MODIFICATION TO EEG MONITOR WITH BIS, MODEL A-2000

K031694 · Aspect Medical Systems, Inc. · OLW · Aug 29, 2003 · Neurology

Device Facts

Record IDK031694
Device NameMODIFICATION TO EEG MONITOR WITH BIS, MODEL A-2000
ApplicantAspect Medical Systems, Inc.
Product CodeOLW · Neurology
Decision DateAug 29, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1400
Device ClassClass 2
AttributesReal-World Evidence, Pediatric

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K031694 · Aug 29, 2003MODIFICATION TO EEG MONITOR WITH BIS, MODEL A-2000Aspect Medical Systems, Inc.Historical clinical cohort (Safe-II study)The Safe-II study utilized a historical control group of patients who did not receive BIS monitoring to compare against a prospective cohort of patients monitored with the device to determine the reduction in the incidence of awareness.Historical control; Anesthesia awareness; Cohort study

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Safe-II; Prospective cohort study with historical controlPatients undergoing anesthesia; Sample Size: 5057 (BIS monitored) vs 7801 (historical control)Historical group with no BIS monitoringReduction in the incidence of awareness

Indications for Use

The Aspect A-2000 EEG Monitor with BIS is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The BIS Monitor is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.

Device Story

A-2000 EEG Monitor with BIS is a microprocessor-based, 2-channel EEG monitoring system; acquires EEG signals to monitor brain state; processes EEG data to calculate BIS (Bispectral Index) value; used in hospitals/medical facilities by trained personnel or licensed practitioners; provides real-time BIS output to assist clinicians in monitoring anesthetic agent effects; helps guide anesthetic administration; reduces incidence of intra-operative awareness and recall in adults; provides clinical benefit by mitigating preventable medical errors during anesthesia and sedation.

Clinical Evidence

Evidence includes multiple prospective studies: B-Aware (n=2503, randomized, double-blinded) showed 82% reduction in awareness; Safe-II (n=5057 vs 7801 historical) showed 78% reduction; Kaplan study showed 78% reduction during ICU sedation; AIM study (n=19576) established baseline awareness incidence of 1-2 cases per 1000. Data supports clinical benefit of BIS monitoring in reducing awareness incidence.

Technological Characteristics

Microprocessor-based 2-channel EEG monitor; processes EEG signals to derive BIS index; intended for adult and pediatric use; hospital/medical facility environment.

Indications for Use

Indicated for adult and pediatric patients in hospital or medical facilities to monitor brain state via EEG signal acquisition; used as an aid in monitoring anesthetic agent effects and reducing incidence of awareness with recall during general anesthesia and sedation.

Regulatory Classification

Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K031694 October 10, 2003 #### VII 510(k) SUMMARY Date prepared: 10//09/03 ### Company Name and Address Aspect Medical Systems, Inc. 141 Needham St. Newton, MA 02464 Contact People: Christine Vozella Regulatory Affairs 1775 Rockies Ct. Lafayette, CO 80026 Telephone (303) 926-5624 Fax #: (303) 604-6477 Paul J. Manberg, PhD Vice President, Clinical and Regulatory Affairs Aspect Medical Systems 141 Needham St. Newton, MA 02464 Telephone (direct dial): (617) 559-7076 Fax #: 617-559-7400 # Device Name | Proprietary Name: | A-2000™ EEG Monitor with BIS | |-------------------|------------------------------| |-------------------|------------------------------| Common Name: EEG Monitor # Predicate Devices Aspect Medical Systems EEG Monitor with BIS, Model A-2000 510(k) #K011534 and GE Medical Systems BIS Module 510(k) #K012466. # Description The Aspect Medical Systems, Inc. A-2000 EEG BIS Monitor (hereafter referred to as the BIS Monitor) is an easy to use, microprocessor- based, 2 channel maximum EEG monitoring system. It is used to monitor the state of the brain by data acquisition of EEG. The BIS, a processed EEG variable, may be used as an aid in monitoring the effects of certain anesthetic agents. There are no changes to the Monitor. {2}------------------------------------------------ #### Indications for Use The Aspect A-2000 EEG Monitor with BIS is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The BIS Monitor is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation. ## Brief Description of Clinical Tests The B-Aware study was a prospective, randomized double blinded multi-center study to determine if the BIS technology reduces the incidence of awareness. 1227 were assigned to BIS monitoring, and 1228 to routine care (no BIS was used). Results of this study (n=2503) are that BIS guided anesthesia significantly reduced the incidence of awareness by 82%. The Safe-II was a prospective cohort study undertaken to determine if BIS monitoring would result in a reduction of awareness. This study compared the patients having had BIS monitoring (n=5057) with a historical group (n=7801) that had no BIS monitoring. Results are that use of BIS significantly reduced the incidence of awareness by78%. The Kaplan study showed the risk of awareness and recall was reduced by 78% during sedation in the ICU setting. The AIM (Awareness Incidence and Monitoring) study was a prospective, observational, cohort, multi-center study conducted in 7 hospitals in the US. This study was undertaken to determine the current incidence of intra-operative awareness in the US. Results of this study (n=19576) are that 1-2 cases of awareness per 1000 occur in the US, and this is similar to other non-US findings (i.e. Australia and Sweden). The fact is that there are 20 million cases of anesthesia every year in the US, and there are 20,000 - 40,000 cases of awareness per year. Study results show that this number can be reduced by approximately 80% with the use of BIS monitoring, and can thus offer clinically meaningful benefit. A reduction in awareness provides a public health benefit, in that BIS technology can now provide anesthesiologists with a way to reduce this often debilitating, yet preventable medical error. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its wing and body. The eagle is oriented towards the upper right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Aspect Medical Systems, Inc. c/o Ms. Christine Vozella Regulatory Affairs 1775 Rockies Court Lafayette, Colorado 80026 Re: K031694 Trade/Device Name: EEG Monitor with BIS Model A2000 Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: OLW, ORT, OMC, OLT Dated (Date on orig SE ltr): May 30, 2003 Received (Date on orig SE Itr): June 3, 2003 APR - 9 2012 Dear Ms. Vozella: This letter corrects our substantially equivalent letter of October 10, 2003. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Image /page/3/Picture/7 description: The image shows a stylized signature in black ink on a white background. The signature appears to begin with a large, looping letter, possibly an 'f' or 'j', which then transitions into a smaller, more cursive script. The overall impression is that of a quick, fluid, and somewhat abstract autograph. Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page I of I : :: K031694 510(k) Number (if known): Device Name: A-2000 EEG Monitor with BIS Indications For Use: The Aspect A-2000 EEG Monitor with BIS is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The BIS Monitor is intended for use on adult and pediatic patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. ... ... The BIS may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Optional Format 3-10-98) Cilla all (Division Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Number 031694
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...