The Audera Pro is intended to be used for the stimulation, recording and measurement of auditory evoked potentials, vestibular evoked myogenic potentials, auditory steady state responses and otoacoustic emissions. The device is indicated for use in the evaluation, identification, documentation and diagnosis of auditory and vestibular disorders. The device is intended to be used on patients of any age. The Audera Pro is intended to be used by qualified medical personnel such as an audiologist, physician, hearing healthcare professional, or trained technician. The Audera Pro is intended to be used in a hospital, clinic, or other healthcare facility with a suitable quiet testing environment. The anatomical sites of contact for auditory evoked potential (AEP) testing are the patient's ear canal (with the contact object being a sound delivery eartip or headphone, or an ear probe and eartip) and the patient's scalp and possibly other body sites (with the contact object being a bone transducer or electrodes that are capable of measuring bio-potentials). The anatomical sites of contact for vestibular evoked myogenic potential (VEMP) testing are the patient's ear canal (with the contact object being a sound delivery eartip or headphone, or an ear probe and eartip) and the patient's head and neck and possibly other body sites (with the contact object being a bone transducer or electrodes that are capable of measuring bio-potentials). The anatomical sites of contact for otoacoustic emission (DPOAE, TEOAE) testing are the patient's ear canal (with the contact object being an ear probe and eartip).
Device Story
Audera Pro is a configurable platform for auditory/vestibular testing; inputs include bio-potentials from scalp/neck electrodes and acoustic responses from ear canal probes. System comprises a laptop PC (Windows 10 Pro) connected to a specialized hardware interface. Device delivers auditory stimuli (clicks, tones, chirps) via headphones, insert earphones, or bone vibrators; records evoked responses (AEP, VEMP, ASSR, OAE). Used in hospitals/clinics by audiologists, physicians, or technicians. Software controls hardware, collects/analyzes signals, and displays results for clinician review. Output aids in diagnosing sensory-neural/hearing conditions and estimating behavioral hearing thresholds. Benefits include objective assessment of cochlear/auditory nerve/brainstem function.
Clinical Evidence
No clinical data. Bench testing only. Performance verified using simulators to compare Audera Pro output against predicate (K163326) across all modules (EP, ASSR, DPOAE, TEOAE). Electrical safety, EMC, and usability testing performed per IEC 60601-1, 60601-1-2, 60601-2-40, 60645-1/3/6/7, and 62366 standards.
Technological Characteristics
PC-based system with external hardware interface. Connectivity: USB. Power: AC line. Display: Laptop LCD. Software: Windows 10 Pro. Sensing: EEG amplifier (2 channels, >10 MΩ impedance, ≥110 dB CMRR). Stimuli: Clicks, tones, chirps (125Hz-12kHz). Transducers: Headphones, insert earphones, bone vibrators. Sterilization: N/A (disposable electrodes).
Indications for Use
Indicated for evaluation, identification, documentation, and diagnosis of auditory and vestibular disorders in patients of any age.
Regulatory Classification
Identification
An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Grason-Stadler Inc. % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct. Naples, Florida 34114
Re: K193033
Trade/Device Name: GSI Audera Pro Regulation Number: 21 CFR 882.1900 Regulation Name: Evoked Response Auditory Stimulator Regulatory Class: Class II Product Code: GWJ Dated: March 27, 2020 Received: March 31, 2020
Dear Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193033
Device Name GSI Audera Pro
### Indications for Use (Describe)
The Audera Pro is intended to be used for the stimulation, recording and measurement of auditory evoked potentials, vestibular evoked myogenic potentials, auditory steady state responses and otoacoustic emissions. The device is indicated for use in the evaluation, identification, documentation and diagnosis of auditory and vestibular disorders. The device is intended to be used on patients of any age.
The Audera Pro is intended to be used by qualified medical personnel such as an audiologist, physician, hearing healthcare professional, or trained technician. The Audera Pro is intended to be used in a hospital, clinic, or other healthcare facility with a suitable quiet testing environment.
The anatomical sites of contact for auditory evoked potential (AEP) testing are the patient's ear canal (with the contact object being a sound delivery eartip or headphone, or an ear probe and eartip) and the patient's scalp and possibly other body sites (with the contact object being a bone transducer or electrodes that are capable of measuring bio-potentials). The anatomical sites of contact for vestibular evoked myogenic potential (VEMP) testing are the patient's ear canal (with the contact object being a sound delivery eartip or headphone, or an ear probe and eartip) and the patient's head and neck and possibly other body sites (with the contact object being a bone transducer or electrodes that are capable of measuring bio-potentials). The anatomical sites of contact for otoacoustic emission (DPOAE, TEOAE) testing are the patient's ear canal (with the contact object being an ear probe and eartip).
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | × |
|----------------------------------------------|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(K) Summary
Image /page/3/Picture/2 description: The image shows the logo for Grason-Stadler, also known as GSI. The logo features a stylized blue sound wave graphic to the left of the letters "gsi" in bold, black font. Below the letters "gsi" is the text "Grason-Stadler" in a smaller, thinner font, also in black.
#### SUBMITTER 1.
Grason-Stadler Inc. 10395 West 70th Street Eden Prairie, MN 55344
Tel: 952 278-4402
Contact Person: Brent Nissly, General Manager/COO Date Prepared: February 28, 2020
#### DEVICE II.
Name of Device: GSI Audera Pro™ Common or Usual Name: Audera Pro Classification Name: Evoked Response Auditory Stimulator (21 CFR 882.1900) Regulatory Class: Class II Product Code: GWJ
#### III. PREDICATE DEVICE
K163326, Predicate for: Hardware platform for all modules, Software platform for VEMP and AEP modules Manufacturer: Intelligent Hearing Systems Trade/Device Name: SmartEP (Duet platform) Classification Name: Evoked Response Auditory Stimulator (21 CFR 882.1900) Regulatory Class: Class II Product Code: GWJ, GWF, GWF, GWE, ETN
K061443, Predicate for: ASSR, DPOAE, and TEOAE modules Trade/Device Name: Smart USBLite (with SmartEP, SmartScreener, SmartOAE, SmartTrOAE, & SmartEP-ASSR) Classification Name: Audiometer (21CFR 874.1050) Regulatory Class: Class II Product Code: GWJ; EWO; GWL
#### IV. DEVICE DESCRIPTION
The device is a configurable platform used to aid in the screening and diagnosis of sensory-neural and hearing conditions. It is capable of performing the following procedures: Auditory Evoked Potentials (EP), Auditory Steady-State Response (ASSR), Distortion Products Otoacoustic Emissions (DPOAE), Transient Evoked Otoacoustic Emissions (TEOAE), and vestibular evoked myogenic potentials (VEMP). The device system consists of a laptop PC with Windows 10 Pro, placed on top of or beside a specialized hardware implementation interface (i.e. platform) for the procedures. Software in the laptop controls the specialized hardware and collects and analyzes the resulting signals. Transducers and various accessories connect to the specialized hardware via connectors on the back of the hardware package.
#### INDICATIONS FOR USE V.
The Audera Pro is intended to be used for the stimulation, recording and measurement of auditory evoked potentials, vestibular evoked myogenic potentials, auditory steady state responses and otoacoustic emissions. The device is
{5}------------------------------------------------
indicated for use in the evaluation, identification, documentation and diagnosis of auditory and vestibular disorders. The device is intended to be used on patients ofany age.
The Audera Pro is intended to be used by qualified medical personnel such as an audiologist, physician, hearing healthcare professional, or trained technician. The Audera Pro is intended to be used in a hospital, clinic, or other healthcare facility with a suitable quiet testing environment. The anatomical sites of contact for and the title (AEP) testing are the patient's ear canal (with the contact object being a sound delivery eartip or headphone, or an ear probe and eartip) and the patient's scalp and possibly other body sites (with the contact object being a bone transducer or electrodes that are capable of measuring bio-potentials).
The anatomical sites of contact for vestibular evoked myogenic potential (VEMP) testing are the patient's ear canal (with the contact object being a sound delivery eartip or headphone, or an ear probe and eartip) and the patient's head and neck and possibly other body sites (with the contact object being a bone transducer or electrodes that are capable of measuring bio-potentials). The anatomical sites of contact for otoacoustic emission (DPOAE, TEOAE) testing are the patient's ear canal (with the contact object being an ear probe and eartip)
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The following technological differences exist between the GSI Audera Pro™ MEG and the predicate devices. In addition, changes made to the USB Jr. Duet hardware platform include:
- Changes to the connectors on the chassis have been changed to mate with accessories;
- . Changes to the equalization circuit for the OAE probe to match probe model; and
- Reduction of earphone and bone vibrator impedance (from 300 to 10Ω)
| | Predicate for AEP and VEMP:<br>K163326<br>Predicate for ASSR,<br>DPOAE, and TEOAE: K061443 | Audera Pro: K193033 |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Photo | Image: Predicate for AEP and VEMP | Image: Audera Pro |
| | Predicate for AEP and VEMP:<br>K163326<br>Predicate for ASSR,<br>DPOAE, and TEOAE: K061443 | Audera Pro: K193033 |
| Indications for<br>Use: | SmartEP is an evoked response testing and<br>diagnostic device, that is capable of eliciting,<br>acquiring, and measuring auditory,<br>somatosensory, visual, and vestibular evoked<br>myogenic potential data, as well as providing<br>nerve stimulation and monitoring. The intended<br>use of the SmartEP device is to objectively record<br>evoked responses from patients of all ages upon<br>the presentation of sensory stimuli. The product is<br>indicated for use as a diagnostic aid and adjunctive<br>tool in sensory related disorders (i.e., auditory,<br>somatosensory, visual, and vestibular) and in<br>surgical procedures for inter-operative nerve<br>monitoring. The SmartEP system is intendedto be<br>used by trained personnel in a hospital, nursery,<br>clinic, audiologist's, EP technologist's, surgeon's,<br>or physician's office, operating room, or other<br>appropriate setting. The intended use of the Smart<br>USBLite device system is for the recording of<br>auditory evoked potential, otoacoustic emissions,<br>& auditory steady-state evoked potential data. The<br>product is intended to be used as a diagnostic aid<br>in auditory and hearing related disorders, as an<br>objective measure of cochlear function, and as an<br>adjunctive tool in the estimation of behavioral<br>hearing thresholds on patients of all ages | The Audera Pro is intended to be used for the<br>stimulation, recording and measurement of<br>auditory evoked potentials, vestibular evoked<br>myogenic potentials, auditory steady state<br>responses and otoacoustic emissions. The device<br>is indicated for use in the evaluation,<br>identification, documentation and diagnosis of<br>auditory and vestibular disorders. The device is<br>intended to be used on patients of any age.<br>The Audera Pro is intended to be used by<br>qualified medical personnel such as an<br>audiologist, physician, hearing healthcare<br>professional, or trained technician. The Audera<br>Pro is intended to be used in a hospital, clinic, or<br>other healthcare facility with a suitable quiet<br>testing environment.<br>The anatomical sites of contact for auditory<br>evoked potential (AEP) testing are the patient's<br>ear canal (with the contact object being a sound<br>delivery eartip or headphone, or an ear probe and<br>eartip) and the patient's scalp and possibly other<br>body sites (with the contact object being a bone<br>transducer or electrodes that are capable of<br>measuring bio-potentials). The anatomical sites of<br>contact for vestibular evoked myogenic potential<br>(VEMP) testing are the patient's ear canal (with<br>the contact object being a sound delivery eartip or<br>headphone, or an ear probe and eartip) and the<br>patient's head and neck and possibly other body<br>sites (with the contact object being a bone<br>transducer or electrodes that are capable of<br>measuring bio-potentials). The anatomical sites of<br>contact for otoacoustic emission (DPOAE, TEOAE)<br>testing are the patient's ear canal (with the<br>contact object being an ear probe and eartip).<br>DIFFERENCE: Does not support Somatosensory<br>and Visual Evoked Potential and Nerve Stimulation<br>modules |
| Tests Performed/<br>Associated<br>predicate<br>Clearance | AEP Auditory Evoked Potentials K163326 ASSR Auditory Steady-State Response K061443 VEMP Vestibular Evoked Myogenic<br>Potential K163326 TEOAE Transient Evoked Otoacoustic<br>Emissions K061443 DPOAE Distortion Products Otoacoustic<br>Emissions K061443 Plus: The SEP, VEP, and nerve stimulation | AEP Auditory Evoked Potentials K163326 ASSR Auditory Steady-State Response K061443 VEMP Vestibular Evoked Myogenic Potential<br>K163326 TEOAE Transient Evoked Otoacoustic Emissions<br>K061443 DPOAE Distortion Products Otoacoustic<br>Emissions K061443 DIFFERENCE: Does not support Somatosensory and<br>Visual Evoked Potential and Nerve Stimulation<br>modules |
| | Predicate for AEP and VEMP:<br>K163326 | Audera Pro: K193033 |
| | Predicate for ASSR,<br>DPOAE, and TEOAE: K061443 | |
| Configuration | PC-based system with external hardware<br>platform and external hardware peripherals<br>(USB interface) | SAME |
| Hardware<br>Implementation | PC-based system with external hardware<br>platform and peripherals (USB interface) | SAME |
| | Technological Characteristics | |
| Preamplifier/<br>Amplifier | Gain: 5K to 200K (8 steps)<br>HPF cutoffs: 0.1 Hz to 300 Hz (8 steps)<br>LPF cutoffs: 30 Hz to 500 Hz (8 steps)<br>Impedance test capable | SAME |
| EEG Amplifier<br>Channels | TWO | SAME |
| Stimulator<br>Frequencies<br>(In Hz) | 125, 250 500 750 1000 2000 3000<br>4000 6000 8000, 16,000, Tone Burst, Click | SAME |
| Stimulator<br>transducers | Headphones Insert<br>Earphones Bone<br>Conductor Probe Ear<br>Tips | SAME |
| Patient<br>electrode | Self stick single use disposable | SAME |
| OAE<br>Probe | Two channels of acquisition; Two speakers | One channel of acquisition;<br>One speaker |
| | | DIFFERENCE: Can perform<br>same testing, but with each<br>ear individually |
| Interface<br>Connectors | Headphones DIN Bone<br>oscillator DIN Speakers DIN<br>OAE Probe DIN<br>Patient Connection 6 pin custom USB:<br>USB A | Headphones 2 x ¼ inch<br>phone (DIFFERENCE)<br>Bone oscillator 1 x ¼ inch<br>phone (DIFFERENCE) |
| | | Speakers 2 x RCA phono<br>(DIFFERENCE) OAE Probe HDMI<br>(DIFFERENCE) |
| | | Patient Connection 6 pin<br>custom (SAME) |
| |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.