The Mitsar-EEG is intended to acquire, display and store the electrical activity of a patient's brain obtained by placing two or more electrodes on the head to aid in diagnosis.
Device Story
Mitsar-EEG is a diagnostic electroencephalograph system; acquires, displays, and stores brain electrical activity via scalp electrodes. System comprises biosignal amplifier, USB cable, USB security dongle, and EEG Studio software; operates on stationary PC or laptop. Used in hospitals, clinics, and research institutes by qualified medical personnel. Device does not contact patients directly; electrodes/caps are user-supplied. System provides raw EEG data for clinician review; does not provide automated diagnostic conclusions or treatment recommendations. Results must be interpreted alongside other clinical findings. Benefits include high-fidelity signal acquisition with low noise and high input impedance, aiding clinical diagnosis of neurological conditions.
Clinical Evidence
No clinical data or animal studies were required. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1, ANSI/AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), and mechanical/acoustic testing (push, impact, drop tests). Software verification and validation were performed per FDA guidance.
Technological Characteristics
Class II electroencephalograph; frequency band 0.16-70 Hz; high input impedance; low noise floor; high offset tolerance. System includes biosignal amplifier, USB connectivity, and security dongle. Non-sterile. Software-based signal processing. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-26.
Indications for Use
Indicated for patients requiring acquisition, display, and storage of brain electrical activity to aid in diagnosis. No specific age or gender restrictions; intended for use by qualified medical personnel.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
Predicate Devices
NicoletOne System V32 Amplifier with Oximetery (K061908)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked one behind the other, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 13, 2015
Mitsar Co., LTD c/o Leslie Sherlin Nova Tech EEG 8503 E. Keats Avenue Mesa, AZ 85209
Re: K143233
Trade/Device Name: Mitsar-EEG Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWO Dated: May 7, 2015 Received: July 16, 2015
Dear Dr. Sherlin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR [SELECT ONE: Part 801 [or, for IVDs only] Parts 801 and
{1}------------------------------------------------
8091); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S 同心
Carlos Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K143233
Device Name Mitsar-EEG
Indications for Use (Describe)
The Mitsar-EEG is intended to acquire, display and store the electrical activity of a patient's brain obtained by placing two or more electrodes on the head to aid in diagnosis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(K) SUMMARY
## 5.1 Submitter
| Date: | August 01, 2014 |
|---------------|----------------------------------------------------------------------------|
| Name: | Mitsar Co. Ltd. |
| Address: | Novorossiyskaya str. 21-2<br>194021 St. Petersburg<br>Russian Federation |
| Post address: | Politekhnicheskaya str., 6D<br>194021 St. Petersburg<br>Russian Federation |
| Contact person in the Russian Federation: | Ales Semchenkov |
|-------------------------------------------|-------------------|
| Telephone: | +7-812-297-90-13 |
| FAX: | +7-821-297-17-33 |
| E-Mail: | aas@mitsar-eeg.ru |
| Contact person in the U.S.: | Leslie Sherlin |
|-----------------------------|-------------------------------------------------------------|
| Address: | NOVA TECH EEG, INC.<br>8503 E. Keats Avenue, Mesa, AZ 85209 |
| Telephone: | (480) 219-3048 |
| FAX: | (419) 858-0407 |
| E-Mail: | lsherlinphd@novatecheeg.com |
## 5.2 Device
| Trade Name: | Mitsar-EEG |
|-------------------------|---------------------------------------------|
| Common Name: | Full-Montage Standard Electroencephalograph |
| Regulation Description: | Electroencephalograph |
| Product Code: | GWQ |
| Regulation Number: | 882.1400 |
| Device Class: | Class II |
| Classification Panel: | Neurology |
# 5.3 Predicate Devices
| 510(K)<br>Number | Classification<br>Product Code | Trade or Proprietary<br>or Model Name | Manufacturer |
|------------------|--------------------------------|------------------------------------------------------|-------------------------------------------------------------|
| K061908 | GWQ | NicoletOne System<br>V32 Amplifier with<br>Oximetery | VIASYS NeuroCare, Inc.<br>(NATUS, Inc.)<br>5225 Verona Road |
{4}------------------------------------------------
| | | | Madison, WI 53711 |
|---------|-----|---------------------|------------------------------------------------------------------------------|
| K946094 | GWQ | Cadwell Easy II EEG | Cadwell Laboratories Inc.<br>909 North Kellogg Street<br>Konnewick, WA 99336 |
### 5.4 Device Description
The device Mitsar-EEG is intended to acquire, display and store the electrical activity of a patient's brain obtained by placing two or more electrodes on the head to aid in diagnosis.
The medical device "Mitsar EEG" consists of biosignal amplifier, USB cable, USB dongle and software.
The USB cable is part of the device. It is a standard USB-cable, substantially equivalent to the legally marketed predicates.
The USB dongle is also part of the device. It is a standard security token to authenticate software.
The software is supplied by means of information media and intended for device functioning.
The medical system includes "Mitsar EEG" device and computer (stationary PC with uninterruptible power supply (UPS) or laptop with internal battery). The system may include a medical cart and isolation transformer.
The device does not come in direct contact with patients. Accessories that contact patients, such as electrodes and cap-type electrodes, are the same as used with the legally marketed devices or are comprised of the same components materials as legally marketed accessories. Electrodes and cap-type electrodes are not supplied together with the Mitsar-EEG.
The device is intended for use in functional diagnostics wards and departments at out-patient clinics, hospitals, health research institutes, health centers and other medical institutions. Also, investigations can be performed outside of healthcare facilities, as long as they are led by qualified medical personnel.
The device does not draw any diagnostic conclusion. Recorded data is intended for use as an aid to diagnosis only. No medication or treatment is applied based on this data only. The results have to be considered only in conjunction with other clinical findings.
The device is not sterile. The device is intended for use by qualified medical personnel only and qualifies for exemption per 21 CFR 801 Subpart D «Prescription devices».
Required Components
- . EEG amplifier
{5}------------------------------------------------
- USB-cable
- USB dongle
- EEG Studio software ●
- User manuals ●
### 5.5 Indications for use
The Mitsar-EEG is intended to acquire, display and store the electrical activity of a patient's brain obtained by placing two or more electrodes on the head to aid in diagnosis.
Intended use of the Mitsar-EEG is substantially equivalent to the intended use of legally marketed predicate devices.
#### 5.6 Comparison of technological characteristics with the predicate devices
Comparison of Mitsar-EEG and Cadwell Easy II K946094, and NicoletOne System V32 Amplifier with Oximetery K061908. The primary predicate device is NicoletOne System V32 Amplifier with Oximetery.
#### Channels
The device Mitsar-EEG contains fewer EEG-channels than the predicate devices do. However, according to Minimum Technical Requirements for Performing Clinical Electroencephalography by American Clinical Neurophysiology Society found in the ACNS Guideline 1: "All 21 electrodes and placements recommended by the International Federation of Clinical Neurophysiology (IFCN; Jasper HH, 1958, 1983) should be used". It is sufficient for the 10-20 System of electrodes placement officially recommended by IFCN: "The 10-20 System is the only one officially recommended by the IFCN. It is the most commonly used existing system, and it should be used universally."
#### Frequency band
Frequency band of Mitsar-EEG (0.16 - 70 Hz) is the same as frequency band of Cadwell Easy II (0,16 - 70 Hz), but more narrow than NicoletOne System V32 (0,053 - 500 Hz). The frequency band 0.16 - 70 Hz is sufficient for registration of EEG.
#### Input impedance
The input impedance of the device is greater than predicate devices, therefor Mitsar device introduces less distortion in the measured signal.
The difference in this characteristic does not affect the safety of the device.
#### Noise
The noise level of Mitsar-EEG lower than predicates have therefore the recording on this device will be better.
The difference in this characteristic does not affect the safety of the device.
{6}------------------------------------------------
### Offset tolerance
The offset tolerance of the device is greater than predicates, which allows using the device at a high electrode polarization.
The difference in this characteristic does not affect the safety of the device.
The safety class of the Mitsar-EEG is Class II and is equal to safety class of primary predicate device NicoletOne System V32 Amplifier.
Value of sterility, presence of calibration signal, the LED impedance indicators and touch proof connectors are the same for all devices.
The Mitsar-EEG is substantially equivalent to legally marketed predicate devices and is safe and effective for its intended use, and performs as well as, or better than the predicate devices.
#### 5.7 Performance data
The following performance data were provided in support of the substantial equivalence.
#### Biocompatibility
The device doesn't come into direct or indirect contact with patients and it supplied without electrodes.
#### Electrical safety and electromagnetic compatibility (EMC)
The tests have been performed by the accredited laboratories and show full compliance with standards below.
The device under consideration has passed the tests according to IEC 60601-1.
Analysis of deviations from IEC 60601-1:2005 in ANSVAAMI ES60601-1:2005 do not affect safety or effectiveness of the system «Mitsar-EEG». Therefore, the conformation with ANSI/AAMI ES60601-1:2005 can be declared.
The device under consideration has passed the tests according to IEC 60601-1-2 and IEC-60601-2-26.
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, «Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices» and «General Principles of Software Validation».
#### Mechanical and acoustic testing
Push test, Impact test, Drop test according to IEC 60601-1.
{7}------------------------------------------------
## Animal Study
This submittal does not include data on animal testing. Therefore, this section is not applicable.
#### Clinical Studies
Clinical studies were not required to demonstrate that this device is at least as safe and effective as the identified predicate device.
#### 5.8 Conclusions
Based on the above, Mitsar Co. Ltd. concludes, that Mitsar-EEG is substantially equivalent to legally marketed predicate devices and is safe and effective for its intended use, and performs as well as, or better than, the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.