Ninety-nine subject clinical study comparing Zmachine algorithm to consensus of human scorers
>1 (certified polysomnographic technologists)
Indications for Use
The CRI Zmachine is a single-channel, EEG acquisition and analysis system, designed for use in the home or clinical environments. This device is intended to be used by qualified healthcare practitioners to monitor the wake and sleep states of adult patients and as an adjunct to their diagnosis of sleep disorders.
Device Story
Zmachine is a battery-operated, single-channel EEG acquisition and analysis system. It records differential-mastoid EEG signals to determine patient wake/sleep states every 30 seconds. Used in home or clinical settings by healthcare practitioners to assist in sleep disorder diagnosis and treatment evaluation. The device processes EEG data in real-time using a proprietary adaptive algorithm. Output includes sleep statistics and wake/sleep classification, which clinicians review to inform diagnostic and treatment decisions. Benefits include objective, ambulatory sleep monitoring.
Clinical Evidence
Clinical study of 99 subjects compared Zmachine algorithm output to consensus human scoring (Rechtschaffen & Kales rules). Results: 95.8% sensitivity (sleep detection), 91.6% specificity (wake detection). Overall Kappa agreement 0.8275, exceeding the 0.75 threshold for high association. Data confirms accurate wake/sleep detection from single-channel mastoid EEG.
Technological Characteristics
Single-channel EEG acquisition system; battery-operated; mastoid electrode placement. Proprietary adaptive algorithm using time and frequency domain features. Complies with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for adult patients to monitor wake and sleep states and as an adjunct to the diagnosis of sleep disorders. Intended for use by qualified healthcare practitioners in home or clinical environments.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{1}------------------------------------------------
K101830
r
t
# Zmachine® 510(k) Summary
| Submitter | Consolidated Research of Richmond, Inc. | |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Address | 26250 Euclid Avenue, Suite 024<br>Euclid, Ohio 44132 | |
| Contact Person | Richard F. Kaplan, Ph.D., President<br>Phone: (216) 289-2331<br>Fax: (216) 393-0079<br>E-mail: kaplan@cri-systems.com | |
| Date Prepared | March 28, 2011 | |
| Trade Name | Zmachine® | |
| Common Name | Sleep monitoring system | |
| Classification<br>Name | Electroencephalograph | |
| Classification | 21 CFR (882.1400) | |
| Product Codes | OLV and OMC | |
| 510(k) Number | K101830 | |
| Indications<br>for Use | The CRI Zmachine is a single-channel, EEG acquisition<br>designed for use in the home or clinical environment<br>intended to be used by qualified healthcare practitioners | |
ition and analysis system, ts. This device is ners to monitor the wake and sleep states of adult patients and as an adjunct to their diagnosis of sleep disorders.
MAR 3 1 20i1
{2}------------------------------------------------
#### Device Description
---
The CRI Zmachine is a battery-operated, single-channel, EEG acquisition and analysis system. The Zmachine system includes the Zmachine device, disposable EEG sensor cable, and a wall charger. The device operates on data from the differential-mastoid EEG channel to determine the wake and sleep states of the patient every 30 seconds.
#### Substantial Equivalence
The Zmachine is substantially equivalent to the Oxford Biosomnia device, manufactured by Oxford Biosignals Limited (K021485). The table below summarizes the technological characteristics of the Zmachine in comparison to the predicate device.
| Device | Zmachine | BioSomnia |
|----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | 21 CFR (882.1400) | 21 CFR (882.1400) |
| Product Code(s) | OLV and OMC | GWQ |
| Indications for Use | The CRI Zmachine is a single-<br>channel, EEG acquisition and<br>analysis system, designed for use in<br>the home or clinical environments.<br>This device is intended to be used<br>by qualified healthcare practitioners<br>to monitor the wake and sleep states<br>of adult patients and as an adjunct<br>to their diagnosis of sleep disorders. | The Oxford BioSignals BioSomnia<br>is a single channel ambulatory EEG<br>with a software package designed<br>for use as an adjunct for the<br>physician in their diagnosis of sleep<br>disorders and to quantify sleep and<br>wakefulness on a second by second<br>basis. The device is designed to be<br>used overnight in the patient's home<br>environment to assist the physician<br>in diagnosing sleep disorders. |
| Patient Population | Adults | Adults |
| Type of Device | EEG-based sleep monitor | EEG-based sleep monitor |
| No. of EEG Channels | 1 | 1 |
| Electrode Placement | Mastoid | Mastoid<br>Frontal<br>EOG Upper<br>Central |
| Analyzes EEG Data in<br>Real Time | Yes | Yes |
| Sleep Classification | Wake and Sleep | Wake and Sleep (light sleep, deep<br>sleep, and REM sleep) |
| EEG Analysis<br>Methodology | Proprietary adaptive algorithm<br>using time and frequency domain<br>features | Proprietary artificial neural<br>network-based algorithm |
| Can Provide<br>Information to Assist<br>the Physician in<br>Diagnosis of Sleep<br>Disorders? | Yes | Yes |
| Can Provide<br>Information to Assist<br>the Physician in the<br>Application of<br>Treatment? | Yes | Yes |
{3}------------------------------------------------
| Device | Zmachine | BioSomnia |
|--------------------------------------------------------------------------------------------|----------|-----------|
| Provides Information<br>to Assist Physician in<br>the Evaluation of<br>Treatment Efficacy? | Yes | Yes |
| Calculates Summary<br>Sleep Statistics? | Yes | Yes |
| Battery Powered? | Yes | Yes |
# Performance Testing - Nonclinical
The Zmachine system has been assessed against the following standards:
Electrical safety in accordance with IEC 60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety: Safety Requirements for Medical Electrical Systems;
Electromagnetic compatibility in accordance with IEC 60601-1-2, Medical Electrical Equipment -Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility --Requirements and Tests. (General).
In addition, software testing was performed in accordance with FDA guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices- Document issued on: May 11, 2005" regarding a moderate level of concern device:
## Performance Testing - Clinical
A ninety-nine subject (99) clinical study was conducted in which polysomnographic (PSG) data was acquired and analyzed by human scorers, and simultaneously data from the mastoid was acquired and processed by the Zmachine wake/sleep algorithm. Each PSG record was scored independently by at least two (2) certified polysomnographic technologists (3 records scored by 2 technologists, 16 records scored by 3 technologists, and 80 records scored by 4 technologists) using the Rechtschaffen & Kales (R&K, 1968) visual sleep scoring rules. The performance of the Zmachine was evaluated by comparing wake/sleep as determined by the Zmachine algorithm to the consensus of human scorers. The analysis demonstrates substantial agreement for wake/sleep determination, on an epoch-by-epoch basis, between the Zmachine algorithm and the consensus of human scorers. Of the 16,908 epochs indicated by the consensus of human scorers as wake, the Zmachine algorithm correctly identified 91.6% as wake (Zmachine specificity). Similarly, of the 65.873 epochs indicated by the consensus of human scorers as sleep, the Zmachine algorithm correctly identified 95.8% as sleep (Zmachine sensitivity). The overall Kappa agreement between the consensus of human scorers and the Zmachine algorithm is 0.8275, exceeding the 0.75 threshold for "high association". In conclusion, the Zmachine algorithm demonstrates accurate performance in detecting wake and sleep from a single channel (mastoid) of EEG data.
{4}------------------------------------------------
## Conclusion
,
」
S
This submission demonstrates that the Zmachine is substantially equivalent to the predicate device based on descriptive information, non-clinical and clinical performance testing.
.
.
:
.
.
:
:
:
.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three overlapping profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
Consolidated Research of Richmond, Inc. President % Richard Kaplan, Ph.D. 26250 Euclid Ave. Suite 024 Euclid, OH 44132
Re: K101830
MAR 3 1 2011
Trade/Device Name: ZMachine, Model DT-100 Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLV and OMC Dated: February 1, 2011 Received: February 4, 2011
Dear Dr. Kaplan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Richard Kaplan, Ph.D.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm117800 html fou the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Indications for Use Statement
510(k) Number: K101830
Device Name: Zmachine
Indications for Use: The CRI Zmachine is a single-channel, EEG acquisition and analysis system, designed for use in the home or clinical environments. This device is intended to be used by qualified healthcare practitioners to monitor the wake and sleep states of adult patients and as an adjunct to their diagnosis of sleep disorders.
Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
KRISTEN BOWSHER
(Division Sign-Off) (Division Sign-Orr)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
KIO| 830 510(k) Number,
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.