Phantom XL Insulated Dilators are indicated for use during surgery of the spine to deliver an electrical stimulus to the tissue and nerves at the operative site, to assist in locating those nerves at risk during the surgical procedure.
Device Story
Phantom XL Insulated Dilators are surgical instruments used to deliver electrical stimulation to tissues and nerves during spinal surgery. The device functions as a needle electrode, allowing surgeons to identify nerves at risk of injury during the procedure. It is available in monopolar configuration with multiple diameter sizes (8mm, 13mm, 18mm, 22mm, and 22mm grooved). The 8mm version is stainless steel (fluoroscopy-visible), while larger sizes are aluminum alloy (radiolucent). The device is supplied sterile, non-pyrogenic, and for single use. It is operated by surgeons in an operating room environment. By delivering electrical stimuli, the device helps the surgeon map nerve locations, potentially reducing the risk of iatrogenic nerve damage during spinal access and instrumentation.
Clinical Evidence
Bench testing only. Performance testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993). Electrical testing confirmed low impedance, dielectric strength, temperature safety, and waveform integrity. Biocompatibility testing (cytotoxicity, irritation, sensitization, systemic toxicity) confirmed the device is non-cytotoxic, non-irritant, and non-toxic. Sterilization validation (ANSI/AAMI/ISO 11135-1) achieved a Sterility Assurance Level of 1x10^-6.
Technological Characteristics
Needle electrode for electrical stimulation. Materials: Stainless steel (8mm) and aluminum alloy (13-22mm). Dimensions: 8, 13, 18, 22, and 22mm grooved diameters. Sterilization: Ethylene Oxide (ANSI/AAMI/ISO 11135-1). Electrical safety: IEC 60601-1. EMC: IEC 60601-1-2. Biocompatibility: ISO 10993. Single-use, sterile, non-pyrogenic.
Indications for Use
Indicated for patients undergoing spinal surgery to assist in locating nerves at risk during the procedure.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a caduceus, possibly representing health and well-being. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 25, 2014
TeDan Surgical Innovations, LLC Ms. Dionicia Reblando 12615 W. Airport Blvd., Suite 200 Sugar Land, TX 77478
Re: K140088
Trade/Device Name: Phantom XL Insulated Dilators Regulation Number: 21 CFR 874.1820 Regulation Name: Neurosurgical Nerve Locator Regulatory Class: Class II Product Code: PDQ Dated: August 25, 2014 Received: August 27, 2014
Dear Ms. Reblando:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Carlos L. Pena -SFD/△
Carlos L. Peña, Ph.D. M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K140088
Device Name Phantom XL Insulated Dilators
Indications for Use (Describe)
Phantom XL Insulated Dilators are indicated for use during surgery of the spine to deliver an electrical stimulus to the tissues and nerves at the operative site, to assist in locating those nerves at risk during the surgical procedure.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <span style="font-family: Arial;"> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> </span> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <span style="font-family: Arial;"> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| Assigned 510(k) number: | K140088 |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company: | TeDan Surgical Innovations, LLC<br>12615 W Airport Blvd, Suite 200<br>Sugar Land, TX USA 77478<br>Phone: 650-631-4515<br>Fax: 650-631-4555 |
| Contact: | Dionicia Reblando<br>Phone: 858-472-5160<br>Email: jreblando@yahoo.com |
| Date Prepared: | July 23, 2014 |
| Proprietary Names: | Phantom XL Insulated Dilators |
| Classification Name: | Needle electrode |
| Classification: | 21 CFR 882.1350, Class II, Product Code GXZ |
| Predicate Device: | K110419 AVS® ARIA Probe Dilators |
| Device Description: | Phantom XL Insulated Dilators are used as instruments to deliver<br>electrical stimulation to tissue during intraoperative neurological<br>monitoring. The Phantom XL Insulated Dilators are available in<br>monopolar configuration and four diameter sizes (8, 13, 18, 22 and<br>22mm Grooved). They are supplied sterile, are non-pyrogenic, and<br>are intended for single use only. The 8 mm dilators are made of<br>stainless steel and are visible under fluoroscopy, while the<br>remaining sizes are made of aluminum alloy and are radiolucent. |
| Intended Use: | Phantom XL Insulated Dilators are indicated for use during<br>surgery of the spine to deliver an electrical stimulus to the tissue<br>and nerves at the operative site, to assist in locating those nerves at<br>risk during the surgical procedure. |
| Technological<br>Comparison to Predicate<br>Device: | Phantom XL Insulated Dilators are similar to the predicate device.<br>Both are needle electrodes used to deliver an electrical stimulus to<br>tissues and nerves at the operative site. Additionally, both use the<br>same fundamental scientific technology and principle of operation.<br>Performance test results confirm that design, material, and |
{5}------------------------------------------------
sterilization differences do not pose new issues of safety or effectiveness. Both devices attain a minimum Sterility Assurance Level of 1 X 10°
#### ELECTRICAL SAFETY TESTING Performance Testing:
Electrical safety testing was performed for the Phantom XL Insulated Dilators in accordance with the standard: IEC 60601-1 (2012) Ed 3.1; Medical electrical equipment – Part 1: General requirements for basic safety and essential performance. The following applicable tests were performed:
- 1. Impedance of current carrying connections. This test was performed to demonstrate that the Phantom XL Insulated Dilators have sufficiently low impedance so that they will not interfere with any stimulator signal passed through them.
- 2. Dielectric Strength test. This test was performed to demonstrate that the insulated portions of the Phantom XL Insulated Dilators are sufficiently separated from the conductive portions to prevent any stimulator signal from reaching any part of the Phantom XL Insulated Dilators except the intended delivery point.
- 3. Temperature test. This test was performed to demonstrate that the Phantom XL Insulated Dilators will not introduce any additional heating to the patient as part of the stimulation process.
- 4. Waveform capture. Waveform of the stimulator output was capture on an oscilloscope in 3 modes of operation. The tests were performed to demonstrate that the Phantom XL Insulated Dilators will not significantly alter the waveform.
The Phantom XL Insulated Dilators met applicable requirements set forth in the referenced standard. Electrical Safety Report 261304 for IEC 60601-1 compliance, is provided in Attachment 4.
## ELECTROMAGNETIC COMPATIBILITY
The Phantom XL Insulated Dilators underwent electromagnetic compatibility (EMC) testing in accordance with IEC 60601-1-2:2007, Medical electrical equipment - Part 1-2:2007 General requirements for basic safety and essential performance -Collateral standard: Electromagnetic compatibility - Requirements and tests. The following applicable tests were performed:
- 1. Emissions tests
- 2. Immunity tests
The above tests were conducted to evaluate and verify that the Phantom XL Dilators/accessories do not impact the emission
{6}------------------------------------------------
profile or increase the sensitivity to adverse conditions of electrical stimulus and medical monitoring equipment.
The Phantom XL Insulated Dilators met applicable requirements set forth in the referenced standard. EMC Test Report Number 2014 261304 EMC for IEC 60601-1-2:2007 compliance, is provided in Attachment 5.
## BIOCOMPATIBILITY
The biocompatibility tests for the Phantom XL Dilators were selected in accordance with blue book memorandum G95-1 entitled Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing". Biocompatibility tests were conducted in compliance with U.S. Food and Drug Administration Good Laboratory Practice (GLP) regulations set forth in 21 CFR 58. The Table below summarizes the results of the biocompatibility testing.
| Test | Results | Conclusions |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity-ISO<br>Elution Method<br>(ISO 10993-5:2009) | No evidence of causing<br>lysis or toxicity. Reactivity<br>is less than grade 2 | Non-Cytotoxic.<br>(acceptance criteria:<br>biological reactivity<br>must be less than or<br>equal a grade 2 |
| ISO Intracutaneous<br>Study- Irritation<br>(ISO 10993-10:2010) | Observations for<br>erythema and edema. The<br>difference between each<br>test article extract overall<br>mean score and<br>corresponding control<br>extract overall mean score<br>was 0.0 and 0.2 | Non-irritant.<br>(acceptance criteria: the<br>difference between the<br>test extract overall<br>mean score and<br>corresponding control<br>overall mean score is<br>1.0 or less). |
| ISO Guinea Pig<br>Maximization<br>Sensitization Test<br>(ISO 10993-10:2010) | The test article extracts<br>showed no evidence of<br>causing delayed dermal<br>contact sensitization | Non-Sensitizer |
| ISO Systemic<br>Toxicity (acute)<br>(ISO 10993-11:2010) | There was no mortality or<br>evidence of systemic<br>toxicity from the extracts | Non-Toxic (acute-<br>systemic) |
{7}------------------------------------------------
## STERILIZATION, BACTERIAL ENDOTOXIN AND ETHYLENE OXIDE RESIDUALS
A sterilization validation of the Phantom XL insulated dilators based on the requirements of ANSI/AAMI/ISO 11135-1:2007 was completed to assure a minimum Sterility Assurance Level of 1X106. The Table below summarizes the results of the Bacterial Endotoxin and the residual levels
| Test | Results | Conclusions |
|----------------------------|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Bacterial<br>Endotoxin | 0.720 EU/Device | a. The results passes FDA criteria for<br>both: medical device - maximum of<br>20 EU/device |
| | 0.880 EU/Device | b. device contacting cerebrospinal<br>fluid-maximum of 2.15 EU/device<br>Note: This device does not contact<br>cerebrospinal fluid. |
| | < 0.200 EU/Device' | |
| Ethylene Oxide<br>residual | Range: <0.12 – 0.15<br>mg/device | Pass the requirements of 4 mg/device |
| | Ethylene<br>Chlorohydrin<br>residual | Range: <0.10 -<br><0.15 mg/device |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.