MSAS PROFESSIONAL

K993824 · Biomeridian, Int. · GZO · Jan 20, 2000 · Neurology

Device Facts

Record IDK993824
Device NameMSAS PROFESSIONAL
ApplicantBiomeridian, Int.
Product CodeGZO · Neurology
Decision DateJan 20, 2000
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 882.1540
Device ClassClass 2

Indications for Use

The MSA-21 intended use is for the measurement of Galvanic Skin Respons

Device Story

The MSA-21 (Meridian Stress Assessment) is a Galvanic Skin Response (GSR) measurement device. It functions by measuring skin electrical conductance to assess stress. The device utilizes specialized probes and electrodes connected via distinct, non-interchangeable connectors (3-pin and 7-pin mini-DIN) to ensure safe operation. It is intended for use by healthcare professionals to obtain physiological measurements. The output provides data on skin response, which clinicians use to evaluate patient stress levels. The device is designed for clinical settings.

Clinical Evidence

Bench testing only. The device underwent safety and performance testing, including electrical safety (UL 2601-1, CSA C22.2 No 601.1, EN60601-1, IEC 601-1) and emissions/immunity testing (EN60601-1-2:1993). No clinical data provided.

Technological Characteristics

Galvanic Skin Response measurement device. Features 3-pin mini-DIN electrode connector and 7-pin mini-DIN probe connector to prevent connection errors. Complies with UL 2601-1, CSA C22.2 No 601.1, EN60601-1, IEC 601-1, and EN60601-1-2:1993 standards. Class II device.

Indications for Use

Indicated for the measurement of Galvanic Skin Response. Patient population and contraindications not specified.

Regulatory Classification

Identification

A galvanic skin response measurement device is a device used to determine autonomic responses as psychological indicators by measuring the electrical resistance of the skin and the tissue path between two electrodes applied to the skin.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 882.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K 993824 Image /page/0/Picture/2 description: The image is a logo for "The Human Calibration Company BioMeridian". The logo features a geometric design in the center, surrounded by the text "The Human Calibration Company" in a circular arrangement. Below the geometric design, the word "BioMeridian" is written in a bold font. ## SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION APPLICANT'S NAME/ADDRESS Contact Person COMMON/USUAL NAME: CLASSIFICATION NAME: ESTABLISHMENT REGISTRATION NUMBER: CLASSIFICATION: PERFORMANCE STANDARD: BioMeridian, Inc. Draper, Utah 84020 12411South 265 West Suite F Joe Galloway Galvanic Skin Response Device Galvanic Skin Response Device ### 1723429 The Galvanic Skin Response Measurement Device is classified into class II as described in Title 21, code of Federal Regulation 82.1540. BioMeridian Inc. is not aware of any special controls or performance standards established for this device under sections 513 or 514, respectively, of the Food, Drug and Cosmetics Act. ## SUBSTANTIAL EQUIVALENCE: BioMeridian, Inc. believes the MAS-21 is substantially equivalent to the J&J G-25a GSR, and Global Corp.'s C-29 CORPORATE OFFICES - - - - TRAINING AND EDUCATION (801) 501- - (801) 501-7518 BioMeridian International, Inc. CUSTOMER SUPPORT (801) 501 - - (801) 501-7518 13526 South 110 West WWW.BIOMERIDIAN.COM · GENERAL INFORMATION mail@biomeridian.com - THAINING AND EDUCATION - training@biomeridian.com JUSTOMER SUPPORT helpdesk@biomeridian.com Draper, Utah 84020 ાં તિપ #### BIOMERIDIAN INTERNATIONAL, INC s a wholly owned subsidiary of - Magellan Technology, Inc. (OTC: MGL MAGELLAN PHONE - (801) 495-2211 - (801) 495 1100 - MAGFILLAN SHARFHOI DER BELATIONS nagellan-info@biomeridian.co {2}------------------------------------------------ Image /page/1/Picture/0 description: The image is a logo for "The Human Calibration Company BioMeridian". The logo features a geometric design in the center, resembling interlocking shapes. The text "THE HUMAN CALIBRATION COMPANY" is arranged in a circular fashion around the geometric design. Below the design, the word "BioMeridian" is written in a larger, bold font. The MSA-21 has undergone an extensive battery of tests to insure safety. The unit design was safety intense from the outset. Small details such as using a 3 pin mini Din connector for the electrode and a 7 pin mini Din connector for the probe so they could not be inserted into the wrong plug are just one example. This quest for safety included all MSA-21 accessories. BioMeridian, Inc. also submitted the MSA-21 to Intertek Testing Services for test and evaluation. The MSA-21 fully passed the following test: Medical Electrical Equipment (see attachment # 3) a) UL 2601-1 - b) CSA C22.2 No 601.1 Medical Electrical Equipment, Part 1: General Requirements for Safety (see attachment # 4) - a) EN60601-1 - b) IEC 601-1 Emissions and Immunity Testing (see attachment # 5) a) EN60601-1-2:1993 In summary the MSA-21 meets or exceeds all the safety requirements for a medical device in its class. Our dedication to safety is evidenced in the many extra steps we have taken to insure a safe product. Also the MSA-21 has met all the requirements and has received a registered CE mark. | <b>CORPORATE OFFICES</b> | | |--------------------------|----------------| | • TOLL-FREE PHONE | (888) 224-2337 | | • TELEPHONE | (801) 495-1188 | | • CORPORATE FAX | (801) 495-1199 | | • SALES FAX | (801) 495-2266 | | <b>TRAINING AND EDUCATION</b> | | |-------------------------------|----------------| | • TOLL-FREE PHONE | (888) 765-4665 | | • TELEPHONE | (801) 501-7517 | | • FAX | (801) 501-7518 | BioMeridian International, Inc. | <b>CUSTOMER SUPPORT</b> | | |-------------------------|----------------| | • TOLL-FREE PHONE | (877) 404-2378 | | • TELEPHONE | (801) 501-7517 | | • FAX | (801) 501-7518 | 13526 South 110 West | <b>WWW.BIOMERIDIAN.COM</b> | | |----------------------------|--------------------------| | • GENERAL INFORMATION | mail@biomeridian.com | | • TRAINING AND EDUCATION | training@biomeridian.com | | • CUSTOMER SUPPORT | helpdesk@biomeridian.com | Draper, Utah 84020 USA | <b>BIOMERIDIAN INTERNATIONAL, INC</b> | | |---------------------------------------|---------------------------------------| | | is a wholly owned subsidiary of | | | Magellan Technology, Inc. (OTC: MGLN) | | • MAGELLAN PHONE | (801) 495-2211 | | • MAGELLAN FAX | (801) 495-1199 | | • MAGELLAN SHAREHOLDER RELATIONS | magellan-info@biomeridian.com | {3}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is an abstract image of an eagle. Public Health Service JAN 20 2000 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Joe Galloway Vice President of Compliance Biomeridian, Inc. 12411 South 265 West Suite F Draper, Utah 84020 K993824 Re: Trade Name: MSA-21 Regulatory Class: II Product Code: GZO Dated: December 21, 1999 Received: December 21, 1999 Dear Mr. Galloway: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed {4}------------------------------------------------ Page 2 - Mr. Joe Galloway predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Styt Rhode James E. Dillard III Acting Director Division of General Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use Statement K 993824 510 (k) Number. 25;cm イ ﺘﺮ Device Name. M SA-21 (Meridian Stress Assessment) Indications for Use. The MSA-21 intended use is for the measurement of Galvanic Skin Respons ## PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDE Concurrence of CDRH, Office of Device Evaluation (ODE) Stupt Rhordu Division of General Restorative Devices 953824 510(k) Number Prescription Use _ OR Over- The Counter Use (Per 21 CFR 801.109)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...