The Nicolet Wireless EEG Amplifier is intended to be used as a front end amplifier to acquire, store, and transmit electrophysiological signals in a wired or wireless mode for the Nicolet Neurodiagnostic system.
Device Story
Nicolet Wireless EEG Amplifier is a portable, battery or AC-powered device for acquiring, storing, and transmitting electrophysiological signals (EEG, EKG, EMG, EP, temperature, blood pressure, pulse). Available in 32 or 64-channel models; units can be cascaded for higher channel counts. Operates in wired (Ethernet) or wireless (802.11b/g) modes; stores data locally for backup during out-of-range conditions. Used in clinical settings including OR, ICU, LTM, and ambulatory/home environments. Operated by clinicians to facilitate neurological and sleep diagnostics. Data is transmitted to a Nicolet Neurodiagnostic system for review, aiding clinical decision-making by providing continuous or mobile physiological monitoring.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation testing confirmed the device meets design inputs and user needs for an EEG amplifier. Testing included basic safety (IEC 60601-1), EMC (IEC 60601-1-2), usability (IEC 60601-6), risk management (ISO 14971), biocompatibility (ISO 10993-1), and FCC radiated energy/SAR standards.
Technological Characteristics
Portable amplifier (32/64 channels); wired (10/100baseT) or wireless (802.11b/g) connectivity; battery or 110/220VAC power. Local data storage for backup. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26, and ISO 10993-1. IP addressable for network integration.
Indications for Use
Indicated for use as a front-end amplifier to acquire, store, and transmit electrophysiological signals (EEG, EKG, EMG, EP, temperature, blood pressure, pulse) for neurological and sleep testing in ambulatory, OR, LTM, EEG, and ICU settings.
Regulatory Classification
Identification
A physiological signal amplifier is a general purpose device used to electrically amplify signals derived from various physiological sources (e.g., the electroencephalogram).
{0}------------------------------------------------
K103140
## 510(k) SUMMARY
ー。
t
APR - 8 2011
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| 1 | Submitter Information | | | | |
|----|--------------------------------------|--------------------------------------|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | Name | CareFusion 209, Inc. | | | |
| | Address | 1850 Deming Way, Middleton, WI 53562 | | | |
| | Phone number | 608-829-8655 | | | |
| | Fax number | 608-829-8769 | | | |
| | Establishment<br>Registration Number | 3008289288 | | | |
| | Name of contact<br>person | Robert J Burdge | | | |
| | Date prepared | | 6 April 2011 | | |
| 2 | Name of Device | | | | |
| | Trade or proprietary<br>name | Nicolet Wireless EEG Amplifier | | | |
| | Common or usual<br>name | Amplifier, Electroencephalograph | | | |
| | Classification name | Electroencephalograph | | | |
| | Classification Panel | Neurology | | | |
| | Device Class | Class II | | | |
| | Regulation | 21 CFR 882.1400 | | | |
| | Product Code(s) | GWQ, GWL | | | |
| ਤੇ | Legally marketed device(s) | 510(k) #: | Cleared on: | Name | Code: |
| | to which equivalence is<br>claimed | K061908 | 11/06/2006 | Nicoletone V32 Amplifier with Oximetry | GWQ |
| | | K090957 | 05/28/2009 | Blackrock Neuroport Biopotential Signal Processing System | GWL |
| | | K060523 | 03/28/2006 | NeuroPort Instrument | GWQ |
| | | K060803 | 05/02/2006 | gUSBamp | GWL |
| | | K040113<br>K033475 | 02/18/2004<br>01/07/2004 | Sandman SD20 Amplifier<br>Nihon Kohden WEE-1000 Wireless Amplifier | GWL<br>GWQ |
| | | K003175 | 12/27/2000 | Siesta Compumedics | GWQ |
| | | | | | |
| | Reason for 510(k)<br>submission | | NicoletOne base system<br>Battery or Mains power | The changes addressed in this submission include:<br>Addition of wireless communication capability between the amplifier and the<br>Increased channels and sampling rates | |
| 4 | Device Description | | | The Nicolet Wireless Amplifiers are used in a wide variety of EEG applications<br>including: Ambulatory (at home or in transit to home), OR, LTM, EEG, Sleep and<br>ICU. The proposed amplifiers are small portable devices that can run on main<br>electrical A/C power or use a portable battery pack. | |
{1}------------------------------------------------
| Device Description (cont) | | The proposed device consists of two models designated as the 32 channel and the<br>64 channel amplifier.<br><br>These units can acquire from up to 32 or 64 channels, they can be grouped or<br>"cascaded" together for more channels. The amplifiers are capable of acquiring a<br>variety of electrophysiological signals at variable sampling frequencies. These<br>signals include EEG, EKG, EMG, EP, temperature, blood pressure, pulse, and<br>other signals standard to neurological and sleep testing.<br><br>The amplifiers have wireless capability and a battery pack for mobile data<br>acquisition. Wireless access points collect the wireless data transmissions. The<br>amplifiers can be connected to the Ethernet by a cable and also to AC power. The<br>amplifiers are IP addressable and can be connected directly to a network device.<br>In all situations the amplifiers store a copy of the data locally to allow for data<br>back-up.<br><br>This amplifier provides storage and subsequent transmission of data that is not<br>transferred live when the amplifier is in out of range situations. | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 5 | Intended use of the device | The Nicolet Wireless EEG Amplifier is intended to be used as a front end amplifier<br>to acquire, store, and transmit electrophysiological signals in a wired or wireless<br>mode for the Nicolet Neurodiagnostic system. | |
| 6 | Summary of the technological characteristics of the device compared to the predicate device | | |
| | Characteristic | New Device<br>Nicolet Wireless EEG Amplifier | Predicate [Device Name]<br>[510(k) number] |
| | Interface to Amplifier | Wired Ethernet (10/100baseT) or<br>Wireless Ethernet (802.11b/g) | Wired Ethernet (10/100baseT)or Wireless<br>Ethernet (802.11b)<br>Substantially equivalent to the Nihon Kohden<br>WEE-1000 Wireless Amplifier (K033475) |
| | Power | Battery or 110/220VAC | 110 VAC, Battery and USB Port<br>Substantially equivalent to the Nihon Kohden<br>WEE-1000 Wireless Amplifier (K033475) |
| | System components | Patient Cable, Amplifier, CPU and a<br>Display Monitor; Telemetry unit and<br>access point | System components comprised of telemetry<br>unit, Electrode junction box and access point<br>to operate with the current NK commercially<br>available EEG devices<br>Substantially equivalent to the Nihon Kohden<br>WEE-1000 Wireless Amplifier (K033475) |
| | Number of Signal Recording<br>Channels | Up to 32 or 64 channels | 32 and up to 128 with one device; Up to 256<br>by cascading two devices |
| PERFORMANCE DATA | | | |
| 7 | SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL<br>EQUIVALENCE | | |
| | Performance Test Summary-New Device | | |
| | Characteristic | Standard/Test/FDA Guidance | Results Summary |
| | 1. Basic safety | IEC 60601-1 | The proposed device passes the applicable tests and standards. |
| | 2. Programmable Electrical<br>Medical Systems | IEC 60601-4 | The proposed device passes the applicable tests and standards. |
| | 3. Usability | IEC 60601-6 | The proposed device passes the applicable tests and standards. |
| | 4. EMC Compatibility | IEC60601-1-2 | The proposed device passes the applicable tests and standards. |
| | 5. Collateral Safety | IEC 60601-1-1 | The proposed device passes the applicable tests and standards. |
| | 6. Electroencephalographs | IEC 60601-2-26 | The proposed device passes the applicable tests and standards. |
| | 7. Risk Management | ISO 14971 | The proposed device passes the applicable tests and standards. |
| | 8. Biocompatibility- non contact<br>device | ISO 10993-1 | The proposed device passes the applicable tests and standards. |
| | 9. FCC Specific Absorption<br>Ratio (SAR) | FCC Part 15C; FCC OET Bulletin 65,<br>Supplement C | The proposed device passes the applicable tests and standards. |
| | 10. Battery safety | UN/DOT 38.3 Transportation testing<br>for lithium batteries | The proposed device passes the applicable tests and standards. |
| | Summary Discussion of Bench Performance Data | | |
| | The Nicolet EEG Wireless Amplifier device passed all specified test requirements. | | |
| | The validation and verification testing confirmed this device meets user needs and design inputs for an EEG amplifier. | | |
| | Testing also confirmed physical attributes and device performance meet the requirements of the standards listed in the<br>performance testing summary above. These standards address safety, biocompatibility, EMC compatibility, risk,<br>usability, and radiated energy. | | |
| 8 | SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL | | |
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{2}------------------------------------------------
ノ
i
{3}------------------------------------------------
| | EQUIVALENCE AND/OR OF CLINICAL INFORMATION |
|---|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Clinical Performance Data/Information<br>Clinical testing was not performed with this device. |
| 9 | CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA |
| | The Nicolet Wireless EEG Amplifiers meet the functional claims and intended use as described in the product labeling.<br>The safety and effectiveness are substantially equivalent to The K061908- NicoletOne V32 Amplifier with Oximetry, the<br>K090957 Blackrock Neuroport Biopotential Signal Processing System, the K060523 NeuroPort Instrument, the<br>K060803 gUSBamp, the K040113 Sandman SD20 Amplifier, the K033475-Nihon Kohden WEE-1000 Wireless<br>Amplifier and the K003175 Siesta Compumedics described in the submission. |
..
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three overlapping wings or feathers.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Carefusion 209. Inc. c/o Mr. Mark Job Regulatory Technology Services LLC 1394 25" Street, NW Buffalo, MN 55313
Re: K103140
APR - 8 2011
Trade/Device Name: Nicolet Wireless EEG Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWL and GWO Dated: January 20, 2011 Received: January 21, 2011
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
signature
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): __K103140_
Device Name: Nicolet Wireless EEG Amplifier
The Nicolet Wireless EEG Amplifier is intended to be used as a front end amplifier to acquire, store, and transmit electrophysiological signals in a wired or wireless mode for the Nicolet Neurodiagnostic system.
Prescription Use _ XXX_ (Part 21 CFR 801 Subpart D)
.
:
AND/OR
Over-The-Counter use_ (21 CFR Part 801 Subpart C)
(PLEASE DO NO WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Quynh Hoang
(Division S Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K103140
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.