SPARK Scan

K231457 · Spark Neuro, Inc. · GWQ · Aug 18, 2023 · Neurology

Device Facts

Record IDK231457
Device NameSPARK Scan
ApplicantSpark Neuro, Inc.
Product CodeGWQ · Neurology
Decision DateAug 18, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1400
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The SPARK Scan is intended to acquire, display, and store the electrical activity of a patient's brain obtained by placing two or more electrodes on the head to aid in diagnosis.

Device Story

SPARK Scan is an SaMD system for EEG data acquisition, display, and storage. It consists of Edge software (desktop application) and a cloud-based Data Storage and Communication Platform. The Edge software manages interaction with third-party EEG hardware (amplifier and cap), facilitates recording, and provides limited review access. The cloud platform manages user authentication, data storage, and access. Used by EEG technicians, nurses, or medical assistants in healthcare facilities. The device does not perform automated analysis or diagnostic interpretation; it serves as a conduit for clinicians to view and store raw EEG signals to support clinical decision-making.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical testing, including software verification and validation, cybersecurity risk analysis, and human factors use-related risk analysis.

Technological Characteristics

SaMD comprising Edge desktop software and cloud-based storage/communication platform. Compatible with third-party EEG amplifiers (K172312) and EEG caps (K110223). Sampling rate: 500Hz. Supports up to 32 recording channels. Includes impedance testing functionality. Connectivity: Networked/Cloud-hosted. No direct patient contact; no hardware components.

Indications for Use

Indicated for adult patients requiring EEG data acquisition, display, and storage to aid in diagnosis. No specific contraindications listed.

Regulatory Classification

Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ August 18, 2023 Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. SPARK Neuro Inc. % John Doucet Vice President, Neurology Regulatory Affairs MCRA. LLC 803 7th Street NW Washington, DC 20001 Re: K231457 Trade/Device Name: SPARK Scan Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWO Dated: May 18, 2023 Received: May 19, 2023 Dear John Doucet: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Patrick Antkowiak -S for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) TBD Device Name SPARK Scan Indications for Use (Describe) The SPARK Scan is intended to acquire, display, and store the electrical activity of a patient's brain obtained by placing two or more electrodes on the head to aid in diagnosis. X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) SUMMARY | Device Trade Name: | SPARK Scan | | | |------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--| | Manufacturer: | SPARK Neuro Inc.<br>212 W. 18th Street #17A<br>New York, NY 10011 USA | | | | Contact: | John Doucet<br>Vice President, Neurology Regulatory Affairs<br>MCRA, LLC<br>803 7th Street NW<br>Washington, DC 20001<br>Office: 202.552.5835<br>Jdoucet@mcra.com | | | | Prepared by: | MCRA, LLC<br>803 7th St NW<br>Washington, DC 20001<br>Office: 202.552.5800 | | | | Date Prepared:<br>Regulation:<br>Class:<br>Product Code:<br>Classification Name:<br>Common Name:<br>Primary Predicate: | May 18, 2023<br>21 CFR 882.1400<br>II<br>GWQ<br>Full-Montage Standard Electroencephalograph<br>Electroencephalograph<br>K143233 Mitsar-EEG | | | #### Indications For Use: The SPARK Scan is intended to acquire, display, and store the electrical activity of a patient's brain obtained by placing two or more electrodes on the head to aid in diagnosis. ## Device Description: The SPARK Scan is intended for the acquisition, display, and storage of electroencephalogram (EEG) data. The SPARK Scan is intended to be used by EEG technicians, or appropriately trained Nurses and Medical Assistants practicing in any medical setting where EEG data collection may be required. The SPARK Scan is an SaMD product that consists of the Edge Software and Data Storage and Communication Platform. {4}------------------------------------------------ The Edge software runs locally on the user's device and consists of a User interface and the EEG hardware interface, including the electroconductive gel (K1117). The User Interface is a desktop application that (a) manages interaction with the EEG hardware interface, (b) facilitates set up and recording of EEG data, and (c) provides limited access to review of patient records according to the user's permissions. The Data Storage and Communication Platform (Cloud Software): The Cloud software runs on a server managed by SPARK Neuro and contains no user interface. The Cloud Software is responsible only for managing (a) authorization and authentication of users and (b) storage, validation, and access to all data collected on the system. The SPARK Scan is compatible with three 3rd-party accessory devices: an FDA-cleared EEG hardware system, a standard off-the-shelf-laptop, and FDA-cleared EEG Recording Viewing Platforms. | | Subject Device | Primary Predicate | Comparison | |----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------| | Trade Name | SPARK Scan | Mitsar-EEG | N/A | | Company | SPARK Neuro | Mitsar Co., LTD | N/A | | K Number | K23xxxx | K143233 | N/A | | Regulation<br>Number | 21 CFR 882.1400 | 21 CFR 882.1400 | Identical | | Product Codes | GWQ | GWQ | Identical | | Indications for<br>Use Statement | The SPARK Scan is<br>intended to acquire,<br>display, and store the<br>electrical activity of a<br>patient's brain obtained<br>by placing two or more<br>electrodes on the head to<br>aid in diagnosis. | The Mitsar-EEG is<br>intended to acquire,<br>display, and store the<br>electrical activity of a<br>patient's brain obtained<br>by placing two or more<br>electrodes on the head to<br>aid in diagnosis. | Identical | | Intended User | Medical Staff | Medical Staff | Identical | | Target Population | Adults | Adults | Identical | | Use Environment | Healthcare Facilities | Healthcare Facilities | Identical | | System<br>Components | Edge Software<br>Data Storage and<br>Communication Platform<br>Off-The-Shelf EEG<br>Amplifier (K172312)<br>Off-The-Shelf EEG Cap<br>(K110223) | EEG Amplifier<br>USB-Cable<br>USB-Dongle<br>EEG Studio Software | Similar.<br>The subject<br>and the<br>predicate<br>devices each<br>comprise<br>components<br>that allow for<br>its proper<br>functionality. | | Acquires signals | Yes, EEG | Yes, EEG | Identical | ### Comparison to the Predicate Device: {5}------------------------------------------------ | Number of Signals<br>Recording<br>Channels<br>Compatibility | up to 32 | up to 21 | Similar.<br>Subject device<br>is similar to<br>the number of<br>recording<br>channels of<br>the predicate<br>device. | |-------------------------------------------------------------|----------|----------|----------------------------------------------------------------------------------------------------------------------| | Impedance Test | Yes | Yes | Identical | | Sampling Rate | 500Hz | 500Hz | Identical. | | Interface | PC | PC | Identical | ### Performance Testing Summary: Non-Clinical Testing - . Software verification and validation testing - Cybersecurity Risk Analysis and Testing ● - Human Factors Use Related Risk Analysis ● ### Biocompatibility The subject device does not come into direct or indirect contact with the patient. This is not applicable. #### Electrical Safety and Electromagnetic Compatibility The subject device does not contain hardware. This is not applicable. Animal Testing Animal testing is not required to support substantial equivalence. Clinical Testing Clinical testing is not required to support substantial equivalence. Conclusion: The subject device and the predicate device have the same intended use and have similar technological characteristics. The data included in this submission demonstrate substantial equivalence to the predicate device listed above. SPARK Scan is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...