K130013 · Advanced Brain Monitoring, Inc. · OMC · Jan 31, 2013 · Neurology
Device Facts
Record ID
K130013
Device Name
X4 SYSTEM
Applicant
Advanced Brain Monitoring, Inc.
Product Code
OMC · Neurology
Decision Date
Jan 31, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.1400
Device Class
Class 2
Indications for Use
The X4 System is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, record, transmit and display physiological signals from adult patients. The X4 System acquires, records, transmits and displays electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG), accelerometer, acoustical and photoplethesmographic signals. The X4 system only acquires and displays physiological signals, no claims are being made for analysis of the acquired signals with respect to the accuracy, precision and reliability.
Device Story
X4 System acquires physiological signals (EEG, EOG, ECG, EMG, PPG, acoustic, 3-D accelerometer) via head-worn sensors. Device records data to internal memory or transmits wirelessly to PC/cloud server. Operated by patient (self-application) or technician. Firmware monitors signal quality during acquisition. Software (PC or web portal) manages study initiation, patient tracking, data upload, and signal visualization. Output is raw physiological data in European Data Format (EDF) for review by healthcare providers. Device facilitates remote or clinical monitoring; no automated diagnostic analysis or clinical interpretation claims made.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical software verification and validation testing confirming identical performance of the new web-based Device Manager module compared to existing PC-based software.
Technological Characteristics
Physiological signal acquisition system. Components: head strip, sensors, 3-D accelerometer, acoustic microphone. Connectivity: wireless transmission, PC/cloud-based storage. Data format: European Data Format (EDF). Software: firmware for signal quality monitoring; PC/web-portal software for study management and visualization.
Indications for Use
Indicated for adult patients in home, healthcare facility, or clinical research settings for the acquisition, recording, transmission, and display of physiological signals (EEG, EOG, ECG, EMG, accelerometer, acoustic, and PPG). No contraindications specified.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
Apnea Risk Evaluation System (ARES), Model 610 (K112514)
Submission Summary (Full Text)
{0}------------------------------------------------
K130013
#### Advanced Brain Monitoring, Inc. X4 System
## 510(k) Summary
# JAN 3 1 2013
In accordance with 21 CFR 807.92 the following summary of information is provided:
DATE: December 28, 2012
## SUBMITTER:
Advanced Brain Monitoring 2237 Faraday Avenue, Suite 100 Carlsbad, CA 92008 T 760.720.0099 F 760.720.3337
## PRIMARY CONTACT PERSON:
Adrienne Lenz, RAC Founder Pathway Regulatory Consulting, LLC T 262-290-0023
## SECONDARY CONTACT PERSON:
Dan Levendowski President and Co-founder Advanced Brain Monitoring, Inc.
#### DEVICE:
TRADE NAME: X4 System
COMMON/USUAL NAME: X4
CLASSIFICATION NAMES: 882.1400 Electroencephalograph
PRODUCT CODE: OMC
#### PREDICATE DEVICE(S):
K120447 X4 System
K112514 Apnea Risk Evaluation System (ARES), Model 610
{1}------------------------------------------------
#### DEVICE DESCRIPTION:
The X4 system is used for configurable acquisition of physiological signals. Model X4-E provides for acquisition of three channels of electroencephalography (EEG) and one photoplethesmographic (PPG) signal from a head strip, with an optional channel connected to two sensors via a dual-lead connector with twice the gain. Model X4-M provides four channels of EEG with the dual-lead connector providing the input for reference sensors. Both models measure sound via an acoustic microphone, and movement and position measured via a 3-D accelerometer. The device is designed so it can be affixed by the patient and to record data. Alternatively, a technician can affix the device and display the signals via a wireless connection during acquisition. The X4 system firmware monitors signal quality to ensure that the sensors are properly applied and that high quality signals are being acquired.
The X4 software provides a means to: a) initiate a study and track patient information, b) acquire and save signals to the memory of the device, c) acquire and wirelessly transmit signals from the device, d) upload data saved in the memory of the device to a PC, and e) visually inspect the signal quality.
The acquired signals are saved in a universal data format (European Data Format - EDF). The study record, once saved on the PC, is available for analysis by Advanced Brain Monitoring's Sleep Profiler software application. The X4's downloaded study will reside on either or local PC or a cloud server, which can be a physical or virtual server. Software on the cloud server is accessed via web portal software.
#### INTENDED USE:
The X4 System is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, record, transmit and display physiological signals from adult patients. The X4 System acquires, records, transmits, and displays electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG); and accelerometer, acoustical, and The X4 system only acquires and displays physiological signals, no photoplethesmographic signals. claims are being made for analysis of the acquired signals with respect to the accuracy, precision and reliability.
#### TECHNOLOGY:
The X4 System is identical to the previously cleared X4 System (K120447). A Device Manager Module was developed to run via a web portal as an alternative to the identical functions provided in the PC software. The Device Manager module running on the cloud server uses the same fundamental technology as the ARES Model 610 portal (K112514). The technologies used in the X4 are used in the same manner as the predicate products and do not raise new questions of safety and effectiveness.
{2}------------------------------------------------
### DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
#### SUMMARY OF NON-CLINICAL TESTS:
Support for the substantial equivalence of the X4 System was provided as a result of risk management and software testing.
The X4 software has been thoroughly tested through verification of specifications and validation, including software validation. The key metric for software verification/validation was confirmation of identical performance using either the desktop or portal for the key functions associated with the Device Manager software:
- . format the device for a new patient,
- . enter and upload study identification information to the device,
- . download study information to data storage, and
- . upload new firmware.
The results of the verification and validation activities that have been performed demonstrate that the software meets requirements for safety, function, and intended use.
#### SUMMARY OF CLINICAL TESTS:
The modifications to the X4 System that are the subject of this premarket submission did not require clinical studies to support substantial equivalence. The functionality of the modified device was completely evaluated by performing nonclinical verification and validation.
#### CONCLUSION:
Advanced Brain Monitoring considers the X4 System to be as safe, as effective, and substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 31 2013
Advanced Brain Monitoring, Incorporated C/O Ms. Adrienne Lenz, RAC Pathway Regulatory Consulting, Limited Liability Company W324 S3649 County Road East DOUSMAN, WI 53118
Re: K130013
Trade/Device Name: X4 System Regulation Number: 21 CFR 882.1400 Regulation Name: Neurological Diagnostic Devices Regulatory Class: II Product Code: OMC Dated: December 31, 2012 Received: January 29, 2013
Dear Ms. Lenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
hor
Harshitio.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Advanced Brain Monitoring, Inc. X4 System
510(k) Number (if known): K130013
Device Name:
X4 System
#### Indications for Use:
The X4 System is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, record, transmit and display physiological signals from adult patients. The X4 System acquires, records, transmits and displays electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG), accelerometer, acoustical and photoplethesmographic signals. The X4 system only acquires and displays physiological signals, no claims are being made for analysis of the acquired signals with respect to the accuracy, precision and reliability.
Prescription Use X AND/OR
(Part 21 CFR 801 Subpart D)
Over-The-Counter Use
(Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Lester W. Schultheis Jr 2013.01.31 08:33:39 -05'00'
Division Sign-Off) Division of Anesthesiology, General Hospital intection Control, Dental Devices
510(k) Number: K130013
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.