K171124 · Nihon Kohden Corporation · GWQ · Aug 25, 2017 · Neurology
Device Facts
Record ID
K171124
Device Name
Nihon Kohden Wireless Input Unit WEE-1200
Applicant
Nihon Kohden Corporation
Product Code
GWQ · Neurology
Decision Date
Aug 25, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The WEE-1200 Wireless Input Unit is intended to acquire, store, record, and transmit the cerebral and extracerebral activity for EEG and sleep studies. The device should be used together with Nihon Kohden specified electroencephalograph via wired or wireless communication to display EEG on the electroencephalograph screen. These data may be used by the clinician in sleep disorder, epilepsies and other related disorders as an aid in diagnosis. The WEE-1200 Wireless Input Unit is designed for use by qualified medical personnel in a medical facility such as a hospital or clinic on all patient populations including adult, neonate, infant, child and adolescent subgroups.
Device Story
WEE-1200 Wireless Input Unit acquires, stores, and transmits physiological signals (EEG, ECG, EMG, EOG, respiration, SpO2) for EEG and sleep studies. Electrodes connect to a mini junction box linked to a patient-worn or bedside telemetry unit (ZB-120A). Data transmits via wired or wireless LAN (IEEE 802.11a/b/g/n) to a Nihon Kohden electroencephalograph (e.g., EEG-1200A) for display. SpO2 data is also displayed on the telemetry unit LCD. Used by qualified medical personnel in hospitals or clinics. Output allows clinicians to visualize waveforms to aid in diagnosing neurological and sleep disorders. Enhanced transmission performance and dual-band operation compared to predicates.
Clinical Evidence
No clinical testing performed. Substantial equivalence established through bench testing, including software verification/validation (unit, integration, system tests), wireless performance validation (coexistence, network, operation), and compliance with safety standards (AAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-2-26, ISO 80601-2-61).
Technological Characteristics
Telemetry unit (ZB-120A) with mini junction box (JE-125AK/225AK/922A). Wireless: IEEE 802.11a/b/g/n (2.4/5 GHz), WPA/WPA2 security. SpO2 adapter (JL-570T) uses two-wavelength light absorption. 16-bit A/D resolution. Input impedance 200M ohm. Battery-operated (10-30 hours). Standards: AAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-2-26, ISO 80601-2-61.
Indications for Use
Indicated for acquisition, storage, recording, and transmission of cerebral and extracerebral activity for EEG and sleep studies. Used for patients of all ages (adult, neonate, infant, child, adolescent) in medical facilities for diagnosis of sleep disorders, epilepsies, and related conditions.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
Predicate Devices
Nihon Kohden Wireless Input Unit WEE-1000A (K033475)
Nihon Kohden EEG-1200A with JE-120A Multi Channel Electrode Junction Box (K113117)
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or movement. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 25, 2017
Nihon Kohden Corporation % Tom Bento Sr. Vice President, Quality and Regulatory Nihon Kohden America, Inc. 15353 Barranca Parkway Irvine, California 92618
Re: K171124
Trade/Device Name: Nihon Kohden Wireless Input Unit WEE-1200 Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWQ, OLV Dated: July 31, 2017 Received: August 1, 2017
Dear Mr. Bento:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171124
Device Name Nihon Kohden Wireless Input Unit WEE-1200
#### Indications for Use (Describe)
The WEE-1200 Wireless Input Unit is intended to acquire, store, record, and transmit the cerebral and extracerebral activity for EEG and sleep studies. The device should be used together with Nihon Kohden specified electroencephalograph via wired or wireless communication to display EEG on the electroencephalograph screen. These data may be used by the clinician in sleep disorder, epilepsies and other related disorders as an aid in diagnosis. The WEE-1200 Wireless Input Unit is designed for use by qualified medical facility such as a hospital or clinic on all patient populations including adult, neonate, infant, child and adolescent subgroups.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Nihon Kohden Wireless Input Unit WEE-1200
| Submitter: | Nihon Kohden Corporation |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 1-31-4 Nishiochiai, Shinjuku-Ku<br>Tokyo, Japan 161-8560 |
| Phone number: | 81-3-5996-8020 |
| Contact person: | Tom Bento |
| Phone number: | (949) 680-9048 |
| Fax number: | (913) 273-0732 |
| Date prepared: | April 14, 2017 |
| Device name: | Nihon Kohden Wireless Input Unit WEE-1200 |
| Common name: | Electroencephalograph (EEG) |
| Primary product codes: | GWQ, OLV |
| Regulation numbers: | 21 CFR 882.1400 |
| Substantial equivalence claimed to: | Nihon Kohden Wireless Input Unit WEE-1000A (K033475)<br>Nihon Kohden EEG-1200A with JE-120A Multi Channel<br>Electrode Junction Box (K113117) |
#### Description:
The WEE-1200 Wireless Input Unit is a system that can be used in routine EEG testing, epilepsy monitoring, or PSG testing to send the measured EEG waves, ECG waves, EMG waves, EOG waves, and respiration waves from the electroencephalograph by wired or wireless LAN. The electrodes are connected to a mini junction box which is connected to the telemetry unit that is worn by patient or placed near patient.
The measured data can also be stored in the telemetry unit of the WEE-1200 Wireless Input Unit.
Additionally, by connecting a SpO2 sensor and SpO2 adapter to the telemetry unit, SpO2 value and waveform can be measured and displayed on both the electroencephalograph screen and the telemetry unit LCD display.
{4}------------------------------------------------
#### Indications for Use:
The WEE-1200 Wireless Input Unit is intended to acquire, store, record, and transmit the cerebral and extracerebral activity for EEG and sleep studies. The device should be used together with Nihon Kohden specified electroencephalograph via wired or wireless communication to display EEG on the electroencephalograph screen. These data may be used by the clinician in sleep disorder, epilepsies and other related disorders as an aid in diagnosis. The WEE-1200 Wireless Input Unit is designed for use by qualified medical personnel in a medical facility such as a hospital or clinic on all patient populations including adult, neonate, infant, child and adolescent subgroups.
## Technological Characteristics - Substantial Equivalence Discussion
The WEE-1200, just like its predicate WEE-1000A (K033475), is a device intended to be used together with Nihon Kohden electroencephalographs such as EEG-1200A (K080546) to transmit the measured data via wired LAN. The WEE-1200 has same measuring parameters as the predicates.
The new SpO2 adapter, JL-570T, that can be used with the subject device for SpO2 measurement is substantially equivalent to the previously cleared SpO2 adapter. JL-550T1 (K113117, EEG-1200A with JE-120A), with only a few minor differences that do not raise any new safety and effectiveness issues.
When compared with the WEE-1000A (K033475), the subject device has an enhanced transmission performance: the maximum transmission throughput is increased and dualband operation (IEEE802.11 a/b/g/n) is available on the subject device. The wireless LAN security adopted on the predicate device was Nihon Kohden original protocol while the subject device utilizes 802.11 WPA/WPA2. The wireless transmission performance of the subject device is validated according to the FDA guidance "Radio Frequency Wireless Technology in Medical Devices" for its co-existence features and network performance.
| | EEG-1200A with JE-<br>120A Junction Box<br>Predicate Device<br>(K113117) | WEE-1000A<br>Wireless Input Unit<br>Predicate Device<br>(K033475) | WEE-1200<br>Wireless Input Unit<br>New Device |
|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | EEG-1200 series<br>Neurofax is intended to<br>record, measure and<br>display cerebral and<br>extraccrebral activity for<br>EEG and Sleep Studies.<br>These data may be used<br>by the clinician in sleep<br>disorder, epilepsies and<br>other related disorders as<br>an aid in diagnosis.<br>The device is intended for<br>use by medical personnel<br>in any location within a | The device is intended to<br>acquire, store, and<br>transfer biophysical<br>parameters to EEG<br>machines for the purpose<br>of assisting the diagnosis<br>of neurological and sleep<br>disorders, measurement<br>and display of cerebral<br>and extracerebral activity<br>for EEG and sleep<br>studies. These data may<br>be used by clinician in | The WEE-1200 Wireless<br>Input Unit is intended to<br>acquire, store, record, and<br>transmit the cerebral and<br>extracerebral activity for EEG<br>and sleep studies. The<br>device should be used<br>together with Nihon Kohden<br>specified<br>electroencephalograph via<br>wired or wireless<br>communication to display<br>EEG on the<br>electroencephalograph<br>screen. These data may be |
| | EEG-1200A with JE-<br>120A Junction Box | WEE-1000A<br>Wireless Input Unit | WEE-1200<br>Wireless Input Unit |
| | Predicate Device<br>(K113117) | Predicate Device<br>(K033475) | New Device |
| | medical facility,<br>laboratory, clinic or<br>nursing home or outside<br>of a medical facility under<br>direct supervision of a<br>medical professional. | sleep disorders,<br>epilepsies and other<br>related disorders as a<br>diagnostic tool.<br>The device is intended for<br>use by medical personnel<br>in any location within a<br>medical facility,<br>physician's office,<br>laboratory, clinic or<br>nursing home or outside<br>of a medical facility under<br>supervision of a medical<br>professional. The device<br>will be available in all<br>populations, including<br>pediatrics. | used by the clinician in sleep<br>disorder, epilepsies and<br>other related disorders as an<br>aid in diagnosis.<br>The WEE-1200 Wireless<br>Input Unit is designed for use<br>by qualified medical<br>personnel in a medical facility<br>such as a hospital or clinic on<br>all patient populations<br>including adult, neonate,<br>infant, child and adolescent<br>subgroups. |
| Product used with | Nihon Kohden EEG-<br>1200A (K080546) | Nihon Kohden EEG-<br>1200A (K080546) | Nihon Kohden EEG-1200A<br>(K080546) |
| Measurement Parameters | EEG, ECG, EMG,<br>EOG, SpO2 and CO2 | EEG, ECG, EMG, EOG,<br>Respiration, and SpO2 | EEG, ECG, EMG, EOG<br>respiration, and SpO2 |
| Telemetry Unit | | | |
| Telemetry Unit Model | N/A | ZB-101AA or ZB-102AA | ZB-120A |
| Wireless LAN | N/A | IEEE 802.11 b/g | IEEE 802.11a/b/g/n |
| Frequency band | N/A | 2.4GHz | 2.4 GHz or 5 GHz |
| Number of channels | N/A | 2.4 GHz: 11 channels (1<br>to 11ch) | 2.4 GHz: 11 channels (1 to<br>11ch)<br>5 GHz: 24 channels (36, 40,<br>44, 48, 52, 56, 60, 64, 100,<br>104, 108, 112, 116, 120, 124,<br>128, 132, 136, 140, 149, 153,<br>157, 161, 165 ch) |
| Modulation | N/A | DSSS (BPSK/QPSK):<br>1 & 2 Mbps<br>DSSS (CCK):<br>5 & 11 Mbps | Auto<br>(DSSS (BPSK/QPSK/<br>CCK)): 802.11b<br>OFDM (BPSK/QPSK/<br>16QAM/64QAM):<br>802.11a/g/n<br>(6/9/12/18/24/36/48/54<br>Mbps, MCS0/1/2/3/4/5/6<br>7) |
| Transmission power | N/A | < 5mW/MHz | 2.4GHz<br>< 10mW/MHz<br>5GHz<br>< 10mW/MHz |
| Wireless LAN security | N/A | NK Original | WEP, WPA, WPA2 |
| Battery operation time | N/A | 18 hours<br>(Using 10s Intermittent<br>transfer mode) | 10 hours (using internal<br>battery only)<br>30 hours (using internal<br>Battery and Extension unit) |
| | EEG-1200A with JE-<br>120A Junction Box | WEE-1000A<br>Wireless Input Unit | WEE-1200<br>Wireless Input Unit |
| | Predicate Device<br>(K113117) | Predicate Device<br>(K033475) | New Device |
| Applied wireless<br>transmission standards<br>and guidance | N/A | FCC 47CFR Part 15 | FCC 47CFR Part 15 Subpart C |
| | | Subpart B<br>FCC 47CFR Part 15<br>Subpart C | FCC 47CFR Part 15 Subpart B |
| | | | (FCC ID: PVH0941) |
| | | | FCC 47CFR Part 2 1093 |
| | | | (FCC ID: B6B0941) |
| | | | |
| Electrode Junction Box | | | |
| Junction Box Model | • Amplification unit<br>JE-120A | • Amplification unit<br>ZB-101AA/102AA | • Amplification unit<br>ZB-120A |
| | • Mini Junction Box<br>JE-125AK or JE-225AK<br>And JE-226AK/227AK/<br>228AK | • Mini Junction Box<br>JE-011A or JE-012A or<br>JE-013A | • Mini Junction Box<br>JE-125AK or JE-225AK<br>or JE-922A |
| Electrode jacks | Up to 64 for EEG<br>Up to 4 for bipolar input<br>Z electrode<br>COM | Up to 62 for EEG<br>Z electrode | Up to 64 for EEG<br>Up to 4 for bipolar input<br>Z electrode<br>COM |
| Reference electrode | A5 and A6 (C3 and C4)<br>or COM | A5 and A6 (C3 and C4) | A5 and A6 (C3 and C4)<br>or COM |
| Input impedance | 200M ohm | 200M ohm | 200M ohm |
| Noise | Less than 1.5 micro Vp-p | Less than 3.0 micro Vp-p | Less than 2.0 micro Vp-p |
| Rejection ratio | More than 110dB | More than 105dB | More than 105 dB |
| Low cut filter | 0.016Hz (TC 10 seconds)<br>or 0.08Hz (TC 2 seconds) | 0.08Hz (TC 2 seconds) | 0.08Hz (TC 2 seconds) |
| High cut filter | 3000Hz (-18dB/oct) | 60Hz (-18dB/oct) | 1200Hz (-18dB/oct) |
| Electrode impedance<br>check | Yes | Yes | Yes |
| A/D resolution | 16 bit (97nV/LSB) or<br>16 bit (388nV/LSB) | 16 bit (97nV/LSB) | 16 bit (97nV/LSB) or<br>16 bit (388nV/LSB) |
| Maximum Sampling<br>frequency | 10000 Hz | 200Hz | • 4000Hz<br>(Wired mode: Without any<br>condition<br>Wireless mode: limit of storage<br>of 40ch electrode )<br>• 2000Hz<br>(without any condition) |
| Main Unit | | | |
| Interface with EEG | 100Base-TX LAN<br>(On the QI-123AK) | 100Base-TX LAN<br>(On the ZR-101AK) | 100Base-TX LAN |
## Device comparison:
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
## Comparison of SpO2 adapters:
| | JL-550T1<br>(K113117) | JL-570T<br>(New) |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Measurement Parameters | Saturated oxygen (SpO2), Pulse rate (PR) | |
| SpO2 measuring specification | | |
| Measuring method | Two wavelength light absorption method | |
| Measuring Range | 0 to 100%SpO2 | |
| Measuring accuracy (rms) | ±2 %SpO2 (80 %SpO2 ≤ SpO2 ≤ 100 %SpO2 )<br>±3 %SpO2 (70 % SpO2 ≤SpO2 < 80 %SpO2 )<br>SpO2 accuracy is guaranteed at surrounding temperature of 18 to<br>40°C (64.4 to 104°F). | |
| Pulse rate measuring specification | | |
| Measuring Range | 30 to 300 beats/min | |
| Measuring accuracy (rms) | ±3% ± 1 1/min (30 to 300 beats/min) | |
| Dimensions and Weight | Width 34±10mm × Height 18±10mm × Depth 117±10mm<br>2500±100mm<br>95g±10% | 600mm±100mm<br>55±10g |
| Degrees of protection provided by enclosure | IPX1 | |
| Temperature<br>(Operating) | 10~40°C | |
| Temperature<br>(Storage/Transportation) | -20~65°C | |
| Humidity<br>(Operating) | 30~85%RH | |
| Humidity<br>(Storage/Transportation) | 10~95%RH | |
| Voltage input | 5V±5% | 3.3V±5% |
| Power consumption | 250mWTyp. | 30mWTyp. |
| Hardware specification of SpO2 measurement board | | |
| 1) LED driver | Peak current 0 to 60 mA | Peak current 0 to 40 mA |
| 2) I/V converter | Transform current from the photodiode to voltage | |
| 3) Filter | Band pass filter | |
| 4) Amplifier | Variable DC amplifier | |
| 5) Demodulation | Demodulate R signal and IR signal separately with low pass filter | |
| 6) A/D converter | Convert analog DC signals to digital ΔΣ-A/D 24 bit | |
| 7) CPU | ARM 7 | |
## Test Summary:
- i) Non-Clinical Testing
The software verification and validation testing for the WEE-1200 includes the following tests –
- Software Unit Test: Code inspection and static analysis .
- . Software Integration Test: transfer of data and control across a program's internal and external interfaces
{8}------------------------------------------------
- System Test: Comprehensive testing of all required functionality of the device (hardware and software)
The following tests on wireless technology performance of the device were performed following FDA Guidance, "Radio Frequency Wireless Technology in Medical Device".
- Wireless Coexistence Test: Performance validation in the presence of wireless . emitters in the intended use environment (hardware and software)
- Wireless Network Test: Verify and validate the ability to securely operate in a mixed . client enterprise 802.11a/b/g/n network, and to demonstrate the correct quality of service (hardware and software)
- Wireless Operation Test: Verify and validate the Wi-Fi operation of the device (hardware and software)
Standards compliance testing on the device includes:
- AAMI/ANSI ES 60601-1:2005/(R)2012 and A1:2012 Medical electrical equipment -● Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: General requirements . for safety and essential performance collateral standard: Electromagnetic compatibility - requirements and tests
- . IEC 60601-2-26:2012 Medical electrical equipment – Part 2-26: Particular requirements for the basic safety and essential performance of electroencephalographs
- . ISO 80601-2-61:2011 Medical electrical equipment – Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
- ii) Clinical Testing
No clinical testing on the subject device was necessary to show substantial equivalence to the predicate.
## Conclusion:
Device comparison and the results of the above listed performance testing indicate that the Nihon Kohden Wireless Input Unit WEE-1200 is substantially equivalent to the predicate devices, the Nihon Kohden Wireless Input Unit WEE-1000A (K033475) and Nihon Kohden EEG-1200A with JE-120A Multi Channel Electrode Junction Box (K113117), and raises no safety or effectiveness issues.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.