← Product Code [JXE](/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/JXE) · K060584

# NC-STAT (K060584)

_Neurometrix, Inc. · JXE · Jul 31, 2006 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/JXE/K060584

## Device Facts

- **Applicant:** Neurometrix, Inc.
- **Product Code:** [JXE](/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/JXE.md)
- **Decision Date:** Jul 31, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1550
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The NeuroMetrix NC-stat is intended to stimulate and measure neuromuscular signals that are useful in diagnosing and evaluating systemic and entraponent neuropathies.

## Device Story

NC-stat is a nerve conduction testing system; stimulates and measures neuromuscular signals. Components: handheld instrument, docking station, disposable biosensors, onCall Information System. Instrument controls nerve conduction study; determines supramaximal stimulus intensity; acquires/analyzes motor and sensory waveforms; displays latency, conduction velocity, amplitude, and response parameters on LCD. Biosensors (single-use) contain electrodes for stimulation, response detection, and skin temperature measurement. Docking station downloads data via infra-red link to onCall Information System via analog phone line. onCall system generates hardcopy reports comparing results to normal ranges; reports sent via fax or email. Used by clinicians to diagnose neuropathies; provides objective data for clinical decision-making; benefits patients through standardized, automated nerve conduction assessment.

## Clinical Evidence

Clinical data submitted demonstrates that nerve conduction measurements obtained using the NC-stat are comparable to those obtained using conventional nerve conduction measurement equipment.

## Technological Characteristics

System includes handheld instrument, docking station, and disposable biosensors. Biosensors include electrodes for stimulation, response detection, and skin surface temperature measurement. Connectivity via infra-red link and analog phone line. Instrument features LCD readout for real-time data display. Software controls stimulus intensity, waveform acquisition, and analysis.

## Regulatory Identification

A nerve conduction velocity measurement device is a device which measures nerve conduction time by applying a stimulus, usually to a patient's peripheral nerve. This device includes the stimulator and the electronic processing equipment for measuring and displaying the nerve conduction time.

## Predicate Devices

- NC-stat ([K982359](/device/K982359.md), [K000565](/device/K000565.md), [K003508](/device/K003508.md), [K013459](/device/K013459.md) and [K041320](/device/K041320.md))
- TECA TD-10/TD-20 ([K802637](/device/K802637.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a rectangular box with the text 'K060584' written inside. The text is written in a bold, handwritten style. The box is outlined with a thick black line.

Page 6 of 3

# 510(k) Summary for NeuroMetrix NC-stat

JUL 3 1 2006

#### 1. SPONSOR

NeuroMetrix, Inc. 62 Fourth Avenue Waltham, MA 02451

Contact Person: Rainer Maas Telephone: (781) 314-2781

Date Prepared: July 26, 2006

#### 2. DEVICE NAME

| Proprietary Name:    | NC-stat                                    |
|----------------------|--------------------------------------------|
| Common/Usual Name:   | Nerve Conduction Testing System            |
| Classification Name: | Nerve Conduction Velocity Measuring Device |

#### 3. PREDICATE DEVICES

- NC-stat (K982359, K000565, K003508, K013459 and K041320) .
- TECA TD-10/TD-20 (K802637) .

#### 4. INTENDED USE

The NeuroMetrix NC-stat is intended to stimulate and measure neuromuscular signals that are useful in diagnosing and evaluating systemic and entraponent neuropathies.

#### ട. DEVICE DESCRIPTION

The NC-stat system consists of the following four components:

- An instrument that contains the electronic circuitry and software required 5.1 to control a nerve conduction study, including determination of the supramaximal stimulus intensity, acquisition and analysis of motor and -

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the text "page 2 of 3" written in cursive. The number 2 is circled. The text appears to be handwritten.

sensory waveforms, and display of latency/conduction velocity, amplitude and response configuration parameters. The instrument saves patient and study data for optional transmission to a docking station for real-time hard copy report generation. The instrument displays nerve conduction data in real-time on the LCD readout, including the distal motor latency (DML), Compound Muscle Action Potential (CMAP) amplitude, CMAP duration, CMAP area, motor conduction velocity (CV), F-wave latency, distal sensory latency (DSL), the Sensory Nerve Action Potential (SNAP) amplitude, SNAP duration, sensory CV, limb indicator (left or right), low battery indicator, the memory slot being used to store the test data, and user messages (menu selections, sensor serial numbers, device status, operator instructions, and error conditions).

- A stationary docking station is used to download the test data from the 5.2 instrument via an infra-red communication link to the onCall Information System via an analog phone line. The docking station is non-transportable and connected to both a power outlet and an analog phone line.
- Single-use, disposable biosensors are available for the median motor, 5.3 ulnar motor, median motor & sensory, ulnar motor & sensory, deep peroneal, posterior tibial, and sural nerves. The biosensors contain electrodes for nerve stimulation, response detection, and skin surface temperature measurement.
- The onCall Information System for automatic generation of the hardcopy 5.4 patient test report, which includes test results (DML, Compound Muscle Action Potential - CMAP- amplitude, DSL, SNAP amplitude, conduction velocity, F-wave latency, and associated response waveforms) and comparison of patient results to normal ranges. Reports are sent to the user by facsimile or e-mail.

#### BASIS FOR SUBSTANTIAL EQUIVALENCE 6.

The NC-stat that is the subject of this 510(k) Premarket Notification is substantially equivalent to the NC-stat as previously cleared for marketing and to the TECA TD-10/TD-20 EMG and nerve conduction velocity measurement device. The purpose of this 510(k) is to describe changes made to the NC-stat, including:

- Addition of the modified median motor-sensory biosensor for use in 6.1 diagnosing neuropathies affecting the median motor-sensory nerves.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the text "Page (3) of 3" written in cursive. The number 3 is circled. The text appears to be handwritten, and it is likely part of a document or report indicating the page number.

### 6.2 Modifications to the labeling.

None of the changes to the NC-stat have altered the intended use of the device, the basic device design or the device operation. Clinical data submitted in the 510(k) demonstrates that nerve conduction measurements obtained using the NCstat are comparable to those obtained using conventional nerve conduction measurement equipment.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo consists of a circular seal with the words "U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with three wavy lines above them.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 3 1 2006

NeuroMetrix, Inc. % Rainer Maas Director, Quality Assurance and Regulatory Affairs Associate 62 Fourth Avenue Waltham, Massachusetts 02451

Re: K060584 Trade/Device Name: NC-stat Regulation Number: 21 CFR 882.1550 Regulation Name: Nerve conduction velocity measurement device Regulatory Class: II Product Code: JXE, IKN Dated: July 5, 2006 Received: July 6, 2006

Dcar Rainer Maas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device. subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{4}------------------------------------------------

### Page 2 - Rainer Maas

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely vours.

Helmut Lachenmann

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Intended Use

510(k) Number (if known): K060584

Device Name: NC-stat

Intended Use:

The NeuroMetrix NC-stat is intended to to stimulate and measure neuromuscular signals that are useful in diagnosing and evaluating systemic and entrapment neuropathies.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE If NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Hulbert Brewer

(Division Simmal 317 Division of General, Restorative, and Neurologica. Devices

510(k) Number Ko 60584

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/JXE/K060584](https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/JXE/K060584)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
