← Product Code [GXY](/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/GXY) · K121816

# SENSUS ELECTRODE (K121816)

_Neurometrix, Inc. · GXY · Nov 26, 2012 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/GXY/K121816

## Device Facts

- **Applicant:** Neurometrix, Inc.
- **Product Code:** [GXY](/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/GXY.md)
- **Decision Date:** Nov 26, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1320
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The NeuroMetrix SENSUS Electrode is intended for use as disposable, conductive, adhesive interface between the patient's skin and a transcutaneous electrical nerve stimulator.

## Device Story

SENSUS Electrode serves as a disposable, conductive, adhesive interface between a transcutaneous electrical nerve stimulator and patient skin. Device features a multi-layer design: Mylar outer layer; conductive silver traces/pads; medical-grade biocompatible hydrogel (KM-10G) for skin contact. Four individual electrodes (36x46 mm each) are arranged in a pre-set geometry (50x280 mm total) and connected in two pairs via male snap connectors. Device is non-sterile and intended for single-patient, multiple-application use. Healthcare providers use the electrode to facilitate electrical stimulation therapy; the pre-configured array provides convenience in application. Benefits include consistent electrical current dispersion and ease of use compared to individual electrodes.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including electrical performance (AC impedance, current dispersion), mechanical testing (snap connector retention force), skin adhesion, and accelerated aging shelf-life studies. Biocompatibility established via testing of related hydrogel formulations (KM-10B and KM-10P) per ISO 10993-5 and ISO 10993-10.

## Technological Characteristics

Multi-layer construction: Polyethylene substrate, silver conductive traces, Katecho KM-10G hydrogel. Dimensions: 50x280 mm (four 36x46 mm pads). Connectivity: Two male snap connectors. Non-sterile. Complies with ISO 10993-5 and ISO 10993-10.

## Regulatory Identification

A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.

## Predicate Devices

- Lead-Lok Reusable Neurostimulation Electrodes ([K010431](/device/K010431.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

112/816

# 510(k) Summary for NeuroMetrix SENSUS Electrode

#### SPONSOR

NeuroMetrix, Inc. 62 Fourth Avenue Waltham, MA 02451 USA

NOV 2 6 2012

Contact Person: Rainer Maas Telephone: (781) 314-2781 Date Prepared: November 7, 2012

#### DEVICE NAME

Proprietary Name: Common/Usual Name: Classification Name:

SENSUS Electrode Transcutaneous Electrical Nerve Stimulator Electrode 882.1320 GXY Cutaneous Electrode

#### PREDICATE DEVICE

Lead-Lok Reusable Neurostimulation Electrodes (K010431)

#### INTENDED USE

The NeuroMetrix SENSUS Electrode is intended for use as disposable, conductive, adhesive interface between the patient's skin and a transcutaneous electrical nerve stimulator.

#### DEVICE DESCRIPTION

The SENSUS Electrode provides an electrically conductive interface between a transcutaneous electrical nerve stimulator and a patient's skin. It is provided non-sterile, is designed and intended for single patient use only, and to be disposable.

The SENSUS Electrode is comprised of four individual electrodes, each of size 36 by 46 mm. The overall dimensions are 50 by 280 mm. The individual electrodes are electrically connected in pairs such that the two outer electrodes constitute one pair and the two inner electrodes constitute a second The SENSUS Electrode contains two conventional male snap connectors for electrical pair. connection to a transcutaneous electrical nerve stimulator.

The SENSUS Electrode has a multi-layer design. The first and outermost layer is a sheet of Mylar. The second layer contains conductive silver traces and silver electrode pads. Where the silver traces are not covered by hydrogel, they are covered by a dielectric mask. The third layer is the four individual electrodes which consist of a medical grade, self adhering, biocompatible hydrogel (KM-10G, Katecho, Inc., Des Moines, IA). When not in use, the hydrogel is covered by a Mylar release Two male type snap connectors interface the SENSUS Electrode to the transcutaneous liner. electrical nerve stimulator. The patient facing surface of the snap connectors are either under hydrogel or covered by a laminated dielectric polypropylene layer. Therefore in both cases the snap connectors do not directly contact the patient's skin.

{1}------------------------------------------------

K12/816

# COMPARISON TO PREDICATE

The SENSUS Electrode and Lead-Lok Neurostimulation Electrodes have the same intended use, Both are intended as single-patient/multiple application use, conductive, adhesive interfaces between electrical stimulators, such as transcutaneous electrical nerve stimulators, and a patient's skin. Both are prescription devices to be used under the direction of a medical professional.

The table below compares the intended use and characteristics of the SENSUS Electrode and the predicate Lead-Lok Reusable Neurostimulation Electrodes. The two devices have intended use, utilize similar materials, and have similar technological characteristics. The primary difference between the two devices is that the SENSUS Electrode consists of four individual electrodes arranged in a pre-set geometry whereas the predicate electrodes are not pre-combined into a specific geometry. The combination of individual electrodes into a pre-configured array is a convenience feature that does not alter the basic function of the individual electrodes, which is to conduct electrical currents from a stimulator to the patient's skin. Furthermore, the SENSUS Electrode shape and area fall within the range of the predicate. As a result, the technological characteristics of the SENSUS Electrode do not raise new questions of safety or effectiveness.

This comparison of the SENSUS Electrode and predicate Lead-Lok Neurostimulation Electrodes, along with testing data provided in this submission, support a finding of substantial equivalence for the SENSUS Electrode.

|                                                     | NeuroMetrix SENSUS Electrode                                                                                                                                                                  | Lead-Lok Reusable Neurostimulation<br>Electrodes (K010431)                                                                                                                               |
|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General Characteristics                             |                                                                                                                                                                                               |                                                                                                                                                                                          |
| Intended Use                                        | The NeuroMetrix SENSUS Electrode is<br>intended for use as a disposable, conductive,<br>adhesive interface between the patient's skin<br>and a transcutaneous electrical nerve<br>stimulator. | Lead-Lok Reusable Neurostimulation<br>Electrodes are intended for use as the<br>disposable, conductive adhesive interface<br>between the patient's skin and the Electrical<br>Stimulator |
| Number of Electrodes                                | 4, two pairs of two electrodes                                                                                                                                                                | 1                                                                                                                                                                                        |
| Sterility Status                                    | Non-sterile                                                                                                                                                                                   | Non-sterile                                                                                                                                                                              |
| Single-Patient Use                                  | Yes                                                                                                                                                                                           | Yes                                                                                                                                                                                      |
| Multiple Applications                               | Yes                                                                                                                                                                                           | Yes                                                                                                                                                                                      |
| Reusable (across patients)                          | No                                                                                                                                                                                            | No                                                                                                                                                                                       |
| Shelf-life                                          | 24 months, shelf-life to be confirmed with<br>accelerated aging test prior to commercial<br>distribution                                                                                      | Unknown                                                                                                                                                                                  |
| Technical Characteristics                           |                                                                                                                                                                                               |                                                                                                                                                                                          |
| Substrate (1st layer)                               | Polyethylene                                                                                                                                                                                  | Polyethylene, polypropylene                                                                                                                                                              |
| Electrical Conductor (2nd<br>layer)                 | Ag                                                                                                                                                                                            | Carbon coated with Ag/AgCl                                                                                                                                                               |
| Hydrogel (3rd layer)                                | Katecho KM10G                                                                                                                                                                                 | Katecho KM10 series gels                                                                                                                                                                 |
| Connector                                           | Male snap connector                                                                                                                                                                           | Male snap connector                                                                                                                                                                      |
| Overall Dimensions                                  | 5 x 28 cm, total area 140 $cm^2$ , includes two<br>electrodes                                                                                                                                 | Various shapes (square, circle, oval,<br>butterfly) with total area 7.9 - 154.8 $cm^2$                                                                                                   |
| Conductive Surface (single<br>electrode) Dimensions | Two rectangular 3.6 x 4.6 cm connected pads,<br>total area 33.1 $cm^2$                                                                                                                        | Various shapes (rectangular, circle, oval)<br>with total area 7.9 - 154.8 $cm^2$                                                                                                         |

#### GUIDANCE DOCUMENT

The recommendations of the FDA's "Draft Guidance for Industry and Staff: Class II Special Controls Guidance Document: Cutaneous Electrode (April 5, 2010)" were taken into account in preparing this 510(k) submission. The draft guidance addresses requirements for cutaneous electrodes which are

{2}------------------------------------------------

K12/816

defined as "an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation." This definition includes electrodes intended for use with transcutaneous electrical nerve stimulators. NeuroMetrix believes that SENSUS Electrode complies with the special controls as outlined in the draft guidance, thereby providing additional assurance of safe and effective use of the SENSUS Electrode.

#### NON-CLINICAL TESTING

Verification testing of the SENSUS Electrode includes electrical and mechanical tests to show that it meets its target specifications over a range of operating and storage conditions. Verification and performance testing further demonstrate that it meets user needs as reflected in the functional specification.

## Biocompatibility Testing

The hydrogel pads, forming the individual electrodes, maintain contact between the SENSUS Electrode and the skin. This contact is established by the hydrogel's self-adhesives properties. The SENSUS Electrode hydrogel is KM-10G manufactured by Katecho, Inc. The KM-10G is a member of the Katecho KM-10 series that has been used in electrodes previously cleared for use with transcutaneous electrical nerve stimulators (K000870). Katecho has established biocompatibility of the KM-10G hydrogel through testing of two related hydrogel formulations. The KM-10G hydrogel has essentially the same formulation as the Katecho KM-10B hydrogel with the exception of a higher concentration of diacetone acrylamide (DAA). Another Katecho hydrogel, KM-10P, has a similar concentration of DAA. Therefore the biocompatibility of the KM-10G hydrogel was established by combining the test results for the KM-10B and KM-10P hydrogels.

#### Electrical and Mechanical Testing

Electrical performance was assessed by measuring the AC impedance of electrodes in a gel to gel configuration and evaluating current dispersion across the surface of the electrodes. Adhesive performance was evaluated by performing a skin adhesion test. Electrode stability was evaluated by assessing electrical performance under normal use conditions. Cutaneous electrode materials should be stable and resist physical and chemical breakdown as a result of conducting electrical current. To assess this, SENSUS Electrodes were tested to ensure they met the electrical impedance specifications after 1 hour of conventional use. Accelerated aging shelf-life testing was performed to ensure that the SENSUS Electrode performs as intended over the course of its labeled shelf-life. The electrical and mechanical properties of the snap connectors were assessed by measuring the resistance of the connection and the connection retention force.

The SENSUS Electrode conforms to the following standards:

- . ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- 1SO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and . skin sensitization

{3}------------------------------------------------

K12/816

## CLINICAL TESTING

NeuroMetrix determined that non-clinical (i.e., bench) testing was sufficient to demonstrate that the SENSUS Electrode is as safe and effective as the predicate.

## CONCLUSION

The verification and performance data presented in this 510(k) submission demonstrate that the SENSUS Electrode is substantially equivalent to the predicate.

{4}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The logo is simple and monochromatic.

November 26, 2012

NeuroMetrix, Inc c/o Mr. Rainer Maas Director of QA/RA 62 Fourth Avenue Waltham, MA 02451

Re: K121816

Trade/Device Name: SENSUS Electrode Regulation Number: 21 CFR 886.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: November 7, 2012 Received: November 8, 2012

Dear Mr. Maas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

Public Health Service

ﭘﮍ

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

{5}------------------------------------------------

Page 2 - Mr. Rainer Maas

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may-obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

# Deboran L. Falls

Malvina B. Eydelman, M.D. Director Division of Opthalmic, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and

Radiological Health

Enclosure

{6}------------------------------------------------

# 4. Indications for Use Statement

## Indications for Use

510(k) Number (if known):

Device Name: NeuroMetrix SENSUS Electrode

Indications for Use:

The NeuroMetrix SENSUS Electrode is intended for use as disposable, conductive, adhesive interface between the patient's skin and a transcutaneous electrical nerve stimulator.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Eyaluation (ODE)

(Division Sign-Off)

Division of Neurological and Physical Medicine Devices 510(k) Number

Page 1 of 1

NeuroMetrix, Inc. SENSUS Electrode 510(k)

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/GXY/K121816](https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/GXY/K121816)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
