K984580 · Biosys AB (Publ) · OLV · Mar 12, 1999 · Neurology
Device Facts
Record ID
K984580
Device Name
BIOSACA SYSTEM
Applicant
Biosys AB (Publ)
Product Code
OLV · Neurology
Decision Date
Mar 12, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Indications for Use
The intended use for the BIOSACA is to record, monitor, display, print, store and transfer bioparameters such as brain, heart and muscle activity, eye movement, blood pressure, breathing and body movements. The BIOSACA unit is intended to be used on an Adult population and for use at home or in health care facilities. It is not intended for use as life support equipment such as vital signs monitoring in intensive care units. The unit may be used for electrocardiography, electroencephalography, electromyography, electrooculography ballistocardiography
Device Story
BIOSACA is a multi-functional, ambulatory biological signal recorder; records up to 22 bioparameters (brain, heart, muscle activity, eye movement, blood pressure, breathing, body movements) via two headboxes, pulse oximeter, and sensor pad. Portable system designed for stationary or mobile operation in home or healthcare facilities. Operator (clinician) views signals in real-time during recording; data stored for later review. Facilitates diagnosis of neurological, cardiology, and sleep disorders. Not for life support or ICU vital signs monitoring. Benefits include minimal patient interaction requirements and silent operation for home use.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Multi-functional ambulatory recorder; 22-channel input capacity (16 from headboxes, 3 from pulse oximeter, 3 from sensor pad). Supports AC/DC signal acquisition. Portable form factor for stationary or mobile use. Connectivity includes data storage and transfer capabilities. No specific materials or software architecture details provided.
Indications for Use
Indicated for adult patients for recording, monitoring, displaying, printing, storing, and transferring bioparameters including EEG, EMG, EOG, ballistocardiography, ECG, breathing, and blood pressure for neurological, cardiology, and sleep disorder investigations. Suitable for home or hospital use. Contraindicated for use as life support equipment, such as vital signs monitoring in intensive care units.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
ं की मौजूद में बाद में बाद में कि महीने की मौजूद में बाद में बाद में कि महीने की मांग की में बाद में कि महीने की मांग की में बाद में कि महीने की मांग की में बाद में कि महीने
1
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
## 510(k) Summary K984580
Safety and effectiveness information concerning the Biosaca from Biosys AB is summarized below.
| Date Prepared | December 18, 1998 | | |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|----------|
| Applicant | BIOSYS AB (publ)<br>Vasaplatsen 8<br>SE-411 34 Göteborg<br>Telephone: +46 31 774 21 25<br>Fax: +46 31 13 98 54<br>e-mail: info@biosys.se | | |
| Contact | Anne Mari Nedevska, Technical Administration Manager | | |
| Device Name | BIOSACA | | |
| Common Name | Biological Signal Recorder | | |
| Classification | No formal classification name or number has been assigned for devices such as the Biosaca. Stand-alone devices for the same use as the Biosaca include the following: | | |
| Panel | Product Code | Name | 21 CFR |
| 84 | GWQ | Electroencephalograph | 882.1400 |
| 84 | GYE | Physiological telemetry system | 882.1855 |
| 84 | GWP | Electromyography | 882.5050 |
| 84 | LEL | Sleep Assessment Device | |
| 80 | FLS | Breathing frequency | 868.2375 |
| 73 | MNR | Ventilatory effort recorder | 868.2375 |
| 74 | DPS | Electrocardiography (ECG) | 870.2340 |
| 74 | DQA | Pulse Oximeter | 870.2700 |
| 73 | DRX | Electrocardiograph electrode | 870.2360 |
Błosys AB (publ).
Vasaplatsen B SE-411 34 Goteborg Telephone +46(0)31 774 21 25 Fax +46(0)31 13 98 54 E-mall info@blosys.se
{1}------------------------------------------------
.
. ...
. . . . . . .
ﺤ
ー・ー
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
| Submission | |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Correspondent | Jane B. Campbell<br>J. & D. Campbell Associates, Inc.<br>485 LaRoe Road<br>Chester, New York 10918<br>Tel. 914-469-4289<br>Fax. 914-469-4212<br>e-mail: jdca@warwick.net |
| Predicate Devices | Compumedics Sleep Monitoring System, Compumedics<br>Sleep Pty. Ltd (K955841)<br>Embla, Flaga hf. (K971813) |
| Device Description | The BIOSACA is a multi-functional and an ambulatory<br>recording device. It is a portable system for the recording,<br>monitoring, storage and transfer of up to 22 bioparameters<br>such as brain, heart and muscle activity, eye movement,<br>blood pressure, breathing, body movements etc. There are<br>applications for the BIOSACA in neurological, cardiology<br>and sleep disorder diagnoses. |
| Function | The BIOSACA is a biological signal recorder able to receive<br>and record up to 22 bioparameters - 16 from two headboxes,<br>AC and/or DC, three from the pulse oximeter and three from<br>the sensor pad. |
| Intended Use | The intended use for the BIOSACA is to record, monitor,<br>display, print, store and transfer bioparameters such as brain,<br>heart and muscle activity, eye movement, blood pressure,<br>breathing and body movements.<br>The BIOSACA unit is intended to be used on an Adult<br>population and for use at home or in health care facilities. It<br>is not intended for use as life support equipment such as vital<br>signs monitoring in intensive care units. The unit may be<br>used for electrocardiography, electroencephalography,<br>electromyography, electrooculography ballistocardiography |
**54**
E-mall info@blosys.se
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -W066-G609 Silver Spring, MD 20993-0002
Biosys AB c/o Jane B.Campbell J. D. Campbell Associates 485 Laroe Road Chester, New York 10918
APR - 9 2012
Re: K984580
Trade/Device Name: BIOSACA Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: OLV, DQA, GWQ, MNR Dated (Date on orig SE ltr): December 18, 1998 Received (Date on orig SE ltr): December 23, 1998
Dear Ms. Campbell:
This letter corrects our substantially equivalent letter of March 12, 1999.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Jane B.Campbell
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Image /page/3/Picture/8 description: The image shows the handwritten word "for" in cursive script. The letter 'f' has a large, sweeping loop that extends both above and below the baseline. The letters 'o' and 'r' are connected and smaller in size compared to the 'f'.
Enclosure
{4}------------------------------------------------
## Indications for Use Statement
K984580 510(k) Number (if known):
Device Name:
BIOSACA
Indications for Use:
:
. ..
The BIOSACA equipment is indicated for use in the recording, displaying, The BIOSACA equipment is materials increases and non-biological signals for sleep monitoring, printing and storage of otoroge of the conceptalography (EEG), disorder and epilepsy investigations as a visuality (EMG), Electrooculography (EOG), Ballistocardiography, etc. BIOSACA is designed for both stationary and (DOO), Ballisiocal diography, cood in either the patient's home or the hospital. mobile operation and may be toos an which can be recorded directly from the Blological siguals are encourted organis are recorded using indirect methods such as pation. Non-of air (breathing), blood pressure, etc.
The equipment is suitable for home use as the patient needs only a minimal amount I he equipment is surations for none setsem has no lights or sounds that could confuse the patient. Signals may be stored for later or may be viewed directly in contuse the patient. Sights may over . When monitoring, the operator sees the signals at the same time as they are recorded.
The Biosaca unit is intended for use on an adult population and at home or I he Diosaca alle is not intended for use as life support equipment such as vital signs monitoring in intensive care units.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation
General Restorative Devices 9845 Am
Prescription Use (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.