Catalog Numbers 2110-XXX: The Subdural Cortical Electrodes (Dual Side Macro Electrodes, Dual Side Micro Electrodes, Dual Side Macro-Micro Electrodes, Dual Sided Hemispherical Macro Electrodes, Dual Sided Hemispherical Micro Electrodes, Dual Side Hemispherical Macro-Micro Elcctrodes, Macro Grids, Macro Strip, Macro-Micro Grids, Macro-Micro Strips, Micro Grids, and Micro Strips.) are intended for temporary (<30 day) use with recording, monitoring, and stimulation equipment for the recording, monitoring and stimulation of electrical signals on the surface of the brain. This device is restricted to the sale by or on the order of a physician. .
Device Story
Invasive subdural cortical electrodes; used intraoperatively or during monitoring for epilepsy; placed directly on surgically exposed brain surface. Device consists of platinum-iridium (90:10) alloy contacts molded into silicone rubber matrix; insulated wires extend to external EEG recording/stimulation equipment. Macro electrodes (2-4.5mm discs) capture general activity; micro electrodes (exposed wire tips) provide precise definition of epileptic activity. Used by neurosurgeons/neurologists to localize seizure foci; enables high-confidence mapping of electrical activity; guides clinical decision-making for surgical intervention. Benefits include improved localization of small or deep seizure foci compared to scalp EEG.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Materials: 90:10 platinum-iridium alloy contacts, silicone rubber matrix. Form factor: Grids and strips with macro (2mm, 3mm, 4.5mm discs) and micro (wire tip) electrodes. Connectivity: Wired connection to external EEG monitoring/stimulation equipment. Sterilization: Not specified. Software: None.
Indications for Use
Indicated for patients requiring temporary (<30 day) intracranial, subdural recording, monitoring, and stimulation of electrical brain activity, specifically for localizing seizure foci, determining seizure frequency, severity, and electro-clinical characteristics when foci are too small or deep for standard EEG.
Regulatory Classification
Identification
A cortical electrode is an electrode which is temporarily placed on the surface of the brain for stimulating the brain or recording the brain's electrical activity.
Predicate Devices
Cortac Cortical Electrode (k964224)
AD-TECH Subdural Electrode (k053363)
Submission Summary (Full Text)
{0}------------------------------------------------
K082474
EXHIBIT 5
MAY - 8 2009
# PMT Corporation 510(k) Summary
### Submitter's Name, Address, and Date of Submission
Eric Caillé General Manager PMT Corporation . 1500 Park Road Chanhassen, MN
Phone: 952-470-0866 Fax: 952-470-0865
Submitted:
Device Name
Trade Name:
PMT Subdural Cortical Electrodes
PMT Subdural Cortical Electrodes (Dual Side Macro Electrodes, Dual Side Micro Electrodes, Dual Side Macro-Micro Electrodes, Dual Sided Hemispherical Macro Electrodes, Dual Sided Hemispherical Micro Electrodes, Dual Side Hemispherical Macro-Micro Electrodes, Macro Grids, Macro Strip, Macro-Micro Grids, Macro-Micro Strips, Micro Grids, and Micro Strips.)
Classification Name: Common/Usual Name: GYC, 21CFR882.1310 Electrode, Cortical
#### Predicate Device
Cortac Cortical Electrode (k964224) AD-TECH Subdural Electrode (k053363)
#### Indication for Use
Catalog Numbers 2110-XXX: The Subdural Cortical Electrodes (Dual Side Macro Electrodes, Dual Side Micro Electrodes, Dual Side Macro-Micro Electrodes, Dual Sided Hemispherical Macro Electrodes, Dual Sided Hemispherical Micro Electrodes, Dual Side Hemispherical Macro-Micro Electrodes, Macro Grids, Macro Strip, Macro-Micro Grids, Macro-Micro Strips, Micro Grids, and Micro Strips.) are intended for temporary (<30 day) use with recording, monitoring, and stimulation equipment for the recording, monitoring and stimulation of electrical signals on the surface of the brain.
#### Device Description
PMT Subdural Cortical Electrodes (Dual Side Macro Electrodes, Dual Side Micro Electrodes, Dual Side Macro-Micro Electrodes, Dual Sided Hemispherical Macro Electrodes, Dual Sided Hemispherical Micro Electrodes, Dual Side Hemispherical Macro-Micro Electrodes, Macro Strip, Macro-Micro Grids, Macro-Micro Strips, Micro Grids, and Micro Strips.)
The Subdural Cortical Electrode is used intraoperatively for monitoring recordable electrical brain activity. This invasive intracranial, subdural electrode recording is performed directly on the surgically exposed brain. This method is necessary when the seizure focus is too small and or too deep within the brain to produce a recordable EEG seizure. The Subdural Cortical Electrode strip and grid electrode are used in cases where it is necessary to establish a high degree of confidence in the electrical localization (foci), seizure frequency, severity type and other electro-clinical characteristics.
{1}------------------------------------------------
Macro Electrodes use 2mm, 3mm or 4.5mm diameter platinum circular recording discs. Micro Electrodes are single wires with the distal tip exposed and flush with the surface of the silicone. They are made of the same material used in the assembly of the Electrode. The Micro electrode allow for more precise definition of the Epileptic activity. They can be used solo or in combination with Macro recording discs.
The PMT device consists of one or more electrically contacts. The model number 2110 indicates a platinum alloy wire and platinum alloy electrode contacts. The electrodes' contacts are molded into a silicone rubber matrix in a fixed pattern. Insulated wires extend from each electrode through a flexible silicone tube to connector for EEG monitoring.
The electrode material for the 2110-XXX is 90:10 platinum iridium
### Technological Characteristics and Performance
The technological characteristics of the product are not affected in this submission.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white and appears to be a simple, graphic representation of the department.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
PMT Corporation % Mr. Eric Caille 1500 Park Road Chanhassen, Minnesota 55317
Re: K082474
Trade Name: PMT Subdural Cortical Electrodes Regulation Number: 21 CFR 882.1310 Regulation Name: Cortical Electrode Regulatory Class: II Product Code: GYC Dated: April 22, 2009 Received: April 23, 2009
Dear Mr. Caille:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
MAY - 8 2009
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2- Mr. Eric Caille
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Daniel Krane
- Mark N. Melkerson Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## EXHIBIT 3
## Indications for Use Statement
K082474
### 510(k) Number (if known)
Device Name
PMT Subdural Cortical Electrodes LL-
#### Indications for Use
Catalog Numbers 2110-XXX: The Subdural Cortical Electrodes (Dual Side Macro Electrodes, Dual Side Micro Electrodes, Dual Side Macro-Micro Electrodes, Dual Sided Hemispherical Macro Electrodes, Dual Sided Hemispherical Micro Electrodes, Dual Side Hemispherical Macro-Micro Elcctrodes, Macro Grids, Macro Strip, Macro-Micro Grids, Macro-Micro Strips, Micro Grids, and Micro Strips.) are intended for temporary (<30 day) use with recording, monitoring, and stimulation equipment for the recording, monitoring and stimulation of electrical signals on the surface of the brain. This device is restricted to the sale by or on the order of a physician. .
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Per 21 CFR 801.109 OR
Over the Counter Use
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K082474
(Optimal Format 1-2-96)
PMT Corporation
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.