WISE Cortical Strip (WCS)

K221123 · Wise S.R.L. · GYC · Nov 22, 2022 · Neurology

Device Facts

Record IDK221123
Device NameWISE Cortical Strip (WCS)
ApplicantWise S.R.L.
Product CodeGYC · Neurology
Decision DateNov 22, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1310
Device ClassClass 2

Indications for Use

The WISE Cortical Strip is intended for intraoperative (≤24 hours) use with recording, monitoring and stimulation equipment for the recording, monitoring, and stimulation of electrical signals on the surface of the brain. The WCS is indicated as an aid to IntraOperative Neurophysiological Monitoring (IONM) during brain surgeries.

Device Story

The WISE Cortical Strip (WCS) is a sterile, single-use cortical electrode strip featuring 4 platinum electrodes on a flexible silicone substrate; manufactured via proprietary Supersonic Cluster Beam Implantation (SCBI) to enhance conformability to the brain surface. The device is used intraoperatively (≤24 hours) by neurosurgeons or IONM technicians during brain surgery. It connects to external, FDA-cleared IONM recording/stimulation equipment via a 4-channel cable. The device captures electrical signals from the brain surface and delivers stimulation pulses; these signals are processed by the external IONM system to assist in real-time neurophysiological monitoring. The WCS provides a flexible interface for signal acquisition, aiding clinicians in monitoring brain activity during surgical procedures.

Clinical Evidence

Bench testing only. Testing included performance verification (dielectric strength, impedance ≤ 10 kΩ @ 10 Hz, insulation resistance, compatibility with EEG/ECOG recorders and neurostimulators), biocompatibility (ISO 10993-1: cytotoxicity, sensitization, intracutaneous reactivity, pyrogenicity, acute systemic toxicity, indirect hemolysis), and shelf-life/sterilization validation (18-month accelerated aging, SAL 10⁻⁶). Testing confirmed the device maintains integrity and performance under worst-case stimulation conditions and environmental stress.

Technological Characteristics

4-electrode cortical strip; platinum electrodes; silicone substrate; SCBI manufacturing process; 2.3 mm electrode diameter; 10 mm spacing; 1.5 mm diameter cable; EO sterilization (ANSI/AAMI/ISO 11135:2014); 18-month shelf life; connectivity via external connecting cable to IONM equipment; max stimulation charge density < 30 µC/cm².

Indications for Use

Indicated for intraoperative (≤24 hours) recording, monitoring, and stimulation of electrical signals on the brain surface as an aid to IntraOperative Neurophysiological Monitoring (IONM) during brain surgeries.

Regulatory Classification

Identification

A cortical electrode is an electrode which is temporarily placed on the surface of the brain for stimulating the brain or recording the brain's electrical activity.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ November 22, 2022 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. WISE S.r.l. % Terri Bogucki Regulatory Consultant Decus Biomedical 2342 Shattuck Ave #333 Berkeley, California 94704 Re: K221123 Trade/Device Name: WISE Cortical Strip (WCS) Regulation Number: 21 CFR 882.1310 Regulation Name: Cortical Electrode Regulatory Class: Class II Product Code: GYC Dated: October 24, 2022 Received: October 26, 2022 Dear Terri Bogucki: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, S for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Patrick Antkowiak - Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K221123 Device Name WISE Cortical Strip Indications for Use (Describe) The WISE Cortical Strip is intended for intraoperative (≤24 hours) use with recording, monitoring and stimulation equipment for the recording, monitoring, and stimulation of electrical signals on the surface of the brain. The WCS is indicated as an aid to IntraOperative Neurophysiological Monitoring (IONM) during brain surgeries. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------| | <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | Submitter's Name: | WISE S.r.l. | |--------------------------|-------------------------------------------------------| | Address: | Via Michelangelo Buonarroti, 38 | | | 20093 Cologno Monzese (MI) - Italy | | Contact Person: | Terese Bogucki | | Title: | Regulatory Consultant | | Telephone Number: | 650-488-7799 | | Fax Number: | 650-227-2264 | | Email: | terri@decusbiomedical.com | | Date Summary Prepared: | 10 November 2022 | | Device Proprietary Name: | WISE Cortical Strip | | Model Number: | WCS04S10A-00-001 | | Common Name: | Cortical electrode | | Regulation Number: | 21 CFR 882.1310 | | Product Code: | GYC | | Device Class: | II | | Predicate Devices | Manufacturer: Ad-Tech Medical Instrument | | | Corporation | | | Address: 400 West Oakview Parkway | | | Oak Creek, Wisconsin 53154 | | | Regulation Number: 21 CFR 882.1310 | | | Regulation Name: Cortical electrode | | | Device Class: Class II | | | Product Code: GYC | | | Primary Predicate | | | Trade name: Ad-Tech Intraoperative Strip Electrode | | | 510(k) Number: K191186 | | | 510(k) Clearance Date: 15 January 2020 | | | Secondary Predicate | | | Trade name: Ad-Tech Intraoperative Subdural Electrode | | | 510(k) Number: K053363 | {4}------------------------------------------------ #### l Description of the Device The WISE Cortical Strip is medical device composed of: - A strip containing 4 electrodes that is positioned during surgery on the . exposed surface of the brain - . A cable composed of 4 conductive channels intended to transfer electrical signals to and from a commercial connecting cable. The device is intended to be used only for intraoperative monitoring during brain surgery and is not implanted. The WCS is a sterile, single-use device, individually packaged with a shelf life of 18 months. Manufacture of the device uses a WISE proprietary patented technology called "Supersonic Cluster Beam Implantation" (SCBI). This technology produces flexible conductors by enabling thin conducting metal layers to be deposited on a silicone substrate. This enables the strip to be flexible and conform to the brain's surface. The WCS is intended to be used with FDA-cleared IONM equipment and a reusable connecting cable. The device does not include (as an accessory or component) these separate devices. #### 2 Indications for Use Both the WISE Cortical Strip (subject device) and Ad-Tech Intraoperative Strip Electrode (predicate device) are designed to be used intraoperatively for the recording, monitoring and stimulation of electrical signals on the surface of the brain. The Indications for Use for the subject and predicate devices are: {5}------------------------------------------------ | WISE Cortical Strip<br>Subject Device | AD-TECH Subdural<br>Electrodes Predicate<br>Device, K191186 | AD-TECH Subdural<br>Electrodes Predicate<br>Device, K053363 | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | The WISE Cortical Strip<br>is intended for<br>intraoperative (≤ 24 hours)<br>use with recording,<br>monitoring and<br>stimulation equipment for<br>the recording, monitoring,<br>and stimulation of<br>electrical signals on the<br>surface of the brain.<br>The WCS is indicated as<br>an aid to IntraOperative<br>Neurophysiological<br>Monitoring (IONM)<br>during brain surgeries. | The Ad-Tech Subdural<br>Electrodes<br>(Strip/Intraoperative Strip,<br>Grid/Intraoperative Grid,<br>Dual-Sided Interhemispheric,<br>Multi-Strip and Split Grid,<br>Intraoperative Mapping Grid)<br>are intended for temporary (<br>30 days) use with recording,<br>monitoring, and stimulation<br>equipment for the recording,<br>monitoring, and stimulation<br>of electrical signals on the<br>surface of the brain. The<br>recording of electrical<br>activity supports definition of<br>the location of epileptogenic<br>foci and brain mapping. | The Ad-Tech Subdural<br>Electrodes (Dual-Sided<br>Interhemispheric, Grid,<br>Intraoperative, Strip,<br>Wyler) are intended for<br>temporary (< 30 days) use<br>with recording, monitoring,<br>and stimulation equipment<br>for the recording,<br>monitoring, and stimulation<br>of electrical signals on the<br>surface of the brain. The<br>recording of electrical<br>activity supports definition<br>of the location of<br>epileptogenic foci and brain<br>mapping. | #### 3 Summary of Technological Characteristics Comparison The primary and secondary predicate devices have regulatory clearance as part of a family of electrodes provided in multiple shapes and configurations that support a range of intraoperative and subdural (implanted <30 day) use. Specifically, the Ad-Tech Intraoperative Strip Electrode is claimed as the primary predicate device. The technologic characteristics of the subject and predicate devices are similar in the following ways: - . They are subdural cortical electrodes used intraoperatively to transfer electrical signals to and from the brain for the same duration of use, in the same environment of use, by the same users. - They are sterilized using EO and provided sterile, for single-use . onlv. - . They come in configurations that have four numbered 2.3 mm diameter electrodes spaced 10 mm apart in a silicone substrate ending in a. - The strips end in a 1.5 mm diameter electrode cable that attaches to a ● separate connecting cable for use. - Consistent with the clinical practice safety limit, they support a ● maximum stimulation charge density of < 30 µC/cm². The main technological differences between the subject and predicate devices are: {6}------------------------------------------------ - The contact materials are similar but not identical; the subject ● device electrodes are Platinum only, whereas the predicate device electrodes are made of 90:10 Platinum:Iridium or stainless steel. - The subject device is labeled "non-pyrogenic" while the predicate is . not. - . The subject device is manufactured using a unique Supersonic Cluster Beam Implantation process which produces a very flexible electrode strip that conforms with the cortex surface. #### 4 Non-Clinical Performance Data Testing was performed with the WCS to support substantial equivalence to the predicate device. Testing included performance testing, biocompatibility, and shelf-life testing. Bench testing was conducted on WCS devices that were exposed to 2x sterilization, 18 months of accelerated aging, environmental conditioning and transport simulation prior to testing. In addition to ensure their integrity even under conditions of misuse, many of the tests were conducted after the devices were exposed to soaking, linear elongation, and bending conditioning. The following standards were utilized in the evaluation: - ASTM D1980-16, Standard Guide for Accelerated Aging of Sterile Barrier . Systems for Medical Devices - ASTM D4332-14 Standard Practice for Conditioning Containers, Packages, or . Packaging Components for Testing - ASTM D4169-16, Standard Practice for Performance Testing of Shipping ● Containers and Systems | Test Case Name | Purpose | Conclusion | |---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Labelling and packaging<br>integrity verification | Prove integrity of device labeling, retail box<br>and sterile barrier system following<br>conditioning | Pass | | Label indelibility test | Prove that product labels preserve their<br>integrity and readability after conditioning | Pass | | Dielectric strength test | Prove the device cable insulation is able to<br>withstand high voltage electrical stresses<br>beyond what would be encountered during<br>normal use | Pass | | Device integrity<br>verification | Prove that the device keeps its integrity<br>before and after exposure to stress<br>conditions | Pass | | Test Case Name | Purpose | Conclusion | | AC impedance test | Prove the device is able to conduct signals between the electrodes on the strip and the corresponding electrodes on the cable with a total impedance ≤ 10 kΩ @ 10 Hz*<br>Prove the device is compatible with the Cabrio L-SRL-4DIN connecting cable<br>Prove the performance of the device with the Cabrio L-SRL-4DIN and WCS Link connecting cables is equivalent | Pass | | Insulation test | Prove that the device insulation resistance meets requirements following 1 hour immersion in saline solution | Pass | | Compatibility with recorders test | Prove the device is compatible with EEG/ECOG recorders measuring the values of offset voltage and combined offset instability and internal noise | Pass | | Compatibility with stimulators test | Prove the device is compatible with commercially available neurostimulators (compliant with IEC60601-1 and IEC60601-2-40) by demonstrating the electrode's ability to deliver worst-case condition stimulating pulses | Pass | | Stimulation and metal release test | Prove the device is able to deliver stimulation pulses for 48 hours in simulated conditions<br>Prove the electrodes keep integrity following 48 hours of stimulation<br>Prove that the platinum released from the electrodes during 48 hours stimulation is within safe limits | Pass | | Bias current tolerance test | Prove the device is compatible with EEG/ECOG recorders measuring the values of the DC offset voltage does not exceed the limits of the input voltage when a DC bias current is applied to a pair of electrodes for 24 hours | Pass | {7}------------------------------------------------ Direct comparison testing of the subject device and the secondary predicate demonstrates that both devices meet the ≤ 10 kΩ acceptance criteria at 10 Hz. In addition, the subject device maintains this performance level even after exposure to worst-case stimulation conditions. {8}------------------------------------------------ #### Biocompatibility Testing 5 Biocompatibility testing was performed based on the nature and duration of patient contact outlined in ISO 10993-1: "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and the FDA guidance document: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (September 2020). The device was tested per the ISO 10993-1 requirements for an externally communicating device with tissue contact and cerebrospinal contact for a limited duration (≤ 24 hours). | Test | Result | Conclusion | |---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------| | Cytotoxicity | The full strength EMEM10 test article extract showed no cytotoxic potential to L-929 mouse fibroblast cells | Non-cytotoxic | | Sensitization | Topical application of the 0.9% NaCl extract and the sesame extract did not induce delayed sensitization in the guinea pig | Non-sensitizer | | Intracutaneous reactivity | The 0.9% NaCl extract of the test article and the sesame extract did not induce any erythema or edema reactions after injection by intracutaneous route in the rabbit | Non-irritant | | Pyrogenicity | Sum of temperature rises = 0.69°C<br>No rabbit showed an individual temperature rise higher or equal to 0.5°C above its initial temperature | Non-pyrogenic | | Acute Systemic Toxicity | There was no evidence of significant systemic toxicity or mortality after test article extracts injection | Non-toxic | | Indirect Hemolysis | Based on the hemolytic index of the assay sample above the negative control, the WCS is demonstrated to be non-hemolytic for indirect contact | Non-hemolytic | Results of testing demonstrate that the materials used in the construction of the WCS are safe for their intended use. {9}------------------------------------------------ #### Sterilization and Shelf-Life Testing 6 Sterilization of the subject device using Ethylene Oxide was validated following ANSI/AAMI/ISO 11135:2014 and demonstrated a SAL of 10-6. A shelf-life of 18 months has been established based on accelerated aging. #### 7 Substantial Equivalence Conclusion The WCS is substantially equivalent to the Ad-Tech Intraoperative Strip Electrode based on the intended use and technological characteristics as well as the intended users and intended use environment. Both devices have similar physical characteristics and stimulation parameters. The manufacturing differences are demonstrated through the testing to not adversely impact the subject device's safety or effectiveness. Therefore, one can conclude that the WISE Cortical Strip is substantially equivalent to the Ad-Tech Intraoperative Strip Electrode predicate device and is safe for its intended use.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...