← Product Code [OLT](/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/OLT) · K142834

# DiscoverEEG System, Model DE-401 (K142834)

_Neurowave Systems, Inc. · OLT · Jun 23, 2015 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/OLT/K142834

## Device Facts

- **Applicant:** Neurowave Systems, Inc.
- **Product Code:** [OLT](/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/OLT.md)
- **Decision Date:** Jun 23, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1400
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The DiscoverEEG System, Model DE-401 is intended to be used for measuring and recording the electrical activity of a subject's brain, obtained by placing non-invasive electrodes on the head. The DiscoverEEG, DE-401 is indicated for use in acquiring electroencephalographic (EEG) signals in the OR, ICU, ER, clinical settings and at home and for clinical research. The medical use of data acquired by the DiscoverEEG is to be performed under the direction and interpretation of a licensed medical professional. The DiscoverEEG, Model DE-401 does not provide any diagnostic conclusion about the subject's condition.

## Device Story

Wearable, medical-grade EEG monitor; acquires/stores up to 4 channels of EEG signals via non-invasive disposable electrodes. Components: Acquisition Module (signal digitization), Memory Module (processing, storage, power), Disposable Electrode Array, and PC-based Data Viewer Software. Operates by digitizing analog EEG signals; performs impedance checks, lead-off detection, and artifact removal. Produces raw EEG waveforms and processed spectral variables (power spectrum, MEF, SEF, frequency bands). Used in OR, ICU, ER, clinical, and home settings by clinicians or researchers. Data transferred to PC for review; supports conversion to standard formats (EDF+, tab-delimited). Provides signal quality metrics (impedance, artifact flags) to assist clinicians in data interpretation. Benefits include portable, continuous EEG monitoring for clinical research and patient assessment without providing automated diagnostic conclusions.

## Clinical Evidence

No clinical data. Bench testing only; device performance evaluated against design specifications and safety standards (IEC 60601-1, 60601-1-11, 60601-2-26, 60601-1-2).

## Technological Characteristics

Wearable EEG monitor; 4-channel acquisition; silver-silver chloride disposable electrodes. Battery-powered; standalone or wireless data transfer. PC-based software for data retrieval/review. Complies with IEC 60601-1, 60601-1-11, 60601-2-26, and 60601-1-2. Amplifier: ≥110 dB CMRR, ≥100 Meg Ohm input impedance.

## Regulatory Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

## Predicate Devices

- NeuroFAST Monitoring System, Model NF-201 ([K092477](/device/K092477.md))
- Zmachine ([K101830](/device/K101830.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized caduceus-like symbol with three human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 23, 2015

NeuroWave Systems Inc. Tatjana Zikov President 2490 Lee Blvd., Suite 300 Cleveland Heights, OH 44118

Re: K142834

Trade/Device Name: DiscoverEEG System, Model DE-401 Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLT, OMC Dated: May 10, 2015 Received: May 11, 2015

Dear Tatjana Zikov:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Felipe Aguel -S

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K142834

Device Name DiscoverEEG System, DE-401

### Indications for Use (Describe)

The DiscoverEEG System, Model DE-401 is intended to be used for measuring and recording the electrical activity of a subject's brain, obtained by placing non-invasive electrodes on the head. The DiscoverEEG, DE-401 is indicated for use in acquiring electroencephalographic (EEG) signals in the OR, ICU, ER, clinical settings and at home and for clinical research. The medical use of data acquired by the DiscoverEEG is to be performed under the direction and interpretation of a licensed medical professional. The DiscoverEEG, Model DE-401 does not provide any diagnostic conclusion about the subject's condition.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| 区 Prescription Use (Part 21 CFR 801 Subpart D)  | Over-The-Counter Use (21 CFR 801 Subpart C) |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### Contact Details 1.

| Applicant Name: | NeuroWave Systems Inc.<br>2490 Lee Blvd., Suite 300<br>Cleveland Heights, OH 44118<br>(216) 361-1591 (telephone)<br>(216) 361-1554 (fax) |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:        | Tatjana Zikov,<br>President<br>tzikov@neurowavesystems.com                                                                               |
| Date Prepared:  | September 29, 2014                                                                                                                       |

#### 2. Device Name

| Trade Name:           | DiscoverEEG System, Model DE-401 |
|-----------------------|----------------------------------|
| Common/Usual Name:    | EEG Monitor                      |
| Classification Name:  | Electroencephalograph            |
| Product Code:         | OMC                              |
| Regulation Number:    | 21 CFR 882.1400                  |
| Device Class:         | Class II                         |
| Classification Panel: | Neurology                        |

#### Legally Marketed Predicate Device 3.

| 510(k)<br>Number | Product<br>Code | Trade Name                                   | Manufacturer                               |
|------------------|-----------------|----------------------------------------------|--------------------------------------------|
| K092477          | OMC             | NeuroFAST Monitoring<br>System, Model NF-201 | NeuroWave Systems, Inc.                    |
| K101830          | OLV,<br>OMC     | Zmachine                                     | Consolidated Research of<br>Richmond, Inc. |

#### 4. Description of Device

The DiscoverEEG System, Model DE-401, is a wearable, medical-grade EEG device that acquires and stores up to four electroencephalograms (EEGs) obtained from noninvasive electrodes placed on a subject's head. The acquired EEG waveforms, as well as, processed EEG spectral variables are continuously stored by the system for later retrieval. The data can be transferred from the DiscoverEEG hardware to a computer for review.

The DiscoverEEG System, Model DE-401 has four main components:

{4}------------------------------------------------

- . Acquisition Module: contains the circuitry for acquisition of the EEG signals through connections to the electrodes placed on the subject's head. The acquired signals are digitized and then passed to the Memory Module, for signal processing and storage.
- . Memory Module: provides power to the system and contains memory and EEG processing circuitry (impedance check, lead off check, pre-processing incl. artifact removal, processed spectral variables). Includes LED indicators for assurance of proper operation and other user interface elements. May contain a wireless option for data transfer to Data Viewer Software in realtime.
- . Disposable Electrode Array: contains noninvasive, disposable, single patient electrodes (5 electrodes, 4 EEG channels) and skin preparation materials. The packaging will guide the user on how to prepare the subject's skin and how to place the electrodes.
- . Data Viewer Software: PC based software used for retrieving and reviewing acquired raw EEG waveforms (as well as processed data) from the DiscoverEEG hardware. This software will also be capable of converting the retrieved EEG data files to standard formats (i.e. EDF+, tab-delimited). Such converted data files can then be reviewed using off-the-shelf bio-signal review software that accepts the converted format.

#### 5. Intended Use

The intended use of DiscoverEEG System, Model DE-401 is consistent with the classification 21 CFR 882.1400, Electroencephalograph:

"An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head."

The DiscoverEEG System, Model DE-401 is intended to be used for measuring and recording the electrical activity of a subject's brain, obtained by placing non-invasive electrodes on the head. The DiscoverEEG, DE-401 is indicated for use in acquiring electroencephalographic (EEG) signals in the OR, ICU, ER, clinical settings and at home and for clinical research. The medical use of data acquired by the DiscoverEEG is to be performed under the direction and interpretation of a licensed medical professional. The DiscoverEEG, Model DE-401 does not provide any diagnostic conclusion about the subject's condition.

#### 6. Substantial Equivalence Comparison

The technology of the DiscoverEEG DE-401 device is equivalent to other EEG monitor devices. The EEG signal is acquired in analog format, digitized, processed and saved for later user interpretation. An Electrode (Sensor) Array is provided with the system for the convenience of the user in applying the acquisition electrodes. The Electrode Array is made of the materials that have been shown to pass cytotoxicity, primary skin irritation and sensitization as required (refer to section 15 for addition details).

{5}------------------------------------------------

| Feature                                                      | DiscoverEEG System, Model<br>DE-401                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | NeuroFAST Monitoring<br>System,<br>Model NF-201<br>(092477)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Zmachine<br>(101830)                                                                                                                                                                                                                                                                                                                                           |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use                                       | The DiscoverEEG System,<br>Model DE-401 is intended to<br>be used for measuring and<br>recording the electrical<br>activity of a subject's brain,<br>obtained by placing non-<br>invasive electrodes on the<br>head. The DiscoverEEG, DE-<br>401 is indicated for use in<br>acquiring<br>electroencephalographic<br>(EEG) signals in the OR, ICU,<br>ER, clinical settings and at<br>home and for clinical research.<br>The medical use of data<br>acquired by the DiscoverEEG<br>is to be performed under the<br>direction and interpretation of<br>a licensed medical<br>professional. The<br>DiscoverEEG, Model DE-401<br>does not provide any<br>diagnostic conclusion about<br>the subject's condition. | The NeuroFAST Monitor NF-<br>201 is intended to be used for<br>measuring and recording the<br>electrical activity of a<br>patient's brain, obtained by<br>placing electrodes on the<br>head. The NeuroFAST NF-<br>201 is indicated for use in<br>acquiring<br>electroencephalographic<br>(EEG) signals in the OR, ICU,<br>ER, clinical settings and for<br>clinical research. The<br>NeuroFAST Monitor is to be<br>used under the direction and<br>interpretation of a licensed<br>medical professional. The<br>NeuroFAST Monitor NF-201<br>does not provide any<br>diagnostic conclusion about<br>the patient's condition. | The CR1 Zrnachine is a<br>single-channel EEG<br>acquisition and analysis<br>system, designed for use in<br>the home or clinical<br>environments. This device<br>is intended to be used by<br>qualified healthcare<br>practitioners to monitor the<br>wake and sleep states of<br>adult patients and as an<br>adjunct to their diagnosis<br>of sleep disorders. |
| Modalities                                                   | EEG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | EEG                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | EEG                                                                                                                                                                                                                                                                                                                                                            |
| Environment<br>of Use                                        | Operating room, intensive care<br>unit, emergency room, clinical<br>settings and At Home where<br>EEG monitoring is used.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Operating room, intensive<br>care unit, emergency room,<br>and clinical settings where<br>EEG monitoring is used.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | In the home or clinical<br>environment                                                                                                                                                                                                                                                                                                                         |
| Power Source                                                 | Battery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 120 Volt 60Hz AC power                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Li Ion Battery                                                                                                                                                                                                                                                                                                                                                 |
| System<br>Components                                         | Electrode (Sensor) Array,<br>Acquisition and Memory<br>Modules, Data Viewer<br>Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Patient Module, Patient Cable,<br>Display Module, Data Cable<br>and EasyPrep Electrode Kit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Zmachine device, EEG<br>sensors, Sensor cable, wall<br>charger                                                                                                                                                                                                                                                                                                 |
| Sensing<br>Electrodes                                        | Silver-silver chloride<br>disposable EEG electrodes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Silver-silver chloride<br>disposable EEG electrodes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Silver-silver chloride<br>disposable electrodes                                                                                                                                                                                                                                                                                                                |
| Feature                                                      | DiscoverEEG System, Model<br>DE-401                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | NeuroFAST Monitoring<br>System,<br>Model NF-201<br>(092477)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Zmachine<br>(101830)                                                                                                                                                                                                                                                                                                                                           |
| Screen<br>Display<br>Details                                 | Data Viewer Software:<br>1) Raw EEG Waveforms<br>2) Signal Quality<br>information: Channel<br>Connection flags,<br>Impedance values,<br>Artifact Detection flags,<br>Noise Level<br>3) Spectral Parameters: EEG<br>power spectrum and<br>frequency bands, 50%<br>50% Median Edge<br>Frequency (MEF), and<br>95% Spectral Edge<br>Frequency (SEF)                                                                                                                                                                                                                                                                                                                                                             | Displays:<br>1) Raw EEG Waveforms<br>2) Signal Quality<br>information: Channel<br>Connection flags,<br>Impedance values,<br>Artifact Detection flags,<br>Noise Level<br>3) Spectral Parameters: EEG<br>power spectrum and<br>frequency bands, 50%<br>50% Median Edge<br>Frequency (MEF), and<br>95% Spectral Edge<br>Frequency (SEF)                                                                                                                                                                                                                                                                                        | Displays:<br>Sleep/Wake Statistics                                                                                                                                                                                                                                                                                                                             |
| Stored EEG<br>data available                                 | Yes - SD Card                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes – through removable<br>storage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Yes - microSD                                                                                                                                                                                                                                                                                                                                                  |
| EEG<br>Channels/<br>Montage                                  | Up to 4 channels (2 fronto-<br>temporal channels, and 2<br>temporal channels)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 2 bilateral fronto-temporal<br>channels viewed concurrently                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 1 channel, differential<br>mastoid                                                                                                                                                                                                                                                                                                                             |
| Real Time<br>EEG Display                                     | Yes, if wireless option is<br>present                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | No                                                                                                                                                                                                                                                                                                                                                             |
| Processed<br>EEG<br>Bandwidth                                | User Selectable in Data<br>Viewer software<br>Low Filter: 0.125 or 0.5 Hz<br>High Filter: 30 or 70 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | User Selectable<br>Low Filter: 0.125 or 0.5 Hz<br>High Filter: 30 or 70 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 0.5-380 Hz                                                                                                                                                                                                                                                                                                                                                     |
| Automatic<br>Artifacting                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Unknown                                                                                                                                                                                                                                                                                                                                                        |
| Amplifier<br>Common<br>Mode<br>Rejection<br>Ration<br>(CMRR) | ≥110 dB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | ≥100 dB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Unknown                                                                                                                                                                                                                                                                                                                                                        |
| Amplifier<br>Input<br>Impedance                              | ≥100 Meg Ohm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | ≥50 Meg Ohm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 10 Gig Ohms                                                                                                                                                                                                                                                                                                                                                    |
| Electrode<br>Impedance<br>Test                               | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Yes<br>Continuous and on demand by<br>the user                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                            |
| Feature                                                      | DiscoverEEG System, Model<br>DE-401                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | NeuroFAST Monitoring<br>System,<br>Model NF-201<br>(092477)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Zmachine<br>(101830)                                                                                                                                                                                                                                                                                                                                           |
| Derived /<br>Processed<br>EEG Measures                       | Power spectrum parameters<br>derived from FFT:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Power spectrum parameters<br>derived from FFT:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Wake/Sleep Determination                                                                                                                                                                                                                                                                                                                                       |
|                                                              | 1) Power spectrum displayed<br>as Density Spectral Array<br>(DSA)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 1) Power spectrum<br>displayed as Density<br>Spectral Array (DSA)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                |
|                                                              | 2) 95% Spectral Edge<br>Frequency (SEF)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 2) 95% Spectral Edge<br>Frequency (SEF)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                |
|                                                              | 3) Median Edge Frequency<br>(MEF)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 3) Median Edge Frequency<br>(MEF)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                |
|                                                              | 4) Spectral Powers in EEG<br>frequency bands<br>traditionally used to<br>quantify EEG signals (α,<br>β1, β2, δ, θ and γ)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 4) Spectral Powers in EEG<br>frequency bands<br>traditionally used to<br>quantify EEG signals (α,<br>β1, β2, δ, θ and γ)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                |
|                                                              | Signal Quality information<br>(Channel connection flags,<br>Impedance values, Artifact<br>Detection flags, Noise Level)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Signal Quality information<br>(Channel connection flags,<br>Impedance values, Artifact<br>Detection flags, Noise Level)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                |
| Contains<br>patient<br>isolation                             | Battery powered, thus no<br>connection between patient<br>and mains                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Yes - Patient module<br>(analog-to digital converter)<br>provides 6kV electrical<br>isolation of the patient from<br>the monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Battery powered, thus no<br>connection between patient<br>and mains                                                                                                                                                                                                                                                                                            |
| Display<br>Interface                                         | Yes, LEDs to ensure proper<br>system connection and<br>operation.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Yes - High-resolution, color,<br>graphical user interface and<br>touch screen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Yes, To ensure proper<br>system connection and<br>operation.                                                                                                                                                                                                                                                                                                   |
| Event Markers                                                | Yes, artifact detection and<br>signal quality events<br>User selectable events<br>included in Data Viewer<br>software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Yes, user selectable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Yes, Sleep/Wake Events                                                                                                                                                                                                                                                                                                                                         |

{6}------------------------------------------------

{7}------------------------------------------------

#### Non-clinical Testing 7.

Laboratory testing, performed on identical hardware to the DiscoverEEG subject of this submission, demonstrated that the DiscoverEEG, Model DE-401 meets its design and functional requirements. Actual device functions and features were evaluated against the device specifications and in all instances the DiscoverEEG, Model DE-401 performed as expected and no unexpected behavior was observed.

Prior to commercial introduction, the device will meet the requirements of UL medical electrical equipment standards for safety and the IEC particular standard for electroencephalographs.

{8}------------------------------------------------

To ensure safety, the DiscoverEEG, Model DE-401 will comply with all applicable requirements of IEC60601-1 and IEC60601-1-11, and will also comply with the particular requirements for the safety of electroencephalographs established in IEC60601-2-26.

To ensure electromagnetic compatibility, the DiscoverEEG, Model DE-401 will comply with all applicable requirements of IEC60601-1-2, and will also comply with the particular requirements for the safety of electroencephalographs established in IEC60601-2-26.

#### 8. Clinical Testing

The DiscoverEEG, Model DE-401 device is an electroencephalographic device comprised of hardware that has been bench tested to assess safety and effectiveness and to establish substantial equivalence with the predicate devices. We believe further clinical data is not required to demonstrate performance for the DiscoverEEG, Model DE-401 for the indication for use subject to this submission.

#### 9. Conclusion

The DiscoverEEG, Model DE-401, when compared to its predicate devices, has the similar intended use and technological characteristics and principles of operation. The nonclinical tests demonstrate that the device is safe, as effective, and performs at least as safely and effectively as the legally marketed devices. Thus, in indications, intended use and technology, the DiscoverEEG, Model DE-401 is substantially equivalent to the predicated devices. Minor technological difference as documented in the Substantial Equivalence Comparison table above, raise no new questions of safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/OLT/K142834](https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/OLT/K142834)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
