K143440 · Natus Medical Incorporated · GWQ · Apr 20, 2015 · Neurology
Device Facts
Record ID
K143440
Device Name
Natus Quantum
Applicant
Natus Medical Incorporated
Product Code
GWQ · Neurology
Decision Date
Apr 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Natus Quantum Amplifier is intended to be used as an electroencephalograph: to acquire, display, store and archive electrophysiological signals. The amplifier should be used in conjunction with Natus NeuroWorks™/SleepWorks™ software to acquire scalp and intracranial electroencephalographic (EEG) signals as well as polysomnographic (PSG) signals. The amplifier is designed to facilitate functional mapping using a Digital Switch Matrix. The Digital Switch Matrix portion of the headbox is a combination of hardware relays and software controls allowing the user (physician or technologist) to switch electrode pairs between the EEG recording amplifier and the external cortical stimulator for stimulus delivery. The Natus Quantum Amplifier is intended to be used by trained medical professionals, and is designed for use in clinical environments such as hospital rooms, epilepsy monitoring units, intensive care units, and operating rooms. It can be used with patients of all ages, but is not designed for fetal use.
Device Story
Natus Quantum Amplifier acquires electrophysiological signals (scalp/intracranial EEG, PSG, SpO2, pulse rate, pleth) via electrodes and sensors; digitizes and transmits data to PC running Natus NeuroWorks/SleepWorks software for display, storage, and review. Features internal Digital Switch Matrix (hardware relays/software controls) to toggle electrode pairs between EEG amplifier and external cortical stimulator for functional mapping. Used in hospitals, epilepsy monitoring units, ICUs, and ORs by physicians or technologists. Provides real-time signal visualization and impedance testing in patient room. Benefits include high-density channel recording (up to 256) and integrated stimulus delivery control, facilitating clinical diagnosis and monitoring of neurological conditions.
Clinical Evidence
Bench testing only. Verification and validation performed per design control processes, including signal quality and functionality verification tests. Compliance with IEC 60601-1, 60601-2-26, and 80601-2-61 standards confirmed.
Technological Characteristics
128-256 referential channels; 24-bit sampling resolution; 16,384 Hz max sampling frequency; >1000 MOhm input impedance. Connectivity via TCP/IP and USB. Includes Digital Switch Matrix for cortical stimulation. Modular wearable breakout box. Complies with IEC 60601-1, IEC 62366, and IEC 62304 standards.
Indications for Use
Indicated for patients of all ages (excluding fetal use) requiring acquisition, display, storage, and archiving of electrophysiological signals, including scalp/intracranial EEG and PSG, and functional mapping via digital switch matrix in clinical settings.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around a central graphic. The graphic consists of three stylized human profiles facing to the right, with flowing lines that suggest movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 20, 2015
Natus Medical Incorporated DBA Excel-Tech Ltd. (Xltek) Sanjay Mehta Quality and Regulatory Affairs Manager 2560 Bristol Circle Oakville, Ontario, Canada, L6H5S1
Re: K143440
Trade/Device Name: Natus Ouantum Amplifier Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWO. OLV. GYC Dated: March 18, 2015 Received: March 19, 2015
Dear Mr. Mehta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aquel -S
Carlos Peña, PhD, MS for Division Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K143440
Device Name Natus Quantum Amplifier
#### Indications for Use (Describe)
The Natus Quantum Amplifier is intended to be used as an electroencephalograph: to acquire, display, store and archive electrophysiological signals. The amplifier should be used in conjunction with Natus New Yorks™ software to acquire scalp and intracranial electroencephalographic (EEG) signals as well as polysomnographic (PSG) signals. The amplifier is designed to facilitate functional mapping using a Digital Switch Matrix. The Digital Switch Matrix portion of the headbox is a combination of hardware relays and software controls allowing the user (physician or technologist) to switch electrode pairs between the EEG recording amplifier and the external cortical stimulator for stimulus delivery.
The Natus Quantum Amplifier is intended to be used by trained medical professionals, and is designed for use in clinical environments such as hospital rooms, epilepsy monitoring units, intensive care units, and operating rooms. It can be used with patients of all ages, but is not designed for fetal use.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Natus Neurology. The word "natus" is written in blue, block letters, with a registered trademark symbol next to it. Below "natus" is the word "neurology" in smaller, italicized, black letters.
## 510K Summary
Date: April 15, 2015
Submitted by: Natus Medical Incorporated DBA Excel-Tech Ltd. (XLTEK) 2568 Bristol Circle Oakville, Ontario Canada L6H 5S1
Contact Person: Sanjay Mehta Senior Manager QA/RA Natus Medical Incorporated Tel .: (905) 829-5300 ext 388 Fax .: (905) 829-5304 E-mail: sanjay.mehta@natus.com
Proprietary Name: Natus Quantum
Common Name: Electroencephalograph
Regulation Number : 21CFR 882.1400
Classification Name: Full-montage standard electroencephalograph
Product code: GWQ, OLV
Device Class: II
Predicate Device: EMU128 (K040360); Neurolink IP 256 (K100683)
#### Description:
#### 1. Overview: Natus Quantum Amplifier
The Natus Quantum Amplifier is intended to be used as an electroencephalograph: to acquire, display, store and archive electrophysiological signals. The amplifier should be used in conjunction with Natus NeuroWorks™ software to acquire scalp and intracranial electroencephalographic (EEG) signals as well as polysomnographic (PSG) signals. The amplifier is designed to facilitate functional mapping using a Digital Switch Matrix. The Digital Switch Matrix portion of the headbox is a combination of hardware relays and software controls allowing the user (physician or technologist) to switch electrode pairs between the EEG recording amplifier and the external cortical stimulator for stimulus delivery.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for "natus. neurology". The word "natus" is in a larger, bold, blue font on the top line. The word "neurology" is in a smaller, italicized, black font on the second line.
The Natus Quantum Amplifier is intended to be used by trained medical professionals. and is designed for use in clinical environments such as hospital rooms, epilepsy monitoring units, intensive care units, and operating rooms. It can be used with patients of all ages, but is not designed for fetal use.
#### 2. Operating Principle of the Quantum Amplifier
The Natus Quantum amplifier is comprised of a base unit and several breakout boxes. It is part of a system that is made up of a personal computer, a photic stimulator, an isolation transformer, video and audio equipment, networking equipment, and mechanical supports. The amplifier also contains an internal switch matrix to allow for a connection to an external cortical stimulator.
EEG and other physiological signals, from scalp electrodes, grid or needle electrodes, and other accessories such as pulse oximeters can be acquired by the Natus Quantum amplifier. These signals are digitized and transmitted to the personal computer running the Natus NeuroWorks software. The signals are displayed on the personal computer and can be recorded to the computer's local storage or to remote networked storage for later review.
Image /page/4/Picture/6 description: The image shows three pieces of Natus Neurology equipment. On the left is a large, black, rectangular device with rounded corners and a small silver knob on the top. To the right are two smaller, square devices with gray tops and black sides. The Natus Neurology logo is visible on all three devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Natus Neurology. The word "natus" is in blue, block letters, with a small registered trademark symbol after the "s". Below that, the word "neurology" is in a smaller, italicized, black font. The logo is simple and clean, with a focus on the company name.
### Quantum features include
- . 128 referential channels (cascade-able up to 256).
- . 16 referential inputs per 128 can be configured as 8 differential inputs
- 16 DC channels,
- Pulse Oximeter including SpO2, Pulse Rate and Pleth signals .
- . Digital Trigger Input
- . Ability to initiate an impedance test, change the threshold, and view the results in the patient room
- . Digital Switch Matrix
- A small and lightweight wearable breakout box .
- . TCP/IP and USB connectivity
- . Patient-event switch interface on both the breakout box and base units
- . Photic stimulator interface for EEG applications
- . Modular pouch for belt or backpack harness
## System Setup Overview
The Natus Quantum amplifier connects to the Natus NeuroWorks (K090019) & Natus SleepWorks (K090277) software for the acquisition storage, analysis, and review of Electroencephalographic & Polysomnographic data in conjunction with synchronized digital video. The system overview is as follows.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for "natus neurology". The word "natus" is in a larger, blue font, while "neurology" is in a smaller, black font. The logo is simple and clean, with a focus on the company name.
Image /page/6/Figure/2 description: This image shows a diagram of a system for recording and digitizing a patient's physiological data. The diagram includes an acquisition PC connected to a network, which is connected to Natus Quantum Breakout Boxes. The breakout boxes are connected to electrodes and sensors placed on the patient's scalp, as shown in the "Patient Electrode Placement International 10-20 System (example)" diagram.
### Device-patient interaction Accessories List:
The table below lists all accessories to the subject device. Accessories (1) to (9) enter in contact with the patient. These sensors guarantees acquisition of the physiological signals and passively transfer them to the head box. Characteristics of the sensors vary and are described (cleared) under their respective 510K submissions (see table).
| | Description | Body contact location | Device connection |
|---|--------------------------------------------------------------------------|-----------------------------------------------|--------------------------------------------------------------|
| 1 | Reusable gold disk electrode<br>(K982053) | Scalp (according to 10-<br>20 & 10-10 system) | Referential and differential inputs<br>(labeled numerically) |
| 2 | Single Use Intracranial Grids, Strips<br>and Depth Electrodes. (K082474) | Intracranial recordings | Referential and differential inputs<br>(labeled numerically) |
| 3 | Body position sensor (K923033) | Thorax | DC input |
| 4 | Xactrace (K043132) | Respiratory belts | Differential input |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Natus Neurology. The word "natus" is written in blue, block letters. Below that, the word "neurology" is written in a smaller, italicized font. The logo is simple and professional.
| | | Thorax/Abdomen | |
|----|-----------------------------------|---------------------------------------|---------------------------------------------|
| 6 | Thermistor (K922112) | Nasal/Oral | DC input |
| 7 | Pulse Oximeter Sensor (K092101) | Finger | Channel labeled "oximeter/photic" |
| 8 | Snoring microphone (K941759) | Nasion/cheek/chin or side of the neck | DC input |
| 9 | Airflow Pressure Sensor (K922112) | Nasal | DC input |
| 10 | Photic Stimulator (K991903) | None | Channel labeled "oximeter/photic" |
| 11 | Cortical Stimulator (K072964) | Intracranial contacts | Amplifier Digital Switch Matrix input port. |
### Indications for Use
The Natus Quantum Amplifier is intended to be used as an electroencephalograph: to acquire, display, store and archive electrophysiological signals. The amplifier should be used in conjunction with Natus NeuroWorks™/SleepWorks™ software to acquire scalp and intracranial electroencephalographic (EEG) signals as well as polysomnographic (PSG) signals. The amplifier is designed to facilitate functional mapping using a Digital Switch Matrix. The Digital Switch Matrix portion of the headbox is a combination of hardware relays and software controls allowing the user (physician or technologist) to switch electrode pairs between the EEG recording amplifier and the external cortical stimulator for stimulus delivery.
The Natus Quantum Amplifier is intended to be used by trained medical professionals, and is designed for use in clinical environments such as hospital rooms, epilepsy monitoring units, intensive care units, and operating rooms. It can be used with patients of all ages, but is not designed for fetal use.
| Specification | Predicate Device<br>EMU128S (K040360) | Predicate Device<br>NeuroLink IP (K100683) | Subject Device<br>Natus Quantum |
|---------------------------------------------------------------------|--------------------------------------------|--------------------------------------------|--------------------------------------------------------------------------------------|
| Manufacturer | Excel-Tech Ltd.(Xltek) | Excel-Tech Ltd.(Xltek) | Excel-Tech Ltd.(Xltek) |
| Technological Characteristics<br>Analog Specifications EEG Channels | | | |
| EEG Channels | 128 | 64-256 | 64-256 |
| Reference<br>Channels | Dedicated separate<br>reference and ground | Dedicated separate reference<br>and ground | Dedicated separate reference<br>and ground |
| Input Impedance | >47 MOhms | > 100 MOhms | >1000 MOhm |
| Input Noise | < 2uV pk to pk @ full<br>bandwidth | < 1uV rms @ 1....70Hz<br>bandwidth | < 1.5uV pk to pk @ .1....100Hz<br>bandwidth<br>(<0.53uV rms@1....100Hz<br>bandwidth) |
### Comparison to Predicate Device
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Natus Neurology. The word "natus" is in blue, block letters, and the word "neurology" is in black, italicized letters. The logo is simple and modern.
| Maximum<br>Operational DC<br>input voltage<br>electrode offset | ±300V | ±300V | ±300V |
|----------------------------------------------------------------|----------------------|-------------------|-----------------------------------------------|
| Input Bias Current | < 20pA | < 1 nA | <1nA |
| Common mode<br>Rejection Ratio | >110dB@60Hz | >40dB@60Hz | >110dB@60Hz |
| Digital Specifications | | | |
| Sampling<br>Frequency | 256, 512, 1024, 2048 | 256, 512, 1024 Hz | 256, 512, 1024, 2048, 4096,<br>8192, 16384 Hz |
| Sampling<br>Resolution - EEG<br>channels | 22 bits | 16 bits | 24 bits |
| Sampling<br>Quantization -<br>EEG channels | 310 nV | 179 nV | 305nV |
| Storage Resolution<br>- EEG Channels | 16 bits | 16 bits | 16 bits |
For a detailed discussion covering any differences between devices and why they do not raised any new safety or effectiveness issues, the reviewer is referred to VOL_007_Substantial Equivalence discussion.
### Brief Summary of Performance Tests
#### Non-clinical:
Testing of the Natus Quantum was performed in compliance with Natus Corporation design control process. The validation was carried out as part of verification. Testing included:
| Test | Results |
|----------------------------------|---------|
| Signal Quality Verification Test | Pass |
| Functionality Verification Test | Pass |
It was found that the Natus Quantum system meets the design specification and performs as specified.
The device is in compliance with the following industrial standards
{9}------------------------------------------------
## Safety Standard of Compliance and normative references
| Standards | Title |
|---------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| CAN /CSA-C22.2 No. 60601-1: 08(R2013) +<br>C2:2011 | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential<br>performance |
| ANSI/AAMI ES60601-1:2005/(R)2012 +<br>C1:2009/(R)2012 and A2:2010/(R)2012 | |
| IEC 60601-1:2005 + C1:2006 and C2:2007,<br>Third Edition | |
| CENELEC EN 60601-1:2006 + A1:2013 | |
| IEC 60601-1-6:2010, Edition 3.0 | Medical electrical equipment - Part 1-6:<br>General requirements for basic safety and<br>essential performance - Collateral standard:<br>Usability |
| IEC 62366:2007, Edition 1.0 | Medical devices - Application of usability<br>engineering to medical devices |
| IEC 60601-2-26:2012, Edition 3<br>CENELEC EN 60601-2-26L2003, Edition 2 | Medical electrical equipment - Part 2-26:<br>Particular requirements for the safety of<br>electroencephalographs |
| EN ISO 80601-2-61:2011, Edition 1 | Medical electrical equipment - Part 2-61:<br>Particular requirements for basic safety and<br>essential performance of pulse oximeter<br>equipment |
## EMC Standard of Compliance and normative references
| Standards | Title |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2:2007, Edition 3.0 | Medical electrical equipment - Part 1-2:<br>General requirements for safety - collateral<br>standard: electromagnetic compatibility –<br>requirements and tests |
| IEC 61000-4-2:2008, ed 2.0 | Electromagnetic Compatibility (EMC) Part 4-2:<br>Testing and Measurement Techniques -<br>Electrostatic Discharge Immunity Test |
| IEC 61000-4-3 ed 3.0 with A1:2007+A2:2010 | Electromagnetic Compatibility (EMC) Part 4-3:<br>Testing and Measurement Techniques -<br>Radiated, Radio-frequency, Electromagnetic<br>Field Immunity Test |
| IEC 61000-4-4:2012, ed 3.0 | Electromagnetic Compatibility (EMC) Part 4-4:<br>Testing and Measurement Techniques -<br>Electrical Fast Transient/Burst Immunity Test |
| IEC 61000-4-5:2014, ed 3.0 | Electromagnetic Compatibility (EMC) Part 4-5:<br>Testing and Measurement Techniques - Surge<br>Immunity Test |
| IEC 61000-4-6 ed 2.0 with A1:2004 + A2:2006 | Electromagnetic Compatibility (EMC) Part 4-6:<br>Testing and Measurement Techniques -<br>Immunity to Conducted Disturbances, Induced<br>by Radio-frequency Fields |
| IEC 61000-4-8:2009, ed 2.0 | Electromagnetic Compatibility (EMC) Part 4-8:<br>Testing and Measurement Techniques -<br>Power Frequency Magnetic Field Immunity<br>Test |
| IEC 61000-4-11:2004, ed 2.0 | Electromagnetic Compatibility (EMC) Part 4-<br>11: Testing and Measurement Techniques -<br>Voltage Dips, Short Interruptions and Voltage<br>Variations Immunity Tests |
| IEC 61000-3-2:2014, ed 4.0 | Electromagnetic Compatibility (EMC) Part 3-2:<br>Limits - Limits for Harmonic Current Emissions |
| IEC 61000-3-3:2013, ed 3.0 | Electromagnetic Compatibility (EMC) Part 3-3:<br>Limits - Limitation of Voltage Changes,<br>Voltage Fluctuations and Flicker in Public Low-<br>voltage Supply Systems |
| CISPR 11 ed 5.0 with A1:2010 | Industrial, Scientific and Medical (ISM) Radio-<br>Frequency Equipment - Electromagnetic<br>Disturbance Characteristics - Limits and<br>Methods of Measurement |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Natus Neurology. The word "natus" is in blue, blocky, sans-serif font. Below it, in a smaller, italicized, sans-serif font, is the word "neurology" in black. The logo is simple and modern.
Quality System Compliance:
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Natus Neurology. The word "natus" is in a larger, blue, sans-serif font. Below it, the word "neurology" is in a smaller, italicized, blue, sans-serif font. The logo is simple and clean, with a focus on the company name.
- . ISO 13485:2003 Medical devices - Quality management systems - Requirements for regulatory purposes.
- SOR/98-282 Canadian Medical Device Regulations .
- . 21 CFR Part 820 US Food and Drug Administration's Quality System Requilation
- . 93/42/EEG European Medical Device Directives
- . ISO 14971:2007 Medical Devices - Application of Risk Management to Medical Devices
- EN 980:2008 Medical Devices Symbols for Use of labeling of medical device .
- EN 1041:2008 Information supplied by the manufacturer of medical devices .
- ISO 15223-1:2012,Medical Devices Symbols to be Used With Medical Device Labels, Labeling, . and Information to be Supplied - Part 1: General requirements
- . IEC 62304:2006 Medical device software - Software life-cycle processes
### Conclusions
The substantial equivalence of the Natus Quantum with EMU128S and Neurolink IP products was demonstrated by testing in compliance with the Design Control process. The intended use and technology of the Natus Quantum is similar to that of the predicate device XLTEK-EMU128S and Neurolink IP. Verification and Validation was performed to ensure no new questions of safety or effectiveness are raised.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.