BWIII EEG, BWIII EEG PLUS, BWIII PSG, BWIII PSG PLUS
K112107 · Neurovirtual USA, Inc. · GWQ · Oct 28, 2011 · Neurology
Device Facts
Record ID
K112107
Device Name
BWIII EEG, BWIII EEG PLUS, BWIII PSG, BWIII PSG PLUS
Applicant
Neurovirtual USA, Inc.
Product Code
GWQ · Neurology
Decision Date
Oct 28, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Indications for Use
The BWIII system may be used for electroencephalography (EEG) and sleep recordings (polysomnography) in research and clinical environments. It acquires displays and archives EEG and PSG data for on-screen review, annotation, and event-marking by the user. The BWIII requires competent user input, and its output must be reviewed and interpreted by a trained physician who will exercise professional judgment in using this information. The BWIII does not make any judgment of normality of the displayed signals or the results of an analysis. In no way are any of the functions represented as being in and of themselves diagnostic.
Device Story
Multi-channel system (up to 50 channels) for EEG and PSG recording; components include amplifier, head box, flash stimulator, and power module. Inputs: patient leads/electrodes and sensors (snore, flow, effort, position). Operation: connects to PC via Ethernet; data acquired, displayed, and archived for user review. Used in sleep labs, hospitals, or clinical settings under physician supervision. Output: raw signal visualization, user-applied annotations, and event-marking. Device performs no automated diagnostic judgment or normality assessment; output requires interpretation by trained physician to inform clinical decision-making. Benefits: facilitates standardized data acquisition and review for neurological and sleep studies.
Clinical Evidence
No clinical data provided. Safety and effectiveness supported by bench testing and compliance with international standards (IEC 60601-1, 60601-1-2, 60601-2-26, 60601-1-4). Literature cited regarding clinical/electrographic aspects of nonconvulsive status epilepticus provided as general support for EEG/PSG equipment utility.
Technological Characteristics
Hard plastic enclosure; 31-50 channel capacity; 16-bit AD resolution; Ethernet connectivity; 100-240VAC power. Filters: 0.16-10Hz (low), 15-100Hz (high). Flash stimulator 1-30Hz. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26, IEC 60601-1-4, EN ISO 14971, EN ISO 13485.
Indications for Use
Indicated for patients requiring electroencephalography (EEG) or polysomnography (PSG) sleep recordings in clinical or research environments. No specific age or gender restrictions. Contraindications not specified.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
OCT 2 8 2011
K 112107 p.1/4
# 510(k) SUMMARY
| A) Submitter's Name: | Neurovirtual USA, Inc. |
|---------------------------------------|-----------------------------------------------------|
| Owner / Operator Registration Number: | 9091724 |
| Manufacture Registration Number: | 3006136239 |
| B) Address: | 2315 NW 107th Ave - Suite 1M27<br>Doral, FL - 33172 |
| C) Phone and Fax Numbers | |
| Phone: | (786) 693-8200 |
| Fax: | (305) 393-8429 |
| D) Contact Person: | Eduardo J. Faria |
| E) Preparation Date: | September 28, 2011 |
| F) Classification Name: | |
| Common / Usual Name: | Full-Montage Standard Electroencephalograph |
| Proprietary Name: | BWIII (Models: EEG, EEG Plus, PSG and PSG Plus) |
| Product Code: | GWQ |
| Additional Product Codes: | OLV |
| Class: | Class II |
| Regulation: | 882.1400 |
#### G) Substantial Equivalence:
The BWIII is equivalent with the following products:
| 510(k) Number | Model | Company |
|---------------|------------------------------|----------------------------|
| K932407 | Easy Write and Easy Reader | Cadwell Laboratories, Inc. |
| K991900 | XLTEC PSG-40 Polysomnography | Excel Tech Ltd. |
| K062533 | BWII Digital | Neurovirtual USA Inc. |
#### Technological Characteristics:
The predicate devices used to establish substantial equivalence for the BWIII are outlined below. This section of this submission will provide a comparison of functional features and technical specifications of the BWIII to each of the predicate devices stratified by functional modality.
{1}------------------------------------------------
| Model | Neurovirtual BWIII | Neurovirtual BWII | Cadwell Easy Write | Excel Tech PSG-40 |
|-----------------------------|-----------------------------|-------------------|--------------------------------------------|-----------------------------------|
| 510(k) Number | K062533 | | K932407 | K991900 |
| Classification | GWQ | GWQ | GWQ | GWQ |
| Application | EEG | EEG | Easy II EEG | EEG and PSG |
| | EEG Plus PSG | PSG Plus | | |
| Number of Channels | 31 | 47 | 25 | Up to 50 |
| | EEG | PSG | EEG and PSG | |
| | | 29 | 32 | |
| Oximeter | Built-in | | Built-in | Built-in |
| Flash Stimulator | 1-30 Hz | | 1-25 Hz | 1-30 Hz |
| Data Communication | Ethernet / RJ-45 | | Ethernet / RJ-45 | Ethernet / RJ-45 |
| Software Based | MS Windows | | MS Windows | MS Windows |
| Material (External) | Hard Plastic | | Hard Plastic | Hard Plastic |
| Power Source | External | | External | Internal |
| Line Power | 100-240VAC | | 120/250VAC | 130/215VAC |
| AD Resolution | 16 Bits | | 16 Bits | 12 Bits |
| Common Mode Rejection Ratio | 50-60Hz | | 50-60Hz | 60Hz |
| Sensitivity Selection | 1-500uv/mm | | 1-100uv/mm | 1-200uv/mm |
| Low Frequency filters | 0.16 - 10Hz | | 0.02-10 Hz | 0.25 - 20Hz |
| High Frequency filters | 15 - 100 Hz | | 15 - 100 Hz | 15 - 120Hz |
| Auxiliary DC Inputs | 4 | 8 | 1-6 DC Inputs | 1-6 DC Inputs |
| User Interface | IBM PC | | IBM PC | IBM PC |
| Software | BWAnalysis<br>510(k)K062533 | | EasyWrite and Easy Reader<br>510(k)K932507 | Excel Neuroworks<br>510(k)K980214 |
:
.
{2}------------------------------------------------
K 112107 p.3/4
# H) Description:
BWIII is multi-channel (up to 50 channels) system designed for polysomnography (PSG) and electroencephalograph (EEG) recording application, in sleep lab, hospital or clinical environment under the supervision of a physician, using a laptop or a desktop computer.
The BWIII system consists of four major components: the amplifier unit, head box unit, flash stimulator unit and the power module (all plastic made). The system provides connections for electrodes and sensors, and connects to the computer using an ethernet cable.
The BWIII works with any good quality patient leads / electrodes and sensors snore, flow, effort belts and position) that have the safety touch connectors and are legally marketed in accordance with FDA requirements. As these accessories are already legally in the market from different manufactures, they are not part of this submission.
The BWIII does not make any judgment of normality of the displayed signals or the results of an analysis. In no way are any of the functions represented as being in and of themselves diagnostic.
#### I) Intended Used:
The BWIII system may be used for electroencephalography (EEG) and sleep recordings (polysomnography) in research and clinical environments. It acquires displays and archives EEG and PSG data for on-screen review, annotation, and event-marking by the user.
The BWIII requires competent user input, and its output must be reviewed and interpreted by a trained physician who will exercise professional judgment in using this information.
The BWIII does not make any judgment of normality of the displayed signals or the results of an analysis. In no way are any of the functions represented as being in and of themselves diagnostic.
{3}------------------------------------------------
K112107 p.4/4
## J) Safety and Effectiveness:
The BWIII is in compliance with the applicable clauses of the following standards:
- IEC 60601-1:2005, "Medical Device Equipment: General Requirements for -Safety"
- IEC 60601-1-1:2009, Medical electrical equipment Part 1: General requirements for safety - Section 1: Collateral standard: Safety requirements for medical electrical systems
- IEC 60601-1-2:2008, "Medical Device Equipment General Requirements for -Safety, Collateral Standard: Electromagnetic Compatibility, Requirements and Test"
- IEC 60601-2-26:2002, "Medical Device Equipment Particular requirements for the safety of electroencephalographs"
- . IEC 60601-1-4:2009, "Medical Device Equipment – General Requirement for Safety, Collateral Standard: Programmable Electrical Medical Systems"
- EN ISO 14971:2007, "Medical Devices: Application of Risk Management to -Medical Devices"
- EN ISO 13485:2003, "Medical Devices, Quality Management Systems: ﻨﻪ Requirements for Requlatory Purposes"
- -General Principles of Software Validation: FDA Guidance software validation version 1.1 (June 09, 1997)
### K) Non-clinical Testing:
In order to demonstrate that the use of EEG / PSG equipment is safe and effective, we have compiled one article from the Journal of Epilepsy and Clinical Neurophysiology (J Epilepsy Clin Neurophysiol 2004; 10(4):191-200) that demonstrates the safety and effectiveness regarding the equipment.
The full article "Nonconvulsive Status Epilepticus: Clinical and Electrographic Aspects" is attached on the ATTACHMENT 05.
#### L) Conclusion
The BWIII system is substantially equivalent to devices already on the market (cleared by the 510(k) process) and presents no new concerns about safety and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT 2 8 2011
Neurovirtual USA, Inc. c/o Mr. Eduardo Faria President 2315 NW 107th Avenue, Ste. 1M27 Doral, FL 33172
Re: K112107
Trade/Device Name: BWIII EEG, BWIII EEG Plus, BWIII PSG, BWIII PSG Plus Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWO, OLV Dated: September 28, 2011 Received: October 4, 2011
Dear Mr. Faria:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Peter m.f.
Malvina B. Evdelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Indications for Use
510(k) Number (if known): _ K112107
Device Name: BWIII
Indications for Use:
The BWIII system may be used for electroencephalography (EEG) and sleep recordings (Polysomnography) in research and clinical environments. It acquires displays and archives EEG and PSG data for on-screen review, annotation, and event-marking by the user.
The BWIII requires competent user input, and its output must be reviewed and interpreted by a trained physician who will exercise professional judgment in using this information.
The BWIII does not make any judgment of normality or abnormality of the displayed signals or the results of an analysis. In no way are any of the functions represented as being in and of themselves diagnostic.
Prescription Use × (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
John Grimes
Division Sion-Off Division of Ophthalmic, Neurological and Ear. Nose and Throat Devices
510(k) Number K112107
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.