NIM 3.0

K083124 · Medtronic Xomed, Inc. · GWF · Feb 27, 2009 · Neurology

Device Facts

Record IDK083124
Device NameNIM 3.0
ApplicantMedtronic Xomed, Inc.
Product CodeGWF · Neurology
Decision DateFeb 27, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1870
Device ClassClass 2

Indications for Use

The NIM 3.0 is intended for locating and identifying cranial and peripheral motor and mixed motor-sensory nerves during surgery, including spinal cord and spinal nerve roots. The APS electrode is an accessory intended for providing automatic periodic stimulation to nerves when used with the Medtronic Nerve Monitoring Systems.

Device Story

NIM 3.0 is a multi-channel intraoperative neurophysiological monitor; connects to patient electrodes; supplies electrical stimulus to nerves; monitors evoked responses from electrical or mechanical stimuli. Used in OR by surgeons/clinicians; provides audio and video output to assist in nerve identification/location during surgery. Helps prevent nerve damage by providing real-time feedback on nerve integrity. APS electrode accessory provides automatic periodic stimulation.

Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and intended use compared to predicate devices.

Technological Characteristics

Multi-channel intraoperative neurophysiological monitor; electrical stimulator; audio/video output; connects to patient monitoring electrodes; includes APS electrode accessory for automatic periodic stimulation.

Indications for Use

Indicated for locating/identifying cranial, peripheral motor, and mixed motor-sensory nerves during intracranial, extracranial, intratemporal, extratemporal, neck, thoracic, and extremity surgeries. Also indicated for spinal procedures including degenerative treatments, pedicle screw procedures, fusion cages, rhizotomy, orthopedic surgery, and open/percutaneous lumbar and cervical procedures.

Regulatory Classification

Identification

An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ FEB 2 7 2009 C083124/ # 510(k) SUMMARY SUMMARY OF SAFETY AND EFFECTIVENESS FOR ## Nerve Integrity Monitor 3.0 510(k) Owner Medtronic Xomed, Inc 6743 Southpoint Drive North Jacksonville, Florida 32216-0980 USA 904-296-9600 904-296-2386 (FAX) Contact Name Jayme Wilson Senior Regulatory Affairs Specialist Medtronic Xomed, Inc Date Summary Prepared October 21, 2008 Nerve Stimulator Proprietary Name Common Name Classification Name Evoked response electrical stimulator (21 CFR 882.1870. Product Code GWF. Class II) Surgical nerve stimulator/locator (21 CFR 874.1820. Product Code ETN. Class II) Nerve Integrity Monitor 3.0 (Final name TBD) #### Marketed device claiming equivalence to The Medtronic Nerve Integrity Monitor 3,0 and accessories are substantially equivalent to the Medtronic NIM Spine (K031510) and the NIM PRS (K024316). ### Device Description NIM 3.0 is a multi-channel intraoperative neurophysiological monitor capable of connecting various styles of patient monitoring electrodes and supplying electrical stimulus for evoked responses. The monitoring console uses both video and audio output. Responses monitored with the device may originate from operator applied electrical stimulus or from direct or indirect mechanical stimulus occurring during the course of the surgery. ## Intended Use The NIM 3.0 is intended for locating and identifying cranial and peripheral motor and mixed motor-sensory nerves during surgery, including spinal cord and spinal nerve roots. The APS electrode is an accessory intended for providing automatic periodic stimulation to nerves when used with the Medtronic Nerve Monitoring Systems. {1}------------------------------------------------ ## Indications for Use Indications for NIM 3.0 EMG Monitoring Procedures include: Intracranial, Extracranial, Intratemporal, Extratemporal, Neck Dissections, Thoracic Surgeries, and Upper and Lower Extremities Indications for Spinal procedures which may use NIM 3.0 EMG monitoring include: Degenerative Treatments, Pedicle Screw Procedures, Fusion Cages, Rhizotomy, Orthopedic Surgery, Open and Percutaneous Lumbar and Cervical Surgical Procedures, and Thoracic Surgical Procedures. #### Substantial Equivalence The indications, design, technology, functions, and principle of operation of the NIM 3.0 are substantially equivalent to the NIM Spine (K031510) and the NIM PRS (K024316). {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an emblem of an eagle. #### Public Health Service FEB 2 7 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Medtronic Xomed, Inc. % Ms. Jayme Wilson Senior Regulatory Affairs Specialist 6743 Southpoint Drive North Jacksonville, FL 32216-0980 Re: K083124 > Trade/Device Name: Xomed Nerve Integrity Monitor 3.0 Regulation Number: 21 CFR 882.1870 Regulation Name: Evoked Response Electrical Stimulator Regulatory Class: Class II Product Code: GWF. ETN Dated: February 17, 2009 Received: February 18, 2009 ### Dear Ms. Wilson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above).into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally {3}------------------------------------------------ Page 2 - Ms. Jayme Wilson marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, . Pik Dothan Ave D.e Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### INDICATIONS FOR USE 510(k) Number (if known): Device Name: Nerve Integrity Monitor 3.0 (Final name to be determined) Indications for Use: The NIM 3.0 is intended for locating and identifying cranial and peripheral motor and mixed motor-sensory nerves during surgery, including spinal cord and spinal nerve roots. The APS electrode is an accessory intended for providing automatic periodic stimulation to nerves when used with the Medtronic Nerve Monitoring Systems. Indications for NIM 3.0 EMG Monitoring Procedures include: Intracranial, Extracranial, Intratemporal, Extratemporal, Neck Dissections, Thoracic Surgeries, and Upper and Lower Extremities Indications for Spinal procedures which may use NIM 3.0 EMG monitoring include: Degenerative Treatments, Pedicle Screw Procedures, Fusion Cages, Rhizotomy, Orthopedic Surgery, Open and Percutaneous Lumbar and Cervical Surgical Procedures, and Thoracic Surgical Procedures. Prescription Use X AND/OR Over-The-Counter Use 4 - 1 (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Dlo. Thim Ruma (Division Sign-Off) (Division Sign-On) Division of General, Restorative, Division of General Devices 160 83124 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...