The intended use of the DirectLink Module is to enable the connection of Codman intracranial pressure sensors to an available invasive blood pressure input channel on select commercially available third party patient bedside monitor systems. The DirectLink ICP Extension Cable is intended for use as a connecting cable between the DirectLink ICP Module and a Codman Microsensor ICP Transducer. The Patient Monitor Interface Cable is intended for use as a connecting cable between DirectLink ICP Module, and selected patient monitors available from third party suppliers.
Device Story
DirectLink ICP Module acts as an interface between Codman Microsensor ICP transducers and third-party patient bedside monitors; enables zeroing of ICP sensors; provides zero reference and calibration signals to the monitor; transfers ICP readings for visualization and storage. Powered by the connected patient monitor. Used in clinical settings by healthcare providers. Facilitates integration of specialized ICP sensors into standard bedside monitoring workflows; allows continuous ICP monitoring to support clinical decision-making in neurocritical care.
Clinical Evidence
Bench testing only. Validation and verification included accuracy tests, output impedance, zero short-term drift, sensor signal stability, power consumption, hardware diagnostics, cable mechanical testing (flex cycles, mating force), and system-level usability testing with surgeons and nurses. Compliance with 60601-1 and 60601-1-2 standards was verified. No clinical or animal studies were required.
Technological Characteristics
Interface module and cabling system. Materials compatible with cleaning via quaternary ammonium/isopropyl alcohol or 70% isopropyl alcohol; extension cable autoclavable. Powered via connection to patient monitor. Connectivity: proprietary interface to Codman sensors and standard invasive blood pressure input channels on third-party monitors. Hardware-based signal conditioning and calibration.
Indications for Use
Indicated for patients requiring intracranial pressure (ICP) monitoring via Codman Microsensor ICP Transducers, facilitating signal transmission to third-party bedside patient monitors.
Regulatory Classification
Identification
An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked one in front of the other. The profiles are rendered in a simple, abstract style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 22, 2016
Medos International SARL c/o Dr. Elena Jugo Codman Neuro 325 Paramount Drive Raynham, MA 02767
Re: K152670
Trade/Device Name: DirectLink ICP Module DirectLink Extension Cable Patient Monitor Interface Cables Regulation Number: 21 CFR 882.1620 Regulation Name: Intracranial pressure monitoring device Regulatory Class: Class II Product Code: GWM Dated: December 18, 2015 Received: December 21, 2015
Dear Dr. Elena Jugo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the [kit/tray] have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# William J. Heetderks -S
for
Carlos L. Peña. PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152670
Device Name DirectLink™ ICP Module DirectLink™ ICP Extension Cable Patient Monitor Interface Cable
#### Indications for Use (Describe)
The intended use of the DirectLink Module is to enable the connection of Codman intracranial pressure sensors to an available invasive blood pressure input channel on select commercially available third party patient bedside monitor systems.
The DirectLink ICP Extension Cable is intended for use as a connecting cable between the DirectLink ICP Module and a Codman Microsensor ICP Transducer.
The Patient Monitor Interface Cable is intended for use as a connecting cable between DirectLink ICP Module, and selected patient monitors available from third party suppliers.
| Type of Use (Select one or both, as applicable) | | | | |
|-------------------------------------------------|--|--|--|--|
|-------------------------------------------------|--|--|--|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary 7.
#### Codman Neuro I. Submitter 325 Paramount Drive Raynham, MA 02767
On behalf of: Medos International SARL Chemin-Blanc 38 CH 2400 LeLocle, Switzerland
Phone: 305-265-6802 305-265-6889 Fax:
Contact Person: Elena Jugo Date of Submission: September 16, 2015
## II. Device
| Name of Device | DirectLink ICP Module |
|---------------------|-----------------------------------------------------------|
| | DirectLink Extension Cable |
| | Patient Monitor Interface Cables |
| Common Name | Intracranial pressure monitoring device |
| Classification Name | Intracranial Pressure Monitoring Device (21 CFR 882.1620) |
| Regulatory Class | II |
| Product Code | GWM |
### III. Predicate Device
Codman CU-II (ICP Express), K945585
DirectLink ICP Module IV. Device Description
> The DirectLink ICP Module is a direct interface to connect the CODMAN MICROSENSOR® ICP Transducers to a patient bedside monitor, allowing the user to do the following:
- . Zero the ICP sensor
- Provide zero reference and calibration signals to the patient bedside . monitor
- . Transfer the ICP readings to the patient bedside monitor for visualization and data storage/processing.
The module does not have its own source of power, but is powered through connection to the patient monitor.
{4}------------------------------------------------
## 510(k) Summary (Cont)
IV. Device In order to connect the DirectLink Module to the Microsensor and to the bedside Description monitor two cables (the DirectLink ICP Extension Cable and the Patient Monitor Interface Cable) are needed. These cables are described below. (Cont.) DirectLink ICP Extension Cable The DirectLink ICP Extension Cable is required to connect the DirectLink ICP Module and the CODMAN MICROSENSOR ICP Transducer. The extension cable is reusable and is supplied non-sterile. The cable can be wiped down before each use with combination wipes (Quarternary Ammonium/Isopropyl Alcohol), or 70% isopropyl alcohol, or can be sterilized by autoclave before each use. Patient Monitor Interface Cable The Patient Monitor Interface Cable is used as a connecting cable between the DirectLink ICP Module, and selected patient monitors available from third party suppliers. The interface cable is reusable and is supplied non-sterile. The cable can be wiped down before each use with combination wipes (Quarternary Ammonium/Isopropyl Alcohol), or 70% isopropyl alcohol. The intended use of the DirectLink ICP Module is to enable the connection of V. Indications Codman intracranial pressure sensors to an available invasive blood pressure input for Use channel on select commercially available third party patient bedside monitor systems. The DirectLink ICP Extension Cable is intended for use as a connecting cable between the DirectLink ICP Module, and a Codman Microsensor ICP Transducer. The Patient Monitor Interface Cable is intended for use as a connecting cable between DirectLink ICP Module, and selected patient monitors available from third party suppliers.
{5}------------------------------------------------
## 510(k) Summary (Cont)
| VI.<br>Comparison to<br>Predicate<br>Device | Based upon the intended use, design, function, comparison to the currently<br>marketed device, and testing performed, it is concluded that the DirectLink<br>ICP Module and related cables are substantially equivalent to the predicate<br>Codman CU-II (ICP Express), and therefore, do not raise any new questions<br>of safety and effectiveness. | | |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| VII.<br>Performance<br>Data | The following performance data have been provided in support of the substantial<br>equivalence determination. | | |
| | Bench Testing<br>Validation and verification testing were performed on the DirectLink ICP<br>Module, the ICP Extension Cable, and the Patient Monitor Interface Cables. | | |
| | Testing was performed on each individual component, as well as on the entire<br>system. Table 2 lists the tests performed. | | |
| | Table 2. Validation and Verification Tests | | |
| | DirectLink ICP Module | | |
| | Accuracy Test | | |
| | Output Impedance, Zero Short Term Drift, Sensor Signal Stability Test<br>A - Researces and Campeon - the - Desember Decamers Defenser - Simmel - Part | | |
Accuracy and Symmetry of the Pressure Reference Signals Test Power Consumption and Input Impedance Test
Wiping Test Hardware Diagnostic Test
DirectLink ICP Extension Cable and Patient Monitor Interface Cables
Cable Baseline Tests (mating and demating force, flex cycles) Cable Autoclave Tests (autoclave, wiping, flex cycles, operating cycles) Documents and Drawings Review
Complete System – DirectLink ICP Modules and Cables
Transit Test
Summative Usability Test – Part 1 - Surgeons
Summative Usability Test – Part 2 – Nurses
Functionality and Performance Testing of the DirectLink System Connected to a Patient Monitor including Cables Mating
60601-1 and 60601-1-2 Test PPQ Validation for DirectLink ICP Module and Cables
{6}------------------------------------------------
## 510(k) Summary (Cont)
| VII.<br>Performance<br>Data, cont. | Biocompatibility Testing<br>Biocompatibility testing was not performed as these devices are not intended<br>to contact the patient. |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Animal Studies<br>No animal studies were required as appropriate verification of the new intended use<br>was achieved based on the similarities of the proposed device to the predicate<br>devices, and from results of bench testing. |
| | Clinical Studies<br>Clinical data are not necessary to demonstrate substantial equivalence of the<br>DirectLink ICP Module and associated cables to the predicate device. |
| VIII.<br>Conclusion | Based upon the intended use, design, materials, function, comparison to currently<br>marketed device, and testing performed it is concluded that the DirectLink ICP<br>Module and associated cables are substantially equivalent to the predicate Codman<br>CU II (ICP Express) and therefore, do not raise any new questions of safety and<br>effectiveness. |
| | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.