K170959 · Dixi Medical · GZL · Nov 24, 2017 · Neurology
Device Facts
Record ID
K170959
Device Name
DIXI Medical Microdeep Depth Electrode
Applicant
Dixi Medical
Product Code
GZL · Neurology
Decision Date
Nov 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1330
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DIXI Medical Microdeep Depth Electrodes are intended for temporary (<30 day) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain.
Device Story
Microdeep Depth Electrode is a smooth, 0.8 mm diameter electrode with a hemispherical tip and multiple platinum/iridium contacts. Used by physicians for temporary (<30 day) subsurface brain signal recording, monitoring, and stimulation. Device connects to external user-supplied equipment to assist in brain mapping and defining epileptogenic foci locations. Operates via direct contact with brain tissue to capture electrical signals or deliver stimulation. Benefits include precise localization of seizure-generating areas for clinical decision-making.
Clinical Evidence
No clinical testing performed. Evidence consists of bench testing (mechanical, electrical, functional, biocompatibility, packaging, shelf life, sterilization) and a human cadaver study. Cadaver study compared implantation and tracking accuracy against a predicate; mean difference in implantation error was 2.87mm (within 3.6mm non-inferiority margin). Tracking accuracy mean error was 0.12mm +/- 0.47mm.
Technological Characteristics
0.8 mm diameter electrode with hemispherical tip; platinum/iridium contacts (2 mm length); total exploration length 16 mm to 80.5 mm. Passive device for electrical signal recording/stimulation. EO sterilization. No internal software or active electronics.
Indications for Use
Indicated for temporary (<30 day) subsurface brain electrical signal recording, monitoring, and stimulation in patients requiring brain mapping or localization of epileptogenic foci.
Regulatory Classification
Identification
A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 24, 2017
DIXI Medical % Valerie Defiesta-Ng Senior Director, Regulatory Affairs Experien Group, LLC 224 Airport Parkway, Suite 250 San Jose, California 95110
Re: K170959
Trade/Device Name: DIXI Medical Microdeep Depth Electrode Regulation Number: 21 CFR 882.1330 Regulation Name: Depth Electrode Regulatory Class: Class II Product Code: GZL Dated: October 25, 2017 Received: October 26, 2017
Dear Ms. Defiesta-Ng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the name "Carlos L. Pena -S" in black font. To the right of the name is the acronym "FDA" in a light blue font. The acronym is partially transparent, making it difficult to read. The image is simple and contains only text.
Carlos L. Peña. PhD. MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170959
Device Name DIXI Medical Microdeep Depth Electrode
Indications for Use (Describe)
The DIXI Medical Microdeep Depth Electrodes are intended for temporary (<30 day) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| <div> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Notification
#### I. GENERAL INFORMATION [807.92(a)(1)]
Applicant:
DIXI Medical 2A Route de Pouligney 25640 Chaudefontaine France Phone: +33 3 81 88 98 90 +33 3 81 88 98 99 Fax:
## Contact Person:
Valerie Defiesta-Ng Regulatory Consultant for DIXI Medical Experien Group, LLC 224 Airport Parkway, Suite 250 San Jose, CA 95110 U.S.A. Phone: 408-400-0856 Fax: 408-400-0865 Email: valerie(@experiengroup.com
## Date Prepared: October 25, 2017
#### II. DEVICE INFORMATION [807.92(a)(2)]
Trade/Proprietary Name: DIXI Medical Microdeep Depth Electrode
## Generic/Common Name: Electrode, depth
Classification Name: Depth electrode (21 CFR§882.1330)
Regulatory Class: Class II
Product Code: GZL
{4}------------------------------------------------
#### PREDICATE DEVICES [807.92(a)(3)] III.
- PMT® Platinum Depthalon® Depth Electrodes (K151790) .
- . AD-TECH Medical Instrumentation Corporation Depth Electrodes (K053358)
- Radionics Recording and Stimulating Electrode (K012850) ●
#### DEVICE DESCRIPTION [807.92(a)(4)] IV.
The DIXI Medical Microdeep Depth Electrode ("Microdeep Depth Electrode" or "Electrode") is a smooth electrode that is 0.8 mm in diameter with a rounded (hemispherical) tip. The electrode has multiple platinum/iridium contacts that are 2 mm long, with a total exploration length varying from 16 mm to 80.5 mm according to the electrode reference. Microdeep Depth Electrodes are intended for temporary use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain. Microdeep Depth Electrodes are used under the supervision of a physician (physicians in the areas of biopotential recording, monitoring and stimulation/response studies understand the use of depth electrodes) and are intended to connect to the user's recording, monitoring and stimulation equipment. This information is then used for defining the location of epileptogenic foci and brain mapping.
#### V. INDICATIONS FOR USE [807.92(a)(5)]
The Indication for Use statement for the Microdeep Depth Electrodes is as follows:
The DIXI Medical Microdeep Depth Electrodes are intended for temporary (<30 day) use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain.
{5}------------------------------------------------
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(a)(6)]
The Microdeep Depth Electrodes are substantially equivalent to the primary predicate device, the PMT Platinum Depthalon Depth Electrodes (K151790) as they have the same intended use in the same patient population, utilize similar performance specifications and have comparable technological features to achieve the same mechanism of action: therefore, the Microdeep Depth Electrodes do not raise any different issues of safety or effectiveness. Additional predicate devices have been selected for comparison to technological characteristics, the AD-TECH Medical Instrumentation Corporation Depth Electrodes (K053358) and the Radionics Recording and Stimulating Electrode (K012850).
#### VII. PERFORMANCE DATA [807.92(b)]
Performance testing was performed to assure safety and effectiveness of the Microdeep Depth Electrodes. All necessary bench testing was conducted on the Microdeep Depth Electrodes to ensure conformance to design specifications and to support a determination of substantial equivalence to the predicate devices.
## [807.92(b)(1)]
The nonclinical, bench testing performed included:
- Verification testing (mechanical, electrical and functional testing); ●
- Biocompatibility testing; ●
- . Packaging, shelf life and transit testing; and
- . EO Sterilization Validation.
A human cadaver study was conducted by DIXI Medical. The purpose of this human cadaver study was to demonstrate simulated use of the DIXI Medical Microdeep Depth Electrode and its implantation and tracking, when implanted according to its instructions for use (IFU). The implantation accuracy and tracking accuracy of the Microdeep Depth Electrode were compared to the implantation accuracy and tracking accuracy of a predicate device (Integra Depth Electrode, K012850). Cadaver heads were implanted with equal number of electrodes. Results are summarized herein.
## Implantation Accuracy
The study compared the implantation accuracy, as measured by the final implant location, of the Microdeep Depth Electrodes to a predicate at a variety of Supratentorial targets. The mean difference in errors of the predicate and the Microdeep Depth Electrode was compared with a non-inferiority test (one side paired two-sample t-test). The data demonstrates a mean difference in errors of 2.87mm which is less than the non-inferiority margin of 3.6mm, therefore the implantation accuracy endpoint has been met.
## Tracking Accuracy
The study assessed the ability of the DIXI Medical Microdeep Depth Electrode to travel within the track previously created by the DIXI Medical stylet. In summary
{6}------------------------------------------------
there are three targets with a 1mm difference and 1 target with a 3mm difference, for a mean and standard deviation of 0.12mm +/- 0.47mm. The data demonstrates the ability of the DIXI Medical Microdeep Depth Electrode to travel within the track previously created by the DIXI Medical stylet.
## [807.92(b)(2)]
No clinical testing was performed in support of this premarket notification.
The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the Microdeep Depth Electrodes meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the Microdeep Depth Electrodes do not raise different questions of safety or effectiveness for use as depth electrodes when compared to the predicate devices.
#### CONCLUSIONS [807.92(b)(3)] VIII.
The Microdeep Depth Electrodes do not raise any different questions of safety or effectiveness, as both the Microdeep Depth Electrode and the predicate devices have the same intended use, have been historically cleared for use in the same procedures, and utilize similar performance specifications and comparable technological features to achieve the same mechanism of action: therefore, the Microdeep Depth Electrode is substantially equivalent to the predicate devices.
The indications for use for the predicate devices are substantially equivalent to the proposed indications for use for the Microdeep Depth Electrode. As demonstrated by successful performance testing, any differences in the technological characteristics between the proposed device and the predicate devices do not raise any different questions of safety or effectiveness. Thus, the Microdeep Depth Electrode is substantially equivalent to the predicate devices.
As such, the Microdeep Depth Electrode is substantially equivalent to the predicate devices, namely the PMT® Platinum Depthalon® Depth Electrode (K151790), as well as the AD-TECH Medical Instrumentation Corporation Depth Electrodes (K053358) and Radionics Recording and Stimulating Electrode (K012850).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.