LEADPOINT FOCUS WORKSTSTION, LEADPOINT FOCUS NOTEBOOK SYSTEM

K140680 · Alpine Biomed Aps · GZL · Dec 12, 2014 · Neurology

Device Facts

Record IDK140680
Device NameLEADPOINT FOCUS WORKSTSTION, LEADPOINT FOCUS NOTEBOOK SYSTEM
ApplicantAlpine Biomed Aps
Product CodeGZL · Neurology
Decision DateDec 12, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1330
Device ClassClass 2

Indications for Use

The Leadpoint Focus microelectrode targeting system is intended to assist in neurosurgical procedures where recording of neuronal activity and stimulation of brain neurons will aid in the placement of depth electrodes. The Leadpoint Focus microelectrode targeting system is intended to be used in the operating room by a neurosurgeon, neurologist, or clinical neurophysiologist.

Device Story

Leadpoint Focus is a microelectrode recording (MER) system used during stereotactic neurosurgery. It captures neuronal activity via microelectrodes (not included) and provides macro stimulation to aid target localization. System components include software, hardware, macro stimulator, MER amplifier, and sterile interconnection cables. Operated by neurosurgeons, neurologists, or clinical neurophysiologists in the operating room. The system processes electrical signals from the brain, displaying them on a digital monitor and providing audio feedback via loudspeakers. Clinicians use the visual and auditory output to guide the precise placement of depth electrodes. The device facilitates intraoperative target localization, potentially improving surgical accuracy and patient outcomes.

Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through software testing per FDA guidance, electrical safety testing (IEC 60601-1), and electromagnetic compatibility testing (EN/IEC 60601-1-2, 60601-2-40, 61000-3-2, 61000-3-3).

Technological Characteristics

PC-based workstation or portable notebook system. 6-channel isolated amplifier; 24-bit ADC at 48kHz. Constant current macro stimulator (0-50V, 10mA max). Connectivity via USB. Safety isolation Type BF. Software-driven interface with digital display and audio monitoring. Sterilization for consumable cables.

Indications for Use

Indicated for neurosurgical procedures requiring recording of neuronal activity and stimulation of brain neurons to assist in depth electrode placement. Intended for use by neurosurgeons, neurologists, or clinical neurophysiologists in the operating room.

Regulatory Classification

Identification

A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to suggest community and support. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 12, 2014 Alpine Biomed ApS Seamus O'Connor Tonsbakken 16-18 DK-2740 Skovlunde, Denmark Re: K140680 Trade/Device Name: Leadpoint Focus Regulation Number: 21 CFR 882.1330 Regulation Name: Depth electrode Regulatory Class: Class II Product Code: GZL Dated: November 10, 2014 Received: November 12, 2014 Dear Mr. O'Connor: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## Felipe Aquel -S Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K140680 Device Name Leadpoint Focus Indications for Use (Describe) The Leadpoint microelectrode targeting system is intended to assist in neurosurgical procedures where recording of neuronal activity and stimulation of brain neurons will aid in the placement of depth electrodes. The Leadpoint microelectrode targeting system is intended to be used in the operating room by a neurosurgeon, neurologist, or clinical neurophysiologist. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 5 510(k) Summary Company Confidential Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde Tel: +45 44 57 90 00 12/10/2014 5 510(k) Summary - 1 of 8 {4}------------------------------------------------ Image /page/4/Picture/1 description: The image is a logo for Alpine BioMed. The logo is divided into two sections, with the left section containing a green and blue circular design with a white line running through it. The right section contains the words "ALPINE" in blue, "bioMed" in green, and a green line separating the two words. Below the logo, the words "a division of natus" are written in a smaller font. ## 510(k) Summary 510(k) Number: K140680 O Alpine Biomed ApS Tonsbakken 16-18 DK-2740 Skovlunde Denmark Phone +45 44 57 90 00 +45 44 57 90 10 10 December 2014 #### Submitter's name: Alpine Biomed ApS Tonsbakken 16-18, DK-2740 Skovlunde, Denmark Tel.: +45 4457 9000 #### Contact Person: Seamus O'Connor, Email seamus.oconnor@natus.com Tel .: +353-(0)87-6330532 Trade name: Leadpoint Focus Device Class Class II Classification name: Depth electrode. #### Product code and Regulation: 21CFR 882.1330 GZL #### Predicate devices: The Leadpoint Focus is substantially equivalent to the predicate device K071364 Guideline 4000 from FHC, Inc. #### Company Confidential Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde Tel: +45 44 57 90 00 12/10/2014 5 510(k) Summary - 2 of 8 {5}------------------------------------------------ Image /page/5/Picture/20 description: The image is a logo for Alpine BioMed. The logo is in a blue box with a green and blue symbol on the left. To the right of the symbol is the text "ALPINE" in blue, with a green line underneath it. Below that is the text "bioMed" in green and blue. Below the logo is the text "a division of natus." Alpine Biomed ApS DK-2740 Skovlunde +45 44 57 90 10 #### Intended use / Indications for use: The Leadpoint Focus microelectrode targeting system is intended to assist in neurosurgical procedures where recording of neuronal activity and stimulation of brain neurons will aid in the placement of depth electrodes. The Leadpoint Focus microelectrode targeting system is intended to be used in the operating room by a neurosurgeon, neurologist, or clinical neurophysiologist. #### Device Description: Intraoperative microelectrode recording (MER) has been introduced to improve the target localization during stereotactic surgery. The Leadpoint Focus is a MER system which consists of the Leadpoint Focus Software, Focus Hardware, a Macro Stimulator, a MER amplifier and a sterile interconnection cable (MER Cable) that connects the system to a Microelectrode. The Microelectrode is supplied by another manufacturer and is not part of the Leadpoint Focus system. The recognized upper limit for safe stimulation is 30μC/cm2 per phase. #### Microelectrode Connection A MER Cable is an electrode cable composed of strands of insulated electrical conductors intended to connect a Microelectrode from a patient to the Leadpoint Focus system. The MER Cable is an optional consumable part. The MER cable is for single use only and delivered sterile. #### Company Confidential Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde Tel: +45 44 57 90 00 5 510(k) Summary - 3 of 8 {6}------------------------------------------------ Image /page/6/Picture/1 description: The image is a logo for Alpine BioMed, a division of Natus. The logo is blue and green, and it features a stylized mountain range with a heartbeat line in the negative space. The text "ALPINE bioMed" is written in blue and green, with "bioMed" in a smaller font. Below the logo, the text "a division of natus." is written in a smaller font. O Alpine Biomed ApS Tonsbakken 16-18 DK-2740 Skovlunde Denmark □ Phone +45 44 57 90 00 +45 44 57 90 10 Fax #### Summary of Technological Characteristics comparison | Characteristic | Alpine Biomed ApS<br>Leadpoint Focus<br>(This submission) | FHC, Inc.<br>Guideline 4000<br>(K071364) | Discussion of<br>Differences | |-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------| | Indications for Use | | | | | 1.1 Intended Use /<br>Indications for Use | The Leadpoint Focus<br>microelectrode targeting system is<br>intended to assist in neurosurgical<br>procedures where recording of<br>neuronal activity and stimulation of<br>brain neurons will aid in the<br>placement of depth electrodes.<br>The Leadpoint Focus<br>microelectrode targeting system is<br>intended to be used in the operating<br>room by a neurosurgeon,<br>neurologist, or clinical<br>neurophysiologist. | The microTargeting® Guideline<br>4000 is intended to be used by a<br>neurosurgeon, neurologist or clinical<br>neurophysiologist to accurately<br>position depth electrodes during<br>functional neurosurgical procedures.<br>The Guideline 4000 is intended to<br>be used to accurately position depth<br>electrodes during functional<br>neurosurgical procedures. | Similar | | 1.2 Clinical Users | The Leadpoint microelectrode<br>targeting system is intended to be<br>used by a neurosurgeon,<br>neurologist, or clinical<br>neurophysiologist. | The Guideline 4000 is intended to<br>be used by a neurosurgeon,<br>neurologist or clinical<br>neurophysiologist. | Similar | | 1.3 Clinical<br>Environment | The Leadpoint microelectrode<br>targeting system is intended to be<br>used for neurosurgical procedures<br>in the operating room. | The Guideline 4000 is intended to<br>be used during functional<br>neurosurgical procedures. | Similar | ### Company Confidential Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde Tel: +45 44 57 90 00 {7}------------------------------------------------ Image /page/7/Picture/1 description: The image is a logo for Alpine BioMed, a division of Natus. The logo features a blue square with a green and blue design inside. The words "ALPINE" and "bioMed" are written in blue, with a green line separating the two words. The words "a division of natus" are written in black below the square. O Alpine Biomed ApS Tonsbakken 16-18 DK-2740 Skovlunde Denmark □ Phone +45 44 57 90 00 +45 44 57 90 10 Fax | General Design | | | | |----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------| | 2.1 General systems<br>approach | Standard PC based equipment with<br>dedicated hardware peripherals /<br>components. | Standard PC based equipment with<br>dedicated hardware peripherals /<br>components. | Identical | | 2.2 Configurations | Workstation on cart<br>Portable notebook PC based<br>system | Workstation on cart: Guideline<br>4000TM<br>Portable notebook PC based<br>system: Guideline 4000 LP+TM | Identical | | 2.3 User input device | Window mouse/keyboard driven<br>graphic interface with dedicated<br>control panel. | Window mouse/keyboard driven<br>graphic interface with dedicated<br>control panel. | Identical | | 2.4 User output<br>device | Digital color display, loudspeakers<br>and commercial printers. | Digital color display, loudspeakers<br>and commercial printers. | Identical, both<br>devices uses dual<br>loudspeakers for high<br>fidelity audio<br>monitoring. | | 2.5 Use of standard<br>software platform<br>(Operating System) | Microsoft Windows. | Microsoft Windows. | Identical | | 2.6 System<br>components | Main Unit including power supply,<br>speakers and control panel<br>Amplifier<br>Macro Stimulator<br>Additional for notebook based<br>system: Notebook PC including<br>monitor, keyboard, and mouse<br>Additional for Workstation: PC,<br>monitor, keyboard, mouse, cart and<br>isolating transformer | Main Unit including power supply,<br>speakers and control panel<br>Amplifier<br>Macro Stimulator<br>Notebook PC including monitor,<br>keyboard, and mouse.<br>Additional for Workstation: PC,<br>monitor, keyboard, mouse, cart and<br>isolating transformer. | Similar | | 2.7 System –<br>computer interface | USB | USB | Identical | | 2.8 System power<br>supply | Mains (100-120V~50/60Hz or 200-<br>240V~50/60Hz) through an isolating<br>transformer. | 100/120/220/240VAC~50/60Hz.<br>400VA | Similar | | 2.9 Display: Size /<br>Resolution | Workstation: 22" / 1680 x 1050<br>Notebook based system: 15.6" /<br>1920 x 1080 | Notebook based system: 17" /<br>UXGA (1600 x 1200) | Similar | | 2.10 Safety Isolation | Type BF | Type BF | Identical | ## Company Confidential Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde Tel: +45 44 57 90 00 5 510(k) Summary - 5 of 8 {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the logo for Alpine BioMed, which is enclosed in a blue box. The logo includes a green and blue symbol on the left, followed by the text "ALPINE" in blue, a green line, and "bioMed" in blue. Below the logo, the text "a division of natus." is written in a smaller font. O Alpine Biomed ApS Tonsbakken 16-18 DK-2740 Skovlunde Denmark □ Phone +45 44 57 90 00 +45 44 57 90 10 Fax | Microeletrode Recording | | | | |---------------------------|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 3.1 Patient inputs | 6 channel amplifier, isolated<br>All active channels continuously displayed. | 2 channel amplifier, isolated.<br>Expandable recording channels to a maximum of ten.<br>All active channels continuously displayed. | Standard single sided MER procedures requires simultaneous recording from 1-5 channels.<br>This can be done with both systems. | | 3.2 Bilateral procedure | Supported with up to 2 x 3 channels | Supported with up to 2 x 5 channels. | Standard MER recordings for bilateral procedures are performed with 2 x 1 channels.<br>This can be performed with both systems | | 3.5 Input Impedance | >1000Mohm//30pF with electrode cable capacitance reduction | $10^{13}\Omega$ // 0.2 pF AC, 200M $\Omega$ DC Differential<br>$10^{13}\Omega$ // 7.0 pF AC, 200M $\Omega$ DC Common Mode | Both within the range required for MER. Leadpoint Focus through use of electrode capacitance reduction and the predicate device through unusual high impedances | | 3.6 Filters | High Pass: 0.01 – 3000 Hz (16 steps)<br>Low Pass: 20 Hz - 13 kHz (12 Steps) | High Pass: Analog: 0 or 240 Hz.<br>Digital: 0 – 1000 Hz<br>Low Pass: Analog: 20 kHz. Digital: 3 - 20 kHz<br>The digital filters are continually adjustable. | Both within the range required for MER. Both systems secure minimal noise: Leadpoint Focus through many steps in the filters and the predicate device via continually adjustable filters. | | 3.7 Connection Types | 1.5mm Touch Proof / DIN socket / Micro D-sub | 5 pin DIN socket | Both within the range required for MER. | | 3.8 Signal acquisition | 24 bit analog to digital conversion at 48kHz sample rate | 16 bit analog to digital conversion at 48kHz sample rate | Both within the range required for MER. | | 3.9 Display | Up to six analysis windows. | Up to five analysis windows. | Both within the range required for MER. | | 3.10 Impedance test | Integrated 220 Hz sine wave, max. 1V / max. 0.1μΑ | Adjustable 100 Hz - 5 kHz sine wave. | Both within the range required for MER. | | 3.11 Impedance test range | 10 ΜΩ | Auto ranging 100 kΩ, 1 ΜΩ or 10 ΜΩ. | Both within the range required for MER. | | 3.12 Epoch Recording | Max. 80 s per Epoch | 2, 10 or 30 sec | Both within the range required for MER. | | 3.13 Audio Modes | Raw or clipped sound | Raw or clipped sound | Identical | | 3.14 Stereotactic systems | Most common | Most common | Identical | ### Company Confidential Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde Tel: +45 44 57 90 00 5 510(k) Summary - 6 of 8 {9}------------------------------------------------ Image /page/9/Picture/1 description: The image is a logo for Alpine BioMed. The logo features a blue square with a green circle inside. Inside the green circle is a white line that resembles a heartbeat. To the right of the square is the text "ALPINE bioMed" in blue and green. Below the logo is the text "a division of natus". Alpine Biomed ApS Tonsbakken 16-18 DK-2740 Skovlunde Denmark Phone +45 44 57 90 00 +45 44 57 90 10 | Macro Stimulation | | | | |--------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|---------------------------------------------------------------------------| | 4.1 Type | Constant Current (CC) | Constant Current (CC) | Identical | | 4.2 Max output intensity | 10 mA | 10 mA | Identical | | 4.3 Voltage Range | 0-50 V | 0-50 V | Identical | | 4.4 Intensity resolution | Fine setting: 0.02 mA<br>Coarse setting: 0.1 mA | 0.1 mA | Both within the range required for MER | | 4.5 Stimulus Duration | 20 µs – 1 ms (9 steps) | 50 µs – 1 ms (8 steps) | Both within the range of required stimulation frequencies (typical 50 µs) | | 4.6 Repetition Rate | 0.1 – 200 Hz (19 steps) | 10 - 300Hz (12 steps) | The added range from 0.1 to 10(0) Hz does not add any new risks. | | 4.7 Stimulus Polarity | Monophasic or biphasic stimulation | Monophasic or biphasic | Identical to the Leadpoint Focus system | | 4.8 Remote control | Yes, handheld device with controls for stimulation On/Off and intensity | Yes, handheld device with controls for stimulation On/Off and intensity | Identical | #### Performance data. #### Software Test. The Leadpoint Focus software was tested in accordance with: FDA guidance: The content of premarket submissions for software contained in medical devices, 11 May 05; FDA quidance: Off-the-shelf software use in medical devices, 09 Sep 99. #### Electrical Safety and Electromaqnetic Compatibility (EMC) Test. The Leadpoint Focus was tested for electrical safety. Test results demonstrated that the Leadpoint Focus complies with: IEC 60601-1:1988 + A1:1991 + A2:1995. Electromagnetic Compatibility (EMC) testing was conducted on the Leadpoint Focus according to the applicable standard. Test results indicated that the system complies with the following: EN/(IEC) 60601-1-2:2007 (3rd edition) EN/(IEC) 60601-2-40:1998, section 5 EN/(IEC) 61000-3-2:2006+A1+A2 EN/(IEC) 61000-3-3:2008 #### Company Confidential Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde 12/10/2014 Tel: +45 44 57 90 00 5 510(k) Summary - 7 of 8 {10}------------------------------------------------ Image /page/10/Picture/1 description: The image is a logo for Alpine BioMed, a division of Natus. The logo features a blue square with a green circle inside, and a white line that resembles a heartbeat. The words "ALPINE" and "bioMed" are written in blue and green, respectively, above the words "a division of natus" in black. The logo is simple and professional, and it conveys a sense of trust and reliability. O Alpine Biomed ApS Tonsbakken 16-18 DK-2740 Skovlunde Denmark O Phone +45 44 57 90 00 +45 44 57 90 10 Fax ### Conclusion. The comparison to the predicate devices demonstrates that the Leadpoint Focus system is safe and effective for its intended use and is substantially equivalent to the predicate devices #### Company Confidential Alpine Biomed Aps, Tonsbakken 16-18 DK-2740 Skovlunde Tel: +45 44 57 90 00 12/10/2014 5 510(k) Summary - 8 of 8
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...