← Product Code [GWM](/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/GWM) · K022638

# NEURORECOVERY VENTRICULAR CATHETER & MAIN VALVE ASSEMBLY KIT (K022638)

_Neurorecovery, Inc. · GWM · Oct 31, 2002 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/GWM/K022638

## Device Facts

- **Applicant:** Neurorecovery, Inc.
- **Product Code:** [GWM](/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/GWM.md)
- **Decision Date:** Oct 31, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1620
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Neurorecovery™, Inc., Neurorecovery" Ventricular Catheter & Main Valve Assembly Kit is designed to be used for external monitoring of intracranial pressure (ICP), cerebrospinal fluid (CSF) sampling, and CSF drainage from the lateral ventricles of the human brain. The Main Valve Assembly is designed to facilitate the monitoring, sampling, clearing, and drainage function in an aseptic manner.

## Device Story

The Neurorecovery Ventricular Catheter & Main Valve Assembly Kit is a sterile, single-use system for neurosurgical applications. It consists of a ventricular catheter, stylet, trocar, valves, tubing, connectors, caps, check-valves, needle-free valves, and stopcocks. The device is inserted into the lateral ventricles of the brain to provide a pathway for external ICP monitoring, CSF sampling, and CSF drainage. The Main Valve Assembly facilitates these functions in an aseptic manner. The device is used in clinical settings (e.g., ICU or OR) by healthcare professionals. It provides a mechanical interface for fluid management and pressure monitoring, allowing clinicians to manage intracranial pressure and collect CSF for diagnostic or therapeutic purposes, thereby assisting in the management of patients with neurological conditions.

## Clinical Evidence

Bench testing only. Testing demonstrated compliance with specifications for biocompatibility, sterility, and functional performance.

## Technological Characteristics

System includes ventricular catheter, stylet, trocar, valves, tubing, connectors, caps, check-valves, needle-free valves, and stopcocks. Mechanical device for fluid drainage and pressure monitoring. Sterilized components.

## Regulatory Identification

An intracranial pressure monitoring device is a device used for short-term monitoring and recording of intracranial pressures and pressure trends. The device includes the transducer, monitor, and interconnecting hardware.

## Predicate Devices

- Codman EDS II ([K902257](/device/K902257.md))
- Heyer-Schulte NeuroCare External Drainage Management System ([K972994](/device/K972994.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary (per 21 CFR807.92)

K022638

## Neurorecovery™ Ventricular Catheter & Main Valve Assembly Kit

#### 1. APPLICANT

NeurorecoveryTM, Inc. 325 Queen City Avenue Tuscaloosa, AL 35401

Landon C. G. Miller; JD, RPA, MSA, BA Contact Person: Telephone: (205) 345-8606

Date Prepared: August 7, 2002

#### 2. DEVICE NAME

| Proprietary Name:     | Neurorecovery™ Ventricular Catheter & Main Valve<br>Assembly Kit     |
|-----------------------|----------------------------------------------------------------------|
| Common/Usual Name:    | Ventricular Catheter and Accessories                                 |
| Classification Names: | Ventricular Catheter<br>Intracranial Pressure Monitoring Accessories |

#### 3. DEVICE CLASSIFICATION

Ventricular Catheter (21 CFR 882.4100; ProCode: HCA) and Intracranial Pressure Monitoring Accessories (21 CFR 882.1620; ProCode: GWM) have been classified as Class II devices.

#### 4. PREDICATE DEVICES

- . Codman EDS II (K902257) Johnson & Johnson Professional, Inc.
- . Heyer-Schulte NeuroCare External Drainage Management System (K972994) [now believed to be MiniTorr CSF Drainage Systems marketed by Integra LifeSciences]

{1}------------------------------------------------

#### ર. DEVICE DESCRIPTION

The Neurorecovery", Inc., Neurorecovery" Ventricular Catheter & Main Valve Assembly Kit consists of legally marketed ventricular catheter, stylet, trocar, valves, tubing, associated connectors, caps, check-valves, needle-free valve, and stopcocks.

#### INTENDED USE 6.

The Neurorecovery", Inc., Neurorecovery" Ventricular Catheter & Main Valve Assembly Kit is designed to be used for external monitoring of intracranial pressure (ICP), cerebrospinal fluid (CSF) sampling, and CSF drainage from the lateral ventricles of the human brain. The Main Valve Assembly is designed to facilitate the monitoring, sampling, clearing, and drainage function in an aseptic manner.

#### 7. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The Neurorecovery", Inc., Neurorecovery" Ventricular Catheter & Main Valve Assembly Kit is equivalent to cited predicate devices based on its indications for use, design, materials, and operational characteristics. Neurorecovery", Inc., believes that differences between devices are minor and raise no new issues of safety or effectiveness.

#### PERFORMANCE TESTING 8.

Testing submitted in the 510(k) demonstrates that the Neurorecovery™ Ventricular Catheter & Main Valve Assembly Kit complies with specifications for biocompatibility, sterility, and functional performance.

{2}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract design of three wavy lines that resemble a stylized human figure.

### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# OCT 31 2002

Neurorecovery, Inc. c/o Medical Device Consultants, Inc. Rosina Robinson Senior Staff Consultant 49 Plain Street North Attleboro, Massachusetts 02760

Re: K022638

Trade/Device Name: Neurorecovery™ Ventricular Catheter and Main Valve Assembly Kit Regulation Number: 882.1620 Regulation Name: Intracranial pressure monitoring device Regulatory Class: Class II Product Code: GWM Dated: August 7, 2002 Received: August 8, 2002

Dear Ms. Robinson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 – Ms. Rosina Robinson

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

mirano. Provos

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### 510(k) Number (if known): K 0226

Neurorecovery *, Inc., Neurorecovery * Ventricular Catheter Device Name: & Main Valve Assembly Kit

Indications for Use:

The Neurorecovery™, Inc., Neurorecovery" Ventricular Catheter & Main Valve Assembly Kit is designed to be used for external monitoring of intracranial pressure (ICP), cerebrospinal fluid (CSF) sampling, and CSF drainage from the lateral ventricles of the human brain. The Main Valve Assembly is designed to facilitate the monitoring, sampling, clearing, and drainage function in an aseptic manner.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Muriam C. Provost

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K022638

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/GWM/K022638](https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94neurological-diagnostic-devices/GWM/K022638)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
