SENTIO MMG Pedicle Access Needles

K173134 · Medos International SARL · PDQ · Mar 22, 2018 · Ear, Nose, Throat

Device Facts

Record IDK173134
Device NameSENTIO MMG Pedicle Access Needles
ApplicantMedos International SARL
Product CodePDQ · Ear, Nose, Throat
Decision DateMar 22, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.1820
Device ClassClass 2

Intended Use

The SENTIO MMG Pedicle Access Needles are indicated for use in pedicle pilot hole preparation and stimulation of peripheral motor nerves for location and identification during surgery, including spinal nerve roots.

Device Story

SENTIO MMG Pedicle Access Needles are surgical instruments used during spine surgery for pedicle pilot hole preparation and nerve detection. The device consists of a cannula, trocar, and 2.5m stimulation cable. It functions by delivering electrical stimulation to peripheral motor nerves; the distal tip is non-insulated stainless steel to focus energy delivery. The device is used in the OR by surgeons to locate and identify spinal nerve roots and peripheral motor nerves, helping to avoid nerve injury during pedicle preparation. The cannula features 1cm depth markers for reference. The device is compatible with the Sentio MMG Nerve Mapping and Avoidance System. By providing real-time feedback on nerve proximity via stimulation, the device assists the surgeon in navigating bone safely, potentially reducing the risk of iatrogenic nerve damage.

Clinical Evidence

Bench testing only. Evaluations included durability testing in simulated bone, wire-to-plug tensile force resistance, electrical stimulation compatibility (100V-300V), biocompatibility per ISO 10993-1, and ethylene oxide sterilization validation per ISO 11135-1. All samples met acceptance criteria.

Technological Characteristics

Materials: Biocompatible stainless steel with Parylene C conformal coating. Design: Cannula with 1cm depth markers and trocar; 2.5m stimulation cable. Energy: Electrical stimulation (compatible with 100V-300V sources). Connectivity: Standard connectors for integration with nerve mapping systems. Sterilization: Ethylene oxide (SAL 10⁻⁶).

Indications for Use

Indicated for pedicle pilot hole preparation and stimulation of peripheral motor nerves for location and identification during surgery, including spinal nerve roots.

Regulatory Classification

Identification

A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized eagle. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. March 22, 2018 Medos International SARL % Sheree Geller Regulatory Affairs Specialist DePuy Synthes Spine 325 Paramount Drive Raynham, Massachusetts 02767 Re: K173134 Trade/Device Name: SENTIO MMG Pedicle Access Needles Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: PDO Dated: February 20, 2018 Received: February 21, 2018 Dear Sheree Geller: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Michael J. Hoffmann -S Carlos L. Peña. PhD. MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K173134 Device Name SENTIO MMG Pedicle Access Needles Indications for Use (Describe) The SENTIO MMG Pedicle Access Needles are indicated for use in pedicle pilot hole preparation and stimulation of peripheral motor nerves for location and identification during surgery, including spinal nerve roots. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) SUMMARY | A. | Submitter Information | | |----|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------| | | Manufacturer: | Medos International SARL<br>Chemin-Blanc 38<br>2400 Le Locle, Switzerland | | | Submitter: | DePuy Synthes Spine<br>325 Paramount Drive<br>Raynham, MA 02767 | | | Contact Person:<br>Telephone:<br>Fax:<br>Email: | Sheree Geller<br>325 Paramount Drive<br>Raynham, MA 02767<br>(508) 828-3291<br>(508) 828-3797<br>sgeller1@its.jnj.com | | B. | Date Prepared | March 22, 2018 | | C. | Device Name | | | | Trade/Proprietary Name: | SENTIO MMG® Pedicle Access Needles | | | Common/Usual Name: | Nerve Stimulator/Locator | | | Classification Name: | 21 CFR §874.1820<br>Surgical nerve stimulator/locator | | | Regulatory Class: | II | | | Product Code & Review Panel: | PDQ; Neurology | | D. | Predicate Device Names | | | | Primary Predicate: | NeuralMAS™ (K100992) | | | Additional Predicate: | NuVasive® Stimulation/Dissection<br>Instruments (K112709) | {4}------------------------------------------------ #### E. Device Description The SENTIO MMG Pedicle Access Needles are surgical instruments that incorporate insulated probes with standard connectors to ensure maximum ease of use and compatibility with the Sentio MMG Nerve Mapping and Avoidance System for use during surgeries in which a motor nerve is at risk. The SENTIO MMG Pedicle Access Needles are utilized for pedicle preparation in spine surgery with application of dynamic stimulation for nerve detection. They are provided sterile (ethylene oxide) and have two components: Needle (cannula and trocar) and 2.5m Stimulation Cable. The cannula is marked with equidistant reference depth markers (1cm) relative to the distal [to user] end. #### F. Intended Use The SENTIO MMG Pedicle Access Needles are indicated for use in pedicle pilot hole preparation and stimulation of peripheral motor nerves for location and identification during surgery, including spinal nerve roots. {5}------------------------------------------------ ### Summary of Similarities and Differences in Technological Characteristics, Performance, and Intended Use G. The intended use and technological characteristics, including material, design, and performance, of the SENTIO MMG Pedicle Access Needles are consistent with those of the predicate devices. | | Predicate Devices | | Subject Device | | |-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------| | | NeuralMASTM (K100992) | NuVasive® Stimulation/Dissection<br>Instruments (K112709) | SENTIO MMG® Pedicle Access<br>Needles | Substantially<br>Equivalent | | Intended Use | NeuralMAS is intended for use in<br>surgical procedures to assist in locating<br>and mapping motor nerves through the<br>use of mechanomyographic (MMG)<br>signals and electrical stimulus of nerves.<br>This device is indicated for locating and<br>identifying spinal nerve roots and<br>peripheral motor nerves origination<br>from spinal levels C3-T1 and L2-S2. | The Stimulation/Dissection Instruments<br>are indicated for tissue dissection and<br>stimulation of peripheral motor nerves<br>for location and identification during<br>surgery, including spinal nerve roots. | The SENTIO MMG® Pedicle Access<br>Needles are indicated for use in pedicle<br>pilot hole preparation and stimulation of<br>peripheral motor nerves for location and<br>identification during surgery, including<br>spinal nerve roots. | Yes | | Product Code | ETN | ETN | PDQ | Yes | | Regulation | 21 CFR §874.1820 | 21 CFR §874.1820 | 21 CFR §874.1820 | Yes | | Electrical<br>Insulation | The sterile stimulation instruments have<br>biocompatible electrical insulation<br>applied to select portions. The distal<br>surfaces of the instruments are<br>selectively non-insulated stainless steel<br>to provide for mechanical, manual<br>dissection, probing, cannulation, and<br>tissue stimulation. | Some instruments have electrical<br>insulation on surfaces not intended to<br>provide electrical contact with the<br>patient and connection while others are<br>used with an insulation accessory that<br>provides electrical insulation on surfaces<br>not intended to provide electrical<br>contact with the patient and connection. | The subject devices have biocompatible<br>electrical insulation applied to select<br>portions. The distal surfaces of the<br>instruments are selectively non-<br>insulated stainless steel to provide for<br>mechanical, manual dissection, probing,<br>cannulation, and tissue stimulation. | Yes | | Proximal<br>Stimulator<br>Connector | Yes | Yes | Yes | Yes | | Patient<br>Contacting<br>Material | Biocompatible:<br>Stainless Steel<br>Parylene C | Biocompatible:<br>Stainless Steel,<br>Anodized Aluminum<br>Radel | Biocompatible:<br>Stainless Steel<br>Parylene C | Yes | | Delivery | Sterile for single use | Some instruments are provided sterile<br>for single use only while others are<br>provided as non-sterile reusable<br>instruments. | Sterile for single use | Yes | {6}------------------------------------------------ #### H. Materials The patient contacting materials of the subject devices are biocompatible stainless steel with Parylene C conformal coating. #### I. Performance Data Non-clinical testing was conducted to demonstrate the safety and effectiveness of the subject devices. The SENTIO MMG Pedicle Access Needles were evaluated as follows: | Characteristic | Evaluation Summary | Conclusion | |----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Durability | Durability testing served to determine that materials and<br>mating interfaces of the subject device survive the force<br>of entering the pedicle, the interface between the top and<br>bottom handles remains intact, the stylet is sufficiently<br>sharp to pass through bone, and the needle cannula<br>beveled edge is sufficient to dig into bone. Testing was<br>completed using simulated bone material. | All samples passed the acceptance<br>criteria. The durability of the subject<br>device has been determined<br>substantially equivalent to the<br>predicate. | | Wire to Plug<br>Connection | The wire to plug connection of the subject device was<br>tested for tensile force resistance. The plug from the<br>connection wire component part was inserted into a test<br>sample intended to accommodate a resistance fit. The<br>sample construct was subject to tensile force until the<br>wire disconnected from the sample handle. | All samples passed the acceptance<br>criteria. The subject device<br>connection assembly has been<br>determined substantially equivalent to<br>the predicate. | | System<br>Compatibility | Needle stimulation testing was conducted to verify that<br>the subject device maintains compatibility with the<br>NeuralMAS™ (K100992) predicate, and provides<br>electrical isolation such that energy delivery is focused<br>from only the distal tip of the needle. | Test results demonstrated subject<br>device compatibility with the<br>predicate NeuralMAS™ (K100992)<br>system. | | Electrical<br>Stimulation | The subject device was tested for compatibility with<br>electrical stimulation sources of 100V, 200V, and 300V. | Compatibility with electrical<br>stimulation sources was verified,<br>supporting substantial equivalence of<br>the subject device. | | Biocompatibility | Per ISO 10993-1, the subject device is categorized as an<br>external communicating device for tissue/bone contact of<br>limited duration. The Parylene C conformal coating<br>remains unchanged from the NeuralMAS™ (K100992)<br>predicate. Stainless Steel has a history of clinical use and<br>biocompatibility; cleaning validations were conducted to<br>confirm removal of processing reagents. | The biocompatibility of the subject<br>device has been assessed as<br>substantially equivalent to the<br>predicate. | | Sterility | The ethylene oxide sterilization method was validated per<br>ISO 11135-1 for the subject device. Sterilization residual<br>levels were tested for compliance to ISO 10993-7. | The subject devices will be provided<br>sterile with a SAL of 10⁻⁶. The<br>sterilization residual levels were<br>determined compliant with ISO<br>10993-7 for limited exposure devices.<br>Delivery state of the subject device is<br>determined to be substantially<br>equivalent to the predicate. | {7}------------------------------------------------ #### J. Conclusion Evaluation of the subject device intended use, technological characteristics, and performance data demonstrates substantial equivalence with the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...