Wovyn Depth Electrode

K253970 · Sensomedical Labs, Ltd. · GZL · Apr 17, 2026 · Neurology

Device Facts

Record IDK253970
Device NameWovyn Depth Electrode
ApplicantSensomedical Labs, Ltd.
Product CodeGZL · Neurology
Decision DateApr 17, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1330
Device ClassClass 2

Indications for Use

Sensomedical Wovyn Depth Electrodes are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain.

Device Story

Wovyn Depth Electrode is a single-use, sterile, implantable device for stereoelectroencephalography (SEEG). It consists of a body (0.8-2.0 mm diameter) with 4-16 macro electrode contacts (1.5-3 mm length) and an included stylet for placement. Used by neurosurgeons in professional healthcare facilities to record, monitor, and stimulate subsurface brain electrical signals. The device connects to external recording/stimulation equipment. It is MR unsafe. The device facilitates clinical decision-making for epilepsy surgery or neurological procedures by providing localized electrical data. Benefits include temporary, precise subsurface signal access for diagnostic evaluation.

Clinical Evidence

No clinical studies were conducted. Substantial equivalence is supported by nonclinical bench testing, including mechanical, electrical, and functional verification, biocompatibility testing (cytotoxicity, sensitization, systemic toxicity, brain implantation study), packaging/shelf-life validation, and EO sterilization validation (SAL 10^-6).

Technological Characteristics

Implantable depth electrode; 4-16 Pt90/Ir10 macro contacts; TPU jacket; 0.8-2.0 mm diameter; 360-400 mm length. Passive sensing/stimulation; MR unsafe. Sterilized via Ethylene Oxide (ISO 11135).

Indications for Use

Indicated for temporary (<30 days) recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain in patients undergoing evaluation for epilepsy surgery or other neurological procedures. Contraindicated in patients at risk for infection or where the surgical procedure cannot be performed safely.

Regulatory Classification

Identification

A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION April 17, 2026 Sensomedical Labs, Ltd. Maroun Farah CEO Nazareth Industrial Park Mount Precipice Nazareth, 1612102 Israel Re: K253970 Trade/Device Name: Wovyn Depth Electrode Regulation Number: 21 CFR 882.1330 Regulation Name: Depth electrode Regulatory Class: Class II Product Code: GZL Dated: November 25, 2025 Received: December 11, 2025 Dear Maroun Farah: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253970 - Maroun Farah Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K253970 - Maroun Farah Page 3 Sincerely, Patrick Antkowiak -S for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253970 | ? | | Please provide the device trade name(s). | | ? | | Wovyn Depth Electrode | | | | Please provide your Indications for Use below. | | ? | | Sensomedical Wovyn Depth Electrodes are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} S SENSO MEDICAL # 510(k) SUMMARY ## Wovyn Depth Electrode Premarket Notification — 510(k) Number: K253970 Submitted pursuant to 21 CFR 807.92 # 1. Submitter and Device Identification | Submitter's Name | Sensomedical Labs, Ltd. | | --- | --- | | Address: | Nazareth Industrial Park, Mount Precipice Nazareth 1612102, Israel +972 (0) 4 6800668 | | Contact Person Name: | Maroun Farah | | 510(k) Number: | K253970 | | Device Trade Name: | Wovyn Depth Electrode | | Common or Usual Name: | Electrode, Depth | | Device Classification: | Class II — 21 CFR 882.1330; Product Code: GZL | | Review Panel: | Neurological and Physical Medicine | | Primary Predicate: | SENSO SEEG Depth Electrode — 510(k) K213170 (Sensomedical Labs, Ltd.) | | Secondary Predicate: | PMT Depthalon Electrode — 510(k) K151790 (PMT Medical) | | Date of Summary: | March 18, 2026 | # 2. Device Description The Wovyn Depth Electrode is a single-use, pre-sterilized, implantable depth electrode designed for stereoelectroencephalography (SEEG). It is used by neurosurgeons to record, monitor, and stimulate electrical signals at the subsurface level of the brain in patients undergoing evaluation for epilepsy surgery or other neurological procedures. The Wovyn Depth Electrode is a smooth electrode, the device is available in configurations of 4 to 16 macro electrode contacts, with contact lengths of 1.5 mm to 3 mm along the electrode body and various inter-contact spacing configurations. The electrode body diameter in the brain contact region ranges from 0.8 mm to 2.0 mm, and the overall electrode length ranges from 360 mm to 400 mm. A stylet is included to facilitate placement. The device is supplied sterile (Ethylene Oxide sterilization) and is labeled for single-patient use with a maximum duration of use of 30 days. {5} S SENSO MEDICAL The Wovyn Depth Electrodes are intended for temporary use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The Wovyn Depth Electrodes are connected to the user under the guidance and instruction of a supervising physician (physicians in the areas of biopotential recording, monitoring and stimulation/response studies, knowledgeable in the use of depth electrodes). Environment of use: This device is intended to be implanted into a patient's brain by a trained clinician in a professional healthcare facility only. It is NOT intended for MR use and is MR unsafe. # 3. Indications for Use Sensomedical Wovyn Depth Electrodes are intended for temporary (&lt;30 days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain. # 4. Summary of Technological Characteristics The technological characteristics of the Wovyn Depth Electrode were compared against those of the primary predicate (SENSO SEEG Depth Electrode, K213170) and secondary predicate (PMT Depthalon Electrode, K151790). The comparison is presented below for each characteristic. | Features | New device Wovyn Fabric Depth Electrode (Under Review) | Senso SEEG depth Electrode (K213170) Primary Predicate | PMT DEPTHALON Electrode K151790 Secondary Predicate | Comments | | --- | --- | --- | --- | --- | | Indications for Use | Senso Medical Wovyn Depth Electrodes are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. | Senso Medical Depth Electrodes are intended for temporary (< 30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. | The Platinum /Stainless Steel Depthalon Electrodes are intended for temporary (<30days) use with recording, monitoring, and stimulation equipment for the recording, monitoring, and stimulation of electrical signals in the subsurface of the brain | Same | | Clinical Application | Placed in the subsurface level of the brain to support recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. | Placed in the subsurface level of the brain to support recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. | Not available online | Same | {6} SENSO MEDICAL | | recording, monitoring and stimulation. | monitoring and stimulation. | | | | --- | --- | --- | --- | --- | | Contraindications | These depth electrodes should not be used on any patient who the physician/surgeon considers at risk for infection or for whom the surgical recording and stimulation procedure cannot be performed safely and effectively. | These depth electrodes should not be used on any patient who the physician/surgeon considers at risk for infection or for whom the surgical recording and stimulation procedure cannot be performed safely and effectively. | Not available online | Same | | Single patient use, Disposable | Yes | Yes | Yes | Same | | Provided Sterile | Yes | Yes | Yes | Same | | Environment of use | Intraoperative and Neurological | Intraoperative and Neurological | | Same | | Use | monitoring locations | monitoring locations | | Same | | Duration of Use | < 30 days | < 30 days | | Same | | Electrode Contact Material | Platinum/Iridium | Stainless-Steel | Stainless steel or platinum | Different than 1st predicate, but same as second predicate | | Maximum Stimulation Charge Density | < 30 μC/cm2 | < 30 μC/cm2 | | Same | | Number of electrode contacts | 4-16 | 4-16 | Not available online | Equivalent | | Electrode jacket | TPU | TPU | TPU | Equivalent | | Electrode body diameter (brain contact) | 0.8 mm to 2 mm | 0.8 mm to 2 mm | 0.8mm | Equivalent - Variations of electrode body diameter is provided | | Stylet | Yes | Yes | Yes | Same | | Electrode contact length (along body of the electrode) | 1.5mm – 3mm Macro, Other different configuration available. | 1.5mm –3mm Macro | Standard 2mm, however; Different spacing configuration available | Similar | | Overall length | 360 - 400[mm] | 360 - 400[mm] | 235.24 to 291 mm | Same | Sensomedical concludes that the Wovyn Depth Electrode (K253970) is substantially equivalent to the primary predicate device, the SENSO SEEG Depth Electrode (K213170), for the following reasons: {7} SENSO MEDICAL - Same intended use: Both devices are intended for temporary (&lt;30 days) recording, monitoring, and stimulation of electrical signals at the subsurface level of the brain, in the same patient population, in the same clinical settings. - Same or equivalent technological characteristics for all features except electrode contact material: The devices share identical or equivalent indications, clinical application, contraindications, use environment, duration of use, electrode jacket material, body diameter, contact length, overall length, stylet design, number of contacts, stimulation charge density limits, and single-use sterile supply. - Different electrode contact material does not raise new safety or effectiveness questions: The use of Pt90/Ir10 alloy contacts in place of stainless steel is the sole technological difference. This difference is well-supported by the secondary predicate (K151790), which uses the same Platinum material as many other Depth electrodes, and does not alter intended use, mechanism of action, or device design. Nonclinical testing confirms the safety and performance of the Pt90/Ir10 contacts under the intended conditions of use. ## 5. Brief Discussion of Nonclinical Tests ### Non-clinical testing performed Performance testing was performed to assure safety and effectiveness of the Senso Medical Depth Electrodes. All necessary bench testing was conducted on the Senso Medical Depth Electrodes to ensure conformance to design specifications and to support a determination of substantial equivalence to the predicate devices. The nonclinical, bench testing performed included: - Verification testing (mechanical, electrical, and functional testing), verifying design output meets design input - Biocompatibility Testing including Cytotoxicity, Skin Sensitization, Intracutaneous test, Material-Mediated Pyrogenicity, Acute Systemic Toxicity, Subacute systemic toxicity, Bacterial Reverse Mutation, Chromosomal aberration, Hemolysis test, Brain Implantation Study. - Packaging, shelf life and shipment and transit testing; and - EO Sterilization Validation in accordance with ISO 11135. Demonstrating achievement of a Sterility Assurance Level (SAL) of $10^{-6}$. - Performance testing including electrical (impedance, recording, and stimulation), mechanical, and electrochemical testing. ## 6. Clinical Tests No clinical studies specific to the Wovyn Depth Electrode were conducted or are required to support these 510(k) submissions. The substantial equivalence determination is supported entirely by {8} SENSO MEDICAL nonclinical testing data and predicate device comparison, consistent with the 510(k) regulatory pathway for this device type and classification. ## 7. Conclusion: From the data available we can justify that the " Senso Medical Depth Electrodes " has the same intended use and the same technological characteristics as the already marketed predicate devices identified above. Hence our device can be considered substantially equivalent to the predicates.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...