X10 HEADSET WITH X-SERIES BASIC SOFTWARE, X24 HEADSET WITH X-SERIES BASIC SOFTWARE
K131383 · Advanced Brain Monitoring, Inc. · GWQ · Nov 27, 2013 · Neurology
Device Facts
Record ID
K131383
Device Name
X10 HEADSET WITH X-SERIES BASIC SOFTWARE, X24 HEADSET WITH X-SERIES BASIC SOFTWARE
Applicant
Advanced Brain Monitoring, Inc.
Product Code
GWQ · Neurology
Decision Date
Nov 27, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K131383 · Nov 27, 2013
X10 HEADSET WITH X-SERIES BASIC SOFTWARE, X24 HEADSET WITH X-SERIES BASIC SOFTWARE
Advanced Brain Monitoring, Inc.
Clinical research study data
Data from a research study involving children (ages 6-8) and adults were analyzed to demonstrate that the X-Series System can successfully acquire EEG signals in pediatric patients and that the data quality is equivalent to that of an adult population.
Pediatric performance; Signal quality; Clinical research cohort
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Comparative analysis of clinical research data
Children ages 6-8 (n=7) and adults (n=7); Sample Size: 14
Adult population
EEG signal quality and successful acquisition
Indications for Use
The X-Series System is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, transmit, display and store physiological signals from patients ages 6 and older. The X-Series system requires operation by a trained technician. The X-Series System acquires, transmits, displays and stores electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG), and accelerometer signals. The X-Series System only acquires and displays physiological signals, no claims are being made for analysis of the acquired signals with respect to the accuracy, precision and reliability.
Device Story
X-Series System (X10/X24) is a physiological data acquisition platform. Input: EEG, EOG, ECG, EMG, and 3-D accelerometer signals via battery-powered headset. Operation: Trained technician applies sensors using synapse cream; device transmits signals wirelessly (Bluetooth 2.0) to a receiving unit. Software: X-Series Basic Software monitors signal quality during acquisition to ensure proper sensor application. Output: Signals saved in EDF+ format for analysis by third-party software (e.g., Persyst). Context: Used in home, clinic, or research environments. Benefit: Enables ambulatory physiological monitoring; provides high-quality signal acquisition for subsequent physician review.
Clinical Evidence
Clinical study included 14 subjects (7 children ages 6-8, 7 adults). Technicians performed setup and resolved impedance issues using software indicators. Results: 100% of sessions yielded appropriate EEG recordings; data quality in children was equivalent to adults. No discomfort or agitation reported. Bench testing confirmed equivalence to predicate for EEG, ECG, and actigraphy performance.
Indicated for patients ages 6 and older requiring acquisition, transmission, display, and storage of EEG, EOG, ECG, EMG, and accelerometer signals in home, healthcare, or clinical research settings. Requires operation by a trained technician.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
K131383
### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
DATE: November 26, 2013
## SUBMITTER:
·
Advanced Brain Monitoring 2237 Faraday Avenue, Suite 100 Carlsbad, CA 92008 T 760.720.0099 . F 760.720.3337
**NOV** 27 2013
## PRIMARY CONTACT PERSON:
Adrienne Lenz, RAC Member Pathway Regulatory Consulting, LLC T 262-290-0023
## SECONDARY CONTACT PERSON:
Dan Levendowski President and Co-founder Advanced Brain Monitoring, Inc.
DEVICE:
TRADE NAME: X-Series System
COMMON/USUAL NAME: X10 / X24
CLASSIFICATION NAMES: 882.1400 Electroencephalograph
PRODUCT CODE: GWQ, OMC
PREDICATE DEVICE(S):
K130013 X4 System
{1}------------------------------------------------
#### DEVICE DESCRIPTION:
The X-Series System is indicated for acquiring, transmitting, displaying and storing physiological data in patients. It can be used with ambulatory patients in the home, health care facility, or clinical research environment. The X-Series system requires operation by a trained technician.
The X-Series System.is comprised the X10 and X24 Headsets and accessories, Synapse Cream, X-Series Basic Software and BT receiving unit. The X-Series Basic Software is also compatible with the Models X4-E, and X4-M (K130013) when used in wireless mode.
The X-Series System combines hardware, firmware and software to acquire physiological signals. It acquires physiological data through a battery powered headset worn by the patient and provides a flexible platform for applying sensors using synapse cream and acquiring signals from multiple locations on the head or body, transmitting and recording the signals and providing visual indications to ensure high quality data are obtained.
Model X24 provides for acquisition of twenty channels of electroencephalography (EEG) and four optional channels connected to two sensors via a dual-lead connector. Model X10 provides for acquisition of nine channels of electroencephalography (EEG) and an optional channel connected to two sensors via a dual-lead connector. Both models measure movement and position measured via a 3-D accelerometer. The device is designed so it can be affixed by a technician and displays the signals via a wireless connection during acquisition. The X-Series Basic software monitors signal quality to ensure that the sensors are properly applied and that high quality signals are being acquired.
The X-Series Basic software provides a means to: a) initiate a study and track patient information, b) acquire and wirelessly transmit signals from the device, c) visually inspect the signal quality.
The acquired signals are saved in a universal data format (European Data Format Plus, EDF+) that is intended to be analyzed by a Physician using FDA cleared third party software, i.e. Persyst Software (K011397).
#### INTENDED USE:
The X-Series System is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, transmit, display and store physiological signals from patients ages 6 and older. The X-Series system requires operation by a trained technician. The X-Series System acquires, transmits, displays and stores electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG), and accelerometer signals. The X-Series System only acquires and displays physiological signals, no claims are being made for analysis of the acquired signals with respect to the accuracy, precision and reliability.
{2}------------------------------------------------
## TECHNOLOGY:
The X-Series System uses the same fundamental technology as the X4 System for most features including the electrophysiological (EEG, EOG, ECG, EMG, ), wireless acquisition, and actigraphy. The technologies used in the X-Series are used in the same manner as the predicate products · and do not raise new questions of safety and effectiveness.
| Specification | X-Series System | X4 System (K120447) |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The X-Series System is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, transmit, display and store physiological signals from patients ages 6 and older. The X-Series system requires operation by a trained technician. The X-Series System acquires, transmits, displays and stores electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG), and accelerometer signals. The X-Series System only acquires and displays physiological signals, no claims are being made for analysis of the acquired signals with respect to the accuracy, precision and reliability. | The X4 System is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, record, transmit and display physiological signals from adult patients. The X4 System acquires, records, transmits and displays electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG), accelerometer, acoustical and photoplethesmographic signals. The X4 system only acquires and displays physiological signals, no claims are being made for analysis of the acquired signals with respect to the accuracy, precision and reliability |
| Patient<br>Population | Ages 6 and older | Adults |
| User | Trained Technician | Trained technician or patient for recording of sleep data. |
| Anatomical<br>Sites | Scalp/Chest | Forehead/Chest |
| Environment of<br>Use | Home (data acquisition)<br>Healthcare facility (data acquisition, analysis and reporting)<br>Clinical Research Environment | Home (data acquisition)<br>Healthcare facility (data acquisition, and analysis.)<br>Clinical Research |
## TABLE 1 COMPARISON OF X-SERIES TO PREDICATE DEVICE
{3}------------------------------------------------
| Specification | X-Series System | X4 System (K120447) |
|---------------------------------|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| User Interface | User control, visual indicators | User control, visual and audio<br>indicators |
| Accessories | EEG and ECG electrodes. | EEG and ECG electrodes. |
| Signals<br>Acquired | Scalp EEG 3-D actigraphy Optional channels<br>ECG/EEG/EOG/EMG | Scalp EEG Infra-red (IR) optical signal Microphone 3-D actigraphy Optional channel<br>ECG/EEG/EOG/EMG |
| Acquisition<br>modes | Monitor only | Record or Monitor |
| Power Supply | 2 to 4 240mAH 3.7 Li-Ion batteries | 1 x 600 mAH 3.7V Li-ION battery |
| Battery<br>Charging | Via JED Connector connected to USB<br>port or USB wall charger | Via USB cable connected to USB port<br>or USB wall charger |
| Typical<br>Charging Time | 0.5-5.0 hours | 0.5 - 3.0 hours |
| Operating<br>Time | Monitoring Days after Charge Hours<br>of Use 0-4 Days: 16 to 17 hours 5-10 Days: 14 to 15 hours | Recording Days after Charge Hours<br>of Use 0-4 Days: 13.0 to 15.5 hours 5-10 Days: 11.5 to 14.8 hours Monitoring Days after Charge Hours<br>of Use 0-4 Days: 6.5 to 7.8 hours 5-10 Days: 6.0 to 7.3 hours |
| Data Storage | Not used. The X-Series system does<br>not include a record mode where<br>data is stored on the recorder. | 2 GB Micro-SD memory card |
| File size per 8<br>hr recording | 512 MB | 72 MB |
| Dimensions | 5.0" long, 2.25"wide, 1.0" deep | 2.1" long, 1.5"wide, 0.75" deep |
| Weight | 3.9 ounces with two batteries | 2.5 ounces with battery |
.
:
:
・
·
{4}------------------------------------------------
:
:
・
.
| Specification | X-Series System | X4 System (K120447) |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cleaning | Cleaned and disinfected by rubbing<br>with isopropyl alcohol. | Cleaned and disinfected by rubbing<br>with isopropyl alcohol. |
| Wireless data<br>transfer | Blue tooth 2.0 | Blue tooth 2.0 |
| Maximum<br>Bluetooth<br>wireless<br>transfer<br>distance and<br>rate | Transfer distance 10 meters line of<br>sight, maximum transfer rate 3<br>Mbaud | Transfer distance 10 meters line of<br>sight, maximum transfer rate 3<br>Mbaud |
| EEG | | |
| Definition | Up to 20 differential or referential<br>channels | 3 or 4 differential or referential<br>channels |
| Signal<br>processing<br>techniques<br>(e.g. filtering,<br>etc.) | Sampling Rate: 256 s/s<br><br>0.1 Hz High Pass, hardware<br>100 Hz Low Pass, hardware | Sampling Rate: 256 s/s<br><br>0.1 Hz High Pass, hardware<br>100 Hz Low Pass, hardware |
| Accuracy,<br>variance and<br>error of<br>measurements,<br>in comparison<br>to standard<br>techniques of<br>measuring<br>identical<br>physiologic<br>variables | Sampling rate: 256Hz<br><br>Dynamic range: +/- 1000μV<br>Resolution 0.03 µV<br><br>Peak to peak noise: 3.7 µV (typical)<br><br>110dB Common Mode Rejection<br>Ratio (typically)<br><br>Input Impedance: 100GOhm | Sampling rate: 256Hz<br><br>Dynamic range: +/- 1000μV<br>Resolution 0.03 µV<br><br>Peak to peak noise: 3.7 µV (typical)<br><br>110dB Common Mode Rejection<br>Ratio (typically)<br><br>Input Impedance: 100GOhm |
.
1. September 19.
.
.
.
。
{5}------------------------------------------------
:
| Specification | X-Series System | X4 System (K120447) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Aux - EOG/EMG | | |
| Definition | Up to 4 optional single channel<br>either dual lead electrooculogram<br>(EOG) or electromyogram (EMG) | Optional single channel either dual<br>lead electrooculogram (EOG) or<br>electromyogram (EMG) |
| Signal<br>processing<br>techniques<br>(e.g. filtering,<br>etc.) | Sampling rate 256 Hz<br>0.1 Hz High Pass, hardware<br>100 Hz Low Pass, hardware | Sampling rate 256 Hz<br>0.1 Hz High Pass, hardware<br>100 Hz Low Pass, hardware |
| Accuracy,<br>variance and<br>error of X4<br>System<br>measurements,<br>in comparison<br>to standard<br>techniques of<br>measuring<br>identical<br>physiologic<br>variables | ECG:<br>Dynamic range: +/- 2000μV<br>Resolution: 0.06 μV<br>Peak to peak noise: 4.2 μV<br>EMG/EOG:<br>Dynamic range: +/- 1000μV<br>Resolution 0.03 μV<br>Peak to peak noise: 3.7 μV (typical) | EOG:<br>Dynamic range: +/- 2000μV<br>Resolution: 0.06 μV<br>Peak to peak noise: 4.2 μV<br>EMG:<br>Dynamic range: +/- 1000μV<br>Resolution 0.03 μV<br>Peak to peak noise: 3.7 μV (typical) |
.
:
.
{6}------------------------------------------------
.
| Specification | X-Series System | X4 System (K120447) |
|--------------------------------------------------------|----------------------------------------------|----------------------------------------------|
| Aux.- ECG | | |
| Definition | Dual lead electrocardiogram | Dual lead electrocardiogram |
| | Number: Single Channel (optional) | Number: Single Channel (optional) |
| | Bandwidth: 0.1 to 100Hz -3dB<br>bandwidth | Bandwidth: 0.1 to 100Hz -3dB<br>bandwidth |
| | Notch Filters: 50, 60Hz, software | Notch Filters: 50, 60Hz, software |
| | Common Mode Rejection Ratio:<br>110dB | Common Mode Rejection Ratio:<br>110dB |
| | Input Impedance: 100GOhm | Input Impedance: 100GOhm |
| | Input Range: +/-4mV | Input Range: +/-4mV |
| | Accuracy: 16 bits over input range | Accuracy: 16 bits over input range |
| | Noise: <5μVpp (typically 4.2μVpp) | Noise: <5μVpp (typically 4.2μVpp) |
| | Maximum Electrode DC Offset: +/-<br>125mV | Maximum Electrode DC Offset: +/-<br>125mV |
| Signal<br>processing | Hardware filtering: 0.1Hz -100Hz<br>bandpass | Hardware filtering: 0.1Hz -100Hz<br>bandpass |
| techniques<br>(e.g. filtering,<br>etc.) | Software Filtering: 50, 60Hz,<br>software | Software Filtering: 50, 60Hz,<br>software |
| Accuracy, | Noise: typical 4.2μVpp | Noise: typical 4.2μVpp |
| variance and<br>error of X4 | Dynamic range: +/- 2000μV | Dynamic range: +/- 2000μV |
| System<br>measurements,<br>in comparison | Resolution: 0.06 μV | Resolution: 0.06 μV |
| to standard<br>techniques of<br>measuring<br>identical | | |
| physiologic<br>variables | | |
. '
·
24
{7}------------------------------------------------
| Specification | X-Series System | X4 System (K120447) | | | | | | | | | | | | | | | | | | | | | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|
| Accelerometer | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Definition | Three channels (X x Y x Z) used by<br>software to measure movement and<br>position<br>Dynamic Range: -180° to 180° | Three channels (X x Y x Z) used by<br>software to measure movement and<br>position<br>Dynamic Range: -180° to 180° | | | | | | | | | | | | | | | | | | | | | | | | |
| Signal<br>processing<br>techniques<br>(e.g. filtering,<br>etc.) | -Sampled 100Hz, downsampled to<br>10Hz<br>- X x Y x Z @ 10 s/s<br>- Resolution nominal 10 bit at 2g<br>- Actual output range for -90 to 90<br>degrees is 12 bit | -Sampled 100Hz, downsampled to<br>10Hz<br>- X x Y x Z @ 10 s/s<br>- Resolution nominal 10 bit at 2g<br>- Actual output range for -90 to 90<br>degrees is 12 bit | | | | | | | | | | | | | | | | | | | | | | | | |
| Accuracy,<br>variance and<br>error of X4<br>System<br>measurements,<br>in comparison<br>to standard<br>techniques of<br>measuring<br>identical<br>physiologic<br>variables | Position accuracy typically +/- 3.0<br>degrees, maximum +/- 5.0 degrees<br>in the +/- 60 degrees range<br><br>ParameterX-Series Sensitivity10 bit / 2g Non-linearity+/-0.5 %FS Cross-Axis+/-1 % Zero-Level+/-0.35-0.4 mg Sensitivity<br>change due<br>temperatureq+/-0.01 LSB/°C | | | | | | | | | | | | | Position accuracy typically +/- 3.0<br>degrees, maximum +/- 5.0 degrees<br>in the +/- 60 degrees range<br><br>ParameterX4 Sensitivity10 bit / 2g Non-linearity+/-0.5 %FS Cross-Axis+/-1 % Zero-Level+/-0.35-0.4 mg Sensitivity<br>change due<br>temperature+/-0.01 LSB/°C | | | | | | | | | | | | |
| Software | Software presents waveforms | Software presents waveforms | | | | | | | | | | | | | | | | | | | | | | | | |
#### DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
#### SUMMARY OF NON-CLINICAL TESTS:
Support for the substantial equivalence of the X-Series System was provided as a result of risk management and testing which included electrical and biological safety, performance and software tests. Testing has provided reasonable assurance of safety and effectiveness for the intended use and supports a determination of substantial equivalence.
Bench tests were conducted using an oscilloscope and signal generator to compare performance of the X-Series to the predicate device. These comparison tests confirm the EEG
{8}------------------------------------------------
and optional dual-lead (for ECG) analog signals between the X-Series and predicate devices were equivalent. The X-Series and X4 were placed in a jig that held the devices in eight unique angles to confirm that the measures obtained from the 3D actigraphs were equivalent. While the X-Series System is not a cardiac monitor as defined in the scope of ANSI/AAMI EC 13, the standard was referenced and requirements and test methods applicable to the electrocardiogram acquired and displayed by X-Series was applied. These tests demonstrated the equivalence of the X-Series system to the predicate device for EEG, ECG and actigraph performance.
The cleaning and disinfection procedures of the X-Series acquisition device have been validated by testing in accordance with AAMI TIR 12-94 Designing, Testing, and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A Guide for Manufacturers and AAMI TIR 30: 2003 A Compendium of Processes, Materials, Test Methods, and Acceptance Criteria for Cleaning Reusable Medical Devices. Acceptance criteria established by Advanced Brain Monitoring and the TIR are ≥ log 3.0 reduction in bioburden after cleaning according to the instructions for use. All tests passed and demonstrate that the cleaning methods are appropriate for ensuring the X-Series acquisition device is clean between uses.
Biocompatibility was assessed following FDA Guidance "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing' (Replaces #G87-1 #8294) (blue book memo)". Per Table 1 of this guidance, the X-Series System is a Surface device with limited contact (Category A, less than 24 hours contact). As such, cytotoxicity, sensitization, and irritation or intracutaneous reactivity tests were selected. Test methods were in compliance with the corresponding standard from the ISO 10993 series for Biological Evaluation of Medical Devices. Test articles were comprised of all patient contacting components of the X-Series System, including the X-Series Strip with disposable foam and synapse cream, X-Series Strap and X-Series enclosure. Results of biocompatibility testing are summarized in table 2 and demonstrate that materials of the device in contact with the patient are biocompatible.
{9}------------------------------------------------
| ISO 10993<br>Standard | Test | Results | Conclusions |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| ISO 10993-5 | Cytotoxicity<br>Agar Diffusion Test - ISO Direct<br>Contact<br>The test article was applied<br>directly to the surface of the<br>agar of mammalian cell culture<br>(mouse fibroblast L929) and<br>the cells were incubated for 48<br>$\pm$ 2 hours at 37 $\pm$ 1°C. The<br>biological reactivity of the cells<br>following the exposure to the<br>test article was visually<br>observed with a microscope,<br>and graded on a scale of 0 to 4. | None of the cultures tested<br>with the test articles showed<br>any reactivity (grade 0). | Non-<br>cytotoxic |
| ISO 10993-10 | Sensitization<br>Buehler Sensitization Test - ISO<br>Direct Contact.<br>The test article, X Series<br>System, was evaluated for its<br>potential to produce skin<br>sensitization reactions<br>following topical application to<br>albino guinea pigs. | Irritation (skin reaction scores)<br>was absent from all the animals<br>(albino guinea pigs). All<br>animals gained weight during<br>the course of the study. No<br>overt signs of toxicity were<br>evident in any of the animals<br>during the course of the study.<br>No skin reaction scores were<br>observed in any of the test<br>animals. No skin reaction<br>scores were evident in the<br>negative control group: The<br>incidence of the responses<br>post-challenge were calculated<br>to be 0% at 24 and 48 hours. | Non-<br>sensitizer |
| ISO 10993<br>Standard | Test | Results | Conclusions |
| 10993-10 | Irritation<br>Primary Skin Irritation Test -<br>ISO Direct Contact<br>The X Series System was<br>evaluated for its potential to<br>produce Primary Skin Irritation<br>after a single topical 4 hour<br>application to the skin of New<br>Zealand White rabbits | All test animals gained body<br>weight during the course of the<br>study. No overt signs of<br>toxicity were evident in any of<br>the animals during the course<br>of the study. No signs of<br>erythema or edema were<br>present at the 24, 48, or the 72<br>hour observation point. None<br>of the control sites of any<br>animal at any of the<br>observation periods showed<br>signs of erythema or edema.<br>The Test Primary Irritation Index<br>(PII) was 0.0. | Non-irritant |
## TABLE 2 BIOCOMPATIBILITY RESULTS
27
.
.
{10}------------------------------------------------
The firmware in the X-Series System and the X-Series Basic software have been thoroughly tested through verification of specifications and validation, including software validation. Software documentation is available per FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document (May 11, 2005). The results of the verification and validation activities that have been performed demonstrate that the software meets requirements for safety, function, and intended use.
Electromagnetic compatibility and electrical safety testing of the X-Series was conducted following recognized standards Compliance to IEC 60601-2-26:2002 - Particular requirements for the safety of electroencephalographs was also demonstrated. The X-Series System has additionally been tested per IEC 60601-1-11: 2010 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment to demonstrate operational and mechanical performance is not negatively affected by various mechanical and environmental hazards encountered in the home-use environment. All test results are passing demonstrating the compliance of the X-Series system to FDA recognized standards for electro-medical equipment as defined in Table 3.
The disposable EEG electrodes used with the X-Series System have been tested to demonstrate their electrical performance. While the scope of the standard does not cover EEG electrodes, testing using applicable methods defined in chapter 4 of AAMI/ANS1 EC12:2000 (R)2010 Disposable ECG electrodes has been conducted, similarly to other cleared EEG electrodes
{11}------------------------------------------------
including Model 4310 PSarray2 EEG electrode (K020670). The results demonstrate electrical performance of the EEG electrodes used with the X-Series System is acceptable.
| TABLE 3: FDA RECOGNIZED CONSENSUS STANDARDS AND VOLUNTARY STANDARDS THE |
|-------------------------------------------------------------------------|
| X-SERIES SYSTEM CONFORMS WITH: |
| Standard Number | Standard Title |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-1:1988+A1:<br>1991+A2: 1995 | Medical Electrical Equipment - Part 1: General requirements for safety |
| IEC 60601-1-2: 2007 | Medical Electrical Equipment Part 1-2: Collateral standard: Electromagnetic<br>compatibility - requirements and tests |
| ISO 10993-1: 2009 | Biological evaluation of medical devices Part 1 |
| IEC 60601-1-11: 2010 | Medical electrical equipment - Part 1-11: General requirements for basic<br>safety and essential performance - Collateral Standard: Requirements for<br>medical electrical equipment and medical electrical systems used in the<br>home healthcare environment |
| IEC 60601-2-26:2002 | Particular requirements for the safety of electroencephalographs |
#### SUMMARY OF CLINICAL TESTS:
To provide additional support for use in ages 6 and older, a subset of data from a research study was analyzed to establish use of the X-10 and X-24 headsets to acquire EEG in this patient population. Data from seven children ages 6-8 were compared to data from seven adults. The subjects were set-up by a technician with the X-Series strip appropriate based on their head measurements as described in the labeling. Technicians were able to quickly set-up subjects and resolve any high impedance issues as displayed by the X-Series Software. None of the subjects complained of discomfort or agitation in wearing the strips and the sensors remained in contact with the scalp for the duration of testing. 100% of all EEG sessions contained appropriate EEG recordings that could be further examined in third party software, and when data acquired from children are compared to an equivalent adult population, the data quality were equivalent.
#### CONCLUSION:
Based on the results of the non-clinical and clinical tests performed, Advanced Brain Monitoring considers the X-Series System to be as safe, as effective, and substantially equivalent to the predicate devices.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure with three stylized wings or feathers.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 27, 2013
Advanced Brain Monitoring, Inc. c/o Ms. Adrienne Lenz W324 S3649 County Road E Dousman. WI 53118
Re: K131383
Trade/Device Name: X-Series System Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWQ, OMC Dated: November 1, 2013 Received: November 5, 2013
Dear Ms. Lenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{13}------------------------------------------------
Page 2 -Ms. Adrienne Lenz
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Joyce M. Whang -S
for
Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{14}------------------------------------------------
## Indications for Use
510(k) Number (if known): K131383
Device Name: X-Series System
Indications For Use:
The X-Series System is intended for prescription use in the home, healthcare facility, or clinical research environment to acquire, transmit, display and store physiological signals from patients ages 6 and older. The X-Series system requires operation by a trained technician. The X-Series System acquires, transmits, displays and stores electroencephalogram (EEG), electrooculogram (EOG), electrocardiogram (ECG), and/or electromyogram (EMG), and accelerometer signals. The X-Series System only acquires and displays physiological signals, no claims are being made for analysis of the acquired signals with respect to the accuracy, precision and reliability.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
# Joyce M. Whang -S
Page 1 of _ 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.