Retrospective clinical patient records/data from DBS procedures
A retrospective clinical validation study was conducted to assess the accuracy of the HaGuide software module in mapping sub-thalamic nucleus (STN) boundaries by comparing software measurements against expert neurosurgeon/electrophysiologist measurements from routine clinical procedures.
The NeuroNav System / NeuroSmart System, incorporated the installed HaGuide software, is intended to be used in assisting neurosurgeons, in the operating functional neurosurgery, and to record from and stimulate brain motor and sensory neurons and to aid in the placement of depth electrode. The NeuroNav Drive is intended to be used in assisting neurosurgeons, in the operating room during functional neurosurgery, to aid in the placement of depth electrodes.
Device Story
NeuroNav/NeuroSmart systems are intraoperative neurophysiological recording and stimulation devices used by neurosurgeons during functional neurosurgery (e.g., DBS). The system records micro-electrode signals (MER) from brain cells or stimulates target zones. The NeuroSmart sub-model includes HaGuide software, which processes real-time data to detect the Subthalamic Nucleus (STN) region, including entry/exit boundaries, and provides stimulation location recommendations. The system is compatible with various stereotactic frames and microprobes. By providing automated boundary detection and stimulation guidance, the device assists surgeons in accurate electrode placement, potentially improving surgical outcomes for movement disorders.
Clinical Evidence
Retrospective, multi-center clinical validation study (n=81 patients, 105 surgeries, 213 electrodes) evaluated HaGuide software accuracy in mapping STN boundaries compared to expert neurosurgeon/electrophysiologist ground truth. Primary endpoint: percent agreement between software and expert measurements. Results: 90.34% agreement (95% CI: 85.27-93.85%). Correlation coefficients for entry and exit site depth were 0.9492 and 0.9317, respectively. Bland-Altman analysis showed mean differences of 0.18 mm (entry) and -0.15 mm (exit). No adverse events reported.
Technological Characteristics
System includes hardware for MER and stimulation, and software (HaGuide). Electrical safety per IEC 60601-1 (3rd Ed); EMC per IEC 60601-1-2 (3rd Ed). Sterilization: Steam (ISO 17665-1, ST77) for mechanical accessories; STERRAD NX/100NX for electrical accessories. Software development per ISO 62304. Connectivity: Standard AC power (100-240V).
Indications for Use
Indicated for use in adults and children undergoing functional neurosurgery to assist neurosurgeons in recording from and stimulating brain motor and sensory neurons and to aid in the placement of depth electrodes.
Regulatory Classification
Identification
A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The first logo is the Department of Health & Human Services - USA logo, which features a stylized human figure. The second logo is the FDA U.S. Food & Drug Administration logo, with the FDA acronym in a blue square and the full name of the administration in blue text.
February 27, 2018
Alpha Omega Engineering Ltd. Maysana Mousa QA RA Manager Nazareth Industrial Park, Mount Precipice, St. 2015 Nazareth, 1612102 Il
Re: K172042
Trade/Device Name: NeuroNav System, NeuroSmart System, and NeuroNav Drive Regulation Number: 21 CFR 882.1330 Regulation Name: Depth Electrode Regulatory Class: Class II Product Code: GZL Dated: December 27, 2017 Received: January 11, 2018
Dear Maysana Mousa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Hoffmann -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172042
Device Name
NeuroNav System, NeuroSmart System, and NeuroNav Drive
Indications for Use (Describe)
The NeuroNav System / NeuroSmart System, incorporated the installed HaGuide software, is intended to be used in assisting neurosurgeons, in the operating functional neurosurgery, and to record from and stimulate brain motor and sensory neurons and to aid in the placement of depth electrode.
The NeuroNav Drive is intended to be used in assisting neurosurgeons, in the operating room during functional neurosurgery, to aid in the placement of depth electrodes.
| Type of Use (Select one or both, as applicable) | | | | |
|--------------------------------------------------|---------------------------------------------|--|--|--|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Alpha Omega. The logo consists of a stylized letter "A" in maroon color, with a gray bridge-like structure at the bottom. To the right of the "A" is the word "omega" in gray. Below the logo is the text "Defining Neuroscience Technology" in maroon color.
# 510(k) Summary
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990. The contents of the 510(k) summary have been provided in conformance with 21 CFR §807.92
### Submitter Information
Alpha Omega Engineering Ltd. Mount Precipice 2015 P.O. Box 2268 Nazareth 1612102, Israel Tel:+972-4-6563-327 Fax:+972-4-6574-075
### Submission contact person:
Mrs. Maysana Mousa Tel:+972-4-6563-327 Fax:+972-4-6574-075 Mobile: +972-54-9043303
### Device Classification
| Proprietary Device Name: | NeuroNav System, NeuroNav Drive and<br>NeuroSmart System | | | | |
|----------------------------|--------------------------------------------------------------------|--|--|--|--|
| Common name: | Intraoperative neurophysiological recording and stimulating device | | | | |
| Product Code: | GZL | | | | |
| Subsequent Product Code: | | | | | |
| Classification Name: | Depth Electrode | | | | |
| Classification Regulation: | 21 CFR §882.1330 | | | | |
| Regulatory Class: | II | | | | |
### Identification of Legally Marketed Predicate Devices
NeuroNav System and NeuroNav Drive device -K071697
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Alpha Omega. The logo features the letters "Alpha" in a stylized font, with a red line running through the letters. The word "omega" is in gray, and the tagline "Defining Neuroscience Technology" is below the logo. The logo is likely used to represent a company or organization that specializes in neuroscience technology.
#### Device Description 1.
Alpha Omega's NeuroNav System is an accurate electrodes navigation system to aid the neurosurgeon in placement of DBS electrodes. The system is designed to be used in the operating room during neurosurgery.
The NeuroNav Drive system is designed for compatibility with Alpha Omega's Microprobes (NeuroProbes K120098) or with other manufacturers' compatible electrodes, and is compatible with all the existing Stereotactic Frames and Frameless procedures.
NeuroNav system can record signals from the brain cells (Micro-Electrode Recording -MER) or stimulate the brain target zone during movement disorders neurosurgery procedure.
In addition, the NeuroNav system is planned to include additional sub model, will be traded as "NeuroSmart", this sub model is identical to the NeuroNav system, the subject device, and incorporated the installed HaGuide software, will be elaborated hereunder.
The installed HaGuide software is a user option tool, which is a real-time software solution designed to accurately detects the Sub Thalamic Nucleus region, its entrance and exit boundaries as well gives the user a stimulation location recommendation.
The HaGuide software robustly detects intra STN detection of Dorso lateral Oscillatory Region (DLOR) and Ventro Medical Non-Oscillatory Region (VMNR) boundary.
#### Intended Use of Device 2.
The NeuroNav system / NeuroSmart System, incorporated the installed HaGuide software, is intended to be used in assisting neurosurgeons, in the operating room during functional neurosurgery, and to record from and stimulate brain motor and sensory neurons and to aid in the placement of depth electrode.
The NeuroNav Drive is intended to be used in assisting neurosurgeons, in the operating room during functional neurosurgery, to aid in the placement of depth electrodes.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Alpha Omega, a company that defines neuroscience technology. The logo features the words "ALPHA" and "omega" in a stylized font, with a red line connecting the two words. The word "omega" is in gray, and the tagline "Defining Neuroscience Technology" is below the logo.
#### 3. Safety & Effectiveness
The NeuroNav system has been compared to the predicate device, NeuroNav device (K071697), in terms of intended use, indications for use, components, principles of operation, technological characteristics and safety features.
#### Intended Use Comparison .
| # | Comparison parameter | Subject device:<br>Modified NeuroNav System and<br>NeuroNav Drive | Predicate device:<br>NeuroNav System and<br>NeuroNav Drive | Substantial Equivalent discussion |
|---|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| 1 | Legally distribution clearance No. | Subject device | K071697 | |
| 2 | Owner | Alpha Omega Engineering Ltd. | Alpha Omega Engineering Ltd. | |
| 3 | Intended use and indications for use. | NeuroNav System<br>Alpha Omega's NeuroNav system, including the<br>NeuroNav Drive, is intended to<br>be used in assisting<br>neurosurgeons, in the operating<br>room during functional<br>neurosurgery, and to record | NeuroNav System<br>Alpha Omega's NeuroNav system, including the<br>NeuroNav Drive, is intended to<br>be used in assisting<br>neurosurgeons, in the operating<br>room during functional<br>neurosurgery, and to record | Similarities:<br>Same intended use and indications<br>Differences:<br>None |
Page 3 of 13
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Alpha Omega. The logo consists of the letters "Alpha" and "Omega" in a stylized font. The word "Alpha" is in a dark red color, while the word "Omega" is in gray. The "A" in Alpha is also represented by a symbol to the left of the word.
| # | Comparison<br>parameter | Subject device:<br>Modified NeuroNav System and<br>NeuroNav Drive | Predicate device:<br>NeuroNav System and<br>NeuroNav Drive | Substantial Equivalent discussion |
|---|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| | | from and stimulate brain motor<br>and sensory neurons and to aid<br>in the placement of depth<br>electrode.<br><br>NeuroNav Drive<br>Alpha Omega NeuroNav Drive<br>is intended to be used in<br>assisting neurosurgeons, in the<br>operating room during<br>functional neurosurgery, to aid<br>in the placement of depth<br>electrodes. | from and stimulate brain motor<br>and sensory neurons and to aid<br>in the placement of depth<br>electrode.<br><br>NeuroNav Drive<br>Alpha Omega NeuroNav Drive<br>is intended to be used in<br>assisting neurosurgeons, in the<br>operating room during<br>functional neurosurgery, to aid<br>in the placement of depth<br>electrodes. | |
| 4 | Device code and<br>regulation | Product Code: GZL<br>Regulation #: 21CFR882.1330 | Product Code: GZL<br>Regulation #: 21CFR882.1330 | Similarities:<br>Same product code and regulation<br><br>Differences:<br>None |
#### Performance Comparison .
| # | Comparison<br>parameter | Subject device:<br>Modified NeuroNav System<br>and NeuroNav Drive | Predicate device:<br>NeuroNav System and<br>NeuroNav Drive | Substantial Equivalent discussion |
|-----|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Legally | Subject device | K071697 | |
| # | Comparison<br>parameter | Subject device:<br>Modified NeuroNav System<br>and NeuroNav Drive | Predicate device:<br>NeuroNav System and<br>NeuroNav Drive | Substantial Equivalent discussion |
| | distribution<br>clearance No. | | | |
| 2 | Owner | Alpha Omega Engineering<br>Ltd. | Alpha Omega Engineering<br>Ltd. | |
| 3 | Target Area | Deep Brain | Deep Brain | Similarities:<br>Same target body target area<br>Differences:<br>None |
| 4 | Target population | Adults and children | Adults and children | Similarities:<br>Same target population<br>Differences:<br>None |
| 5 | Users | Neurosurgeon and<br>Neurosurgery staff | Neurosurgeon and<br>Neurosurgery staff | Similarities:<br>Same users<br>Differences:<br>None |
| 4 | Use environment | Operating Room | Operating Room | Similarities:<br>Same use environment<br>Differences: |
| # | Comparison<br>parameter | Subject device:<br>Modified NeuroNav System<br>and NeuroNav Drive | Predicate device:<br>NeuroNav System and<br>NeuroNav Drive | Substantial Equivalent discussion |
| 6 | Usability and Human<br>Factors | ISO 62366, IEC 60601-1-6<br>and FDA guidance # 1497 -<br>Incorporating Human Factors<br>in Risk Management dated<br>July 18 2000. | ISO 60601-1-6 and FDA<br>guidance #<br>1497 - Incorporating Human<br>Factors in Risk Management<br>dated July 18 2000. | None<br>Similarities:<br>Conformity to same or newer<br>Human Factors standards<br>Differences:<br>The modified device usability design<br>complies to a newer medical device<br>usability standard (ISO 62366) |
| 7 | AC Power<br>Supply | 100 - 240V AC<br>50/ 60 Hz | 100-120 V AC,220-240V AC<br>50/ 60 Hz | Similarities:<br>Same AC Power input<br>Differences:<br>None |
| 8 | Software<br>applications | GUI, Monitoring, display,<br>Audio, signal processing,<br>Recording, Stimulation | GUI, Monitoring, display,<br>Recording, Stimulation | Similarities:<br>Same application controlled by the<br>device's software<br>Differences:<br>New and improved applications<br>implemented by the software.<br>The differences do not raise concerns<br>in safety and effectiveness |
| # | Comparison<br>parameter | Subject device:<br>Modified NeuroNav System<br>and NeuroNav Drive | Predicate device:<br>NeuroNav System and<br>NeuroNav Drive | Substantial Equivalent discussion |
| 9. | Electrical Safety | IEC 60601-, 3rd Ed. +<br>Risk analysis (ISO 14971) | IEC 60601-1 +<br>Risk analysis (ISO 14971) | Similarities:<br>Conformity to the same or newer<br>safety standard<br>Differences:<br>The modified device was tested for<br>compliance to a newer (3rd edition)<br>medical device safety standard |
| 10. | Particular safety | Risk analysis (ISO 14971) | Risk analysis (ISO 14971) | Similarities:<br>Risk assessment performed<br>according to the same standard<br>Differences:<br>Non |
| 11 | Electromagnetic<br>Compatibility | IEC 60601-1-2, 3rd Ed. +<br>Risk analysis (ISO 14971) | IEC 60601-1-2 +<br>Risk analysis (ISO 14971) | Similarities:<br>Conformity to the same or newer<br>EMC standard<br>Differences:<br>The modified device was tested for<br>compliance to a newer (3rd edition)<br>medical device EMC standard |
| 12 | Sterility and<br>Sterilization standards | Sterility of the sterileable | Sterility of the sterileable | Similarities: |
| # | Comparison<br>parameter | Subject device:<br>Modified NeuroNav System<br>and NeuroNav Drive | Predicate device:<br>NeuroNav System and<br>NeuroNav Drive | Substantial Equivalent discussion |
| | | components is performed by<br>the hospital clinical staff and<br>was validated to comply with<br>the following standards: ISO<br>17665-1, ST77 | components is performed by<br>the hospital clinical staff and<br>was validated to comply with<br>the following standards: ISO<br>17665-1, ISO 11135, ISO<br>10993-7, ST77 | Conformity to the same or newer<br>sterility validation standards<br>Differences:<br>The modified device head-stage<br>electronic components were<br>validated for compliance with<br>different types of STERRAD<br>sterilization processes instead of<br>ETO sterilization process that is not<br>used for electronic components<br>sterilization. |
| 13 | Software development<br>standard | ISO 62304 | ISO 60601-1-4 | Similarities:<br>Conformity to the same or newer<br>medical device software standard<br>Differences:<br>The modified device software<br>development process complies to a<br>newer medical device life cycle<br>software development standard (ISO<br>62304) |
Page 4 of 13
Mount Precipice 2015 P.O. Box 2268 Alpha Omega Engineering Ltd. Nazareth 1612102, Israel Tel:+972-4-6563-327 Fax:+972-4-6574-075 info@alphaomega-eng.com www.alphaomega-eng.com
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Alpha Omega. The logo consists of the letters "Alpha" and "Omega" in a stylized font. The "Alpha" is in a dark red color, and the "Omega" is in gray. The "Alpha" is also stylized with a line running through it.
Page 5 of 13
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Alpha Omega. The word "Alpha" is written in a stylized font with a red line running through the letters. The word "Omega" is written in gray, with the "O" in the shape of the Greek letter omega. The logo is simple and modern, and the colors are eye-catching.
Page 6 of 13
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Alpha Omega. The logo has the letters "Alpha" in a dark red color, with a line going through the letters. The letters "omega" are in gray. The logo is simple and modern.
Page 7 of 13
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows a logo with the words "ALPHA omega" in a stylized font. The word "ALPHA" is in a dark red color and has a line running through it that resembles a heartbeat. The word "omega" is in gray and is positioned below the word "ALPHA". To the left of the words is a symbol that combines the Greek letters alpha and omega.
Page 8 of 13
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image is a logo for Alpha Omega. The logo has the word "ALPHA" in a dark red color, and the word "omega" in gray. Below the words is the phrase "Defining Neuroscience Technology" in black.
Based on the performance results provided in this submission (including test results) and the analysis of similarities and differences presented above, Alpha Omega Technologies Ltd. believes that the proposed device is substantially equivalent to the predicate device without raising new safety and/or effectiveness issues.
### Rational for Substantial Equivalency
The proposed modified NeuroNav system and NeuroNav Drive has been compared to the predicate NeuroNav system and NeuroNav Drive in terms of intended use, indications for use, components, principles of operation, technological characteristics and safety features.
Based on the analysis results provided in this submission, including test results and bench performance tests and the analysis of similarities and differences discussed, Alpha Omega Engineering Ltd. believes that the proposed device is substantial equivalent to the predicate device without raising new safety and/or effectiveness issues.
### Non Clinical and Clinical validation data
Verification and validation bench tests were performed, to demonstrate the safety and effectiveness of the proposed NeuroNav System.
The purposes of the design verification & validation processes were to verify the Device specifications and to verify the proper function of all device components and options.
After comparing the predicates device to the subject device, results show that with the above intended use, the device is equivalent in safety and effectiveness.
Therefore, the subject of this 510(k) notification, the NeuroNav system, did not require clinical studies to the device.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Alpha Omega, a company that defines neuroscience technology. The logo features the letters "ALPHA omega" in a stylized font, with a red line above the letters. Below the logo is the text "Defining Neuroscience Technology" and "Performance Tests".
### For non-clinical performance tests:
| Test | Test Method Summary | Results |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Software<br>Verification | This verification performed on<br>full system, the subject device<br>(SW & HW) including<br>accessories, and checked that<br>the design output meets the SW<br>design input. | NeuroNav SW has been tested<br>under a complete SW verification<br>plan traceable to NeuroNav SRS.<br>All samples passed the acceptance<br>criteria which determines the<br>effectiveness of NeuroNav System,<br>the subject device, with HaGuide<br>software. |
| System<br>Verification | This verification performed on<br>full system, the subject device<br>(SW & HW), and checked that<br>the design output meets the<br>system (HW & MECH) design<br>input | NeuroNav system has been verified<br>under a complete system<br>verification plan traceable to<br>NeuroNav system design input. All<br>samples passed the acceptance<br>criteria which determines the<br>effectiveness of NeuroNav System,<br>the subject device, with HaGuide<br>software. |
| Steam Sterilization<br>Validation | Mechanical accessories of the<br>subject device | Each test article was evaluated to a<br>sterility assurance level (SAL) of ≤<br>10-6 using the biological indicator<br>(BI) overkill method. Geobacillus<br>stearothermophilus, ATCC #7953,<br>was the indicator organism. All test<br>method acceptance criteria were<br>met.<br>In addition to the SAL validation,<br>dry times were validated using full<br>cycle parameters. |
| STERRAD NX<br>sterilization<br>validation | Electrical accessories of the<br>subject device | NeuroNav was validated to a<br>sterility assurance level (SAL) of ≤<br>10-6 using the biological indicator<br>(BI) overkill method. Geobacillus<br>stearothermophilus, ATCC #7953,<br>was the indicator organism.<br>Results validated the individual test |
| | | articles for the STERRAD® NX |
| | | Standard cycle. |
| STERRAD 100NX<br>sterilization<br>validation | Electrical accessories of the<br>subject device | Results from testing validate that<br>the Alpha Omega Engineering, Ltd.<br>Neuro Omega Tray is able to<br>achieve a 10-6 SAL in a<br>STERRAD® 100NX Standard<br>sterilization process. |
## Page 10 of 13
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Alpha Omega. The logo consists of the letters 'A' and 'O' stacked on top of each other, with the word 'ALPHA' written in red above the letters and the word 'omega' written in gray below the letters. The 'A' and 'O' are also in red and gray, respectively.
Page 11 of 13
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image is a logo for Alpha Omega. The logo features the letters "Alpha" in a stylized font, with a red line running through the letters. The letters "Omega" are in gray, and the tagline "Defining Neuroscience Technology" is below the logo. The logo is likely used to represent a company or organization that specializes in neuroscience technology.
### For clinical performance test:
HaGuide Clinical Study:
#### ● Test Method Summary:
A Retrospective, Multi-centre, Clinical Validation of HaGuide Software module accuracy in mapping sub-thalamic nucleus (STN) boundaries in Parkinson's disease patients who underwent Deep Brain Stimulation (DBS) procedure
#### Level of evidence: ●
### Single-arm study with Objective Performance Criteria
#### Location of Study: ●
Both in United States and Outside Unitest States
#### ● Primary Effectiveness Endpoint:
Percent agreement between HaGuide software measurements and expert Neurosurgeon/Electrophysiologist measurements, for STN entry point vs. exit point depth (in mm).
| Stage | Investigational Device<br>Arm total | Control Arm<br>Total | Total |
|-------------------------------------------|-------------------------------------|----------------------|-------|
| Enrollment | 81 | - | 81 |
| Treatment | 81 | - | 81 |
| Primary Effectivness<br>Endpoint Analysis | 81 | - | 81 |
#### Patient Accountability: ●
#### The study met the primary endpoint: ●
The reference against which the accuracy of HaGuide mapping was compared is the gold standard being the qualified user (Neurosurgeon/Electrophysiologist) concluding about the location of the entry and exit points of the sub-thalamic nucleus (STN) from the visual and audio recording patterns.
#### . Results:
In total 81 patients were enrolled. In total 105 surgeries, as 213 electrodes were evaluated for percent agreement between HaGuide software measurements and expert Neurosurgeon/Electrophysiologist measurements about the location of the entry and exit points of the sub-thalamic nucleus (STN).
Page 12 of 13
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the logo for Alpha Omega. The logo consists of the letters "ALPHA" in a dark red color, with a line running through the letters. Below the word "ALPHA" is the word "omega" in gray.
Percent agreement between Haguide and Neurosurgeon/Electrophysiologist expert is 90.34% (95% CI, 85.27, 93.85 %)
Correlation between the HaGuide measurements and and expert Neurosurgeon/Electrophysiologist measurements in depth of STN entry site and exit site and was found very strong (coefficient = 0.9492 and 0.9317 respectively and both lower 95% Confidence Limit >0.91).
Bland-Altman analysis of average against the difference between Haguide and expert measurements for the location of STN entry was 0.18 mm (95% CI, -0.9, 1.2 mm) and exit points was -0.15 mm (95% CI, -1.25, 1 mm).
- . No adverse events and complications observed in the study
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.