K191322 · GE Healthcare Finland Oy · OLT · Jan 22, 2020 · Neurology
Device Facts
Record ID
K191322
Device Name
E-EEGX, N-EEGX
Applicant
GE Healthcare Finland Oy
Product Code
OLT · Neurology
Decision Date
Jan 22, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Indications for Use
The GE EEG module, E-EEGX, the GE EEG headbox, N-EEGX, and accessories are intended to be used with the compatible CARESCAPE monitors for the monitoring of electroencephalograph (EEG), frontal electromyograph (FEMG), and auditory evoked potentials (AEP) of all hospital patients. The device is intended for use by qualified medical personnel only.
Device Story
E-EEGX module and N-EEGX headbox are plug-in components for CARESCAPE Bx50 V3 patient monitors. System measures raw EEG (4 channels), FEMG (1 channel), and AEP (2 channels) via patient-connected accessories. Module digitizes EEG data for host monitor, generates auditory stimuli for AEP, and performs partial AEP data processing. Used in hospital settings by qualified medical personnel. Output displayed on host monitor for clinical assessment of neurophysiological status. Benefits include real-time monitoring of brain activity and evoked potentials to support clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including risk analysis, design reviews, and verification/validation testing against IEC 60601-1, 60601-1-2, 60601-2-26, 60601-2-40, and 60601-2-49 standards.
Technological Characteristics
Plug-in interface module and headbox. Measures EEG, FEMG, AEP. Spectral analysis includes SEF, median frequency, relative power (Delta, Theta, Alpha, Beta), and burst suppression. Connectivity via proprietary interface to CARESCAPE Bx50 V3 monitors. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-26, IEC 60601-2-40, IEC 60601-2-49.
Indications for Use
Indicated for monitoring EEG, FEMG, and AEP in all hospital patients. Intended for use by qualified medical personnel.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
Predicate Devices
Datex-Ohmeda S/5 EEG Module, E-EEG and Datex-Ohmeda S/5 EEG Headbox, N-EEG and Accessories (K051883)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
Joel Kent Senior Regulatory Affairs Manager GE Healthcare Finland Oy Kuortaneenkatu 2 00510 Helsinki, Finland
Re: K191322
Trade/Device Name: E-EEGX module, N-EEGX headbox and accessories Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLT, GWJ, MHX, MLD, OMC, ORT Dated: December 19, 2019 Received: December 23, 2019
Dear Joel Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and probibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191322
Device Name
E-EEGX module, N-EEGX headbox and accessories
Indications for Use (Describe)
The GE EEG module, E-EEGX, the GE EEG headbox, N-EEGX, and accessories are intended to be used with the compatible CARESCAPE monitors for the monitoring of electroencephalograph (EEG), frontal electromyograph (FEMG), and auditory evoked potentials (AEP) of all hospital patients. The device is intended for use by qualified medical personnel only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="margin-right:5px;">☑</span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><span style="margin-right:5px;">☐</span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo, which consists of the letters 'GE' intertwined in a stylized monogram. The monogram is enclosed within a circle, and there are decorative flourishes or 'teardrops' around the circle's perimeter. The logo is presented in black and white, with the monogram and circle being black against a white background.
GE Healthcare Finland Oy Kuortaneenkatu 2, P.O. Box 900 00031 GE
Finland
T: +358 10 39411
F: +358 9 1463310
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Owner/Contact/Date<br>(807.92(a)(1): | | |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Date: | January 16, 2020 | |
| Owner/Submitter: | GE Healthcare Finland Oy.<br>Kuortaneenkatu 2<br>00510 Helsinki<br>FINLAND<br>Phone: +358 10 39411 | |
| Primary Contact Person: | Joel Kent<br>Senior Regulatory Affairs Manager<br>GE Healthcare<br>Phone: 617-851-0943<br>E-mail: joel.kent@ge.com | |
| Secondary Contact Person: | Anna Pehrsson<br>Regulatory Affairs Leader<br>GE Healthcare Finland Oy<br>Kuortaneenkatu 2<br>00510 Helsinki<br>Finland<br>Phone: + 358 40 569 5108<br>E-mail: anna.pehrsson@ge.com | |
| Device names (807.92(a)(2)): | | |
| Trade Name: | E-EEGX module, N-EEGX headbox and accessories | |
| Common/Usual Name: | EEG measurement module, headbox and accessories | |
| Classification Name: | 21 CFR 882.1400 electroencephalograph | |
{5}------------------------------------------------
#### Product Code OLT
Subsequent Product Codes: GWJ, MHX, MLD, OMC, ORT
- Predicate Device(s) K051883 Datex-Ohmeda S/5 EEG Module, E-EEG and Datex-(807.92(a)(3): Ohmeda S/5 EEG Headbox, N-EEG and Accessories
The E-EEGX module is a single-width plug-in interface module Device Description (807.92(a)(4)): to be used with CARESCAPE Bx50 V3 patient monitors. It is used with N-EEGX headbox and accessories for monitoring neurophysiological status of all hospital patients by measuring the electroencephalogram (EEG), frontal electromyogram (FEMG) and auditory evoked potentials (AEP).
Intended Use: (807.92(a)(5): The GE EEG module, E-EEGX, the GE EEG headbox, N-EEGX, and accessories are intended to be used with the compatible CARESCAPE monitors for the monitoring of electroencephalograph (EEG), frontal electromyograph (FEMG), and auditory evoked potentials (AEP) of all hospital patients. The device is intended for use by qualified medical personnel only.
The E-EEGX module is used with the N-EEGX headbox for Technology (807.92(a)(6)): monitoring of EEG. FEMG. to stimulate the brain with auditory stimuli, and to measure AEP. The E-EEGX module connects to a N-EEGX headbox which further connects to accessories that connect to the patient.
> The EEG, FEMG and AEP measurements are performed by the N-EEGX headbox. The N-EEGX headbox measures the raw EEG waveform data from four real-time EEG waveform channels, FEMG from one channel and AEP from two channels. The N-EEGX headbox is connected to the patient with EEG accessories.
The E-EEGX module transfers the digitized EEG data received from the N-EEGX headbox to the host monitor. The module also generates the stimuli used in the AEP measurement and performs part of the AEP measurement data processing.
The fundamental function and operation of the device remains unchanged compared to the predicate. A summary of the main changes compared to the predicate is listed below.
{6}------------------------------------------------
| Specification | Datex-Ohmeda S/5 EEG Module, E-EEG and<br>Datex-Ohmeda S/5 EEG Headbox, N-EEG and<br>Accessories (K051883) | E-EEG X module, N-EEGX headbox and EEG<br>accessories | Discussion of differences |
|-------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The Datex-Ohmeda EEG module, E-EEG and the<br>Datex-Ohmeda EEG headbox, N-EEG and<br>accessories are indicated for monitoring of<br>electroencephalograph (EEG), frontal<br>electromyograph (FEMG) and auditory evoked<br>potentials (AEP) of all hospital patients.<br>The device is indicated for use by qualified<br>medical personnel only. | The GE EEG module, E-EEGX, the GE EEG<br>headbox, N-EEGX, and accessories are intended<br>to be used with the compatible CARESCAPE<br>monitors for the monitoring of<br>electroencephalograph (EEG), frontal<br>electromyograph (FEMG), and auditory evoked<br>potentials (AEP) of all hospital patients. The<br>device is intended for use by qualified medical<br>personnel only. | Equivalent<br>The changes from previously cleared<br>indications are:<br>Company name "Datex" has been<br>replaced by "GE".<br>The indications refer to compatible<br>monitors.<br>Some minor text edits and changes<br>for added clarity.<br>The changes have been made to<br>improve clarity of labeling.<br>Therefore, they do not significantly<br>affect safety and/or effectiveness. |
| Physical Properties<br>Module Size(H x W x<br>D)<br>Module Weight | E-EEG module<br>112 x 37 x 186 mm<br>(4.4 x 1.5 x 7.3 in)<br>0.3 kg (0.7 lb)<br>N-EEG headbox<br>34 x 97 x 174 mm<br>(1.3 x 3.8 x 6.8 in)<br>0.44 kg (1.1 lb) | E-EEGX module<br>112 x 37 x 187 mm<br>(4.4 x 1.5 x 7.3 in)<br>0.3 kg (0.7 lb)<br>N-EEGX headbox<br>34 x 97 x 174 mm<br>(1.3 x 3.8 x 6.8 in)<br>0.5 kg (1.1 lb) | Equivalent<br>The N-EEGX headbox weight has<br>slightly changed due to a new<br>hardware design.<br>This change does not significantly<br>affect safety and/or effectiveness. |
| Host device<br>compatibility | E-EEG module is compatible with S/5 AM,<br>CAM, S/5 CCM, CARESCAPE B450 Monitor<br>with software ESP V1 or V2, CARESCAPE<br>Monitor B650 with software ESP V1 or V2 and<br>CARESCAPE Monitor B850 with software ESP<br>V1 or V2 | E-EEGX module and N-EEGX headbox are<br>compatible with the CARESCAPE Bx50 V3<br>patient monitors | Equivalent<br>The E-EEGX module and N-EEGX<br>headbox are only compatible with<br>CARESCAPE Bx50 V3 patient<br>monitors.<br>This change has been verified and the<br>overall functionality remains<br>equivalent.<br>This change does not significantly<br>affect safety and/or effectiveness. |
| Parameter Specifications | | | |
| Measured parameters | EEG, Auditory Evoked Potentials, EMG | EEG, Auditory Evoked Potentials, EMG | Identical |
| Mode | Referential or Bipolar | Referential or Bipolar | Identical |
| Processing of the EEG | Spectral analysis: Spectral Edge Frequency<br>(SEF), Median frequency, Relative power in<br>Delta, Theta, Alpha and Beta bands<br>Burst suppression detection<br>Total power | Spectral analysis: Spectral Edge Frequency<br>(SEF), Median frequency, Relative power in<br>Delta, Theta, Alpha and Beta bands<br>Burst suppression detection<br>Total power | Identical |
| EEG parameter specifications | | | |
| EEG Measurement<br>method | 1, 2, 3 or 4 channels of EEG | 1, 2, 3 or 4 channels of EEG | Identical |
| Range | $\pm$ 400 μV | $\pm$ 500 μV | Equivalent<br>Range has been expanded to meet the<br>requirements of IEC 60601-2-<br>26:2012.<br>This change has been verified and the<br>overall functionality remains<br>equivalent.<br>This change does not significantly<br>affect safety and/or effectiveness. |
| Measurement range<br>frequency | 0.5 ... 30 Hz | 0.5 ... 50 Hz | Equivalent<br>Range has been expanded to meet the<br>requirements of IEC 60601-2-<br>26:2012.<br>This change has been verified and the<br>overall functionality remains<br>equivalent.<br>This change does not significantly<br>affect safety and/or effectiveness. |
| AEP parameter specifications | | | |
| Evoked potentials | Auditory evoked potentials: brain stem and mid-latency | Auditory evoked potentials: brain stem and mid-latency | Identical |
| Measurement Sampling<br>frequency | Brainstem Auditory Evoked Potentials (BAEP):<br>4800 Hz<br>Middle Latency Auditory Evoked Potential<br>(MLAEP): 2400 Hz | Brainstem Auditory Evoked Potentials (BAEP):<br>4800 Hz<br>Middle Latency Auditory Evoked Potential<br>(MLAEP): 2400 Hz | Identical |
| Frequency range | 0.5 Hz - 1000 hz | 0.5 Hz - 1000 Hz | Identical |
| Stimulation type | Condensating click | Condensating click | Identical |
| Stimulation frequency | 1.1 to 9.1 Hz(1 Hz steps) @ 10 ms sweep<br>1.1 to 8.1 Hz(1 Hz steps) @ 100 ms sweep | 1.1 to 9.1 Hz(1 Hz steps) @ 10 ms sweep<br>1.1 to 8.1 Hz(1 Hz steps) @ 100 ms sweep | Identical |
| EMG parameter specifications | | | |
| EMG measurement<br>frequency range | 60 to 300 Hz | 60 to 300 Hz | Identical |
{7}------------------------------------------------
Determination of Substantial Equivalence (807.92(b)(1))
### Summary of Non-Clinical Tests:
For the E-EEGX module, N-EEGX headbox and accessories the following quality assurance measures were applied to the development of the system.
- 1. Risk Analysis
- 2. Requirements Reviews
- 3. Design Reviews
- 4. Testing on unit level (Module verification)
- 5. Integration testing (System verification)
- 6. Final acceptance testing (Validation)
- 7. Performance testing (Verification)
- Safety testing (Verification) 8.
- 9. The following list contains applicable standards regarding performance testing related to E-EEGX module, N-EEGX headbox and accessories:
- ANSI/AAMI ES60601-1:2005/(R)2012 and ● A1:2012: Medical electrical equipment Part 1: General requirements for basic safety and essential performance (IEC 60601-1: 2012)
- . IEC 60601-1-2: 2014: Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
- IEC 60601-2-26:2012: Medical electrical equipment Part 2-26: Particular requirements for
{8}------------------------------------------------
the safety and essential performance of electroencephalographs
- IEC 60601-2-40: 2016: Medical Electrical ● Equipment Part 2-40: Particular requirements for the basic safety and essential performance of electromyographs and evoked response equipment
- IEC 60601-2-49: 2011: Medical electrical ● equipment Part 2-49: Particular requirements for the safety essential performance of multifunction patient monitoring equipment
## Clinical (807.92(b)(2)): Summary of Clinical Tests:
The subject of this premarket submission, the E-EEGX module, N-EEGX headbox and accessories did not require clinical studies to support substantial equivalence.
- Conclusion (807.92(b)(3)): GE Healthcare considers the E-EEGX module, N-EEGX headbox and accessories to be as safe, as effective, and performance is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.