BWII EEG DIGITAL, MODEL PV 1652; BWII PSG DIGITAL, MODEL PV 1524

K062533 · Neurovirtual USA, Inc. · OLT · Dec 22, 2006 · Neurology

Device Facts

Record IDK062533
Device NameBWII EEG DIGITAL, MODEL PV 1652; BWII PSG DIGITAL, MODEL PV 1524
ApplicantNeurovirtual USA, Inc.
Product CodeOLT · Neurology
Decision DateDec 22, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1400
Device ClassClass 2

Indications for Use

The BWII system may be used for electroencephalography (EEG) and sleep recordings (polysomnography) in research and dinical environments. It acquires, displays, and archives EEG and PSG data for on screen review, annotation and event marking by the user. The BWII system also allows for computer-assisted, usercontrolled quantitative analysis of EEG. The BWII system is not intended to replace conventional devices or methods used for EEG and PSG monitoring in critical care or intraoperative settings. The BWII system requires competent user input, and its output must be reviewed and interpreted by a trained technicians or neurologists who will exercise professional judgment in using this information. The BWII system does not make any judgment of normality or abnormality of the displayed signals or the results of an analysis. In no way are any of the functions represented as being in and of themselves diagnostic.

Device Story

BWII is a multi-channel (up to 25) system for EEG and PSG recording; consists of amplifier unit, power module, and BWAnalysis software installed on a PC. System connects to electrodes/sensors via Ethernet; acquires, displays, and archives physiological signals. Used in sleep labs, hospitals, or clinical environments by physicians, neurologists, or trained technicians. Software provides tools for on-screen review, annotation, event marking, and user-controlled quantitative EEG analysis. Device does not perform automated diagnostic judgments; output requires professional interpretation by trained personnel to assist in clinical decision-making. Benefits include standardized data acquisition and review for sleep and neurological assessment.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing, compliance with international safety standards (EN 60601 series, EN ISO 14971), and software validation per FDA guidance.

Technological Characteristics

Multi-channel (up to 25) EEG/PSG system. Components: hard plastic amplifier and power module. Connectivity: Ethernet to PC. Software: BWAnalysis running on Windows XP. Features: impedance checking, calibration, spectrum analysis (frequency/power), and montage editor. Power: external module. Sterilization: N/A (non-invasive).

Indications for Use

Indicated for patients requiring EEG or polysomnography (PSG) sleep recordings in clinical or research environments. Not indicated for critical care or intraoperative monitoring.

Regulatory Classification

Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ [K062533]/A1 Image /page/0/Picture/1 description: The image shows the logo for Neurovirtual. The logo consists of a stylized "N" symbol on the left, followed by the word "NEUROVIRTUAL" in bold, sans-serif font. The logo is black and white. 23 15NW 107th Avenue - B.133 - WH. 1A16 Doral, FL - USA - 33172 Phone: (786) 693-8200 - Fax (305) 393-8429 29 Page ① of 4 3/300/1100 1 Z ... ## Section S 510(k) Summary | A) Submitter's Name: | Neurovirtual USA, Inc. | DEC 2 2 2006 | |---------------------------------------|----------------------------------------------------------|--------------| | Owner / Operator Registration Number: | | | | Manufacture Registration Number: | | | | B) Address: | 2315NW 107th Ave - B.133 - WH. 1A16<br>Doral, FL - 33172 | | | C) Phone and Fax Numbers | | | | Phone: | (786) 693-8200 | | | Fax: | (305) 393-8429 | | | D) Contact Person: | Eduardo J.M. Faria | | | E) Preparation Date: | August 14, 2006<br>November 28, 2006 (Reviewed) | | | F) Classification Name: | | | | Common / Usual Name: | Electroencephalograph | | | Proprietary Name: | BWII Digital (Amplifier and Power Module) | | | Product Code: | GWQ | | | Class: | Class II | | | Regulation: | 882.1400 | | | Common / Usual Name: | Analyzer, spectrum, electroencephalogram signal | | | Proprietary Name: | BWAnalysis (Software) | | | Product Code: | OLT, GWQ, OLV | | | Class: | Class I | | | Regulation: | 882.1420 | | FDA# G) Substantial Equivalence: | 510(k) Number | Model | Manufacture | |---------------|---------------------------------|-------------------------------| | K932407 | Easy Write and Easy Reader | Cadwell<br>Laboratories, Inc. | | K991900 | XLTEC PSG-40<br>Polysomnography | Excel Tech Ltd. | {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for Neurovirtual. The logo consists of a stylized waveform symbol on the left, followed by the word "NEUROVIRTUAL" in bold, sans-serif font. The "O" in "NEUROVIRTUAL" contains a circular design with a pattern inside. 2315NW 107th Avenue - B.133 - WH. 1A16 Doral, FL - USA - 33172 Phone: (786) 693-8200 - Fax (305) 393-8429 Page ② of 4 And the BWAnalysis software is equivalent with the following: | 510(k) Number | Model | Manufacture | |---------------|----------------------------|-------------------------------| | K980214 | Excel Neuro Works | Excel Tech Ltd. | | K932407 | Easy Write and Easy Reader | Cadwell<br>Laboratories, Inc. | | K974718 | Persyst Prism | Persyst<br>Development Corp. | ## F) Description: BWII is multi-channel (up to 25 channels) system designed for electroencephalograph (EEG) and polysomnography (PSG) recording application, in sleep lab, hospital or dinical environment under the supervision of a physician, using a lap top or a desk top computer. The BWII system consists of three maior components: the Amplifier Unit, the Power Module (Both plastic made) and the BWAnalysis Software (a CD to be installed on a regular PC). The system provides connections for electrodes and sensors, and connects to the computer using an Ethernet cable. The BWII works with any good quality patient leads / electrodes and sensors (snore, flow, effort belts and position) that have the safety touch connectors and are legally marketed in accordance with FDA requirements. As these accessories are already legally on the market from different manufactures, they are not part of this submission. The BWII does not make any judgment of normality of the displayed signals or the results of an analysis. In no way are any of the functions represented as being in and of themselves diagnostic. #### I) Intended Used: The BWII system may be used for electroencephalography (EEG) and sleep recordings (polysomnography) in research and dinical environments. It acquires, displays, and archives EEG and PSG data for on screen review, annotation and event marking by the user. The BWII system also allows for computer-assisted, usercontrolled quantitative analysis of EEG. The BWII system is not intended to replace conventional devices or methods used for EEG and PSG monitoring in critical care or intraoperative settings. The BWII system requires competent user input, and its output must be reviewed and interpreted by a trained technicians or neurologists who will exercise professional judgment in using this information. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for Neurovirtual. The logo consists of a stylized waveform symbol on the left, followed by the word "NEUROVIRTUAL" in bold, sans-serif font. The waveform symbol is a simple, angular line that resembles a brainwave pattern. 2315NW 107th Avenue - B.133 - WH. 1A16 Doral, FL - USA - 33172 Phone: (786) 693-8200 - Fax (305) 393-8429 page 3 of 4 The BWII system does not make any judgment of normality or abnormality of the displayed signals or the results of an analysis. In no way are any of the functions represented as being in and of themselves diagnostic. ## J) Technological Characteristics: | | Neurovirtual<br>BWII | Cadwell<br>Easy Write | Excel Tech<br>PSG-40 | Excel Tech<br>Neuro Works | |--------------------------------------------------|-----------------------|-----------------------|-----------------------|---------------------------| | HARWARE | | | | | | 510(k) Number | K062533 | K932407 | K991900 | K980214 | | Classification | Electroencephalograph | Electroencephalograph | Electroencephalograph | Electroencephalograph | | Application | EEG and PSG | EEG and PSG | EEG and PSG | EEG | | Number of<br>Channels | Up to 25 | 32 | 40 | 40 | | Data<br>Communication | Ethernet | Ethernet | Ethernet | Ethernet | | Material<br>(External) | Hard Plastic | Hard Plastic | Hard Plastic | Hard Plastic | | Power Source | External Power Module | External Power Module | Internal Power Module | Internal Power Module | | SOFTWARE | | | | | | Platform | Windows XP | Windows XP | Windows XP | Windows NT | | Impedance | YES | YES | YES | YES | | Calibration | YES | YES | YES | YES | | Spectrum<br>Analyzer<br>(Frequency and<br>Power) | YES | YES | YES | YES | | Montage Editor | YES | YES | YES | YES | ## K) Safety and Effectiveness: The BWII is in compliance with the applicable clauses of the following standards: - EN 60601-1:1990, "Medical Device Equipment: General Requirements for Safety" - EN 60601-1-1:2000, Medical electrical equipment Part 1: General requirements for safety - Section 1: Collateral standard: Safety requirements for medical electrical systems - EN 60601-2: 2001, "Medical Device Equipment General Requirements l for Safety, Collateral Standard: Electromagnetic Compatibility, Requirements and Test" - -EN 60601-2-26:1994, "Medical Device Equipment - Particular requirements for the safety of electroencephalographs" - -EN 60601-1-4: 1996, "Medical Device Equipment - General Requirement for Safety, Collateral Standard: Programmable Electrical Medical Systems" - -EN ISO 14971:2000, "Medical Devices: Application of Risk Management to Medical Devices" {3}------------------------------------------------ Page (4) of 4 Image /page/3/Picture/1 description: The image shows the logo for Neurovirtual. The logo consists of a stylized waveform symbol on the left, followed by the word "NEUROVIRTUAL" in bold, sans-serif font. The waveform symbol is a series of sharp, angular lines resembling a brainwave pattern. 2315NW 107th Avenue – B.133 – WH. 1A16 Doral, FL – USA – 33172 Phone: (786) 693-8200 – Fax (305) 393-8429 - EN 13485: 2003, "Medical Devices, Quality Management Systems: Requirements for Regulatory Purposes" . - General Principles of Software Validation: FDA Guidance software validation version 1.1 (June 09, 1997) - {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, positioned to the right of a circular emblem. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Neurovirtual c/o Mr. Eduardo J.M. Faria President 2315 NW 107th Avenue Box #133 Warehouse 1A16 Doral, Florida 33172 APR - 9 2012 Re: K062533 Trade/Device Name: BWII Digital EEG and BWII Digital PSG with BWAnalysis Software Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: OLT, GWQ, OLV Dated (Date on orig SE 1tr): December 8, 2006 Received (Date on orig SE ltr): December 19, 2006 Dear Mr. Faria: This letter corrects our substantially equivalent letter of January 16, 2007. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Kesia Alexander Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Neurovirtual. The logo consists of a stylized waveform graphic to the left of the word "NEUROVIRTUAL". The word is in all caps and in a bold sans-serif font. The letter "O" in "NEUROVIRTUAL" contains a circular design inside it. 2315NW 107th Avenue - B.133 - WH. 1A16 Doral, FL - USA - 33172 Phone: (786) 693-8200 - Fax (305) 393-8429 #### Section 4 Indications for Use 510(k) Number (if known): K062533 Device Name: BWII Digital EEG and BWII Digital PSG with BWAnalysis Software Indications for Use: The BWII system may be used for electroencephalography (EEG) and sleep recordings (polysomnography) in research and dinical environments. It acquires, displays, and archives EEG and PSG data for on screen review, annotation and event marking by the user. The BWII system also allows for computer-assisted, usercontrolled quantitative analysis of EEG. The BWII system is not intended to replace conventional devices or methods used for EEG and PSG monitoring in critical care or intraoperative settings. The BWII system requires competent user input, and its output must be reviewed and interpreted by a trained technicians or neurologists who will exercise professional judgment in using this information. The BWII system does not make any judgment of normality or abnormality of the displayed signals or the results of an analysis. In no way are any of the functions represented as being in and of themselves diagnostic. AND/OR **Prescription Use** X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Plo and Neurologics **510(k) Number** L06253
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...