The Trackit T4 EEG Amplifier is intended to be used as a front-end amplifier to acquire, store and transmit electrophysiological signals (wireless or cabled).
Device Story
Trackit T4 EEG Amplifier is a multi-channel electroencephalograph for routine EEG, lab monitoring, and ambulatory applications. Device inputs include electrophysiological signals via electrodes; pulse oximetry data via Nonin interface; patient event markers; and auxiliary DC signals. Device transforms inputs into digital signals stored on internal micro-SD card or transmitted via USB, Bluetooth, or WLAN. Operated by trained clinicians in hospitals, clinics, or home settings; electrodes applied by clinician prior to home use. Device features 32 channels (expandable to 68), built-in calibration, and impedance measurement. Output is reviewed by clinicians on PC software to assist in patient diagnosis. Benefits include compact, portable form factor for ambulatory monitoring without requiring patient intervention during recording.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance verification (calibration, noise, impedance, frequency response) compared to the predicate device.
Technological Characteristics
ABS plastic housing; 32-channel (expandable to 68) referential/poly inputs; 24-bit ADC; 250-16000 Hz sampling rate; DC-4193 Hz bandwidth; USB, Bluetooth, and WLAN connectivity; Li-ion rechargeable battery or external battery pack; micro-SD storage. Complies with IEC 60601-1, IEC 60601-2-26, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, and ISO 14971.
Indications for Use
Indicated for use as a front-end amplifier to acquire, store, and transmit electrophysiological signals in clinical, hospital, research, and home environments for patients undergoing EEG examinations.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. The FDA logo is on the right, with the letters "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 4, 2018
Lifelines Ltd. % Yolanda Smith Consultant Smith Associates 1468 Harwell Ave Crofton, Maryland 21114
Re: K172271
Trade/Device Name: Trackit T4 EEG Amplifier Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWO. GWO. GWL Dated: April 5, 2018 Received: April 10, 2018
Dear Yolanda Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{2}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Hoffmann -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K172271
Device Name Trackit T4 EEG Amplifier
Indications for Use (Describe)
The Trackit T4 EEG Amplifier is intended to be used as a front-end amplifier to acquire, store and transmit electrophysiological signals (wireless or cabled).
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k)Summary
| SPONSOR | |
|----------------------|------------------------------------------------------------------------|
| Company Name:<br>Lif | Lifelines, Ltd. |
| Company Address | 7 Clarendon Court<br>Over Wallop, Hampshire<br>SO20 8HU United Kingdom |
| Telephone: | +44 (0)1483 224 245 |
| Contact Person: | Michael Hulin |
Summary Preparation Date: July 17, 2017
#### DEVICE NAME
| Trade Name: | Trackit T4 EEG amplifier: |
|----------------------|---------------------------|
| Common/Usual Name: | EEG Amplifier |
| Classification Name: | Electroencephalograph |
| Regulation Number: | 21 CFR 882.1400 |
| Product Code: | GWQ, GWL |
| Device Class: | Class II |
#### PREDICATE DEVICE
| Legally Marketed Equivalent Device | | |
|------------------------------------|-------------------|-----------------|
| <b>Company</b> | <b>Product</b> | <b>510(k) #</b> |
| Lifelines, Ltd. | R40 EEG Amplifier | K151600 |
#### DEVICE DESCRIPTION
The Trackit T4 EEG Amplifier is a multi-channel electroencephalograph designed for use in routine EEG and lab monitoring applications and due to its small size, can be used in ambulatory applications. In this situation, the EEG electrodes are fitted to the patient by a trained clinician prior to the patient being sent home. No subsequent intervention is required by the patient. Upon completion of the recording, the data which is stored on a memory card is reviewed by a clinician using review and analysis software on a PC.
lt is a compact USB amplifier which provides 32 channels with internal expansion option) with built-in calibration and electrode impedance measurement. Also provided is a Nonin pulse oximeter interface, a Patient Event input and an Aux DC input. Optional wireless communication is available (Bluetooth and WLAN WiFi).
There are two variants of the Trackit T4 EEG Amplifier:
- Trackit T4-32 providing 24 referential + 8 poly channels. .
- . Trackit T4-68 providing 64 referential + 4 poly channels (using internal expansion board).
Plug-on Patient Connection Units (PCUs) provide 32 channel touchproof inputs (model T4-PCU 24+8) or 68 channels (model T4-PCU 64+4).
The Amplifier is intended to be connected to a USB port on a PC which is powered from a medically
{5}------------------------------------------------
approved power supply. In addition it can be battery powered in ambulatory applications. This equipment is intended only as an adjunct device in patient assessment; it must be used in conjunction with other methods of patient diagnosis. The equipment does not sustain or support life.
#### DEVICE INDICATIONS FOR USE
The Trackit T4 EEG Amplifier is intended to be used as a front-end amplifier to acquire, store and transmit electrophysiological signals (wireless or cabled).
#### COMPARISON OF TECHNICAL CHARACTERISTICS
The following chart compares the new T4 EEG Amplifier with the company's R-40 EEG Amplifier as the predicate device. The new device has evolved from the R-40 with the goal of making it suitable for ambulatory and home use. Differences in characteristics are discussed in the notes in the following section.
| | New device:<br>Lifelines T4 EEG Amplifier | Predicate device:<br>Lifelines R-40 EEG Amplifier | Similarities and<br>Differences |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | GWL and GWQ | GWL and GWQ | Same |
| 510(k) no. | New device | K151600 | |
| Indications for<br>Use | The T4 EEG Amplifier is<br>intended to be used as a<br>front-end amplifier to<br>acquire, store and transmit<br>electrophysiological signals<br>(wireless or cabled). | The R-40 EEG Amplifier is<br>intended to be used as a<br>front-end amplifier to<br>acquire, store and transmit<br>electrophysiological signals<br>(wireless or cabled). | Same |
| Description of<br>device | The T4 EEG Amplifier is a<br>stand-alone 32 channel (24<br>EEG + 8 poly) EEG amplifier<br>with built-in calibration and<br>electrode impedance<br>measurement. It is<br>expandable to 68 channels<br>with an internal expansion<br>board.<br><br>The device can be powered<br>from an external battery<br>pack for ambulatory usage.<br>No intervention is required<br>by the patient. | The R-40 EEG Amplifier is a<br>stand-alone 40 channel (32<br>EEG + 8 poly) EEG amplifier<br>with built-in calibration and<br>electrode impedance<br>measurement. | New device has 32<br>channels instead of 40.<br>Expandable to 68note 5.<br>Added ambulatory usage<br>with external battery<br>packnote 3. |
| | A Nonin XPOD interface is<br>provided for pulse oximetry.<br>A USB connection is used to<br>communicate with the host<br>computer with the option of<br>wireless (Bluetooth and<br>WiFi). | A Nonin XPOD interface is<br>provided for pulse oximetry.<br>A USB connection is used to<br>communicate with the host<br>computer with the option of<br>wireless (Bluetooth and<br>WiFi). | ) Same<br>)<br>)<br>)<br>)<br>)<br>) |
| | There are no physiological<br>alarms. | There are no physiological<br>alarms. | ) |
| | New device:<br>Lifelines T4 EEG Amplifier | Predicate device:<br>Lifelines R-40 EEG Amplifier | Similarities and<br>Differences |
| Clinical<br>Application<br>Environment | For use in research<br>institutions, clinic, hospital,<br>operating room and epilepsy<br>evaluation environments.<br>Home use. | For use in research<br>institutions, clinic, hospital,<br>operating room and epilepsy<br>evaluation environments. | ) Same<br>)<br>)<br>)<br>Addition of home usenote 8 |
| Intended User | A healthcare professional<br>who has the training and<br>knowledge to undertake EEG<br>examinations and is familiar<br>with EEG equipment and<br>practice. | A healthcare professional<br>who has the training and<br>knowledge to undertake EEG<br>examinations and is familiar<br>with EEG equipment and<br>practice. | Same |
| Channels | 24 EEG + 8 bipolar + SpO2 +<br>patient event button.<br>(64 EEG + 4 bipolar with<br>internal expansion board). | 32 EEG + 8 bipolar + SpO2 +<br>patient event button. | Reduced number of<br>channels in order to<br>reduce the size of the<br>device. Optional expansion<br>availablenote 5. |
| ADC<br>resolution | 24 bits | 24 bits | Same |
| Full-scale<br>input | ± 375 mV | ± 375 mV | Same |
| Sampling rate | 250 - 16000 Hz | 250 - 16000 Hz | Same |
| Input noise | < 1.5 µV pk-pk | < 1.5 µV pk-pk | Same |
| Bandwidth<br>(-3dB) | DC to 4193 Hz max. | DC to 4193 Hz max. | Same |
| Calibration | 8000 µV at 1 sec period | 8000 µV at 1 sec period | Same |
| Impedance<br>pass/fail<br>levels | 2, 5, 10, 20, 50 kΩ limits<br>adjustable on host computer<br>during setup. | 2, 5, 10, 20, 50 kΩ limits<br>adjustable on front panel. | Facility removed from front<br>panel to reduce the size of<br>the devicenote 7. |
| SaO2 input | Yes | Yes | Same |
| E-Cap<br>connector | No | Yes | E-Cap connector has been<br>removed to reduce the size<br>of the devicenote 6. |
| Front panel<br>display | Displays battery capacity,<br>elapsed recording time,<br>wireless connection status. | No | Added display shows<br>device status during<br>ambulatory usagenote 4. |
| Host PC<br>communicat-<br>ion | Wired (USB) or Wireless<br>(802.11b/g) or Bluetooth | Wired (USB) or Wireless<br>(802.11b/g) or Bluetooth | Same |
| Power | USB (isolated from patient)<br>or external battery pack | USB (isolated from patient)<br>or battery | New device can be<br>powered from external<br>battery pack for<br>ambulatory usagenote 3. |
| Internal<br>storage | micro-SD flash card | micro-SD flash card | Same |
| | New device:<br>Lifelines T4 EEG Amplifier | Predicate device:<br>Lifelines R-40 EEG Amplifier | Similarities and<br>Differences |
| Internal<br>battery | Li-ion rechargeable | Li-ion rechargeable | Same |
| Patient event<br>button input | Yes | Yes | Same |
| System<br>components | Amplifier, laptop PC, medical<br>grade power supply, Nonin<br>oximeter, patient event<br>button.<br>Bag and straps. | Amplifier, laptop PC, medical<br>grade power supply, Nonin<br>oximeter, patient event<br>button. | ) Same<br>)<br>)<br>Bag for ambulatory<br>usagenote 8. |
| Material | ABS | ABS | Same |
| Patient<br>contact | Device does not directly<br>contact the patient.<br>EEG electrodes (not<br>supplied) are patient-<br>applied. | Device does not directly<br>contact the patient.<br>EEG electrodes (not<br>supplied) are patient-<br>applied. | Same |
| Size | 9 x 17 x 3 cm | 11 x 17 x 4 cm | New device is smallernote 1 |
| Weight | 270 gm | 400 gm | New device is lighternote 1 |
| Compliance/<br>regulatory | IEC 60601-1 + ANSI + CAN<br>IEC 60601-2-26<br>IEC 60601-1-2<br>CE Mark<br>IEC 60601-1-11 | IEC 60601-1 + ANSI + CAN<br>IEC 60601-2-26<br>IEC 60601-1-2<br>CE Mark | ) Same<br>)<br>)<br>Addition of IEC 60601-1-11<br>Home Usenote 8 |
{6}------------------------------------------------
{7}------------------------------------------------
#### Similarities and Differences
The technological characteristics are substantially similar; the new device has evolved from the R-40 with the intention of making it suitable for ambulatory and home use. It is based on existing, well established technologies and is intended for use in the established field of EEG.
The differences between the new device and the predicate device are:
- 1. The new device uses the same internal electronics as the R-40 but its case has been reduced in size to make it more suitable for ambulatory use.
- 2. The Indications for Use for the new device are the same as the R-40.
- 3. The new device can be powered from an external power pack for ambulatory use.
- 4. The new device incorporates a small alpha-numeric display which shows battery capacity, elapsed recording time and wireless connection status during ambulatory use.
- 5. The channel count has been reduced from 40 to 32 in order to reduce the size of the device to optimize it for ambulatory use. The channel count can optionally be increased to 68 with an internal expansion board, for more complex clinical examinations.
- 6. The Electro-cap connector has been removed in order to reduce the size of the device to optimize it for ambulatory use.
- 7. The front-panel impedance limit feature has been removed. Impedance limits can be adjusted on the host computer during set-up.
- 8. The new device is suitable for ambulatory and home use.
{8}------------------------------------------------
These differences raise no new questions concerning safety or effectiveness. The Lifelines T4 EEG Amplifier is substantially equivalent to the predicate device.
#### Pre-clinical Performance Testing
The table below provides details of the bench testing performed on the T4 EEG Amplifier. Comparison is provided with the predicate device (R-40 Amp) in order to establish substantial equivalence.
| | | Test specification | | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|-------------------------------|-------------------------------------------------------------|
| Test | Test Method | T4<br>subject<br>device ¹ | R-40<br>predicate<br>device ² | Conclusion |
| Insulation<br>resistance | Using an insulation tester measure the resistance<br>between the TF1 'Aux. B' socket and the metalwork<br>of the USB plug. Ensure that it is > $200 ΜΩ$ . | 3.1 | 3.1 | Same test<br>as<br>predicate |
| Dielectric<br>strength | Connect a breakdown tester. Increase the voltage<br>slowly over 10s to 1500Vrms and hold it for 1<br>minute. Ensure that there are no signs of flash-over<br>or breakdown. | 3.2 | 3.2 | Same test<br>as<br>predicate |
| Mains on<br>Applied Parts | Measure the Mains on Applied Parts leakage<br>current. Ensure that it is ≤ 5000uA AC. | 3.6 | 3.6 | Same test<br>as<br>predicate |
| Patient<br>leakage | Measure the Patient leakage current. Ensure that it<br>is ≤100uA AC and ≤10uA DC. | 3.8 | 3.8 | Same test<br>as<br>predicate |
| Calibration | Check for 8mVpp ±0.4 square wave signals on all<br>channels at 1Hz frequency. | 4.4 | 4.4 | Same test<br>as<br>predicate |
| Short circuit<br>noise | Check that the short circuit noise for all channels is <<br>1.5uVpp. | 4.8 | 4.7 | Same test<br>as<br>predicate |
| HF response | Check HF response to 80 Hz. | 4.10 | 4.9 | Same test<br>as<br>predicate |
| DC response | Check for a +300mV step and a -300mV step. | 4.11 | 4.10 | Same test<br>as<br>predicate |
| Impedance<br>check | Check all channels indicate an impedance of $3000Ω$<br>±500. | 4.16 | 4.15 | Predicate<br>has front-<br>panel LEDs.<br>Otherwise<br>same |
<sup>4</sup> Clause numbers in T4-32 Amp Final Test Specification
² Clause numbers in R-40 Amp Final Test Specification
{9}------------------------------------------------
#### PERFORMANCE DATA
Testing was undertaken by an independent certification body and provided confirmation that the device performance and physical attributes met the requirements of the standards listed below. These standards address safety, EMC compatibility, risk, usability and home use.
Verification and validation testing confirmed that this device met the design requirements and user needs.
#### Safety Testing
- 12.4.1 IEC 60601-1:2012. Medical electrical equipment – Part 1: General requirements for basic safety and essential performance. Including ANSI/AAMI and CAN/CSA. 12.4.2 IEC 60601-2-26:2012. Medical electrical equipment. Particular requirements for the safety of electroencephalographs. 12.4.3 IEC 60601-1-2:2007. Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromaqnetic compatibility. 12.4.4 IEC 60601-1-6:2010 + A1:2013. Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability. 12.4.5 IEC 62366:2007 + A1:2014. Application of usability engineering to medical devices. 12.4.6 IEC 60601-1-11:2015. Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – Collateral standard: Requirements for medical electrical equipment & medical electrical systems used in the home healthcare environment. Including ANSI/AAMI and CAN/CSA. 12.4.7 ISO 14971:2007. Application of risk management to medical devices. 12.4.8 IEC 62133:2012. Secondary cells and batteries containing alkaline or other non-acid
- electrolytes Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications.
- 12.4.9 UN/DOT 38.3. Transportation testing for lithium batteries.
Clinical testing was not performed with this device.
#### CONCLUSION
The Lifelines Trackit T4 EEG Amplifier meets the functional claims and intended use as described in the product labeling. The safety and effectiveness are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.