Use of the Integra™ Camino® Flex Ventricular Intracranial Pressure Monitoring Kit with Integra™ Camino® Flex Adapter is indicated when direct and continuous intraventricular intracranial pressure (ICP) monitoring and cerebrospinal fluid (CSF) drainage are required.
Device Story
The Integra Camino Flex Ventricular Intracranial Pressure Monitoring Kit is a single-use, disposable system for neurosurgical use. It consists of a flexible catheter with a pressure-sensing tip and a drainage lumen for CSF removal. The catheter is implanted into the lateral cerebral ventricle via a tunneling surgical method. The system includes an adapter and extension cable that interface with the previously cleared Camino Advanced Monitor. The adapter receives transducer signals from the catheter, amplifies them, and transmits the data to the monitor. Clinicians use the monitor to observe real-time ICP, which informs clinical decision-making regarding patient neurological status and intracranial pressure management. The device provides the benefit of simultaneous pressure monitoring and fluid drainage.
Clinical Evidence
Bench testing only. Testing included sterilization validation (AAMI TIR 28:2009, ISO 11135-1:2007, ISO 10993-7:2008, ISO 11737-1:2006, ISO 11737-2:2009, USP <85>, USP <161>) and packaging integrity testing (ISO 11607-1:2006, ASTM F1980-07:2011, ISTA 2A:2008, ASTM F2096-04, ASTM F88, ASTM F1929-98:2004).
Technological Characteristics
Single-use, disposable ventricular catheter with pressure-sensing tip and drainage lumen. Sterilized via ethylene oxide. Connectivity via proprietary adapter and extension cable to Camino Advanced Monitor. Materials and packaging validated per ISO and ASTM standards.
Indications for Use
Indicated for patients requiring direct, continuous intraventricular intracranial pressure (ICP) monitoring and cerebrospinal fluid (CSF) drainage.
Regulatory Classification
Identification
An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.
Ventricular Pressure Monitoring Kit, Model 110-4HM (K914735)
Combined Intracranial Pressure-Temperature Sensing System (K962928)
Submission Summary (Full Text)
{0}------------------------------------------------
# K121159
## Integra LifeSciences Corporation-Traditional 510(k)
Integra™ Camino® Flex Ventricular Intracranial Pressure Monitoring Kit with Integra™ Camino® Flex Adapter
JUL 2 0 2012
## 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| Submitter Information | |
|------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Integra LifeSciences Corporation |
| Address | 311 Enterprise Drive Plainsboro, NJ 08536 USA |
| Phone number | (609) 750-2309 |
| Fax number | (609) 275-9445 |
| Establishment | 3003418325 |
| Registration Number | |
| Name of contact person | Steven R. Peltier, Vice President, Corporate Regulatory Affairs |
| Date prepared | April 9, 2012 |
| Name of device | |
| Trade or proprietary name | Integra™ Camino® Flex Ventricular Intracranial Pressure<br>Monitoring Kit (for Catheter and Accessories) |
| | Integra™ Camino® Flex Adapter (for Adapter and Extension Cable) |
| Common or usual name | Ventricular Catheter<br>Intracranial Pressure Monitoring System |
| Classification name | Device, Monitoring, Intracranial Pressure |
| Classification panel | Neurology |
| Regulation | Class II, under 21 CFR 882.1620 |
| Product Code(s) | GWM |
| Legally marketed<br>device(s) to which<br>equivalence is claimed | Ventrix® Ventricular Tunneling Pressure Monitoring Kit<br>(catalog number: NL950-V; formerly known as the OPTYX<br>catheter)<br>510(k) K904883<br>Codman® MicroSensor™ Ventricular Catheter Kit<br>510(k) K991222<br>Ventricular Pressure Monitoring Kit, Model 110-4HM<br>510(k) - K914735<br>Combined Intracranial Pressure-Temperature Sensing System<br>(with pre-amplification cable, catalog number PAC1)<br>510(k) K962928 |
| Reason for 510(k)<br>submission | New Device |
| Device description | The Integra™ Camino® Flex Ventricular Intracranial Pressure<br>Monitoring Kit with Integra™ Camino® Flex Adapter functions as<br>an entire system that includes the following: Integra Camino Flex<br>Ventricular Catheter Monitoring Kit (Catheter and Accessories)<br>and Integra Camino Flex Adapter (Adapter and Extension Cable),<br>in which the Integra Camino Flex Adapter connects to the<br>previously cleared Camino Advanced Monitor (510(k) K962928<br>Integra LifeSciences Corporation). |
| | The Integra Camino Flex Ventricular Intracranial Pressure<br>Monitoring Kit with Integra Camino Flex Adapter is indicated<br>when direct and continuous intraventricular intracranial pressure<br>(ICP) monitoring and cerebrospinal fluid (CSF) drainage are<br>required. The device is a single use, disposable product. The<br>device is comprised of a highly flexible catheter with a tensile<br>member, a pressure sensing tip, an electrical connection to a<br>monitor, and a drainage lumen to allow fluid connection to an<br>extraventricular drainage (EVD) device. The device is designed for<br>the tunneling surgical method and the kit includes the necessary<br>accessories for access and implantation of the catheter. The<br>following Accessories are referenced in 510(k) K904883: Drill,<br>Drill Stop, Set Screw, Hex Wrench, Trocar, Trocar Sheath, Suture<br>Loops, and Male Luer Cap. |
| | The tip of the Integra Camino Flex Ventricular Catheter is<br>implanted within the anterior horn of the left or right lateral<br>cerebral ventricle. A cylindrical volume with a height of at least 11<br>mm and a diameter of at least 5 mm are required for catheter<br>implantation. |
| | The Integra Camino Flex Adapter performs the functions of<br>receiving the transducer signal from the Integra Camino Flex<br>Ventricular Intracranial Pressure Monitoring Kit catheter,<br>amplifying the signal, and transmitting the signal to the Camino®<br>Advanced Monitor. The Integra Camino Flex Adapter converts the<br>signals produced by the Integra Camino Flex Ventricular<br>Intracranial Pressure Monitoring Kit catheter sensor. This signal is<br>sent to the Adapter by means of the extension cable to the<br>Camino® Advanced Monitor. |
| Sterilization Information | The Integra Camino Flex Ventricular Catheter Monitoring Kit<br>(Catheter and Accessories) is sterilized by ethylene oxide. Testing<br>was conducted with reference to AAMI TIR 28: 2009, ISO 11135-<br>1:2007, ISO 10993-7:2008, ISO 11737-1:2006, ISO 11737-2:2009,<br>USP <85> and USP <161>. A half-cycle sterilization cycle was |
| Intended use of the<br>device | run to confirm that the Integra Camino Flex Ventricular<br>Intracranial Pressure Monitoring Kit could be adopted into the<br>OSVII (flow regulating valve) family. Packaging integrity was<br>conducted with reference to ISO 11607-1:2006, ASTM F1980-<br>07:2011, ISTA 2A:2008, ASTM F2096-04, ASTM F88, and<br>ASTM F1929-98:2004. The sterile kits are labeled as non-<br>pyrogenic.<br>Use of the Integra™ Camino® Flex Ventricular Intracranial<br>Pressure Monitoring Kit with Integra™ Camino® Flex Adapter is<br>indicated when direct and continuous intraventricular intracranial<br>pressure (ICP) monitoring and cerebrospinal fluid (CSF) drainage<br>are required. |
| Indications for use | Use of the Integra™ Camino® Flex Ventricular Intracranial<br>Pressure Monitoring Kit with Integra™ Camino® Flex Adapter is<br>indicated when direct and continuous intraventricular intracranial<br>pressure (ICP) monitoring and cerebrospinal fluid (CSF) drainage<br>are required. |
{1}------------------------------------------------
Integra LifeSciences Corporation-Traditional 510(k)
Integra™ Camino® Flex Ventricular Intracranial Pressure Monitoring Kit with
Integra™ Camino® Flex Adapter
{2}------------------------------------------------
K121159
Integra LifeSciences.Corporation-Traditional 510(k)
Integra™ Camino® Flex Ventricular Intracranial Pressure Monitoring Kit with
Integra™ Camino® Flex Adapter
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Integra LifeSciences Corporation % Mr. Steven R. Peltier Vice President Regulatory Affairs 311 Enterprise Drive Plainsboro, NJ 08536
JUL 2 0 2012
Re: K121159
Trade Name: Integra™ Camino® Flex Ventricular Intracranial Pressure Monitoring Kit with Integra™ Camino® Flex Adapter Regulation Number: 21 CFR 882.1620 Regulation Name: Intracranial pressure monitoring device Regulatory Class: Class II Product Code: GWM Dated: June 15, 2012 Received: June 18, 2012
### Dear Mr. Peltier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-
{4}------------------------------------------------
### Page 2 - Mr. Steven R. Peltier
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
E.A. Kumm, MD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K12iis9
Integra LifeSciences Corporation-Traditional 510(k)
Integra™ Camino® Flex Ventricular Intracranial Pressure Monitoring Kit with Integra™ Camino® Flex Adapter
#### Indications for Use
Page 1 of 1.
#### KIJ | 1 59 510(k) Number (if known):
Device Name:
Integra™ Camino® Flex Ventricular Intracranial Pressure Monitoring Kit with Integra™ Camino® Flex Adapter
#### Indications For Use:
The Integra " Camino" Flex Ventricular Intracranial Pressure Monitoring Kit with Integra Camino Flex Adapter is indicated when direct and continuous intraventricular intracranial pressure (ICP) monitoring and cerebrospinal fluid (CSF) drainage are required.
Prescription Use X (Per 21 CFR 801 Subpart D)
Over-The Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
Q
510(k) Number K121159
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.