DeltaScan Monitor

K222680 · Prolira B.V. · NCG · Feb 2, 2023 · Neurology

Device Facts

Record IDK222680
Device NameDeltaScan Monitor
ApplicantProlira B.V.
Product CodeNCG · Neurology
Decision DateFeb 2, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1440
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K222680 · Feb 2, 2023DeltaScan MonitorProlira B.V.Retrospective clinical calibration dataset (Numan et al., 2019); Expert-labeled EEG recordings from routine clinical practiceThe retrospective dataset was used to calibrate the device's technical 5-point scale and determine the clinical thresholds for the DeltaScan Output (NEGATIVE vs. POSITIVE) by optimizing sensitivity, specificity, and NPV against expert labels.Algorithm calibration; Retrospective clinical data; Expert labeling; Acute encephalopathy

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Numan et al., 2019 (Calibration Dataset); Retrospective analysis of clinical EEG recordingsAdult patients with acute encephalopathy and delirium; Sample Size: 321 EEG recordingsNot applicable for this studyCalibration of 5-point DeltaScan Score and clinical thresholds for acute encephalopathy

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Acute encephalopathy diagnosisAutomated waveshape recognition and signal quality quantification algorithmsNPV >= 0.80 (ICU), NPV >= 0.85 (wards), NPV + PPV >= 1ICU: NPV 0.85, NPV+PPV 1.62; Ward: NPV 0.83, NPV+PPV 1.66; Pooled: NPV 0.84, NPV+PPV 1.63321 EEG recordings with expert labels (Numan et al., 2019)>1 (EEG experts)DeltaStudy (clinicaltrials.gov # NCT03966274): 434 patients (195 ICU, 239 wards)3 (EEG experts) + 3 (clinical delirium experts)

Indications for Use

The DeltaScan Monitor provides the binary DeltaScan Output based on a technical index of polymorphic delta (PMD) waveshape detections made in the EEG from the bipolar Fp2 and Pz channel on adult patients (over 60 years of age) to aid in the diagnosis of acute encephalopathy. DeltaScan should only be used by a healthcare provider as a component of a complete clinical evaluation or as support for the clinician's decision to pursue further testing. The device is NOT to be used as a stand-alone method in the evaluation or diagnosis of acute encephalopathy. The intended patient is a hospitalized, awake adult, who is at risk of acute encephalopathy and delirium as decided by the responsible licensed healthcare physician or a medical professional working under the responsibility of a licensed healthcare physician. The use environment is in hospitals: · non-sterile environments; · ICUs, wards, and other patient evaluation locations; The DeltaScan Monitor is intended to be used in combination with the DeltaScan Patch (K222671) through a proprietary connector design.

Device Story

Device acquires EEG signals via DeltaScan Patch (Fp2, Pz, Fpz electrodes); processes signals to detect polymorphic delta (PMD) waveshapes; uses signal conditioning and artifact rejection (eye movement, muscle, motion) to isolate PMD; calculates 5-point DeltaScan Score; outputs binary result (Negative/Positive for acute encephalopathy). Used in hospitals (ICUs, wards) by clinicians. Output supports clinical decision-making by confirming or challenging clinical assessments of delirium/encephalopathy; helps guide further diagnostic testing. Benefits include objective, consistent assessment of acute encephalopathy in hospitalized patients.

Clinical Evidence

Prospective multi-center study (DeltaStudy, n=434) compared DeltaScan output against expert panel consensus (3 EEG experts for acute encephalopathy; 3 clinical experts for delirium). Primary endpoints (NPV, NPV+PPV) were evaluated. NPV was 0.84 (pooled) and 0.85 (ICU). Sensitivity analysis adjusting for prevalence confirmed performance met pre-specified criteria. Repeatability (ICC) was 0.799 for the 5-point scale and 0.829 for binary output. No adverse events reported.

Technological Characteristics

EEG acquisition device with integrated computer module. Uses DeltaScan Patch (cutaneous electrodes). Components: EEG amplifier, eMMC memory, Embedded Linux OS, battery, charger. Connectivity: proprietary connector for patch. Software: DeltaScan Monitor Application (Class B per EN 62304). Safety: IEC 60601-1, IEC 60601-1-2. Non-invasive, transient skin contact.

Indications for Use

Indicated for hospitalized, awake adults (>60 years) at risk of acute encephalopathy and/or delirium. Used as an adjunct to clinical evaluation by licensed physicians or supervised medical professionals. Contraindicated for patients on sedation (RASS -4 or -5), <60 years old, history of brain injury, using Lithium/Clozapine, or with metal/plastic implants in the upper head hemisphere.

Regulatory Classification

Identification

The neuropsychiatric interpretive electroencephalograph assessment aid is a prescription device that uses a patient's electroencephalograph (EEG) to provide an interpretation of the patient's neuropsychiatric condition. The neuropsychiatric interpretive EEG assessment aid is used only as an assessment aid for a medical condition for which there exists other valid methods of diagnosis.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The technical parameters of the device, hardware and software, must be fully characterized and must demonstrate a reasonable assurance of safety and effectiveness. (i) Hardware specifications must be provided. Appropriate verification, validation, and hazard analysis must be performed. (ii) Software, including any proprietary algorithm(s) used by the device to arrive at its interpretation of the patient's condition, must be described in detail in the software requirements specification and software design specification. Appropriate software verification, validation, and hazard analysis must be performed. (2) The device parts that contact the patient must be demonstrated to be biocompatible. (3) The device must be designed and tested for electrical safety, electromagnetic compatibility, thermal, and mechanical safety. (4) Clinical performance testing must demonstrate the accuracy, precision, reproducibility, of determining the EEG-based interpretation, including any specified equivocal zones (cutoffs). (5) Clinical performance testing must demonstrate the ability of the device to function as an assessment aid for the medical condition for which the device is indicated. Performance measures must demonstrate device performance characteristics per the intended use in the intended use environment. Performance measurements must include sensitivity, specificity, positive predictive value, and negative predictive value per the device intended use. Repeatability of measurements must be demonstrated using interclass correlation coefficients and illustrated by qualitative scatter plot(s). (6) The device design must include safeguards to prevent use of the device as a stand-alone diagnostic. (7) The labeling must include the following information: (i) A warning that the device is not to be used as a stand-alone diagnostic. (ii) A detailed summary of the clinical performance testing, including any adverse events and complications. (iii) The qualifications and training requirements for device users including technicians and clinicians. (iv) The intended use population and the intended use environment. (v) Any instructions technicians should convey to patients regarding the collection of EEG data. (vi) Information allowing clinicians to gauge clinical risk associated with integrating the EEG interpretive assessment aid into their diagnostic pathway. (vii) Where appropriate, validated methods and instructions for reprocessing of any reusable components.

In combination with the general controls of the FD&C Act, the Neuropsychiatric Interpretive Electroencephalograph Assessment Aid is subject to the following special controls:

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ February 2, 2023 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Prolira B.V. % Dr. Paul Manberg Corolla Clin/Reg Consulting 481 Spindrift Trail #696 Corolla, North Carolina 27927 ## Re: K222680 Trade/Device Name: DeltaScan Monitor R2 Regulation Number: 21 CFR 882.1440 Regulation Name: Neuropsychiatric Interpretive Electroencephalograph Assessment Aid Regulatory Class: Class II Product Code: NCG Dated: December 16, 2022 Received: January 3, 2023 ### Dear Paul Manberg: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {2}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerelv. # Jay R. Gupta -S Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K222680 Device Name DeltaScan Monitor R2 #### Indications for Use (Describe) The DeltaScan Monitor provides the binary DeltaScan Output based on a technical index of polymorphic delta (PMD) waveshape detections made in the EEG from the bipolar Fp2 and Pz channel on adult patients (over 60 years of age) to aid in the diagnosis of acute encephalopathy. DeltaScan should only be used by a healthcare provider as a component of a complete clinical evaluation or as support for the clinician's decision to pursue further testing. The device is NOT to be used as a stand-alone method in the evaluation or diagnosis of acute encephalopathy. The intended patient is a hospitalized, awake adult, who is at risk of acute encephalopathy and delirium as decided by the responsible licensed healthcare physician or a medical professional working under the responsibility of a licensed healthcare physician. The use environment is in hospitals: · non-sterile environments; · ICUs, wards, and other patient evaluation locations; The DeltaScan Monitor is intended to be used in combination with the DeltaScan Patch (K222671) through a proprietary connector design. Please refer to the Instructions for Use and the Instructions for Use on the Primary packaging of the DeltaScan Patch for more information. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with 21 CFR 807.92. #### SUBMITTER l. Prolira BV Padualaan 8 Utrecht, The Netherlands NL-3584CH Phone Number: +31 6 47762936 Email: r.vanmerkerk@prolira.com Contact Person: Rutger van Merkerk Date Prepared: February 2nd, 2023 #### II. DEVICE Name of Device: DeltaScan Monitor Common or Usual Name: DeltaScan Monitor R2 Classification Name: Neuropsychiatric interpretive electroencephalograph assessment aid (21 CFR 882.1440) Regulatory Class: II Product Code: NCG #### III. PREDICATE DEVICE | Name | Manufacturer | De novo # | |-------------|------------------|-----------| | NEBA system | NEBA HEALTH, LLC | DEN110019 | #### IV. DEVICE DESCRIPTION Acute encephalopathy and delirium are extremely common in hospitalized patients, for example, one third of general medical patients aged 70 years or older has delirium (Marcantonio et al., 2017, NEJM). Acute encephalopathy and delirium are, by definition, the consequence of the same underlying medical condition (Slooter et al., 2020, Intensive Care Med). Acute encephalopathy can present as subsyndromal delirium, or delirium. Acute encephalopathy can be assessed with EEG (Palanca et al., 2017, BJA). Delirium is determined by clinical examination (Diagnostic and Statistical Manual of Mental Disorders, Fifth Edition). Acute encephalopathy and delirium occur in the same patient. Many patients with acute encephalopathy (i.e., EEG alterations) develop delirium (i.e., behavioral changes). Patients with acute encephalopathy and/or delirium should therefore also be seen as the same {5}------------------------------------------------ patient population. Clinically, there is a clear relation between the two, which is important to understand. The figure below presents the relationship between underlying conditions, acute encephalopathy, and delirium. Image /page/4/Figure/3 description: The image shows a flow chart of medical conditions. The first box at the top says "Underlying condition" and has an arrow pointing down to a box that says "Acute Encephalopathy EEG change". There is an arrow pointing down to a box that says "Delirium Behavioral change". The arrows are labeled "Can cause" and "Can present as". Figure 5-1: the relation between acute encephalopathy and delirium In everyday care, standard 21-lead EEG as well as delirium expert assessments are not routinely feasible. Therefore, checklists were developed for routine screening. The diagnostic performance in routine care settings, however, is found to be low as these tests have subjective elements. Patient can be best cared for when acute encephalopathy is assessed with an objective tool that helps to detect acute encephalopathy as early as possible, followed by further clinical assessment. Specific delta activity (PolyMorphic Delta activity; PMD) occurs in the EEG of patients with acute encephalopathy and delirium (i.e., PMD waves are a known characteristic of acute encephalopathy and delirium). PMD is considered a sensitive and specific biomarker for acute encephalopathy and delirium in awake adults (over 60 years of age) without dementia, neurological trauma, or other causes of PMD in the awake state. The DeltaScan Monitor provides EEG signal acquisition and analysis technology intended for use as an adjunct to clinical judgment. The DeltaScan Monitor provides support in clinical decision-making by providing an assessment for a patient having acute encephalopathy or not, based on a measure of the detected polymorphic delta (PMD) waves in the EEG. {6}------------------------------------------------ The DeltaScan Monitor consists of (see Figure 5-2 below): Image /page/5/Figure/3 description: The image shows a medical device and its components. The device includes a monitor, a patch connector, and a recording button, labeled as 1, 2, and 3 respectively. Additionally, a charger for the monitor is shown and labeled as 4. Figure 5-2 DeltaScan Monitor components The Patch connector contains the EEG amplifier hardware. The Monitor contains electronics for galvanic isolation to the EEG cable with Patch connector, storage of EEG recording and log files (eMMC memory chip), processing capacity to run software (DeltaScan Monitor Application, or DMA), user interface elements (e.g., screen, keys, recording button), battery (FEY PA-IEC-LNB162Q.R001), and the charging circuitry. EEG data is collected by the DeltaScan Monitor using a DeltaScan Patch (see Figure 5-3). The DeltaScan Patch requires separate 510(k) clearance as a Class II device under Classification Name Cutaneous electrode (21 CFR 882.1320). In designing the DeltaScan Patch, the International 10-20 System was used as a basis for electrode placement. Two recording electrodes are placed on the Fp2 and Pz locations, and the reference electrode is placed at the Fpz location. The electrodes are individually packaged and pregelled. The DeltaScan Monitor should only be used in combination with a DeltaScan Patch. {7}------------------------------------------------ Image /page/6/Figure/2 description: The image shows a diagram of a brain scanning device. The device consists of a headset with sensors that is connected to a control unit. The control unit has a display screen and buttons for operation. The headset is worn on the head and the sensors are positioned to measure brain activity. Figure 5-3 Artist impression of the DeltaScan Monitor used with a DeltaScan Patch Collected EEG signals are amplified, digitized, and then processed by the software algorithms to provide the user with the DeltaScan Output. The DeltaScan Monitor Application is stand-alone software running on an Embedded Linux OS. To understand Prolira's technology, first the basic concept of EEG power spectrum analysis is explained. Kooi et al. (2015, Chest) explain that a relative delta measure presents the relative amount of energy that is present in the EEG spectrum between 0.5Hz and 4Hz, divided by the energy in the whole spectrum of interest that lies between 0.5Hz and 30Hz. Kooi et al. used EEG segments that are handpicked by the researcher and that are relatively free of artefacts to perform their analysis. But, as with any automatically derived parameter, artefacts and poor signal quality may lead to inappropriate results. For example, eye movements are commonly present in a typical EEG segment (even in eyes closed EEG data). Furthermore, eye movements can present as waveshapes in the same frequency range as the waveshapes that are characteristic of acute encephalopathy and delirium (i.e., energy in the delta frequency band). Therefore, a general relative delta measure will also count the energy of eye movements when automated algorithms provide the result in a clinical setting. {8}------------------------------------------------ To deal with this situation, Prolira's technology specifically detects EEG waveshapes that are characteristic for acute encephalopathy and delirium, allowing to calculate a specific (rather than general) measure based on the amount of detected PMD waves in the EEG. In this, Prolira uses several artifact detection and segregation approaches: - 1. State-of-the-art signal conditioning and signal quality quantification technology are applied to minimize the disturbance of artefacts, removing as many artefacts as possible, for example due to poor skin contact (high impedance), muscle activity or rigidity, head and body motion, improper sensor placement, and unusual or excessive electrical interference. - 2. Specific EEG waveshapes in the delta frequency range (polymorphic delta waves; PMD) are detected. Using advanced waveshape recognition technology, the chance of detecting the target waveshapes is increased, and the chance to include artefacts in the resulting measure is reduced. In addition, the algorithm also specifically detects eye artifact, which allows for additional scrutiny of detected PMD waves in the case that eye artifact and PMD waves are situated close together (see Figure 5-4 for an example of PMD wave and eye artifact detections in the EEG). Image /page/7/Figure/5 description: The image shows a graph of voltage over time. The x-axis represents time in seconds, ranging from 82 to 92. The y-axis represents voltage in microvolts, ranging from -100 to 100. There are two lines plotted on the graph, one in blue and one in red, showing the voltage fluctuations over time. Figure 5-4: examples of a polymorphic delta (PMD) wave detection (red line) and eye artifact detections (turquoise line) {9}------------------------------------------------ The detected and analyzed PMD waveshapes in the EEG are then translated into the 5-point DeltaScan Score, which is the DeltaScan's calibrated technical index. The DeltaScan Output is NEGATIVE for acute encephalopathy for DeltaScan Score 1 or 2. The DeltaScan Output is POSITIVE for acute encephalopathy for DeltaScan Score 3, 4 or 5. See below in the section Clinical Studies for details on the clinical validation thereof. Table 5-1 below indicates the clinical meaning of the DeltaScan Output and the technical meaning of the 5-point DeltaScan Score. | DeltaScan result | Clinical meaning of<br>DeltaScan Output | Technical meaning of<br>DeltaScan Score | |---------------------------|-------------------------------------------|----------------------------------------------------------------------------------------------| | NEGATIVE<br><br>1 2 3 4 5 | NEGATIVE test for acute<br>encephalopathy | For Score 1 and 2,<br>PMD waveshapes are<br>not or hardly detected in<br>the EEG | | NEGATIVE<br><br>1 2 3 4 5 | | | | POSITIVE<br><br>1 2 3 4 5 | POSITIVE test for acute<br>encephalopathy | For Score 3, 4 and 5,<br>the amount of<br>detected PMD<br>waveshapes in the<br>EEG increases | | POSITIVE<br><br>1 2 3 4 5 | | | | POSITIVE<br><br>1 2 3 4 5 | | | ### Table 5-1 Technical and clinical meaning of a DeltaScan result The technical 5-point scale and clinical threshold for acute encephalopathy are calibrated based on a previous clinical calibration dataset (the dataset is presented in Numan et al., 2019, BJA) containing of 321 EEG recordings with expert labels for acute encephalopathy and delirium. The following calibration steps were performed: - . For each recording in the calibration dataset, the number of PMD detections and the PMD power (detected PMD waveshapes / unit time) was calculated. - . The clinical threshold for acute encephalopathy, NEGATIVE (score 1-2) and POSITIVE (score 3-5), is determined on the calibration dataset by finding an optimum for sensitivity, specificity and NPV using the expert labels for acute encephalopathy and delirium. {10}------------------------------------------------ - The cut-offs between score 3 and 4, and score 4 and 5, were chosen to maximize repeatability of the score points. The number of patients from the calibration dataset with a POSITIVE result are equally divided over the score bins 3, 4, and 5. This sets the boundaries in terms of the PMD power. - . The cut-off between score 1 and 2 is determined by the number of detections; none or one detection always leads to DeltaScan Score 1, two detections always lead to DeltaScan Score 2, regardless of the power in those detections. Note: the measure based on the detected PMD waveshapes underlying the 5-point scale is nonlinear. #### V. INDICATIONS FOR USE The DeltaScan Monitor provides the binary DeltaScan Output based on a technical index of polymorphic delta (PMD) waveshape detections made in the EEG from the bipolar Fp2 and Pz channel on adult patients (over 60 years of age) to aid in the diagnosis of acute encephalopathy. DeltaScan should only be used by a healthcare provider as a component of a complete clinical evaluation or as support for the clinician's decision to pursue further testing. The device is NOT to be used as a stand-alone method in the evaluation or diagnosis of acute encephalopathy. The intended patient is a hospitalized, awake adult (over 60 years of age), who is at risk of acute encephalopathy and/or delirium as decided by the responsible licensed healthcare physician or a medical professional working under the responsibility of a licensed healthcare physician. The use environment is in hospitals: - · in non-sterile environments; - · in departments like ICUs, wards, and other patient evaluation locations; The DeltaScan Monitor is intended to be used in combination with the DeltaScan Patch (K222671) through a proprietary connector design. Please refer to the Instructions for Use and the Instructions for Use on the Primary packaging of the DeltaScan Patch for more information. #### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The DeltaScan Monitor is an EEG acquisition device with an integrated computer module that runs the DeltaScan Monitor Application (DMA) to provide the User Interface and runs automated processing modules to determine the DeltaScan Output on an EEG recording. Although standard EEG devices could also be used by a trained physician to detect elevated slow wave activity (high relative delta measure), the DeltaScan Monitor has been designed to provide a more practical, consistent, and easy to interpret measure of elevated low frequency waveforms associated with acute encephalopathy. In recognition of this more focused clinical application, a standard EEG device may not be the most appropriate predicate device for a 510(k) regulatory pathway. {11}------------------------------------------------ Instead, the NEUROPSYCHIATRIC EEG-BASED ASSESSMENT AID FOR ADHD (NEBA) SYSTEM with De Novo # DEN110019 is suggested as the most appropriate predicate device. A technology comparison between the DeltaScan Monitor and the NEBA System is provided in Table 5-2 below. | NEBA System | DeltaScan Monitor | |--------------------------------------------------|-----------------------------------------------------| | Intended Use | Intended Use | | Neuropsychiatric Interpretive | Neuropsychiatric Interpretive | | Electroencephalograph Assessment Aid. | Electroencephalograph Assessment Aid. | | The Neuropsychiatric Interpretive | The Neuropsychiatric Interpretive | | Electroencephalograph Assessment Aid is a | Electroencephalograph Assessment Aid is a | | prescription device that uses a patient's | prescription device that uses a patient's | | electroencephalograph (EEG) to provide an | electroencephalograph (EEG) to provide an | | interpretation of the patient's neuropsychiatric | interpretation of the patient's neuropsychiatric | | condition. | condition. | | The Neuropsychiatric Interpretive EEG | The Neuropsychiatric Interpretive EEG Assessment | | Assessment Aid is used only as an assessment aid | Aid is used only as an assessment aid for a medical | | for a medical condition for which there exists | condition for which there exists other valid | | other valid methods of diagnosis. | methods of diagnosis. | | Indications for Use | Indications for Use | #### Table 5-2 Technology comparison {12}------------------------------------------------ | NEBA System | DeltaScan Monitor | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | The Neuropsychiatric EEG-Based ADHD | The DeltaScan Monitor provides the binary | | Assessment Aid (NEBA®) uses the theta/beta | DeltaScan Output based on a technical index of | | ratio of the EEG measured at electrode CZ on a | polymorphic delta (PMD) waveshape detections | | patient 6-17 years of age combined with a | made in the EEG from the bipolar Fp2 and Pz | | clinician's evaluation to aid in the diagnosis of | channel on adult patients (over 60 years of age) to | | ADHD. | aid clinicians in the diagnosis of acute | | NEBA should only be used by a clinician as | encephalopathy. | | confirmatory support for a completed clinical | DeltaScan should only be used by a clinician as a | | evaluation or as support for the clinician's | component of a complete clinical evaluation or as | | decision to pursue further testing following a | support for the clinician's decision to pursue | | clinical evaluation. The device is NOT to be used | further testing. The device is NOT to be used as a | | as a stand-alone in the evaluation or diagnosis of | stand-alone method in the evaluation or diagnosis | | ADHD. | of acute encephalopathy. | | | The intended patient is a hospitalized, awake adult | | | (over 60 years of age), who is at risk of acute | | | encephalopathy and/ or delirium as decided by the | | | responsible licensed healthcare physician or a | | | medical professional working under the | | | responsibility of a licensed healthcare physician. | | | The use environment is in hospitals: | | | • in non-sterile environments; | | | • in departments like ICUs, wards, and other | | | patient evaluation locations; | | | The DeltaScan Monitor is intended to be used in | | | combination with the DeltaScan Patch (K222671) | | | through a proprietary connector design. | | | Please refer to the Instructions for Use and the | | | Instructions for Use on the Primary packaging of | | | the DeltaScan Patch for more information | | NEBA System | DeltaScan Monitor | | LIMITATIONS | LIMITATIONS | | For prescription use only. | For prescription use only. | | The NEBA cannot be used in an individual for whom an<br>EEG recording is not valid, specifically a patient with:<br>a history of EEG abnormalities; a history of a seizure disorder; on anticonvulsant medication(s); a metal plate in the head; or a metal device in the head. | The DeltaScan Monitor cannot be used in an individual<br>for whom an EEG recording is not valid, specifically a<br>patient:<br>on sedation (RASS -4 or -5); persons under 60 years of age; with a history of brain injury; using Lithium/Clozapine; with a metal or plastic implant in the upper<br>hemisphere of the head; with an active medical device in the head | | The NEBA system cannot be used in subjects who are<br>unable to remain still for a minimum of 30 seconds for<br>EEG recording. | The DeltaScan system cannot be used in subjects who<br>are unable to follow three measurement instructions;<br>the patient should (1) be awake, (2) be relaxed, and (3)<br>keep the eyes closed. | | | For patients on sedatives with a RASS -3, with dementia<br>(MMSE $\le$ 24), and with (acute) brain injury, the DeltaScan<br>Output can be higher than expected. For more severe<br>and/or acute cases (i.e., of sedation, dementia and brain<br>injury), it becomes less likely that a POSTIVE DeltaScan<br>Output relates to an acute encephalopathic brain state.<br>The meaning of NEGATIVE DeltaScan Output for these<br>patients is most likely not altered. The DeltaScan Output<br>for these patients should be interpreted accordingly. | | The NEBA system should only be used by medical<br>professionals qualified to assess psychiatric disorders<br>and experienced in diagnosing ADHD. To ensure<br>proper device performance, the user must first<br>perform a diagnostic evaluation per the standard of<br>their practice. NEBA interpretations are based on the<br>clinician's initial diagnostic evaluation, the subject's<br>age and the EEG results. | The DeltaScan Monitor should only be used by licensed<br>healthcare physicians or medical professionals qualified<br>to assess inpatient disorders and who are experienced<br>in diagnosing acute encephalopathy and/ or delirium, or<br>by other medical professionals working under the<br>responsibility of a licensed healthcare physician. To<br>ensure proper device performance, the user must<br>perform a diagnostic evaluation per the standard of<br>their practice. DeltaScan interpretations are based on<br>the clinician's initial diagnostic evaluation, the subject's<br>response to questioning and the EEG results. | | The device should not be used as a stand-alone<br>diagnostic device. | The device should not be used as a stand-alone<br>diagnostic device. | | PLEASE REFER TO THE LABELING FOR A MORE<br>COMPLETE LIST OF WARNINGS, PRECAUTIONS AND<br>CONTRAINDICATIONS. | PLEASE REFER TO THE LABELING FOR A MORE<br>COMPLETE LIST OF WARNINGS, PRECAUTIONS AND<br>CONTRAINDICATIONS. | | Primary EEG Feature/Electrode position | | | Theta/Beta Ratio | A binary DeltaScan Output based on a technical<br>index of polymorphic delta (PMD) waveshape | | NEBA System | DeltaScan Monitor | | | detections made in the EEG | | CZ | Fp2 and Pz | | Software | | | Software for the device consisted of both<br>proprietary software and off-the- shelf (OTS)<br>software. | DeltaScan Monitor Application developed<br>according to "EN 62304:2006+A1:2015 Medical<br>device software - Software life-cycle processes". | | | Software class according to<br>EN62304:2006+A1:2015: Class B. | | The software was reviewed and the provided<br>documentation was found adequate and<br>consistent with a 'MODERATE' level of concern.,<br>as discussed in the FDA document, "Guidance for<br>the Content of Premarket Submissions for<br>Software Contained in Medical Devices," issued<br>May 11, 2005. | The software was reviewed and the provided<br>documentation was found adequate and<br>consistent with a 'MODERATE' level of concern., as<br>discussed in the FDA document, "Guidance for the<br>Content of Premarket Submissions for Software<br>Contained in Medical Devices," issued May 11,<br>2005. | | Device output | | | Three categories<br>LOW TBR = Strongly Recommend Further Clinical<br>Testing (other conditions)<br>MODERATE TBR = Suggest Further Clinical Testing<br>(other conditions)<br>HIGH TBR = Confirmatory Support for ADHD as<br>primary diagnosis | Two categories based on a threshold of the<br>amount of PMD waveshapes detected in the EEG:<br>- NEGATIVE for acute encephalopathy<br>- POSITIVE for acute encephalopathy | {13}------------------------------------------------ {14}------------------------------------------------ #### VII. PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination. ### Biocompatibility testing The biocompatibility evaluation for the DeltaScan Monitor was assessed in accordance with the FDA's guidance document titled, "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices Part 1: Evaluation and testing within a risk management process", June 16, 2016. The DeltaScan Monitor, its EEG cable, the Recording button, and the medical-grade Power adapter have transient contact with the patient's and/ or user intact skin (i.e., finger contact while setting up and using the device). If there is contact between the patient and a part of the DeltaScan Monitor at all, this would be with the Patch connector, see Figure 5-2 above. {15}------------------------------------------------ None of the components are "attached" to skin, all contact is handling contact. The material of DeltaScan Monitor is generally used in combination with intact skin of the user, and the materials used are identical as in widespread use applications. Based on the above, there is no biocompatibility testing needed for the DeltaScan Monitor. #### Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the DeltaScan Monitor. The DeltaScan Monitor complies with the IEC 60601-1:2005+A1:2012 standard for safety and the IEC 60601-1-2:2014 standard for EMC. #### Software verification and validation testing Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices". The software for this device was considered a "moderate" level of concern, since a failure or latent flaw in the software could lead to a delay in delivery of appropriate medical care that could lead to minor injury. #### Performance bench testing Thorough bench testing is provided in the file which are considered adequate. This included, but is not limited to: - Usability testing demonstrated compliance to EN 62366-1:2015+A1:2020 - . The labeling was tested to be legible, durable, compliant with regulations, and compatible with the packaging. - . The packaging was tested to verify it is compatible with the labeling and can be shipped without damage. - Mechanical Strength testing demonstrated compliance to IEC 60601-1:2005+A1:2012 (cl.15.3). - Bench testing showed >95% reliability for the technical 5-point DeltaScan Score for increasing amounts of polymorphic delta activity (PMD) present in sample EEGs. These verification and validation activities all show that the device complies to the preset requirements. #### Clinical Studies Diagnostic Performance testing of DeltaScan was done in the multi-center DeltaStudy, which ran between Q1 2018 and Q3-2021. The study has been designed to comply with ISO 14155:2011 Clinical investigation of medical devices for human subjects - Good clinical practice, and in the European Medical Device Regulation (MDR) {16}------------------------------------------------…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...