ECHO-SCREEN III PRO HEARING SCREENER WITH BARCODE SCANNER, ECHO-SCREEN III PRO HEARING SCREENER WITHOUT BARCODE SCANNER,
K141446 · Natus Medical Incorporated · GWJ · Aug 22, 2014 · Neurology
Device Facts
Record ID
K141446
Device Name
ECHO-SCREEN III PRO HEARING SCREENER WITH BARCODE SCANNER, ECHO-SCREEN III PRO HEARING SCREENER WITHOUT BARCODE SCANNER,
Applicant
Natus Medical Incorporated
Product Code
GWJ · Neurology
Decision Date
Aug 22, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.1900
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Echo-Screen III hearing screener models are based upon otoacoustic emission (OAE) and auditory brainstem response (ABR) technology. The device is intended to screen hearing for newborns through adults, including geriatric patients. The device does not measure hearing per se, but helps to determine whether or not a hearing loss may be present. The Echo-Screen III product family consists of handheld, automated OAE and ABR based hearing screening systems which are easy to use. The measurement flow is menu guided and the evaluation is based upon signal statistics. The Echo-Screen III devices are intended to be used by trained personnel in a medical or school environment. The Echo-Screen III models are not intended for fitting assistive listening devices such as hearing aids or cochlear implants.
Device Story
Portable, handheld, battery-operated hearing screener; utilizes OAE and ABR technologies. Inputs: acoustic signals via probe (TEOAE click trains, DPOAE pure tone pairs) or electrical activity via three scalp electrodes (AABR broadband clicks/chirps). Processing: automated signal statistics; binomial statistical tests for OAE/AABR pass/refer determination. Outputs: Pass/Refer result displayed on color screen. Used in medical or school environments by trained personnel. Features Android OS, hardware/touch keyboard, optional barcode scanner, docking station. Benefits: provides rapid, automated hearing screening without requiring manual interpretation of complex waveforms; assists in identifying patients needing further diagnostic evaluation.
Clinical Evidence
No clinical data. Bench testing, design verification, and validation were performed to demonstrate performance specifications and substantial equivalence to the predicate device.
Technological Characteristics
Handheld, battery-operated (Li-ion). OAE sensing: miniature microphone/receiver probe. ABR sensing: three-electrode scalp array. Connectivity: docking station for data transfer. Software: Android OS, C and Java languages. Algorithms: binomial statistical tests for TEOAE and AABR; statistical discrimination for DPOAE. Form factor: color screen, hardware/touch keyboard, optional barcode scanner.
Indications for Use
Indicated for hearing screening in populations ranging from newborns to adults, including geriatric patients, to determine the potential presence of hearing loss.
Regulatory Classification
Identification
An evoked response auditory stimulator is a device that produces a sound stimulus for use in evoked response measurements or electroencephalogram activation.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a silhouette effect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 22, 2014
Natus Medical Incorporated Ms. Judy Buckham Regulatory Affairs Associate 5900 First Avenue South Seattle, WA 98108
Re: K141446
> Trade/Device Name: Echo-Screen III Pro Regulation Number: 21 CFR 882.1900 Regulation Name: Evoked Response Auditory Stimulator Regulatory Class: Class II Product Code: GWJ, EWO Dated: July 23, 2014 Received: July 24, 2014
Dear Ms. Buckham,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141446
Device Name Echo-Screen III
#### Indications for Use (Describe)
The Echo-Screen III hearing screener models are based upon otoacoustic emission (OAE) and auditory brainstem response (ABR) technology.
The device is intended to screen hearing for newborns through adults, including geriatric patients. The device does not measure hearing per se, but helps to determine whether or not a hearing loss may be present.
The Echo-Screen III product family consists of handheld, automated OAE and ABR based hearing systems which are easy to use. The measurement flow is menu guided and the evaluation is based upon signal statistics. The Echo-Screen III devices are intended to be used by trained personnel in a medical or school environment. The Echo-Screen III models are not intended for fitting assistive listening devices such as hearing aids or cochlear implants.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/2/Picture/15 description: The image shows the FDA logo. The logo is in a light gray color. The letters "FDA" are large and bold.
# Ting Zhang -S 2014.08.22 09:47:45 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) SUMMARY Echo-Screen III
| Manufacturer's Name: | Natus Medical Incorporated<br>5900 First Avenue South<br>Seattle, WA 98108 |
|----------------------------------|----------------------------------------------------------------------------|
| Corresponding Official: | Judy Buckham<br>Regulatory Affairs Associate |
| Telephone Number:<br>Fax Number: | 206 268 5187<br>206 268 5104 |
| Summary Date: | August 21, 2014 |
| Trade Name: | Echo-Screen III |
Common or Usual Name: Audiometer and Evoked Response Auditory Stimulator
| Classification Name and Number: | Audiometer<br>21 CFR 874.1050<br>Product Code: EWO<br><br>Evoked Response Auditory Stimulator<br>21 CFR 882.1900<br>Product Code: GWJ |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device: | K013977 Echo-Screen T, TA, TD, TDA, TC |
## Device Description:
The Echo-Screen III hearing screener is a portable, handheld, battery-operated device that can detect hearing loss using Otoacoustic Emission or Auditory Brainstem Response screening technologies. The Echo-Screen III may be configured to support one or any combination of TEOAE, DPOAE, and AABR technologies.
The device represents the next generation of the Echo-Screen product line with key enhancements over the previously cleared predicate Echo- Screen T, TA, TD, TDA, TC [K013977], hereinafter referred to as the Echo-Screen T series; specifically, use of the Android operating system, programming upgrade to C and Java languages, addition of a color screen and built-in full hardware keyboard plus icons and on-screen touch keyboard, optional barcode scanner, Li-ion rechargeable battery, and inclusion of a docking station for battery charging and data transfer.
## Intended Use:
The Echo-Screen III hearing screener models are based upon otoacoustic emission (OAE) and auditory brainstem response (ABR) technology.
{5}------------------------------------------------
The device is intended to screen hearing for newborns through adults, including geriatric patients. The device does not measure hearing per se, but helps to determine whether or not a hearing loss may be present.
The Echo-Screen III product family consists of handheld, automated OAE and ABR based hearing screening systems which are easy to use. The measurement flow is menu guided and the evaluation is based upon signal statistics. The Echo-Screen III devices are intended to be used by trained personnel in a medical or school environment. The Echo-Screen III models are not intended for fitting assistive listening devices such as hearing aids or cochlear implants.
## Technological Characteristics:
Both the Echo-Screen III and the predicate Echo-Screen T series perform transient evoked otoacoustic emissions (TEOAE) and distortion product otoacoustic emissions (DPOAE) tests. The devices produce a controlled acoustic signal in the ear canal and measure the resulting evoked emissions that are generated by the inner ear as a result of normal cochlear function. The stimuli are presented via miniature receivers and the acoustic response in the external ear canal is recorded via a miniature microphone, all embedded in the OAE probe. For TEOAE, the stimulus consists of short duration, transient nonlinear acoustic click trains. For DPOAE, the stimulus consists of pure tone pairs.
. For TEOAEs, the system uses a binomial statistical test to give either a Pass or Refer result. The device calculates the statistical probability that an emission has been recorded at a succession of sampling points ranging from 6 to 12 ms after the stimulus.
For DPOAEs, the system applies a statistical algorithm to discriminate the sound . of the environment (noise) from the cochlear sound response. Based upon criterion set up for four f2 frequencies, a Pass or Refer result is determined.
Automated auditory brainstem response (AABR®) testing involves the placement of three recording electrodes on the patient's head. The electrodes record electrical activity generated by the auditory nervous system that result from the presentation of a stimulus into the ear. The stimulus for ABR is a short duration, transient broadband acoustic stimulus, i.e., acoustic click / ultra-short frequency sweeps or chirps. The AABR technology uses a statistical approach for the Pass criteria that is based on binomial statistics.
The TEOAE, DPOAE, and ABR screening test performance of the Echo-Screen III is equivalent to the performance of the predicate Echo-Screen T series.
## Clinical Tests: N/A
## Nonclinical Tests:
Design verification and validation were performed to assure that the Echo-Screen III meets its performance specifications and demonstrates equivalence to the specified predicate device.
# Conclusions:
The verification and validation summary and risk analysis documentation provided in this 510(k) support the conclusion that the Echo-Screen III is safe and effective.
{6}------------------------------------------------
The Echo-Screen III is substantially equivalent to the Echo-Screen T series cleared under K013977.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.