K170462 · KARL STORZ Endoscopy-America, Inc. · GWG · May 16, 2017 · Neurology
Device Facts
Record ID
K170462
Device Name
KARL STORZ ShuntScope
Applicant
KARL STORZ Endoscopy-America, Inc.
Product Code
GWG · Neurology
Decision Date
May 16, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1480
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The KARL STORZ ShuntScope is intended for visualization purposes during cranial diagnostic and therapeutic procedures.
Device Story
KARL STORZ ShuntScope is a fiberoptic neurological endoscope used for visualization and access during cranial procedures; includes examination and operating sheaths. Device provides direct visualization via fiberoptic imaging; external light source required. Used in clinical settings by neurosurgeons; provides real-time visual feedback to guide surgical instruments during procedures like shunt placement or tumor biopsy. Benefits include minimally invasive access to intracranial structures. No software or automated analysis involved.
Clinical Evidence
Bench testing only. No clinical data required. Verification testing included surface/edge inspection, insertion width, field of view, direction of view, and resolution, all of which passed. Biocompatibility testing performed per ISO 10993-1 and sterilization validated via steam (SAL 10^-6).
Technological Characteristics
Fiberoptic neurological endoscope; 1 mm insertion shaft diameter (1.3 mm with sheath, 2.6 mm with operating sheath); 161.8 mm shaft length; 10,000 pixel fibers; 0° direction of view; external light source; manual cleaning; steam sterilization. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for visualization and access during cranial diagnostic and therapeutic procedures including shunt placement, tumor biopsy/resection, hydrocephalus treatment, ETV/CPC, ETV, cyst fenestration, and aqueduct exploration in pediatric and adult populations.
Regulatory Classification
Identification
A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.
Predicate Devices
Medtronic PS Medical's NeuroPEN Endoscope and Optical Accessories (K003914)
KARL STORZ Endoscopy-America's Neuro-Fiberscope (K002788)
Reference Devices
KARL STORZ Endoscopy-America's Sialendoscopes and Accessories (K012527)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, creating a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 16, 2017
Karl Storz Endoscopy-America, Inc. Leigh Spotten Regulatory Director 2151 E. Grand Avenue El Segundo, California 90245
Re: K170462
Trade/Device Name: KARL STORZ ShuntScope Regulation Number: 21 CFR 882.1480 Regulation Name: Neurological Endoscope Regulatory Class: Class II Product Code: GWG Dated: February 14, 2017 Received: February 15, 2017
Dear Leigh Spotten:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael 原田offmann -S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K170462
Device Name KARL STORZ ShuntScope
#### Indications for Use (Describe)
The KARL STORZ ShuntScope is indicated to provide visualization and access during cranial diagnostic and therapeutic procedures such as shunt placement, tumor biopsy and resection, hydrocephalus treatment, endoscopic third ventriculostomy with choroid plexus cauterization (ETV/CPC), endoscopic third ventriculostomy, cyst fenestration, and aqueduct exploration in pediatrics and adult population.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The word "STORZ" is written in large, bold, blue letters. Below that, the words "KARL STORZ - ENDOSKOPE" are written in smaller, blue letters. The logo is simple and professional.
## 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.
| Submitter: | KARL STORZ Endoscopy-America, Inc.<br>2151 E. Grand Avenue<br>EI Segundo, CA 90245 |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Nozomi Yagi<br>Regulatory Affairs Specialist<br>Phone: (424) 218-8351<br>Fax: (424) 218-8519 |
| Date of<br>Preparation: | May 15, 2017 |
| Type of 510(k)<br>Submission: | Traditional |
| Device<br>Identification: | Trade Name: KARL STORZ ShuntScope<br>Classification Name: Endoscope, Neurological (21 CFR Part 882.1480) |
| Regulatory<br>Class: | II |
| Product Code: | GWG |
| Guidance<br>Document: | Not Applicable for GWG product code |
| Recognized<br>Consensus<br>Standards: | Not Applicable for GWG product code |
| Predicate<br>Device(s): | Primary Predicate Device: Medtronic PS Medical's NeuroPEN Endoscope<br>and Optical Accessories (K003914)<br>Secondary Predicate Device: KARL STORZ Endoscopy-America's Neuro-<br>Fiberscope (K002788)<br>Reference Device: KARL STORZ Endoscopy-America's KSEA |
| | Sialendoscopes and Accessories (K012527) |
| Device<br>Description: | The components subject of this submission are: (1) the KARL STORZ<br>ShuntScope (Model Number: 28164SSA), (2) the examination sheath<br>(Model Number: 11577KA) and (3) operating sheath (Model Number:<br>11577KE) |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The word "STORZ" is in large, bold, blue letters. Below that, in smaller blue letters, is the text "KARL STORZ - ENDOSKOPE".
| Intended Use: | The KARL STORZ ShuntScope is intended for visualization purposes during cranial diagnostic and therapeutic procedures. | | |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Indications For Use: | The KARL STORZ ShuntScope is indicated to provide visualization and access during cranial diagnostic and therapeutic procedures such as shunt placement, tumor biopsy and resection, hydrocephalus treatment, endoscopic third ventriculostomy with choroid plexus cauterization (ETV/CPC), endoscopic third ventriculostomy, cyst fenestration, and aqueduct exploration in pediatrics and adult population. | | |
| Technological Characteristics: | Comparison Table: Subject vs. Primary and Secondary Predicate Devices | | |
| | Subject Device | Primary Predicate Device, K003914 | Secondary Predicate Device, K002788<br>Reference Device, K012527 |
| | KARL STORZ Shuntscope | Medtronic NeuroPEN | KSEA Neuro-Fiberscope<br>Marchal Mini Sialendoscope 0° |
| | Physical Characteristics | | |
| | Type of Scope | Same as the subject device | Flexible<br>Same as the subject device |
| | Insertion Shaft Diameter | 1 mm (wo sheath)<br>1.3 mm (w 11577KA)<br>N/A | 1.14 mm<br>N/A<br>Same as the subject device |
| | | 2.6 mm (w 11577KE) | 3.7 mm<br>Same as the subject device |
| | Insertion Shaft Length | 161.8 mm | 155 or 190 mm<br>340 mm<br>Same as the subject device |
| | Working Channel Diameter | 1.3 mm (w 11577KE) | N/A, No working channel<br>1.2 mm<br>Same as the subject device |
| | Optical Characteristics | | |
| | Type of Imager | Fiberoptic | Same as the subject device<br>Same as the subject device |
| | Direction of View | 0° | Information not publicly available<br>Same as the subject device |
| | Pixel Fibers | 10,000 | Same as the subject device<br>7,000<br>Same as the subject device |
| | Light Source | External | Same as the subject device<br>Same as the subject device |
| | Cleaning and Sterilization Methods | | |
| | Cleaning | Manual | Single-use<br>Same as the subject device |
| | Sterilization | Steam | Sterile<br>Same as the subject device |
| Non-Clinical Performance Data: | Bench Testing Summary | | |
| | Verification Test | Test Method Summary | |
| | Surface and Edges | The surface of the ShuntScope shall be free of pores, cracks, and remainders of tooling agents. | |
| | Maximum insertion portion width | The maximum insertion portion width shall not be larger than 1 mm, stated in the instruction manual. | |
| | Field of view | The deviation of field of view shall not be greater than 15%. | |
| | Direction of view | The deviation of direction of view shall not be greater than 0° ± 10°. | |
| | Resolution | Characterization only. | |
| | | Result | |
| | | Pass | |
| | | Pass | |
| | | Pass | |
| | | Pass | |
| | | Pass | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for STORZ. The word "STORZ" is written in large, bold, blue letters. Below that, the words "KARL STORZ - ENDOSKOPE" are written in smaller, blue letters.
| <b>Biocompatibility Summary</b><br>The biocompatibility evaluation for the patient contacting components of the neuroscope<br>was performed according to ISO 10993-1 and FDA Guidance. The following tests were<br>conducted based contact type and duration:<br>Cytotoxicity (ISO 10993-5) Sensitization (ISO 10993-11) Irritation (ISO 10993-11) Systemic Toxicity (ISO 10993-10) The biological-toxicological safety of the patient is not affected by the materials used in the<br>tested items. | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| <b>Sterilization Validation Summary</b> | | |
| Test | Test Method Summary | Result |
| Steam | SAL=10-6, Half cycle method | Pass |
| <b>Bibliography</b> | The following published literature was provided to support the pediatric use<br>of the subject device:<br>1. Kulkarni, Abhaya V., et al. "International Infant Hydrocephalus Study: initial results<br>of a prospective, multicenter comparison of endoscopic third ventriculostomy (ETV)<br>and shunt for infant hydrocephalus." Child's Nervous System 32.6 (2016): 1039-1048.<br>2. Depreitere, Bart, et al. "Endoscopic biopsy for intraventricular tumors in children."<br>Journal of Neurosurgery: Pediatrics 106.5 (2007): 340-346.<br>3. Stone, Scellig SD, and Benjamin C. Warf. "Combined endoscopic third<br>ventriculostomy and choroid plexus cauterization as primary treatment for infant<br>hydrocephalus: a prospective North American series: Clinical article." Journal of<br>Neurosurgery: Pediatrics 14.5 (2014): 439-446.<br>4. Zada, Gabriel, et al. "Pathogenesis and treatment of intracranial arachnoid cysts in<br>pediatric patients younger than 2 years of age." Neurosurgical focus 22.2 (2007): 1-5. | |
| <b>Clinical<br/>Performance<br/>Data:</b> | Clinical testing was not required to demonstrate the substantial equivalence<br>to the predicate devices. Non-clinical bench testing was sufficient to<br>establish the substantial equivalence of the modifications. | |
| <b>Conclusion:</b> | The conclusions drawn from the nonclinical tests demonstrate that the<br>subject device, KARL STORZ ShuntScope performs as well as or better than<br>the legally marketed predicate devices. | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.