.F.R. §807.92. The submitter of this J. P. Quellette premarket notification Principal is: Custom Microbiology Associates 101 Milk Street Methuen, Massachusetts 01844 Tel: (978) 682-8085 Fax: (207) 912-8085 e-mail: jpouellette01@gmail.com **AUG** 2 3 2013 Date of summary March 19, 2013 D702 Coil Device name Common name Evoked Response Electrical Stimulator | | Classification names Regulation Number | ProCode | Classification Name | |--|-------------------------------------------------|---------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | 882.1870 | GWF | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Evoked Response | | | | | Clootrion Stimulator | Device Description The Magstim D702 coil is designed for use with the Magstim 2002 stimulator or the Magstim Rapid2 stimulator. The Magstim D70 coil provides a method of non-invasively delivering its output to a target such as the peripheral nerves. Discharge current through the coil generates a brief, high intensity magnetic pulse. When the coil is applied to a conductive medium such as the human body, eddy currents are produced in the medium by electromagnetic induction. This stimulation method enables deep and otherwise inaccessible nerves to be stimulated easily and relatively painlessly. In addition, no skin preparation is required and stimulation can be achieved through clothing. The coil is a non-sterile reusable device intended for multipatient use. The coil is transportable and is intended to be used in a laboratory, consulting room or equivalent environment. The coil is intended for short-term patient contact where single or cumulative contact does not exceed 24 hours. Predicate Devices This device is substantially equivalent to the current Magstim Model 2002 with Double 70mm Remote Coil (K060847) and the Magstim Model Rapid2 with Double 70mm Remote Coil (K051864). {1}------------------------------------------------ The primary modification to this device is the overlapping windings in the Modifications coil resulting in improved achievement of motor thresholds with less stimulator output. The modified device has the same intended use as the legally marketed Intended Use predicate devices. The Magstim D70+ coil is a stimulating coil intended for use with the Magstim 2002 Stimulating Unit or the Magstim Rapid-Stimulating Unit for the purpose of peripheral nerve stimulation for diagnostic purposes. The coil is an accessory to the Magstim 200-Stimulating Unit and the coil is an accessory to the Magstim Rapid2 Stimulating Unit. The modified device has the same technological characteristics as the Technological characteristics legally marketed predicate devices. Verification and validation testing activities were conducted to establish the Testing performance and reliability characteristics of the modified device and to assess any of the differences between the subject device and the predicate device. Testing involved safety testing from the risk analysis, including laboratory studies for electrical safety testing, EMC testing, usability testing for consumer accuracy as well as magnetic field measurement testing. Acceptance criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The Magstim Model 2002 with D702 coil and the Magstim Model Rapid2 Standards with D704 coil were designed to comply with the applicable portions of the following product standards: 1. IEC 60601-1:1998 2. IEC 60601-1-2:2007 3. ISO 10993-1:2009 Conclusion This device is substantially equivalent to the predicate device and does not present any new issues of safety or effectiveness. The intended use of the modified device, as described in its labeling, has not changed from that of the predicate device as a result of the modifications. The only difference between the subject device and the predicate device were qualified with appropriate testing. The fundamental scientific technology employed in the operation of this device has not changed from the predicate device. {2}------------------------------------------------ #### 5.0 Substantial Equivalence . · 5.1-Similarities and Differences The similarities and differences between the predicate and subject device are listed in Table 5.1. Table 5.1 Comparison of Similarities and Differences | Comparative<br>Characteristic | Subject Device<br>Magstim D70² Coil<br>P/N 4102-00 | Predicate Device<br>Magstim 2ⁿᵈ Generation<br>Double 70mm Coil<br>P/N 3191-00<br>(Reference K060847) | Predicate Device<br>Magstim 2ⁿᵈ Generation<br>Double 70mm Coil<br>P/N 3191-00<br>(Reference K051864) | |-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------| | Intended Use | The Magstim D70² coil is a<br>stimulating coil intended for<br>use with the Magstim 200²<br>Stimulating Unit or the<br>Magstim Rapid² Stimulating<br>Unit for the purpose of<br>peripheral nerve stimulation<br>for diagnostic purposes. The<br>coil is an accessory to the<br>Magstim 200² Stimulating Unit<br>and the coil is an accessory to<br>the Magstim Rapid²<br>Stimulating Unit. | The Magstim Model 200²<br>is a magnetic nerve<br>stimulator intended for<br>the purpose of peripheral<br>nerve stimulation for<br>diagnostic purposes. | The Magstim Model<br>Rapid², Magstim Super<br>Rapid² is indicated for<br>stimulation of peripheral<br>nerves | | Indications for<br>Use | The Magstim D70² coil is a<br>stimulating coil intended for<br>use with the Magstim 200²<br>Stimulating Unit or the<br>Magstim Rapid² Stimulating<br>Unit for the purpose of<br>peripheral nerve stimulation<br>for diagnostic purposes. The<br>coil is an accessory to the<br>Magstim 200² Stimulating Unit<br>and the coil is an accessory to<br>the Magstim Rapid²<br>Stimulating Unit. | The Magstim Model 200²<br>is a magnetic nerve<br>stimulator intended for<br>the purpose of peripheral<br>nerve stimulation for<br>diagnostic purposes | The Magstim Model<br>Rapid², Magstim Super<br>Rapid² is indicated for<br>stimulation of peripheral<br>nerves | | Average<br>inductance | 13μH | 15.50μH | 15.50μH | | Number of Turns | 11 | 9 | 9 | . {3}------------------------------------------------ | Number of<br>Windings | 2 | 2 | 2 | | |----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|--------------------------|------------|------| | Wire Cross<br>Sectional Area (1) | 12.5 mm2 | 9 mm2 | 9 mm2 | | | Coil Cable<br>Length | 1.8m | 2.0m | 2.0m | | | Manner of<br>stimulation | Peripheral | Peripheral | Peripheral | | | Peak Magnetic<br>Field | 2.1T (2)<br>(Magstim 2003) | 1.2T<br>(Magstim Rapid2) | 1.7T (2) | 1.2T | | Number of<br>pulses the coil<br>can be operated<br>at 80%<br>stimulator output<br>before reaching<br>maximum<br>operating<br>temperature (1) | >290<br>(Magstim 2003) | >600<br>(Magstim Rapid1) | 110 | 110 | | Cooling | No | No | No | | | Multiple Use | Yes | Yes | Yes | | | Sterile | No | No | No | | | Weight | 1.95kg | 1.8kg | 1.8kg | | | Integral adapter | Yes | No | No | | Notes: 5.3 Similarities - (1) Larger cross sectional wire area reduces resistance and heat generation. As a result the coil can run longer before warming up. This does not mean it will be used to deliver more stimuli: the number of stimuli delivered is ultimately up to the physicians conducting the test. - (2) Peak magnetic field strengths were measured using a very thin magnetic field probe (0.8mm) in order to get as close to the coil surface as possible. Since magnetic field strength reduces with distance away from a coil, probe size does affect the measurement outcome. At typical stimulation depths magnetic field figures are, therefore, lower. The predicate device is the Magstim 2nd Generation Double 70mm 5.2 Predicate devices Coil p/n 3191-00 evaluated with the Magstim Rapid² TMS unit in K051864 and with the Magstim 2002 TMS unit in K060847. > The subject device and the predicate device are similar with respect to respect to indications for use and intended use, fundamental scientific technology as well as the number of windings, the lack of the need for cooling, the ability to be used multiple times and are limited to peripheral stimulation only. {4}------------------------------------------------ #### 5.4 Differences The subject device and the predicate devices are different due to the overlapping of the windings in the D70-coil (P/N 4102-00) which results in improved achievement of motor thresholds with less stimulator output. This design feature also manifests different values for peak magnetic field and pulses generated when the subject device is used with one or the other stimulating unit. When the subject device is used with the Magstim Rapid", more pulses are generated than with the Magstim 2002 until the peak temperature is achieved and the coil shuts down, This design feature allows the practitioner the flexibility to use the D70- coil and the Magstim Rapid2 for a longer period until the peak temperature is achieved and the coil shuts down. When the subject device is used with the Magstim 2002, a slightly higher peak magnetic field value is achieved on the coil surface. This is due to the higher efficiency of the coil as well as the thickness of the plastic cover and the inner coil insulation of the subject device. Peak magnetic field figure of 2.1 tesla is below that used inside 3 tesla MRI scanners. Based upon the analysis of the similarities and differences, the Magstim D702 Coil as described in this 510(k) is substantially equivalent to the predicate device and does not present any new issues of safety or effectiveness. #### 5.5 Conclusion {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the seal is a stylized symbol that resembles an abstract caduceus or a person embracing a globe. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W066-G609 Silver Spring, MD 20993-0002 #### August 23, 2013 The Magstim Company Limited c/o Mr. J.P. Quellette, Custom Microbiology Associate 101 Milk Street Methuen, MA 01844 Re: K130403 Trade/Device Name: D702 coil Regulation Number: 21 CFR 882.1870 Regulation Name: Evoked response electrical stimulator Regulatory Class: Class II Product Code: GWF Dated: July 8, 2013 Received: July 11, 2013 Dear Mr. Ouellette: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {6}------------------------------------------------ Page 2 - Mr. J.P. Ouellette forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Victor Krauthamer -S Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K130403 Device Name: D702 Coil Indications For Use: The Magstim D702 coil is a stimulating coil intended for use with the Magstim® 2002 Stimulating Unit or the Magstim® Rapid2 Stimulating Unit for the purpose of peripheral nerve stimulation for diagnostic purposes. The coil is an accessory to the Magstim ® 2002 Stimulating Unit, and the coil is an accessory to the Magstim® Rapid2 Stimulating Unit. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Victor Krauthamer - S 2013.08.23 15:37:34 -04'00' Division Sign Off) Division of Neurological and Physical Medicine Devices (DNPMD) 510(k) Number K130403 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...