The devices are intended to acquire, store, and transfer biophysical signals to separate polysomnographic analysis software available with marketed systems for EEG and Sleep Studies. The devices are intended for use by physicians, technicians, or other medical professionals that are trained in EEG and/or PSG in any location within a medical facility, physician's office, laboratory, clinic or nursing home or outside of a medical facility under supervision of a medical professional. The device will be available on all patient populations, as determined by a trained professional. These devices do not provide alarms and they are not intended for use as an automated apnea monitor. The devices do not make any judgment of normality of the displayed signals.
Device Story
Portable holter/polysomnography recorder; acquires biophysical signals (EEG, flow, respiratory effort, body position, SpO2/HR) via external sensors; digitizes and stores data on Compact Flash memory; transmits data to PC via USB, Ethernet, or Bluetooth for analysis. Used in medical facilities, clinics, or home settings by trained professionals. Device provides no automated analysis, alarms, or diagnostic judgments; serves as a data acquisition/storage front-end for separate, cleared polysomnographic software. Benefits include portable, battery-powered signal recording in electromagnetically noisy environments.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including firmware verification, functional performance testing, and safety/EMC testing (IEC 60601-1/UL2601-1-1, IEC 60601-2-26, and IEC 60601-1-2).
Technological Characteristics
Portable recorder in plastic enclosure; 32 input channels (EEG, DC, flow, respiratory effort, body position, oximetry); powered by 2 AA batteries or external interface (USB/Ethernet); connectivity via USB, Ethernet, Bluetooth; Compact Flash storage; optically isolated communication lines; firmware-controlled; no patient-contacting parts (uses external sensors).
Indications for Use
Indicated for acquisition and storage of physiological signals for EEG and Sleep Studies in all patient populations as determined by a trained professional. Intended for use by physicians, technicians, or trained medical professionals. Not for use as an automated apnea monitor.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
K071782
Image /page/0/Picture/1 description: The image shows the word "micromed" in a stylized, three-dimensional font. The letters are textured with a speckled pattern, giving them a slightly rough appearance. The word is presented in a way that suggests depth and dimension, with a drop shadow effect that makes it appear to pop out from the background.
ELETTROMEDICALI
## 510(k) Summary
### 1.1.1 Submitter
Micromed S.p.A., via Giotto 2 , 31021 Mogliano Veneto (TV) Italy Tel. +39-041-5937000; fax +39-041-5937011; micromed@micromed-it.com
### 1.1.2 Official Contact Person
Monica Camillo
## 1.1.3 Summary preparation date
October 9th , 2007
### 1.1.4 Device identification
BRAIN SPY PLUS holter polygraph MORPHEUS polysomnograph EMBLA TITANIUM polysomnograph
- recommended classification regulation: 21. CFR 882.1400 .
- class II .
- panel: Neurology .
- product code: Oly .
#### Devices to which substantial equivalence is claimed: 1.1.5
- . K042150 Xltek Trex
- . K050425 Grass Telefactor Aura PSG
- K031202 Braebon MEDIPalm .
#### Device description 1.1.6
The device function is the acquisition of bloelectric signals, as is typical for EEG amplifiers and holter recorders. The device receives and amplifies patient input signals, converts them to digital format, packs them by a programmable logic and sends them to an internal microcontroller. The microcontroller stores them in a RAM buffer, from where they are written at defined time to the Compact Flash memory. The data can be viewed directly on the display or downloaded by using elther specific interface modules (BQUSB towards the USB port of the PC or BQNET towards an Ethernet link) or the built-in Bluetooth® link. Data can also be read directly from the Compact Flash card via a memory card reader. In PC a typical polysomnographic analysis software application manages the storage and the on-line or off-line display of the data. In the main version, the device is provided with 32 input channels:
- 24 channels for EEG (BW: 0.15-220 Hz) .
- DC channels (0-330 Hz) .
- 2 channels for the built-in flow sensors .
- 2 channels for external inductive belts (respiratory effort sensors) .
- 1 channel for a built-in body position sensor .
1 channel for digital data from the oximeter (SpO2, heart rate and plethysmographic . signal)
EPS*510k*V - page 3 of 6
Micromed S.p.A. Via Giotto 2, 31021 Mogliano Veneto, Treviso - Italy Tel: +39-041-593 7000 - Fax: +39-041-593 7011 Business Registry of Treviso Registration Number TV03906850262 VAT number IT03906850262
Image /page/0/Picture/30 description: The image contains three logos. The first logo is for IQNet, with the letters "IQ" stylized with a circle of stars. The second logo is a circular seal, also for IQNet, with the letters inside a circle. The third logo is for CSQ MED.
Image /page/0/Picture/31 description: The image shows the logo for ESQ. The logo is a black square with the letters "ESQ" in white. The letters are in a sans-serif font, and the "E" is slightly larger than the "S" and "Q". The logo is simple and modern.
QUALITY SYSTEM CERTIFIED - UNI EN ISO9001 - UNI CEI EN. ISO13485
JAN 1 1 2 3
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "micromed" in a stylized, three-dimensional font. The letters are blocky and have a layered effect, creating a sense of depth. The word is presented in a dark color, possibly black or a dark gray, against a lighter background. The overall design gives the word a bold and somewhat retro appearance.
ELETTROMEDICALI
The device is enclosed in a (2.5 -> 3.5) h x 7.5 x 11.4 cm plastic case and weighs 200g. It is powered by 2 AA 1,5V alkaline batteries, by the BQUSB interface (via the connected PC) or the BQNET module (via separate AC mains adapter). .
The device has a LCD allowing the menu to be display gives information on the recording mode, acquisition time, remaining memory, impedance measure and allows the rough display of the acquired traces both on-line and after the recording. The user interface includes 4 membrane keys enclosed in the label below the display.
The device has no patient contacting parts. Input signals are collected by commercially available accessories like EEG electrodes, oximeter sensors, cannulae and respiratory effort sensors (elastic band). The patient and operator can eventually touch the plastic external enclosure but the device is usually worn above the clothes in a synthetic fabric bag.
The manufacturer does not intend to limit the utility of this device by listing a finite set of commercially available sensors. The device is only a recorder that receives and records input form sensors and subsequently downloads those recordings to a PC resident polysomnological application software for clinical analysis: it is very versatile in the range of input sensors that can be used in various sleep study dinical settings. The sensors, kits and application software are not intended in this application for dearance.
For connection with the sensors and the PC, the devices are provided with:
- . touch proof connectors for patient electrodes (EEG channels)
- touch proof (BRAIN SPY PLUS) or 3-pin connector (MORPHEUS/TITANIUM) connectors for DC . channels
- . 1 screw-lock and 2 hose barb connectors for flow sensor cannulae (MORPHEUS/TTTANIUM version)
- . 2 connector for the resplratory effort elastic band (MORPHEUS/TTTANIUM version)
- 3-pin (MORPHEUS/TITANIUM) or 4-pin (BRAIN SPY PLUS) connectors for the oximeter .
- . 4 pin push-pull connector for the connection to the BQUSB or BQNET module.
The communication lines with the BQUSB or BQNET module are optically isolated.
In the PC data coming from either USB, Ethernet or Bluetooth® ports are then acquired by an acquisition software. The submitted device is supported by PC based application software such as Embla's Somnologica Studio (version 5.1), or Micromed's System Plus (version 1.03). The application software is not intended to be included in this application for clearance.
Battery power is provided to make the device safer, ergonomic and able to acquire good signals even in electromagnetically nolsy environment. Low power consumption allows 24 hours of consecutive recording with a pair of standard alkaline battery. Maximum power consumption is 80-100 mA, plus 30 mA in wireless transmission mode. In stand-by, the power consumption is less than 100µA.
When voltage and communication signals are present on the cable connection, the device is powered by the USB interface modules through a DC/DC converter.
The device can also be connected to the PC through a radio (Bluetooth®) link using the RADIOEEG. built-in module. The microcontroller communicates with the radio module through Its serial port and transmits the data to a Bluetooth® receiver on the PC. .
EPS*510k*V - page 4 of 6
Image /page/1/Picture/20 description: The image contains two logos. The first logo has the word "Net" in bold letters with a dotted circle to the left of the word. There is a horizontal line above and below the word. The second logo is a circular seal with text inside, but the text is too small to read.
Image /page/1/Picture/21 description: The image contains two logos. The first logo is a circle with the word "test" written in the center. The second logo is a square with the letters "CSQ" on top of the word "MED". The letters and words are all in black.
esq
LITY SYSTEM CERTIFIED - UNI EN ISO9001 - UNI CELEN, ISO13485
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "micromed" in a stylized, blocky font with a 3D effect. The word is in all lowercase letters and has a slightly tilted appearance. To the right of "micromed" is the word "ELETTROMEDICALI" in a simpler, sans-serif font. The overall image appears to be a logo or branding for a company or product.
The Compact Flash storage memory capacty determines the duration of the recording to the number of the channels used. The commercially available (1 GB) allow more than 24 hours recording of 32 channels at 128 Hz sampling rate, without data compression. If data compression algorithm Is used, recording time can be increased by 25-30%.
The device family is composed of two basic models all sharing the same case, internal hardware and firmware. The BRAIN SPY PLUS model is obtained by a MORPHEJS model without the internal sensors and with a different input panel (less and different connectors). The Embla TITANIUM model Is identical to MORPHEUS, differing only in labeling (distributor brand and device name).
#### 1.1.7 Devices intended use
The devices are intended to acquire, store, and transfer biophysical signals to separate polysomnographic analysis software available with marketed systems for EEG and Sleep Studies. .
The devices are intended for use by physidans, technicians, or other medical professionals that are trained in EEG and/or PSG in any location within a medical facility, physidan's office, laboratory, clinic or nursing home or outside of a medical facility under supervision of a medical professional. The device will be available on all patient populations, as determined by a trained professional.
These devices do not provide alarms and they are not intended for use as an automated apnea monitor.
The devices do not make any judgment of normality of the displayed signals.
#### Technological characteristics and product performance 1.1.8
The device shares with the predicates many technological characteristics.
The EEG and polysomnography devices on the market have different number of channels and integrated sensors, depending on the Intended use (mainly PSG, mainly EEG or both uses).
When the body position, respiratory effort and flow measurement sensors are not integrated in the devices, external devices with the same functions are used In connection with the recorders. Brain spy Plus, like some of the predicate, can achieve equivalent performance with respect to the major models by acquiring signals from the external sensors.
Also oximeter circuit can be internal or external, but they are based on two major manufacturers technology (Nonin Inc. and Masimo): oximeters based on technology provided from Nonin are used in Braebon and Grass Telefactor products.
The devices are hence comparable for channel numbers, characteristics and features availability (integrated sensors and/or dedicated channels).
All the devices have small size, limited weight and a plastic endosure. Some of these are provided with a fixed base station, other includes all the functions in the portable part.
Power supply is provided by Internal batterles. Some devices use standard disposable batteries which do not require the use of a battery charger, others use rechargeable batteries.
Expected battery life is comparable or greater than those declared for the other devices.
Some of the predicates (e.g. Braebon MEDIPalm and Grass Telefactor Aura-PSG devices) can be used In tethered (on-Ilne) mode, connected to the PC or/and a base station. When used in this mode the devices can be powered directly by the remote power source (USB cable and proper issolation or medical grade power supply). An equivalent degree of safety Is achieved in the device through compliance with IEC 60601-1 standard.
FPS 510k V - nage 5 of 6
Image /page/2/Picture/20 description: The image contains three logos. The first logo is for "IONet" with the letters in bold and a dotted circle above the "IO". The second logo is a circular seal with text inside. The third logo is a black rectangle.
Image /page/2/Picture/21 description: The image contains two logos. The first logo has the letters "CSQ" stacked on top of "MED", with a line underneath "MED". The second logo is partially visible and appears to have the letters "CSQ" in a similar style to the first logo.
QUALITY SYSTEM CERTIFIED - UNI EN ISO6001 - UNI CEI EN. ISO 13485
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a stylized logo of the word "digimod". The letters are in a blocky, three-dimensional font with a black outline and a textured fill. The logo has a shadow effect, making it appear to pop out from the background. The letters are all connected, and the overall design is eye-catching and modern.
ELETTROMEDICALI
All the devices are based on the same working principle: biopotentials and data from patientconnected sensors are acquired, digitized, and stored in the internal compact flash memory.
For all the devices data stored in the flash can be transmitted (through cable and in some devices via a radio link using the ISM 2.4 GHz band and low RF output power) to a remote PC for signal check contemporarily to the acquisition. Data can also be downloaded at the end of the exam through the interface connection to the PC or removing the card.
Card capacity is always sufficient to store whole-night examination data.
All the devices can be remotely controlled by the separate polysomnographic analysis software installed on the PC, which sold separately and not Included In this application for clearance, which also performs the signal analysis. None of the devices directly analyzes the collected data. Verification and validation test on the firmware shows that the device Is correctly and safely controlled through the remote commands.
Integral kevs and display to show menu and check the signals directly from the device are also available in one of the predicate devices. Verification and validation test on the firmware shows that the device is correctly and safely controlled through the local commands
All the devices contact the patient only through the accessories connected to the inputs (electrodes, oximeter sensor, cannulae, and elastic bells for respiratory efforts) which are separate devices. Hence, no questions on patient-contacting material arise for these device can be carrled in a bag intended to be worn over a garment.
#### Performance specifications, including any testing. 1.1.9
Safety Tests have been performed to verify compliance with IEC 60601-1/UL2601-1-1 and IEC 60601-2-26 to ensure that there are no potential hazards on patients, other persons, or the surroundings.
Electromagnetic Compatibility tests according to IEC 60601-1-2 have been performed to ensure no intolerable magnetic disturbances are introduced into its electromagnetic environment.
Immunity tests to IEC 60501-1-2 have been performed to ensure that the EEG equipment has the ability to operate satisfactorily in its electromagnetic environment
No specific guldance document on performance is required for EEG/PSG devices.
The device capability is equivalent to the features spedfied for predicate devices.
Verification of firmware ensures that the system conforms to all the system design requirements.
Functional testing was performed to confirm that the devices are capable of meeting their stated performance specifications.
### 1.1.10 Conclusions
Based on the above discussion, the risk analysis and the table of comparison between the devices, Micromed Morpheus, Micromed Brain Spy Plus and Embla Tltanium devices are substantially equivalent to the predicate devices as identification of Predicate Devices section.
EPS 510k V-page 6 of 6
Image /page/3/Picture/20 description: The image shows the logo for IONet. The logo consists of the word "IONet" in a bold, sans-serif font. To the left of the word is a circle made up of smaller circles. There are horizontal lines above and below the word.
Image /page/3/Picture/21 description: The image contains two distinct logos. On the left, there is a circular logo with a solid black fill. To the right, there is a logo with the letters "CSQ" stacked on top of the letters "MED", with a horizontal line separating the two sets of letters. The letters are in a bold, sans-serif font.
CSQ
QUALITY SYSTEM CERTIFIED - UNI EN ISOB001 - UNI CEI EN. ISO13485
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of stylized lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Micromed S.P.A. c/o Mr. Jan Van Lochem Kema Quality B.V. 4377 County Line Road Chalfont, PA 18914
API - 9 2012
Re: K071782
Trade/Device Name: Micromed Brain Spy Plus Embia Titanium Devices Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: OLV Dated (Date on orig SE ltr): December 21, 2007 Received (Date on orig SE ltr): December 26, 2007
Dear Mr. Van Lochem:
This letter corrects our substantially equivalent letter of January 10, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Jan Van Lochem
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kesia Alexander
for
Malvina B. Eydelman, M.D. Director Division of Ophthalmic. Neurological. and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K071782
Device Name: EMBLA TITANIUM; MICROMED MORPHEUS; MICROMED BRAIN SPY PLUS
Indications for Use:
The devices are intended to acquire and store physiological signals for EEG and Sleep Studies, and to transfer the data to separate polysomnographic analysis software.
The devices are intended for use by physicians, technicians, or other medical professionals that are trained in EEG and/or PSG .
Over-The-Counter Use Prescription Use . YES AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rph
(Division Sign-Off Division of General, Restorative, and Neurological Devices
**Page 1 of 1**
**10(k) Number** k071782
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.