B4C System - Addition of new sensor model (BcSs-PICNIW-2000)
K240821 · Braincare Desenvolvimento E Inovação Tecnológica S.A. · GWM · Jul 18, 2024 · Neurology
Device Facts
Record ID
K240821
Device Name
B4C System - Addition of new sensor model (BcSs-PICNIW-2000)
Applicant
Braincare Desenvolvimento E Inovação Tecnológica S.A.
Product Code
GWM · Neurology
Decision Date
Jul 18, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1620
Device Class
Class 2
Indications for Use
The B4C System is intended for the monitoring of variation in intracranial pressure in patients with suspected alteration of intracranial pressure (ICP) or change in intracranial compliance, by providing surrogate ICP waveforms and associated parameters (estimated P2/P1 ratio, normalized Time-to-Peak, derived useful ICP pulses and cardiac pulses) for interpretation. Refer to device labeling for more information regarding the derivation and interpretation of the output of the device.
Device Story
Non-invasive ICP monitoring system; detects skull deformations via scalp-mounted sensor (piezoelectric or strain gauge). Sensor transmits data via Bluetooth to mobile app and Physio Core analytical software. System outputs surrogate ICP waveforms, trend lines, and parameters (P2/P1 ratio, Time-to-Peak, cardiac pulses). Used in clinical settings by healthcare providers; output visualized on mobile device or compatible patient monitor. Provides preliminary assessment of intracranial hypertension or compliance changes; not a standalone diagnostic tool; does not replace absolute ICP measurement methods.
Clinical Evidence
Bench testing only. No clinical data. Performance demonstrated via stability, repeatability, and reproducibility testing comparing the new piezoelectric sensor to the predicate strain-gauge sensor. Biocompatibility, electrical safety (IEC 60601-1), and EMC (IEC 60601-1-2) testing passed.
Technological Characteristics
Non-invasive sensor (piezoelectric or strain gauge) on headband; silicone/polycarbonate/ABS casing; polypropylene headband. Wireless Bluetooth connectivity to mobile app/receiver. Analytical software processes signals. Complies with IEC 60601-1 and IEC 60601-1-2. Not MRI safe.
Indications for Use
Indicated for adult patients (18+) with suspected alteration of intracranial pressure (ICP) or change in intracranial compliance. Contraindicated for patients with decompressive craniectomy, craniotomy, or cranial defects (missing skull portion).
Regulatory Classification
Identification
An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.
{1}------------------------------------------------
July 18, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Braincare Desenvolvimento e Inovação Tecnológica S.A. % Cherita James Official Correspondent ProPharma MedTech 1129 20th Street NW, Ste 600 Washington, District of Columbia 20036
Re: K240821
Trade/Device Name: B4C System with BcSs-PICNIW-2000 sensor model Regulation Number: 21 CFR 882.1620 Regulation Name: Intracranial Pressure Monitoring Device Regulatory Class: Class II Product Code: GWM Dated: March 21, 2024 Received: March 25, 2024
Dear Cherita James:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Patrick Antkowiak -S
for
Jay Gupta, MS Assistant Director DHT5A: Division of Neurosurgical,
{3}------------------------------------------------
Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K240821
Device Name
B4C System - Addition of new sensor model (BcSs-PICNIW-2000)
Indications for Use (Describe)
The B4C System is intended for the monitoring of variation in intracranial pressure in patients with suspected alteration of intracranial pressure (ICP) or change in intracranial compliance, by providing surrogate ICP waveforms and associated parameters (estimated P2/P1 ratio, normalized Time-to-Peak, derived useful ICP pulses and cardiac pulses) for interpretation.
Refer to device labeling for more information regarding the derivation and interpretation of the output of the device.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) Summary
# B4C System - Addition of BcSs-PICNIW-2000 Sensor
| Sponsor: | Braincare Desenvolvimento e Inovacao Tecnologica S.A<br>Avenida Bruno Ruggiero Filho, 971<br>São Carlos, SP – Brazil<br>13562-420 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Connie Qiu<br>ProPharma MedTech<br>1129 20th Street NW, Suite 600<br>Washington, DC 20036<br>connie.qiu@propharmagroup.com<br>Ph. 347-954-0411 |
| Date Prepared: | March 6, 2024 |
| Proprietary Name: | B4C System (Addition of BcSs-PICNIW-2000 Sensor) |
| Common Name: | Intracranial pressure monitoring device |
| Regulatory Class: | II |
| Regulation: | 882.1620 |
| Product Code: | GWM |
| Predicate Device(s): | B4C System K201989 |
#### Device Description
The B4C System is a non-invasive device intended for monitoring of variation in intracranial pressure including patients with suspected alteration of intracranial pressure (ICP) or change in intracranial compliance. It consists of a sensor with Bluetooth wireless module, headband, mobile device software application, receiver, charger, as well as processing and analytical software. The subject of this 510(k) is to introduce an additional sensor, model BcSs-PICNIW-2000, that is compatible with the existing B4C System (K201989). The BcSs-PICNIW-2000 sensor consists of a piezoelectric fixed on a circular base that is supported on a headband placed over the patient's head. The skull pulsation is sensed by the fixed piezoelectric. Users may use either the existing sensor with the B4C System. During monitoring sessions, either sensor continuously transmits the Mobile App via Bluetooth connection and then to the analytical software component, Physio Core, to perform signal processing. The processed information is then sent back to the Mobile App in the form of minute graphs of waveform derived parameters as well as a report with additional waveform information. Like the predicate
{6}------------------------------------------------
sensor, the proposed sensor does not measure absolute intracranial pressure values, but continues to produce surrogate waveform morphology, its trend, and associated parameters reflecting changes in ICP. The B4C System and surrogate waveform and associated outputs do not substitute ICP monitoring methods when measurement of the absolute value of ICP is required to make a clinical decision.
The sensor is supported on a headband worn by the patient, such that the sensor is in contact with the scalp and is perpendicularly positioned in the temporoparietal transition, 2 inches (5-6 cm) above the entrance of the external auditory canal on the coronal plane. Slight pressure is applied so that the sensor maintains contact with the scalp throughout the monitoring session. The sensor continuously records and transfers acquired data to the B4C analytical and processing software, and back to the mobile device application or to a compatible multi-parameter monitor that has piezoresistive pressure transducer sensitivities of 5uV/Vex/mmHg or greater and automatic amplitude window adjustment capability via a paired receiver. Data is transferred wirelessly via Bluetooth connection between sensor and mobile application and HTTPS protocol between mobile application and analytics software. The clinician may view the visualized waveform on the mobile device along with real-time waveform, minute graphs, intermediate, or final reports of surrogate waveform and associated parameters including surrogate waveform trend line, average waveform per minute and estimated P2/P1 ratio, normalized time to peak, as well as derived useful ICP pulses and cardiac pulses. Alternatively, with a supplied dongle, a paired patient monitor's inherent software interprets the signal received from the B4C System's sensor and displays a surrogate waveform that allows for viewing the same ICP waveform on the monitor's display. Clinicians review the B4C System outputs to assess patients with suspected intracranial hypertension or changes in intracranial compliance based on the characteristics Percussion (P1), Tidal (P2,), and Dicrotic (P3) peaks of the waveform morphology and associated parameters.
The B4C System is not intended to be a standalone diagnostic tool. The surrogate waveform and associated parameter outputs do not replace a comprehensive clinical evaluation, but only provide an element for preliminary assessment. The clinician is responsible for determining the additional clinical information that may be required to make a diagnosis.
The B4C System is intended for use for adult patients ages 18 and older.
Indications for Use: The B4C System is intended for the monitoring of variation in intracranial pressure in patients with suspected alteration of intracranial pressure (ICP) or change in intracranial compliance, by
{7}------------------------------------------------
providing surrogate ICP waveforms and associated parameters (estimated P2/P1 ratio, normalized Timeto-Peak, derived useful ICP pulses and cardiac pulses) for interpretation.
Refer to device labeling for more information regarding the derivation and interpretation of the output of the device.
# Comparison to Predicate Device
The subject of this 510(k) is to introduce the additional BcSs-PICNIW-2000 sensor to the existing B4C System that is marketed under K201989. A comparison of the subject and predicate devices is presented in Table 1. The differences compared to the currently marketed device do not affect the intended use and do not raise new questions of safety and effectiveness.
| | B4C System<br>Addition of Sensor<br>BcSs-PICNIW-2000 | B4C System K201989 | Substantial<br>Equivalence |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | GWM | GWM | Same |
| Indications for Use | The B4C System is<br>intended for the monitoring<br>of variation in intracranial<br>pressure in patients with<br>suspected alteration of<br>intracranial pressure (ICP)<br>or change in intracranial<br>compliance, by providing<br>surrogate ICP waveforms<br>and associated parameters<br>(estimated P2/P1 ratio,<br>normalized Time-to-Peak,<br>derived useful ICP pulses<br>and cardiac pulse) for<br>interpretation.<br><br>Refer to device labeling for<br>more information regarding<br>the derivation and<br>interpretation of the output<br>of the device. | The B4C System is intended<br>for the monitoring of<br>variation in intracranial<br>pressure in patients with<br>suspected alteration of<br>intracranial pressure (ICP) or<br>change in intracranial<br>compliance, by providing<br>surrogate ICP waveforms and<br>associated parameters<br>(estimated P2/P1 ratio,<br>normalized Time-to-Peak,<br>derived useful ICP pulses and<br>cardiac pulse) for<br>interpretation.<br><br>Refer to device labeling for<br>more information regarding<br>the derivation and<br>interpretation of the output of<br>the device. | Same |
| Prescription Device | Yes | Yes | Same |
| Device<br>Description | Non-invasive ICP<br>monitoring device<br>consisting of pressure<br>sensors supported on a<br>headband to detect skull | Non-invasive ICP monitoring<br>device consisting of strain<br>gauge pressure sensors<br>supported on a headband to<br>detect skull deformations in | Similar. The main<br>difference compared to<br>the predicate is that an<br>additional sensor is<br>introduced as an option |
| | deformations in response to ICP changes. System wirelessly transmits acquired signal for processing and analytics. System outputs surrogate ICP waveform and report of waveform's associated parameters on mobile device application and web portal. ICP waveform may also be viewed on compatible monitor via paired wireless receiver. This 510(k) introduces an additional piezoelectric pressure sensor available for use with the existing B4C System in addition to a strain gauge pressure sensor. | response to ICP changes. System wirelessly transmits acquired signal for processing and analytics. System outputs surrogate ICP waveform and report of waveform's associated parameters on mobile device application and web portal. ICP waveform may also be viewed on compatible monitor via paired wireless receiver. | to the user with the existing B4C System. The B4C System's software components have also undergone minor software updates. |
| Clinical<br>Application | Non-invasive application of<br>a sensor on the scalp<br>perpendicularly positioned<br>in the temporoparietal<br>transition, 2 inches (5-6<br>cm) above the entrance of<br>the external auditory canal<br>on the coronal plane | Non-invasive application of a<br>sensor on the scalp<br>perpendicularly positioned in<br>the temporoparietal<br>transition, 2 inches (5-6 cm)<br>above the entrance of the<br>external auditory canal on the<br>coronal plane | Same |
| Contraindications | The B4C System is<br>contraindicated for use in<br>patients who have:<br>• Undergone<br>decompressive<br>craniectomy or<br>craniotomy;<br>• Cranial defects (portion<br>of skull missing);<br>Any other conditions that<br>the health practitioner<br>deems to be unsuitable for<br>use of this device. | The B4C System is<br>contraindicated for use in<br>patients who have:<br>• Undergone decompressive<br>craniectomy or<br>craniotomy:<br>• Cranial defects (portion of<br>skull missing);<br>Any other conditions that the<br>health practitioner deems to<br>be unsuitable for use of this<br>device. | Same |
| Device Materials | • Silicone and<br>polycarbonate/ABS<br>sensor casing<br>• Polypropylene<br>headband | • Polycarbonate sensor<br>casing and contact pin<br>• Silicone base around<br>sensor<br>• Polypropylene headband | Same materials |
| MRI Claim | MR Unsafe | MR Unsafe | Same |
| Sterilization | Not applicable | Not applicable | Same |
| Device dimensions | Sensor case: 250 x 40 x 34 mm<br><br>Receiver case: 94 x 17.5 x 15 mm<br><br>Headband size with turnbuckle:<br>340 x 32 x 20 mm (smallest)<br>460 x 32 x 20 mm (largest) | Sensor case: 75.6 X 51.5 X 27.7 mm<br><br>Receiver case: 94 X 17.5 X 15 mm<br>Receiver cable and connector length: 20 cm<br><br>Headband size with turnbuckle:<br>XXS: 49.5 cm, XS: 52 cm, S: 54.5 cm, M: 57 cm, L: 59.5 cm, XL: 52 cm, XXL: 64.5 cm | Similar.<br>Differences in dimensions do not raise new questions of safety or effectiveness. |
| Biocompatibility | Limited duration contact (≤24) with intact skin<br><br>Non-cytotoxic<br>Non-sensitizing<br>Non-irritating | Limited duration contact (≤24) with intact skin<br><br>Non-cytotoxic<br>Non-sensitizing<br>Non-irritating | Same |
| Energy modality | Sensor contains internal rechargeable battery | Sensor contains internal rechargeable battery and external rechargeable battery pack | Similar<br>Both subject and predicate sensors contain internal rechargeable batteries. The subject sensor does not include an external rechargeable battery pack. This difference does not raise new questions of safety and effectiveness. |
| ICP Waveform<br>Outputs | Waveform displayed on compatible patient monitor<br><br>Analytical software also produces the following associated parameters about the surrogate ICP waveform displayed in a report and on the accompanying mobile medical application:<br>● Surrogate Waveform<br>● Waveform Trend line<br>● Average waveform<br>● Estimated P2:P1 ratio | Waveform displayed on compatible patient monitor<br><br>Analytical software also produces the following associated parameters about the surrogate ICP waveform displayed in a report and on the accompanying mobile medical application:<br>● Surrogate Waveform<br>● Waveform Trend line<br>● Average waveform<br>● Estimated P2:P1 ratio<br>● Normalized Time-to-Peak | Same |
| | Normalized Time-to-Peak Derived useful ICP pulses Derived Cardiac Pulse These associated parameters are derived based on well-established principles in scientific literature and clinical practice. | Derived Cardiac Pulse These associated parameters are derived based on well-established principles in scientific literature and clinical practice. | |
| Sensing element | Piezoelectric | Strain gauge | Different<br>The subject sensor utilizes a piezoelectric while the predicate sensor is strain gauge based. Both sensors follow the same operating principle of detecting changes in ICP through small variations in skull deformation based on well-understood principles of ICP waveform morphology. Performance data demonstrate that this difference does not raise new questions of safety and effectiveness. |
| Functional pressure range | Not applicable as it does not provide absolute values of pressure | Not applicable as it does not provide absolute values of pressure | Same |
| Functional over pressure range without damage | Not applicable as it does not provide absolute values of pressure, and does not have a specified functional pressure range. | Not applicable as it does not provide absolute values of pressure, and does not have a specified functional pressure range. | Same |
| Input/ Output Impedance | The wireless sensor is not physically connected to any device and has an internal resistive bridge of 3000 Ohms (typical values). It also uses a piezoelectric disc that does not require a power supply to generate electric charges, with | The wireless sensor is not physically connected to any device and has an internal resistive bridge with input and output impedance of 1000 Ohms. | Similar.<br>Differences do not raise new questions of safety and effectiveness. |
| | magnitude of 14.5 nF. | | |
| Output signal<br>(sensitivity) | Not applicable for the<br>wireless sensor since it is<br>not physically connected to<br>any device.<br>The receiver can output a<br>maximum signal of 25mV<br>and minimum of -2.5mV. | Not applicable for the<br>wireless sensor since it is not<br>physically connected to any<br>device.<br>The receiver can output a<br>maximum signal of 25mV<br>and minimum of -2.5mV. | Same |
| Zero Drift | Not applicable for the<br>sensor as it does not<br>provide absolute values and<br>brain4care App performs<br>auto scale so that the<br>waveform is always visible.<br>The receiver is also capable<br>of automatically readjusting<br>the signal offset level so<br>that the waveform is always<br>visible on the monitor. | Not applicable for the sensor<br>as it does not provide<br>absolute values and<br>brain4care App performs<br>auto scale so that the<br>waveform is always visible.<br>The receiver is also capable<br>of automatically readjusting<br>the signal offset level so that<br>the waveform is always<br>visible on the monitor. | Same |
| Electrical Safety | Complies with IEC 60601-1 | Complies with IEC 60601-1 | Same |
| Electromagnetic<br>Compatibility | Complies with IEC 60601-1-2 | Complies with IEC 60601-1-2 | Same |
| Software | The B4C System includes a<br>mobile device application,<br>firmware, analytical and<br>processing software, and<br>administrative software<br>components. | The B4C System includes a<br>mobile device application,<br>firmware, analytical and<br>processing software, and<br>administrative software<br>components. | Same |
| Sensor Connection<br>to Monitor | Wireless<br>Bluetooth connection to a<br>receiver or micro-USB<br>connection specific to<br>compatible patient monitors | Wireless<br>Bluetooth connection to a<br>receiver or micro-USB<br>connection specific to<br>compatible patient monitors | Same |
| Wireless Module | Bluetooth | Bluetooth | Same |
# Table 1 Comparison of Subject and Predicate Devices
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
### Differences from Predicate
This 510(k) introduces the addition of the BcSs-PICNIW-2000 sensor to the existing B4C System, the predicate device K201989. The main difference is an additional sensor that uses a piezoelectric based sensor compared to the existing strain gauge based sensor. The B4C System continues to maintain the same indications for use, operating principle of detecting small variations in skull deformation, and clinical utility. There are no significant changes to the B4C System software.
### Discussion of Performance Data
The following performance data in Table 2 are provided in support of demonstrating substantial
{12}------------------------------------------------
equivalence between the subject and predicate devices.
| Table 2 Summary of Non-Clinical Performance Data |
|--------------------------------------------------|
|--------------------------------------------------|
| TEST | TITLE/TEST METHOD SUMMARY | RESULTS |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|
| Biocompatibility | The BcSs-PICNIW-2000 is comprised of the same materials as the predicate device sensor. A biological<br>risk assessment was provided in lieu of repeating testing. | |
| Electrical Safety and Electromagnetic Compatibility | | |
| IEC 60601-1 | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential<br>performance | Pass |
| ANSI AAMI ES<br>60601-1 | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential<br>performance | Pass |
| IEC 60601-1-2 | Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential<br>performance - Collateral | Pass |
| AAMI TIR69 | Risk management of radio-frequency wireless<br>coexistence for medical devices and systems | Testing not required<br>based on risk<br>assessment |
| Disinfection | The BcSs-PICNIW-2000 is intended to be reprocessed according to the same methods as the predicate<br>device sensor. Reprocessing validation was not required. | |
| Bench Testing | | |
| Monitor Compatibility | There are no changes to the predicate device receiver and wireless<br>communication protocol. Monitor compatibility testing was not repeated. | |
| Stability and<br>Reproducibility | Demonstration of stability, repeatability, and<br>reproducibility between the ICP waveform outputs of<br>the wireless and wired sensors. | Pass |
| Software | | |
| Software Verification<br>and Validation | Demonstrate that all software requirements were<br>appropriately implemented in the software. | Pass |
Non-clinical performance data is sufficient to demonstrate substantial equivalence.
## Conclusion
Based on results of the performance testing and substantial equivalence comparison, the B4C System maintains the same intended use as the predicate device and the information presented is sufficient to determine that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.