DIXI Medical Microdeep Micro-Macro Depth Electrodes
K202087 · Dixi Medical · GZL · Nov 18, 2021 · Neurology
Device Facts
Record ID
K202087
Device Name
DIXI Medical Microdeep Micro-Macro Depth Electrodes
Applicant
Dixi Medical
Product Code
GZL · Neurology
Decision Date
Nov 18, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1330
Device Class
Class 2
Indications for Use
The DIXI Medical Microdeep® Micro-Macro Depth Electrodes are intended for temporary (<30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports definition of epileptogenic foci and brain mapping.
Device Story
Invasive depth electrode for temporary (<30 days) intracranial use; records, monitors, and stimulates brain electrical signals. Input: neural electrical signals; Output: electrical signals transmitted to external recording/monitoring equipment. Used in intraoperative and neurological monitoring settings by physicians trained in intracranial neurophysiology. Device features macro-contacts on the exterior and micro-contacts flush at the surface; includes a microdrive for micro-contact adjustment. Output facilitates identification of epileptogenic foci and brain mapping, aiding clinical decision-making for epilepsy treatment. Benefits include precise subsurface neural signal acquisition.
Clinical Evidence
Bench testing only. Testing included dimensional verification, mechanical functionality (microdrive, water tightness, tensile strength), electrical characteristics (continuity, resistance, stimulation integrity), and biocompatibility per ISO 10993. Comparative testing against the predicate confirmed equivalent tensile strength and mechanical functionality after simulated use.
Technological Characteristics
Single-use, sterile, ethylene oxide-sterilized depth electrode. Macro-contacts: Platinum/Iridium 90/10; Micro-contacts: Tungsten. Anchor bolt: Titanium alloy Ti6Al4V (ASTM F136). One-piece design with microdrive. Dimensions: 0.8 mm macro-contact diameter, 20-micron micro-contact diameter. Connectivity: connects to external recording/stimulation equipment via connection system. No software/algorithm.
Indications for Use
Indicated for patients requiring temporary (<30 days) intracranial recording, monitoring, and stimulation of electrical signals to define epileptogenic foci and perform brain mapping. Used by physicians trained in intracranial neurophysiology.
Regulatory Classification
Identification
A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.
Predicate Devices
Ad-Tech Medical Instrument Corporation's Macro Micro Depth Electrodes (K163355)
{0}------------------------------------------------
November 18, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Steven Plymale CEO Dixi Neurolab, Inc. 145 Howland Pines Drive Oxford, MI 48371
Re: K202087
Trade/Device Name: DIXI Medical Microdeep Micro-Macro Depth Electrodes Regulation Number: 21 CFR 882.1330 Regulation Name: Depth Electrode Regulatory Class: Class II Product Code: GZL Dated: October 20, 2021 Received: October 22, 2021
Dear Steven Plymale:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202087
Device Name
DIXI Medical Microdeep® Micro-Macro Depth Electrodes
Indications for Use (Describe)
The DIXI Medical Microdeep® Micro-Macro Depth Electrodes are intended for temporary (<30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports definition of epileptogenic foci and brain mapping.
Type of Use (Select one or both, as applicable)
| <div> <span> <b> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> |
|-----------------------------------------------------------------------------------------------|
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue rectangle on the left, followed by the word "DIXI" in black, bold letters. Below "DIXI" is the word "medical" in a smaller, light gray font. The logo is simple and modern, with a focus on the company name.
# 510(k) Summary
#### I. SUBMITTER
510(k) Owner DIXI Medical 2A route de Pouligney 25640 Chaudefontaine, France Office: 033-3-81889890 Fax: 033-3-81889899 info@diximedical.com
Official correspondent: Lucie Petegnief R&D Manager l.petegnief@diximedical.com
Submission Correspondent Name: Steve Plymale CEO Dixi Neurolab Inc. 145 Howland Pines Drive Oxford, MI 48371 Email: steven.plymale@dixineurolab.com Phone: 514 882 3258
Date Prepared Oct. 20, 2021
> II. DEVICE
| Name of Device: | DIXI Medical Microdeep® Micro-Macro Depth Electrode |
|-----------------------|-----------------------------------------------------|
| Common or Usual Name: | Depth Electrode |
| Classification Name: | 21 CFR §882.1330 Depth Electrode |
| Regulatory Class: | II |
| Product Code: | GZL |
#### III. PREDICATE DEVICE
The predicate device is the Ad-Tech Medical Instrument Corporation's Macro Micro Depth Electrodes (K163355). The DIXI Medical Microdeep® Depth Electrode (K170959) was a reference device for this submission.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue rectangle on the left, followed by the word "DIXI" in black, bold letters. Below "DIXI" is the word "medical" in gray, smaller letters. The logo is simple and modern, with a focus on the company name.
### IV. DEVICE DESCRIPTION
The Microdeep® Micro-Macro Depth Electrode is a single patient use, sterile and disposable device. The Microdeep® Micro-Macro Depth Electrode is invasive as it is placed in contact with nerve tissue (brain) and must only be used during an SEEG procedure with Anchor Bolts. The Microdeen® Micro-Macro Depth Electrode is intended to connect to the user's recording. monitoring and stimulation equipment using the Connection System. This product is intended to be used only by physicians in the area of biopotential recording, monitoring and stimulation / response studies who are trained in intracranial neurophysiology.
The DIXI Medical Microdeep® Micro-Macro Depth Electrode is comprised of the following components:
- Macro-contacts - located on the outside of the device.
- . Micro-contacts - sit flush at the surface of the device between the macro electrodes.
- Microdrive for micro-contacts ●
- . Cap
- Connectors for micro and macro-contacts ●
#### V. INDICATIONS FOR USE
The DIXI Medical Microdeep® Micro-Macro Depth Electrodes are intended for temporary (<30 days) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. The recording of electrical activity supports definition of the location of epileptogenic foci and brain mapping.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue square on the left, followed by the word "DIXI" in black, bold letters. Below the word "DIXI" is the word "medical" in gray, smaller letters. The logo is simple and modern, with a focus on the company name.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE VI.
The technological characteristics comparing to the predicate device are outlined in the Substantial Equivalent table.
| Attribute | DIXI Medical<br>Microdeep® Micro-Macro<br>Depth Electrodes<br>(Subject Device) | Ad-Tech Medical Instrument<br>Corporation<br>Macro Micro Depth<br>Electrodes<br>(Predicate device K163355) | DIXI Medical<br>Microdeep Depth Electrode<br>(Reference Device K170959) |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number | 21 CFR §882.1330 | 21 CFR §882.1330 | 21 CFR §882.1330 |
| Regulation Name | Depth electrode | Depth electrode | Depth electrode |
| Regulatory Class | Class II | Class II | Class II |
| Product Code | GZL | GZL | GZL |
| Indications for Use | The DIXI Medical Microdeep®<br>Micro-Macro Depth<br>Electrodes are intended for<br>temporary (<30 days) use<br>with recording, monitoring<br>and stimulation equipment<br>for the recording, monitoring<br>and stimulation of electrical<br>signals at the subsurface level<br>of the brain. The recording of<br>electrical activity supports<br>definition of the location of<br>epileptogenic foci and brain<br>mapping. | The AD-TECH Depth<br>Electrodes (Depth Electrodes,<br>Foramen Ovale Depth<br>Electrodes, Macro Micro<br>Depth Electrodes, Spencer<br>Probe Depth Electrodes,<br>Wyler Sphenoidal Depth<br>Electrodes) are intended for<br>temporary (< 30 days) use<br>with recording, monitoring<br>and stimulation equipment for<br>the recording, monitoring and<br>stimulation of electrical<br>signals at the subsurface level<br>of the brain. The recording of<br>electrical activity supports<br>definition of the location of | The DIXI Medical Microdeep<br>Depth Electrodes are intended<br>for temporary (<30 days) use<br>with recording, monitoring<br>and stimulation equipment for<br>the recording, monitoring and<br>stimulation of electrical signals<br>at the subsurface level of the<br>brain. |
| Attribute | DIXI Medical<br>Microdeep® Micro-Macro<br>Depth Electrodes<br>(Subject Device) | Ad-Tech Medical Instrument<br>Corporation<br>Macro Micro Depth<br>Electrodes<br>(Predicate device K163355) | DIXI Medical<br>Microdeep Depth Electrode<br>(Reference Device K170959) |
| Environment of Use | Intraoperative and<br>Neurological monitoring<br>locations | Intraoperative and<br>Neurological monitoring<br>locations<br>epileptogenic foci and brain<br>mapping. | Intraoperative and<br>Neurological monitoring<br>locations |
| Provided Sterile | Yes | Yes | Yes |
| Method of sterilization | Ethylene oxide | Ethylene oxide | Ethylene oxide |
| Single Use, disposable | Yes | Yes | Yes |
| Duration of Use | < 30 days | < 30 days | < 30 days |
| Maximum Stimulation<br>Charge Density | $\leq 30 \mu C/cm^2$ | $\leq 30 \mu C/cm^2$ | $\leq 30 \mu C/cm^2$ |
| Electrode Characteristics | | | |
| General characteristic | One piece design: depth<br>electrode that has electrical<br>macro-contacts collars placed<br>on the outside and micro-<br>contacts cut flush at the<br>surface, between macro-<br>contacts, which can be<br>released | Two kinds of Macro Micro<br>Depth Electrodes:<br>1. BEHNKE FRIED DEPTH<br>ELECTRODES, two piece<br>design:<br>- Macro portion: depth<br>electrode that has electrical<br>macro-contacts collars placed<br>on the outside and an inner<br>lumen throughout the entire<br>length<br>- Micro portion: wire bundle | One piece design: depth<br>electrode that has electrical<br>macro-contacts collars placed<br>on the outside, no micro-<br>contacts |
| Attribute | DIXI Medical<br>Microdeep® Micro-Macro<br>Depth Electrodes<br>(Subject Device) | Ad-Tech Medical Instrument<br>Corporation<br>Macro Micro Depth<br>Electrodes<br>(Predicate device K163355) | DIXI Medical<br>Microdeep Depth Electrode<br>(Reference Device K170959) |
| | | (micro-contacts) that passes<br>through the inner lumen of the<br>macro-electrode<br><br>2. MACRO-MICRO DEPΤΗ<br>ELECTRODES: one piece<br>design: depth electrode that<br>has electrical macro-contacts<br>collars placed on the outside<br>and micro-contacts cut flush<br>at the surface, between macro-<br>contacts | |
| Removable stylet in the<br>electrode | No | Yes | No |
| Contacts Material | Platinum/Iridium 90/10<br>(Macro-contact)<br>Tungsten (Micro-contact) | Platinum (Macro-contact)<br>Platinum/Iridium (Micro-<br>contact) | Platinum/Iridium 90/10<br>(Macro-contact) |
| Electrode body diameter<br>(brain contact) | 0.8 mm (Macro-contact)<br>20 microns (Micro-contact) | 1.3 mm (Macro-contact)<br>38 to 51 microns (Micro-<br>contact) | 0.8 mm (Macro-contact) |
| Number of electrode<br>contacts | From 6 to 9 (Macro-contact)<br>From 8 to 12 (Micro-contact) | From 6 to 12 (Macro-contact)<br>From 8 to 24 (Micro-contact) | From 5 to 18 (Macro-contact) |
| Electrode contact length<br>(along body of the<br>electrode) | 2 mm (Macro-contact) | 1.57 mm (Macro-contact) | 2 mm (Macro-contact) |
| Attribute | DIXI Medical<br>Microdeep® Micro-Macro<br>Depth Electrodes | Ad-Tech Medical Instrument<br>Corporation<br>Macro Micro Depth<br>Electrodes | DIXI Medical<br>Microdeep Depth Electrode |
| | (Subject Device) | (Predicate device K163355) | (Reference Device K170959) |
| Overall length | ~ 400 mm | ≤ 660 mm | ≤ 1050 mm |
| Device Accessories - Anchor Bolt | | | |
| Product Code | GZL | GZL | GZL |
| Regulation Number | 21 CFR §882.1330 | 21 CFR §882.1330 | 21 CFR §882.1330 |
| Regulatory Class | II | II | II |
| Material | Titanium alloy Ti6Al4V (ELI,<br>ASTM F136) | Titanium<br>Silicone (inner lumen gasket)<br>Parylene | Titanium alloy Ti6Al4V (ELI,<br>ASTM F136) |
| Length | From 15 mm to 35 mm | From 13 mm to 26 mm | From 15 mm to 35 mm |
| Required Drill Hole<br>Diameter | 2.1 mm | 2.4 mm or 2.8 mm | 2.1 mm |
| Compatible Depth<br>Electrode Body Diameter | 0.8 mm | 0.86 mm to 1.3 mm | 0.8 mm |
| Attachment onto the skull | Threaded into a pre-drilled<br>hole in the skull | Threaded into a pre-drilled<br>hole in the skull | Threaded into a pre-drilled<br>hole in the skull |
| Anchor bolt placement | Long Screwdriver | Anchor bolt placement<br>wrench | Long Screwdriver |
| Anchor bolt removal | Wrench and Short<br>Screwdriver | Anchor bolt removal wrench | Wrench and Short Screwdriver |
| Implantation duration | Up to 30 days | Up to 30 days | Up to 30 days |
| Single use | Yes | Yes | Yes |
| Attribute | DIXI Medical<br>Microdeep® Micro-Macro<br>Depth Electrodes<br>(Subject Device) | Ad-Tech Medical Instrument<br>Corporation<br>Macro Micro Depth<br>Electrodes<br>(Predicate device K163355) | DIXI Medical<br>Microdeep Depth Electrode<br>(Reference Device K170959) |
| Method of sterilization | Ethylene oxide | Ethylene oxide | Ethylene oxide |
| Device Accessories | | | |
| List of Accessories | Instrument guide, Bone<br>starter, Sliding Ruler,<br>Coagulation Electrode and<br>Stop, Flat Screwdriver (for<br>stop), Drill and Stop, Long<br>Screwdriver, Anchor Bolt,<br>Marking Rod, Cap for Anchor<br>Bolt, Depth Report device,<br>Stylet, Stop, Sheath Holder,<br>Extension cable, Adapter and<br>Connection cable, and Wrench<br>and Short Screwdriver for<br>Anchor Bolt. | Guide, Channeled Ruler, Drill<br>Bit with Stop and Wrench,<br>Anchor placement/removal<br>Wrench, Anchor Bolt,<br>Obturator and Cabrio and<br>Tech-Attach Connection<br>Systems | Instrument guide, Bone<br>starter, Sliding Ruler,<br>Coagulation Electrode and<br>Stop, Flat Screwdriver (for<br>stop), Drill and Stop, Long<br>Screwdriver, Anchor Bolt,<br>Marking Rod, Cap for Anchor<br>Bolt, Depth Report device,<br>Stylet, Stop, Extension cable,<br>Adapter and Connection cable,<br>and Wrench and Short<br>Screwdriver for Anchor Bolt. |
| Single-Use Accessories | Coagulation Electrode and<br>Stop, Drill and Stop, Anchor<br>Bolt, Cap for Anchor Bolt,<br>Stylet, Sheath Holder | Drill Bit with Stop and<br>Wrench, Anchor Bolt and<br>Obturator | Coagulation Electrode and<br>Stop, Drill and Stop, Anchor<br>Bolt, Cap for Anchor Bolt,<br>Stylet |
| Method of Sterilization<br>for Single Use Accessories | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
| Reusable Accessories | Instrument guide, Bone<br>starter, Sliding Ruler, Flat<br>Screwdriver (for stop), Long<br>Screwdriver, Marking Rod | Guide, Channeled Ruler,<br>Anchor placement/removal<br>Wrench, and Cabrio and Tech-<br>Attach Connection Systems | Instrument guide, Bone<br>starter, Sliding Ruler, Flat<br>Screwdriver (for stop), Long<br>Screwdriver, Marking Rod |
| Attribute | DIXI Medical<br>Microdeep® Micro-Macro<br>Depth Electrodes<br>(Subject Device) | Ad-Tech Medical Instrument<br>Corporation<br>Macro Micro Depth<br>Electrodes<br>(Predicate device K163355) | DIXI Medical<br>Microdeep Depth Electrode<br>(Reference Device K170959) |
| | Depth Report Device, Stop,<br>Extension Cable, Adapter,<br>Connection Cable, Wrench<br>and Short Screwdriver for<br>Anchor Bolt | | Depth Report Device, Stop,<br>Extension Cable, Adapter,<br>Connection Cable, Wrench and<br>Short Screwdriver for Anchor<br>Bolt |
| Method of Sterilization<br>for Reusable Accessories | Steam Sterilization (with the<br>exception of the Adapter,<br>which does not require<br>sterilization) | Unknown (except for Cabrio<br>and Tech-Attach Connection<br>Systems: Ethylene Oxide or<br>Sterrad) | Steam Sterilization (with the<br>exception of the Adapter,<br>which does not require<br>sterilization) |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue square, followed by the word "DIXI" in black, bold letters. Below "DIXI" is the word "medical" in gray, smaller letters.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for DIXI medical. The word "DIXI" is in bold, black letters, and the word "medical" is in gray letters below it. To the left of the word "DIXI" is a blue square.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue square, followed by the word "DIXI" in black, bold letters. Below "DIXI" is the word "medical" in gray, smaller letters.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue square, followed by the word "DIXI" in black, bold letters. Below "DIXI" is the word "medical" in gray, smaller letters.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue square, followed by the word "DIXI" in black, bold letters. Below "DIXI" is the word "medical" in gray, smaller letters.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for DIXI medical. The logo consists of a blue square on the left, followed by the word "DIXI" in black, bold letters. Below "DIXI" is the word "medical" in gray, smaller letters. The logo is simple and modern, with a clear emphasis on the company name.
#### VII. NON-CLINICAL TESTING
DIXI Medical has performed the following non-clinical laboratory testing to determine substantial equivalence.
| Test | Summary of Verifications | Result |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Dimensional<br>Characteristics | Verification of electrodes dimensional<br>characteristics | Pass |
| | | |
| Mechanical characteristics | - Verification of electrode microdrive<br>functioning | Pass |
| | - Verification of electrode water tightness | |
| | - Verification of cap maintenance on the<br>electrode sheath | |
| | - Verification of the antirotation function of<br>electrode cap | |
| | - Verification of tensile strength of electrode<br>macro-contacts | |
| |…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.