K242094 · Beacon Biosignals, Inc. · OLZ · Nov 22, 2024 · Neurology
Device Facts
Record ID
K242094
Device Name
Dreem 3S
Applicant
Beacon Biosignals, Inc.
Product Code
OLZ · Neurology
Decision Date
Nov 22, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
AI/ML, PCCP
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Sleep staging
Artificial neural network with recurrent and attention layers
—
—
—
—
—
—
Indications for Use
The Dreem 3S is intended for prescription use to measure, record, display, transmit and analyze the electrical activity of the brain to assess sleep and awake in the home or healthcare environment. The Dreem 3S can also output a hypnogram of sleep scoring by 30-second epoch and summary of sleep metrics derived from this hypnogram. The Dreem 3S is used for the assessment of sleep on adult individuals (22 to 65 years old). The Dreem 3S allows for the generation of user/predefined reports based on the subject's data.
Device Story
Headband-worn device; collects EEG signals via 6 dry electrodes and movement data via 3D accelerometer; data transmitted to servers for analysis. Artificial neural network (recurrent/attention layers) processes frequency spectrums of raw data to perform automatic sleep staging per AASM classification; outputs 30-second epoch hypnograms and sleep metrics. Used by clinicians to aid diagnosis of sleep disturbances; provides PDF reports. Includes bone-conduction speaker for notifications. PCCP allows for future updates to signal preprocessing, ML model, and postprocessing.
Clinical Evidence
No new clinical, bench, or animal testing performed for this submission. Performance characteristics remain identical to the predicate device (K223539).
Technological Characteristics
Headband with ABS plastic, foam, and polyester fabric. 6 dry EEG electrodes; 3D accelerometer. Bone-conduction speaker. Connectivity: data transfer to servers. Software: Artificial neural network (recurrent/attention layers). PCCP included for algorithmic updates.
Indications for Use
Indicated for assessment of sleep in adults (22-65 years) in home or healthcare settings; prescription use only.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 22, 2024
Beacon Biosignals, Inc. Alexander Chan VP of Analytics and Machine Learning 22 Boston Wharf Road 7th Floor, Unit 41 Boston, Massachusetts 02210
Re: K242094
Trade/Device Name: Dreem 3S Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLZ, OLV Dated: October 18, 2024 Received: October 21, 2024
Dear Alexander Chan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not
{1}------------------------------------------------
required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
{2}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay R. Gupta -S
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
K242094
Device Name
Dreem 3S
Indications for Use (Describe)
The Dreem 3S is intended for prescription use to measure, record, display, transmit and analyze the electrical activity of the brain to assess sleep and awake in the home or healthcare environment. The Dreem 3S can also output a hypnogram of sleep scoring by 30-second epoch and summary of sleep metrics derived from this hypnogram.
The Dreem 3S is used for the assessment of sleep on adult individuals (22 to 65 years old). The Dreem 3S allows for the generation of user/predefined reports based on the subject's data.
Type of Use (Select one or both, as applicable)
< Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for Beacon Biosignals. The logo consists of the word "BEACON" in bold, black letters above the word "BIOSIGNALS" in smaller, lighter letters. To the right of the text is a stylized graphic of a brain, formed by a series of connected, rounded lines in a light green color.
510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92.
### SUBMITTER
| Applicant | Beacon Biosignals, Inc.<br>22 Boston Wharf Rd.<br>7th Floor, Unit 41<br>Boston, MA 02210 |
|---------------|------------------------------------------------------------------------------------------|
| Contact | Alexander Chan<br>VP of Analytics and Machine Learning<br>Email: alex.chan@beacon.bio |
| Date Prepared | 21 June 2024 |
# DEVICE INFORMATION
### Subject Device
| Name of Device | Dreem 3S |
|----------------|-----------------|
| Regulation | 21 CFR 882.1400 |
| Product Code | OLZ, OLV |
| Device Class | Class II |
| Review Panel | Neurology |
### Predicate Device
| Predicate Manufacturer | Beacon Biosignals, Inc. |
|------------------------|-------------------------|
| Predicate Trade Name | Dreem 3S |
| Predicate 510(k) | K223539 |
## DEVICE DESCRIPTION
The Dreem 3S headband contains microelectronics, within a flexible case made of plastic, foam, and fabric. It includes 6 EEG electrodes and a 3D accelerometer sensor.
The EEG signal is measured by two electrodes in the frontal position) and two at the back of the head (occipital position), along with one reference electrode and one ground electrode.
The 3D accelerometer is embedded in the top of the headband to ensure accurate measurements of the wearer's head movement during the night. The raw EEG and accelerometer data are transferred to Dreem's servers for further analysis after the night is over.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Beacon Biosignals. The logo consists of a stylized graphic of several vertical lines of varying heights, resembling a signal or waveform. To the right of the graphic is the text "BEACONBIOSIGNALS" in a simple, sans-serif font. The color scheme is primarily light blue and white, giving the logo a clean and modern appearance.
The device includes a bone-conduction speaker with volume control to provide notifications to the wearer, and a power button circled by a multicolor LED light
The device generates a sleep report that includes a sleep staging for each 30-second epoch during the night. This output is produced using an algorithm that analyzes data from the headband EEG and accelerometer sensors. A raw data file is also available in EDF format.
The algorithm uses raw EEG data and accelerometer data to provide automatic sleep staging according to the AASM classification. The algorithm is implemented with an artificial neural network. Frequency spectrums are computed from raw data and then passed to several neural network layers including recurrent layers and attention layers. The algorithm outputs prediction for several epochs of 30 seconds at the same time, every 30 seconds. The various outputs for a single epoch of 30 seconds are combined to provide robust sleep scoring.
## INDICATION FOR USE
The Dreem 3S is intended for prescription use to measure, record, display, transmit and analyze the electrical activity of the brain to assess sleep in the home or healthcare environment.
The Dreem 3S can also output a hypnogram of sleep scoring by 30-second epoch and summary of sleep metrics derived from this hypnogram.
The Dreem 3S is used for the assessment of sleep on adult individuals (22 to 65 years old). The Dreem 3S allows for the generation of user/predefined reports based on the subject's data.
## TECHNOLOGICAL CHARACTERISTICS AND COMPARISON
The Dreem 3S is substantially equivalent to the previous version of the Dreem 3S (K223539) based on the technological and performance characteristics as described in the summary table below.
The subject device Dreem 3S and the predicate Dreem 3S (K223539) have identical technological characteristics except for the presence of a Predefined Change Control Plan (PCCP) that is a new characteristic of the subject device. The PCCP allows for the algorithmic components of the Dreem 3S device. Both subject and predicate device are for the assessment of sleep and are both used to measure, record, display, transmit and analyze physiological parameters during sleep and wake in the home and healthcare facility. Both subject and predicate devices are used to aid diagnosis of adult patients with disturbed sleep. Both devices allow for the generation of user/predefined reports based on the subject's data. Both the subject and predicate devices may be used in home and healthcare facilities. Both devices are for prescription use only.
#### It is thus concluded that the intended use of the Dreem 3S is substantially equivalent the previously cleared version of the Dreem 3S (K223539).
The comparison table below is provided as a summary of the most relevant characteristics of the Dreem 3S relative to the predicate device.
| Technological<br>Characteristic | Subject Device | Predicate Device | Comparison to<br>Predicate Device |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| Device Name | Dreem 3S | Dreem 3S (K223539) | N/A |
| Manufacturer | Beacon Biosignals, Inc. | Beacon Biosignals, Inc. | Same |
| Regulation Number | 21 CFR 882.1400 | 21 CFR 882.1400 | Same |
| Class | 2 | 2 | Same |
| Device<br>Classification Name | Automatic Event<br>Detection Software<br>for PSG with EEG | Automatic Event<br>Detection Software<br>for PSG With EEG | Same |
| Product Codes | OLZ, OLV | OLZ, OLV | Same |
| Portable Design | Yes | Yes | Same |
| Patient Worn Device | Yes | Yes | Same |
| Physical dimensions | Head perimeter<br>540mm to 620mm.<br>One size fits all.<br>Adjustable with XS, S,<br>M, L spacers. | Head perimeter<br>540mm to 620mm.<br>One size fits all.<br>Adjustable with XS, S.<br>M, L spacers. | Same |
| Weight | 130g | 130g | Same |
| Materials | ABS Soft polyester<br>fabric | ABS Soft polyester<br>fabric | Same |
| Method of Connection<br>to Patient | 6 dry electrodes for<br>EEG assessment on<br>the headband.<br>3D-accelerometer for<br>movement/body<br>position assessment.<br>Bone conduction<br>audio system. | 6 dry electrodes for<br>EEG assessment on<br>the headband.<br>3D-accelerometer for<br>movement/body<br>position assessment.<br>Bone conduction<br>audio system. | Same |
| Data Analysis | Automatic scoring and<br>derived sleep metrics<br>are provided to the<br>health care provider<br>through a specific<br>report, in a pdf file.<br>Manual analysis and<br>marking are available<br>on raw data. | Automatic scoring and<br>derived sleep metrics<br>are provided to the<br>health care provider<br>through a specific<br>report, in a pdf file.<br>Manual analysis and<br>marking are available<br>on raw data. | Same |
| #Channels of data<br>recorded | - 5 EEG channels<br>from 5 electrodes<br>(4 frontal-occipital<br>derivations, 1<br>frontal-frontal<br>derivation)<br>- 3 Triaxial<br>accelerometer<br>channels | - 5 EEG channels<br>from 5 electrodes<br>(4 frontal-occipital<br>derivations, 1<br>frontal-frontal<br>derivation)<br>- 3 Triaxial<br>accelerometer<br>channels | Same |
| Operating Time | Up to 24 hours | Up to 24 hours | Same |
| Recording Time | Same as the operating<br>time. Up to 24 hours. | Same as the operating<br>time. Up to 24 hours. | Same |
| Predetermined<br>Change Control Plan | Includes PCCP that<br>allows for the update<br>of the signal<br>preprocessing,<br>machine learning<br>model, and probability<br>postprocessing. | No PCCP | Different - Predicate<br>devices does not<br>include a PCCP |
#### Table 1 : Comparison of technological characteristics
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Beacon Biosignals. The logo consists of a stylized graphic of several vertical lines of varying heights, resembling a signal or waveform. To the right of the graphic is the text "BEACONBIOSIGNALS" in a simple, sans-serif font. The text is in a dark gray color, while the graphic is in a light teal color.
{7}------------------------------------------------
## PERFORMANCE TESTING
No bench testing, animal testing, or clinical testing was performed to support this submission. Because this submission only involves the additional inclusion of the Predetermined Change Control Plan, the performance characteristics of the submitted device are the same as previously reported for the predicate device (K223539).
# PREDETERMINED CHANGE CONTROL PLAN
Dreem 3S includes an authorized Predetermined Change Control Plan (PCCP) that allows for planned updates of the machine learning software device function (ML-DSF) and non-ML algorithmic components to improve sleep staging performance within the existing intended use and indications for use. This PCCP allows for the modification of the algorithmic components of Dreem 3S including the signal preprocessing, machine learning model, or postprocessing to achieve increased sleep staging performance. The three modifications are summarized in the table below.
| # | Modification | Description |
|---|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Update of Signal<br>Preprocessing<br>Steps | Dreem 3S's EEG signal preprocessing may be modified for the<br>purposes of improving sleep staging performance within the intended<br>use population by:<br>Updating the parameters of the digital signal processing steps<br>(e.g., filtering) applied to the Dreem 3S headband signals<br>before being input to the machine learning model |
| 2 | Update of Machine<br>Learning Model | Dreem 3S's sleep staging neural network may be modified for the<br>purposes of improving sleep staging performance within the intended<br>use population by:<br>Retraining with an updated training/tuning dataset Retraining with updated hyper-parameters, loss function,<br>optimizer Retraining with updated model selection criteria Retraining with an updated neural network architecture with<br>limitations on model size and type |
| 3 | Update of<br>Probability<br>Postprocessing | Dreem 3S's probability postprocessing may be modified for the<br>purposes of improving sleep staging performance within the intended<br>use population by:<br>Updating the methods by which sleep stages are generated<br>from the model output sleep stage probabilities |
Modifications 1 and 2 above would trigger re-training of the machine learning model, while modification 3 would not trigger re-training of the machine learning model.
The testing of any modification to Dreem 3S within the scope of the PCCP will include comprehensive software verification and validation testing, including repeating unit, integration, and system level tests related to the software components affected by the change. All of these software
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Beacon Biosignals. The logo consists of a stylized graphic of brainwaves in a light teal color, followed by the text "BEACONBIOSIGNALS" in a simple, sans-serif font. The text is in a light gray color.
verification tests must pass for a modification to be considered valid. In addition, clinical performance validation will also be repeated, and will require that the performance of any modification to Dreem 3S to be non-inferior to the all previously released versions of the Dreem 3S device. Non-inferiority marqins were selected based on the level of human variability estimated from comparison of expert scoring from 284 American Academy of Sleep Medicine (AASM) compliant polysomnography recordings.
Recordings that are used for any purpose (e.g., training, tuning, failure analysis, etc.) that might lead to direct or indirect insight regarding the performance of a modified sleep staging algorithm on this recording, other than execution of the clinical performance validation per the methods specified in the PCCP, are excluded from the test dataset. Quality checks will ensure that the test data are sufficiently high quality and representative of the intended use population.
Upon a release of an updated version of Dreem 3S based on this PCCP, communication will be provided to all clinical users of Dreem 3S, informing them that a new version of Dreem 3S is available, with a description of the release and its updated performance.
# CONCLUSIONS
The Dreem 3S is substantially equivalent to the previous version of the legally marketed Dreem 3S (K223539) and presents no new concerns about safety or effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.