ECLIPSE NEUROLOGICAL WORKSTATION

K050798 · Axon Systems, Inc. · GWF · Aug 15, 2005 · Neurology

Device Facts

Record IDK050798
Device NameECLIPSE NEUROLOGICAL WORKSTATION
ApplicantAxon Systems, Inc.
Product CodeGWF · Neurology
Decision DateAug 15, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1870
Device ClassClass 2

Indications for Use

The Eclipse neurological workstation is intended for use to monitor sensory and motor pathways in the operating room and critical care areas. The instrument uses electroencephalography (EEG), electromyography (EMG), motor and sensory evoked potentials and nerve potentials to provide health care professionals with information to help assess a patient's neurological status. Transcranial stimulation techniques for motor evoked potentials are used to assess for acute dysfunction in axonal conduction of the corticospinal tract.

Device Story

Eclipse Neurological Workstation provides continuous intraoperative or ICU neurological monitoring. Inputs: electrophysiological signals from recording electrodes (EEG, EMG, evoked potentials); external vital signs/physiological monitors; video input (e.g., surgical microscope). System components: computer, controller, digital preamplifiers, electrical/audio/visual stimulators, LED goggles, insert earphones, pulse oximeter module. Operation: electrodes detect signals; preamplifiers amplify/filter/digitize; DSP processes data and controls stimulation timing. Output: processed neurological data displayed on screen for clinician review. Clinicians (physicians/technicians) use output to correlate neurological function with systemic measurements, aiding intraoperative decision-making and patient status assessment. Network-compatible for hospital-wide review and secure remote access.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

System includes digital preamplifiers, electrical/audio/visual stimulators, and pulse oximeter module. Features optical isolation for signal integrity. Supports high-impedance electrode recording. Connectivity includes network compatibility for data review and secure internet access. Software-based signal processing and user interface control.

Indications for Use

Indicated for patients requiring intraoperative or critical care neurological monitoring of sensory and motor pathways, including assessment of corticospinal tract axonal conduction via transcranial motor evoked potentials.

Regulatory Classification

Identification

An evoked response electrical stimulator is a device used to apply an electrical stimulus to a patient by means of skin electrodes for the purpose of measuring the evoked response.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K050798 Image /page/0/Picture/1 description: The image shows the text "Page 1 of 2" in cursive handwriting. The text indicates that it is the first page of a two-page document. The handwriting is clear and legible, with the words spaced evenly. ## 15 510(k) Summary AUG 1 5 2005 This 510(k) summary of safety and effectiveness is being submitted in t and effective and effect of OED 807 00 This 910(K) Summary of Sammary of CFR 807.92. | Name: | Howard Bailin<br>Vice President, C.O.O. | |----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Address: | Axon Systems, Inc.<br>400-2200 Oser Ave<br>Hauppauge, NY 11788 | | Phone: | 631 436 5112 | | Fax: | 631 436 5141 | | Email: | hbailin@axonsystems.com | | Proprietary Name: | Eclipse Neurological Workstation<br>Eclipse Lite Neurological Workstation | | Common Name: | Electroencephalograph (EEG Monitor), Evoked<br>Potential (SEP, BAEP, AEP, VEP, MEP) System, EMG<br>Monitor | | Classification Name: | Electroencephalograph, Evoked Response, Electromyograph | | Classification: | Class II (Performance Standards)<br>Panels: Neurology, Anesthesiology<br>Number: 882.1400 Electroencephalograph<br>882.1420 Electroencephalograph (EEG)<br>Signal Spectrum Analyzer<br>Electromyograph Monitor<br>Stimulator, Electrical, Evoked Response<br>Stimulator, Photic, Evoked Response<br>Stimulator, Sonic, Evoked Response<br>Procodes: GWQ, GWS, GWF, GWE, GWJ, CAB | | Predicate Devices | Axon Systems - EpochXP (K032741)<br>Digitimer - D185 (K020400)<br>NuVasive - Neurovision JJB (K032083)<br>Grass Telefactor - AS40 Amplifier (K021807) | | Description: | The Eclipse Neurological Workstation provides continuous<br>monitoring of brain and neural pathways intraoperatively or in<br>the intensive care unit. The system has been designed to<br>meet the requirements for comprehensive neurological<br>monitoring in the operating room and critical care areas.<br><br>The Eclipse Neurological Workstation can be used to monitor<br>neurological data using either individual or multimodality EEG,<br>EMG and evoked potential test protocols. | | | The main Eclipse system components include: computer,<br>controller, digital preamplifiers, direct nerve, sensory and<br>motor evoked potential electrical stimulators, stimulator<br>extension modules, LED goggles and insert earphones. The<br>Eclipse also provides support for the Nonin XPod pulse<br>oximeter module and high impedance preamplifier module to<br>allow recording from high impedance electrodes. | | | Recording electrodes detect spontaneous or stimulus evoked<br>electrophysiological activity and are used as inputs to the<br>digital preamplifier. The electrophysiological signals are<br>amplified, filtered, optically isolated and digitized. The<br>digitized data is then routed to the digital signal processor<br>(DSP) located in the Eclipse controller. The DSP processes<br>the data and controls timing for the electrical, audio and visual<br>stimulators. The computer controls the user interface for<br>setting parameters and the display of processed data. | | | A built-in pulse oximeter provides pulse rate and oxygen<br>saturation measures. Data from external devices, such as<br>vital signs or other physiological monitors, can be imported to<br>the Eclipse display screen, allowing the operator to correlate<br>changes in neurological function with the patient's systemic<br>measurements. In addition, a display window may be opened<br>to observe the surgeon's microscope view or other video<br>input. The Eclipse is network compatible for data review<br>within the hospital and permits secure information access over<br>the Internet. | | | Based on the technical information provided in this 510(k) and<br>the safety and effectiveness criteria of the design and<br>development process, validated and verified, we claim the<br>Eclipse Neurological Workstation to be safe, effective and<br>substantially equivalent to the predicate device(s) noted. | {1}------------------------------------------------ K050798 Image /page/1/Picture/1 description: The image shows the handwritten text "page 2 of 2". The text is written in cursive and is slightly slanted. A horizontal line is drawn below the text. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with stylized feathers. ## Public Health Service AUG 1 5 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Howard Bailin Vice President, C.O.O. Axon Systems, Inc. 400-2200 Oser Avenue Hauppauge, New York 11788 Re: K050798 Trade/Device Name: Eclipse Neurological Workstation Regulation Number: 21 CFR 882.1870 Regulation Name: Evoked response electrical stimulator Regulatory Class: II Product Code: GWF Dated: August 1, 2005 Received: August 2, 2005 Dear Mr. Bailin: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Section 9 rotty premier is substantially equivalent (for the indications felerenced above and nave develomes and and marketed predicate devices marketed in interstate for use stated in the encrosale) to regars and the Medical Device Amendments, or to commerce prior to May 20, 1978, the enaordance with the provisions of the Federal Food, Drug, devices mat have been recuire approval of a premarket approval application (PMA). and Cosment Act (Act) that do not requent of the general controls provisions of the Act. The 1 ou may, mercrore, market the act include requirements for annual registration, listing of general controls provisions of ractice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is olassified (600 wor 70) als. Existing major regulations affecting your device can Inay of subject to sublications, Title 21, Parts 800 to 898. In addition, FDA may oe found in the Our cements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean i that FDA has made a determination that your device complies with other requirements of the Act that I Dri has made a availations administered by other Federal agencies. You must or any i caeral statutes and registents, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI It Fat 607); accems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. . {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin marketing your artial equivalence of your devertire your dever devi premarket nothication. The PDA miding of backand for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (2 CFR Part 801), please If you desire specific advice for your accessed on one and the regulation entitled, and contact the Office of Complaned at (210) = i & cation" (21CFR Part 807.97). You may obtain " Misoranung by Icierchee to premainters with the Act from the Division of Small other general information on your responsibilities under the Act from the 1990s C29 AA other general information on your responses Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Octable so http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely vours, Barbara Buehmd for Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Koso798 ## Indications for Use Statement | 510(k) Number | K050798 | |---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | Eclipse Neurological Workstation | | Indications for Use | The Eclipse neurological workstation is intended for use to monitor sensory and motor pathways in the operating room and critical care areas. The instrument uses electroencephalography (EEG), electromyography (EMG), motor and sensory evoked potentials and nerve potentials to provide health care professionals with information to help assess a patient's neurological status. Transcranial stimulation techniques for motor evoked potentials are used to assess for acute dysfunction in axonal conduction of the corticospinal tract. | Prescription Use _____________________________________________________________________________________________________________________________________________________________ Use (Part 21 CFR 801 Subpart D) . 1. 1. 1. 1. 1. : AND/OR Over-The-Counter Barbara for (Division Sign-Off) Melkers Division of General, Restoratio and Neurological Devices 510(k) Number: K050798
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...