The Eclipse with VEMP (Vestibular Evoked Myogenic Potential) is intended for vestibular evoked myogenic potential testing to assist in the assessment of vestibular function. The target population for Eclipse with VEMP includes patients aged from 8 years and up. The device is to be used only by qualified medical personnel with prior knowledge of the medical and scientific facts underlying the procedure.
Device Story
Eclipse with VEMP is a PC-based audiometric system for performing Vestibular Evoked Myogenic Potential (VEMP) testing. It utilizes a hardware platform, preamplifier, stimulation transducers, and recording electrodes to measure evoked potentials from sternocleidomastoid (cVEMP) or inferior oblique (oVEMP) muscles. The system provides auditory stimulation to assess otolith organs (saccule/utricle) and afferent pathways. Operated by qualified medical personnel in hospitals or clinics, the device includes an integrated EMG monitor to guide electrode placement and ensure muscle activation is within range for recording. The software displays waveforms for manual, subjective interpretation by the clinician. By facilitating VEMP testing, the device assists medical practitioners in diagnosing balance disorders. It does not provide an automated diagnosis.
Clinical Evidence
Two internal clinical evaluations (test-retest reliability and reproducibility) were conducted with 14 normal adults. cVEMP and oVEMP waveforms were recorded using 500Hz tone bursts at 95dBnHL. Test-retest correlation means were 0.9374 (cVEMP) and 0.9095 (oVEMP). Reproducibility study correlation means were 0.8794 (cVEMP) and 0.8923 (oVEMP). Results indicate high reliability and reproducibility, consistent with or exceeding performance metrics reported for the predicate device (K143670).
Technological Characteristics
PC-based system with external hardware platform, preamplifier, stimulation transducers, and recording electrodes. Connectivity via USB interface. Energy source: AC power. Features integrated EMG monitoring for VEMP data collection. Software-controlled protocols. No changes to electrical safety or EMC compared to the cleared Eclipse platform (K090406).
Indications for Use
Indicated for vestibular evoked myogenic potential (VEMP) testing to assist in the assessment of vestibular function in patients aged 8 years and older.
Regulatory Classification
Identification
An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.
Eclipse with EP15, EP25 software modules (K090406)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 23, 2017
Interacoustics A/S Mr. Erik Nielsen Director, Regulatory & Compliance Audiometer Allé 1 5500 Middelfart, Denmark
Re: K162037
Trade/Device Name: Eclipse with VEMP Regulation Number: 21 CFR 882.1900 Regulation Name: Evoked Response Auditory Stimulator Regulatory Class: Class II Product Code: GWJ Dated: February 08, 2017 Received: February 21, 2017
Dear Mr. Nielsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162037
Device Name Eclipse with VEMP
Indications for Use (Describe)
The Eclipse with VEMP (Vestibular Evoked Myogenic Potential) is intended for vestibular evoked myogenic potential testing to assist in the assessment of vestibular function. The target population for Eclipse with VEMP includes patients aged from 8 years and up.
The device is to be used only by qualified medical personnel with prior knowledge of the medical and scientific facts underlying the procedure.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(K) SUMMARY
### Eclipse with VEMP
### Submitter Information:
| Company Name | Interacoustics A/S |
|-----------------------|------------------------------------|
| Address | Audiometer Allé |
| | 5500 Middelfart |
| | Denmark |
| Phone | +45 6371 3555 |
| e-mail | erni@dgs.com |
| Contact Person | Erik Nielsen, |
| | Director, Regulatory & Compliance, |
| Date Summary Prepared | July 18 2016 |
## Device Identification:
| Trade Name | Eclipse with VEMP |
|---------------------|-----------------------------------------|
| Common Name | Auditory Evoked Response Stimulator |
| Classification Name | Stimulator, Auditory, Evoked Response |
| Product Code Class | GWJ |
| Panel | Neurology |
| Device Class | Class II (According to 21 CFR 882.1900) |
#### Predicate Device 1:
Predicate Device Manufacturer 510(k) No. Date Cleared
ICS CHARTR EP 200 WITH VEMP GN OTOMETRICS A/S7 K143670 10/23/2015
## Predicate Device 2:
Predicate Device Manufacturer 510(k) No. Date Cleared
Eclipse with EP15, EP25 software modules Interacoustics A/S K090406 07/22/2009
{4}------------------------------------------------
### Device Description
The Eclipse with VEMP is audiometric equipment intended to perform various Otoacoustic Emissions (OAEs) and Auditory Evoked Potential evaluations. The Eclipse is operated solely from PC based software modules. The Eclipse platform performs the physical measurements. The protocols are created in the software modules. The Eclipse consists of a hardware platform, a preamplifier, stimulation transducers and recording electrodes.
VEMP evaluations are tests of the vestibular portion of the inner ear and acoustic nerve, evoked with an auditory stimulation. The evoked response results in a potential recorded from the sternocleidomastoid (neck) muscles or the inferior oblique (eye) muscles. VEMP is not a test of the neck or eye musculature directly; the clinician is interested in the vestibular anatomy which triggers the response. The cervical Vestibular Evoked Myogenic Potential (cVEMP) is an evoked potential measured from the sternocleidomastiod (SCM) muscle and the ocular VEMP (oVEMP) is an evoked potential measured from the inferior oblique muscle. Both tests are used to assess the otolith organs (saccule and utricle) of the vestibular system and their afferent pathways and assist medical practitioners in the diagnosis of various balance disorders.
Summary: VEMP is Auditory Evoked Potentials like ABR obtained using any commercially available EP system. The addition of the VEMP module to Eclipse will make it possible for clinicians to conduct VEMP tests while using EMG (electromyography) monitoring and scaling. The VEMP function of the Eclipse with VEMP does not make a diagnosis. It only assists the medical professional.
## Indications for Use
The Eclipse with VEMP (Vestibular Evoked Myogenic Potential) is intended for vestibular evoked myogenic potential testing to assist in the assessment of vestibular function. The target population for Eclipse with VEMP includes patients aged from 8 years and up.
## Technological characteristics
The Eclipse with VEMP consists of a hardware platform, a preamplifier, stimulation transducers and recording electrodes. The Eclipse VEMP is solely operated from a PC with VEMP software
{5}------------------------------------------------
### Product Comparisons charts summary
We have chosen to compare Eclipse with VEMP with the currently cleared Eclipse with EP15, EP25 and the currently cleared Chartr EP200 for the following reasons.
- Chartr EP200 has the same medical purpose as Eclipse with VEMP and is intended as the . primary predicate for VEMP functionality
- . Eclipse with EP15, EP25 supports identical hardware as Eclipse with VEMP and provide identical technical characteristics
| Description | Predicative K143670<br>Chartr EP 200 with VEMP | Modified K162037<br>Eclipse with VEMP | Equivalence |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for use | The ICS Chartr EP 200 with VEMP is<br>indicated for auditory evoked<br>potential testing as an aid in<br>assessing hearing loss and lesions in<br>the auditory pathway. The<br>Vestibular Evoked Myogenic<br>Potential is indicated for vestibular<br>evoked potential testing as an aid in<br>assessing vestibular function in<br>adult patients. The device is to be<br>used only by qualified medical<br>personnel with prior knowledge of<br>the medical and scientific facts<br>underlying the procedure. | The Eclipse with VEMP (Vestibular<br>Evoked Myogenic Potential) is<br>intended for vestibular evoked<br>myogenic potential testing to assist in<br>the assessment of vestibular function.<br>The target population for Eclipse with<br>VEMP includes patients aged from 8<br>years and up. | Same intended<br>use of assisting<br>in the<br>assessment of<br>vestibular<br>function, but<br>the Eclipse with<br>VEMP is<br>indicated for<br>children 8 years<br>and older and<br>substantial<br>equivalence<br>was established<br>using clinical<br>literature. |
| Tests<br>Performed | Auditory Brainstem Response (ABR)<br>Electrocochleography (ECochG)<br>Auditory Middle Latency Response<br>(AMLR)<br>Auditory Late Response (ALR)<br>P300 - optional<br>Auditory Steady State Response<br>(ASSR) — Optional<br>Vestibular Evoked Myogenic<br>Potential (VEMP) — optional | Auditory Brainstem Response (ABR)<br>Electrocochleography (ECochG)<br>Auditory Middle Latency Response<br>(AMLR)<br>Auditory Late Response (ALR)<br>P300/Mismatch Negativity (MMN)<br>Auditory Steady State Response (ASSR)<br>Vestibular Evoked Myogenic Potential<br>(VEMP) — optional | Same<br>The addition of<br>VEMP is the<br>subject of this<br>510(k |
| Anatomical<br>sites | Examination of Ear and hearing<br>nerves and vestibular functions | Examination of Ear, hearing nerves and<br>vestibular functions | Same |
| Environment<br>of use | Hospitals and clinics | Hospitals and clinics | Same |
| Evaluation of<br>results | Manually / subjective | Manually / subjective | Same |
| Description | Predicative K143670<br>Chartr EP 200 with VEMP | Modified K162037<br>Eclipse with VEMP | Equivalence |
| Photo | Image: Chartr EP 200 with VEMP | Image: Eclipse with VEMP | Similar<br>The physical<br>differences are<br>not appraised<br>to raise any<br>issues about<br>safety and<br>efficiency |
| System | Chartr EP 200 connected to a PC<br>(PC-based system with external<br>hardware platform and peripherals<br>(USB interface)) | Eclipse platform connected to PC (PC-<br>based system with external hardware<br>platform and peripherals (USB<br>interface)) | Similar<br>The physical<br>differences are<br>not appraised<br>to raise any<br>issues about<br>safety and<br>efficiency |
| VEMP<br>Monitor | Image: VEMP Monitor | A visual or acoustical VEMP monitor is<br>available. The visual VEMP monitor is<br>displayed on an external screen and<br>indicates below range (low), in range<br>for recording (good) and outside range<br>(high). | Similar.<br>Both methods<br>have three<br>ranges for<br>assessing the<br>EMG<br>monitoring. We<br>appraise the<br>difference in<br>presentation<br>does not raise<br>any issues<br>about safety<br>and efficiency |
| Electrode<br>montage | 2x on each SCM muscle one for<br>recording and one for EMG<br>monitoring on external device, 1 at<br>clavicle junction, 1 low forehead.<br>In total 6 electrodes | 1 on each SCM which is for both<br>recording and monitoring via the built<br>in software EMG monitor, 1 at clavicle<br>junction, 1 at low forehead.<br>In total 4 electrodes | Similar.<br>Both methods<br>measures at the<br>same anatomic<br>place. The<br>difference in<br>configurations<br>are not<br>appraised to<br>raise any issues<br>about safety<br>and efficiency |
| Description | Predicative K143670<br>Chartr EP 200 with VEMP | Modified K162037<br>Eclipse with VEMP | Equivalence |
| Preamplifier | 2 channels | 1 or 2 channels | Similar<br>(subset)<br>The differences<br>are appraised<br>not to raise any<br>issues about<br>safety and<br>efficiency |
{6}------------------------------------------------
{7}------------------------------------------------
| Description | Predictive K090406<br>Eclipse with EP15, EP25 | Modified K162037<br>Eclipse with VEMP | Equivalence |
|----------------------|------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Target<br>population | All ages and genders | From age 8 and up both genders. | Similar<br>(Subset)<br>The VEMP<br>functionality<br>has a more<br>narrow age<br>range |
| Stimulus | Click, Tone Burst, NB CE-Chirp, CE-<br>Chirp | Click, Tone Burst, NB CE-Chirp, CE-<br>Chirp | Same |
| Energy<br>Source | AC Power | AC Power | Same |
| Photo | Image: Predictive K090406 Eclipse with EP15, EP25 | Image: Modified K162037 Eclipse with VEMP | Same |
| System | Eclipse platform connected to PC (PC-<br>based system with external hardware<br>platform and peripherals (USB<br>interface)) | Eclipse platform connected to PC (PC-<br>based system with external hardware<br>platform and peripherals (USB<br>interface)) | Same |
{8}------------------------------------------------
#### Brief discussion of non-clinical tests
Bench verification and testing demonstrates compliance with system and software specifications. There are no changes to the hardware including electrical safety or EMC compared to the already cleared Eclipse platform (K090406). Software verification including risk analysis was performed to demonstrate compliance with the specifications
#### Brief discussion of clinical evaluations
VEMP, like ABR, is an evoked potential obtained using any commercially available EP system. The Eclipse with VEMP module has the ability to perform EMG monitoring during VEMP data collection. The VEMP module of the Eclipse does not make a diagnosis is made by a qualified medical professional. Testing the vestibular evoked myogenic potential (VEMP) is welldocumented in the literature.
Two internal clinical evaluations have been conducted to show the test-reliability and the reproducibility of the VEMP using the Eclipse with VEMP module
A summary of the studies follows:
Testing was conducted at two sites (one in Denmark and one in the United States). Test subjects included 14 normal adults for cVEMP and oVEMP.
| Min number of runs per ear | 2 repeatable waveforms |
|---------------------------------------|----------------------------------------------------|
| Transducer | Insert phones |
| Stimulus | 500Hz tone burst (Blackman) |
| Stimulus rate | 5.1/second |
| Stimulus level | 95dBnHL |
| High Pass Filter | 10Hz |
| Low Pass Filter | 1000Hz |
| # of sweeps | ~150 samples per waveform |
| EMG controlled stimulus (c VEMP only) | 50.0uV RMS to 150uV RMS (subject to aim for 100uV) |
Collection parameters for both the cVEMP and oVEMP tests:
The correlation results for the test-retest reliability study were as follows:
#### Correlation of the cVEMP waveform (5 – 35ms)
| Mean | 0.9374 |
|------------------------------|--------|
| Median | 0.9562 |
| IQR (75th – 25th percentile) | 0.0547 |
#### Correlation of the oVEMP waveform (4 – 20ms)
| Mean | 0.9095 |
|------------------------------|--------|
| Median | 0.9280 |
| IQR (75th – 25th percentile) | 0.0738 |
{9}------------------------------------------------
Mean and median values indicate a high degree of test-retest repeatability of the cVEMP and oVEMP measurements. The IQR (75th – 25th percentile) values are low, indicating a large number of the correlation coefficients are clustered near the median. In practical terms, the spread in correlation coefficients is not so large indicating, alongside the high median value, that test-retest measurements are typically highly correlated with each other for both the cVEMP and oVEMP measurements.
The correlation results for the reproducibility study were as follows:
| Mean | 0.8794 |
|------------------------------|--------|
| Median | 0.9235 |
| IQR (75th – 25th percentile) | 0.0955 |
Correlation of the cVEMP waveform (5 – 35ms)
Correlation of the oVEMP waveform (4 – 20ms)
| Mean | 0.8923 |
|------------------------------|--------|
| Median | 0.9100 |
| IQR (75th - 25th percentile) | 0.1112 |
Mean and median values indicate a high degree of reproducibility for the cVEMP and oVEMP measurements recorded on different days. The IQR (75th – 25th percentile) values are higher than for the test-retest cVEMP and oVEMP measurements. They are still relatively low, indicating a large number of the correlation coefficients are clustered near the median. The reason for the increase in IQR is probably to be expected, as each of the cVEMP and oVEMP measurements were made on different days with slightly different electrode placements, by different testers.
In the GN Otometrics K143670 (510(k)) document, mean correlation values for both cVEMPs and oVEMPs are reported for a reproducibility study (see table below). The mean correlation values recorded with the Eclipse with VEMP module for both the test-retest reliability study and the reproducibility study are similar to or higher than those reported by GN Otometrics in their 510(k).
| Correlation values reported in K143670 (510(K)) | | |
|-------------------------------------------------|-------------------------|----------------------|
| | cVEMP (5 - 35ms) | oVEMP (4 - 20 ms) |
| Mean | 0.9146 (R) / 0.9162 (L) | 0.926 (R) / 0.93 (L) |
## Conclusion
We appraise that the results of this comparison, the clinical evaluations and literature review, demonstrates that the Eclipse with VEMP (the modified indications for use) is safe and efficient and is substantially equivalent to the marketed predictive devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.