The Next Generation Access Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
Device Story
Single-lumen, flexible, variable stiffness composite catheter; distal tip includes marker band; proximal end features luer fitting for accessory attachment and liquid infusion. Distal shaft has external hydrophilic coating for lubricity; visible under fluoroscopy. Provided with peelable split introducer sheath to facilitate vascular entry. Used in clinical settings by physicians for navigating peripheral, coronary, and neuro vasculature to deliver interventional devices. Device provides access path; enables physician to perform procedures; benefits patient by facilitating minimally invasive interventions.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993 series), mechanical performance (tensile strength, kink resistance, leakage, burst, torque), and usability in a clinically relevant anatomical model.
Technological Characteristics
Single-lumen composite catheter; medical grade plastics and stainless steel; hydrophilic coating. Dimensions: 6F-7F OD, 1.80-2.06 mm ID, 110-132 cm length. Sterilization: Ethylene Oxide (ISO 11135). Standards: ISO 10993 (biocompatibility), ISO 10555-1 (catheter performance), ISO 80369-7 (hub), ASTM F1980 (shelf-life).
Indications for Use
Indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature in patients requiring such access.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
July 2, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
Balt USA, LLC Catherine Chiou Senior Regulatory Affairs Specialist 29 Parker Irvine, California 92618
Re: K234074
Trade/Device Name: Next Generation Access Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP, DQY Dated: May 31, 2024 Received: June 3, 2024
Dear Catherine Chiou:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K234074
Device Name
Next Generation Access Catheter
Indications for Use (Describe)
The Next Generation Access Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
Type of Use (Select one or both, as applicable)
> | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary: K234074
| Applicant: | Balt USA, LLC<br>29 Parker<br>Irvine, CA 92618<br>Registration No.: 3014162263 |
|-----------------|-------------------------------------------------------------------------------------------|
| Contact Person: | Catherine Chiou<br>Specialist, Regulatory Affairs<br>Email: Catherine.chiou@baltgroup.com |
| Date Summary<br>Prepared: | June 28, 2024 |
|---------------------------|----------------------------------------------------|
| Trade Name: | Next Generation Access Catheter |
| Common Name: | Catheter, Percutaneous |
| Review Panel: | Neurology, Cardiovascular |
| Product Code: | QJP, DQY |
| Regulation Number: | 21 CFR 870.1250 |
| Device Classification: | Class II |
| Predicate Device: | BENCHMARK BMX81 Access System<br>510(k) #: K221822 |
#### Device Description:
The Next Generation Access Catheter is a single-lumen, flexible, variable stiffness composite catheter. The distal tip of the catheter shaft includes a marker band while the proximal end of the catheter has a luer fitting to allow attachment of accessories and infusion of liguids through the system. The Next Generation Access Catheter distal shaft has an external hydrophilic coating which provides a lubricious surface during use. The Next Generation Access Catheter is visible under fluoroscopy. The Next Generation Access Catheter dimensions are included on the individual device label. The Next Generation Access Catheter is offered in various sizes to accommodate physician preferences and anatomical variations.
A peelable split introducer sheath is provided in the package to provide support and facilitate the introduction of the distal tip of the Next Generation Access Catheter into an appropriate vascular sheath.
The catheter and introducer sheath are provided sterile, non-pyrogenic, and are intended for single use only.
#### Indications for Use:
The Next Generation Access Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
{4}------------------------------------------------
Comparison of Technological Characteristics:
| | Predicate Device | Subject Device |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | BENCHMARK BMX81<br>Access System<br>(K221822) | Next Generation Access Catheter<br>(K234074) |
| Indications for Use | The BENCHMARK BMX81 Access<br>System is indicated for the introduction of<br>interventional devices into the peripheral,<br>coronary, and neuro vasculature. | The Next Generation Access Catheter is<br>indicated for the introduction of<br>interventional devices into the<br>peripheral, coronary, and neuro<br>vasculature. |
| Device Classification /<br>Product Code | Class II / QJP, DQY<br>(Percutaneous Catheter) | Same as K221822 |
| Dimensional Specifications | | |
| Catheter Outer<br>Diameter (OD) | 7F (2.46 mm, 0.097") | 6.F (2.11 mm, 0.083")<br>7F (2.36 mm, 0.093") |
| Catheter Inner<br>Diameter (ID) | 0.081" (2.06 mm) | 0.071" - 0.081" (1.80 mm - 2.06 mm) |
| Effective Length | 95 cm, 105 cm, 115 cm | 110 cm - 132 cm |
| Coating Length | 18 cm | 60 cm |
| Device Attributes | | |
| Catheter Materials | Commonly used medical grade plastics &<br>stainless steel | Same as K221822 |
| Coating | Hydrophilic | Same as K221822 |
| Packaging Materials | Commonly used medical device<br>packaging materials | Same as K221822 |
| Sterilization | | |
| How Supplied | Sterile, Single Use | Same as K221822 |
| Method | Ethylene Oxide | Same as K221822 |
{5}------------------------------------------------
### Performance Testing Summary:
### Biocompatibility:
The following biocompatibility testing was conducted for the Next Generation Access Catheter:
| Test | Test Method Summary | Results |
|-----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Cytotoxicity - MEM Elution | Tested in accordance with ISO<br>10993-5 | Pass<br>The test article is considered<br>non-cytotoxic. |
| Hemocompatibility -<br>Hemolysis (Direct Contact &<br>Extract Method) | Tested in accordance with ISO<br>10993-4 | Pass<br>The test article is considered<br>non-hemolytic. |
| Hemocompatibility -<br>Complement Activation<br>SC5b-9 Assay | Tested in accordance with ISO<br>10993-4 | Pass<br>Requirements have been met by<br>the test article. |
| Hemocompatibility - Partial<br>Thromboplastin Time (PTT) | Tested in accordance with ASTM<br>F2382-18 | Pass<br>Requirements have been met by<br>the test article. |
| Hemocompatibility – Blood<br>Platelet and Leukocyte Count<br>(PLC) | Tested in accordance with ASTM<br>F2888-19 | Pass<br>Requirements have been met by<br>the test article. |
| Hemocompatibility -<br>Comparative Surface<br>Assessment | The test article was visually<br>inspected at more than 40x<br>magnification. | Pass<br>Requirements have been met by<br>the test article. |
| Hemocompatibility -<br>Thrombogenicity in a Canine<br>Model | Tested in accordance with ISO<br>10993-4 | Pass<br>The test article performed<br>similarly to the comparator. |
| Pyrogenicity - Material-<br>Mediated Rabbit Pyrogen | Tested in accordance with <usp<br>151></usp<br> | Pass<br>Requirements have been met by<br>the test article. |
| Sensitization - Guinea Pig<br>Maximization Sensitization | Tested in accordance with ISO<br>10993-10 | Pass<br>The test article did not elicit a<br>sensitization response. |
| Systemic Toxicity - Acute<br>Systemic Injection | Tested in accordance with ISO<br>10993-11 | Pass<br>Requirements have been met by<br>the test article. |
| Irritation - Intracutaneous<br>Reactivity | Tested in accordance with ISO<br>10993-23 | Pass<br>Requirements have been met by<br>the test article. |
| Test | Test Method Summary | Results |
| Dimensional<br>Verification | The catheter outer diameter, inner diameter, usable<br>length, tip length, and coating length were measured. | Pass |
| Surface<br>Contamination | Visual inspection completed for surface defects. | Pass |
| Tensile Strength | The peak tensile force was evaluated per ISO 10555-1<br>after preconditioning in a simulated use model. | Pass |
| Kink Resistance | Kink resistance was evaluated after preconditioning in a<br>simulated use model. | Pass |
| Liquid Leakage | The device was exposed to a liquid pressure for 30<br>seconds. The device was inspected for leakage per ISO<br>10555-1. | Pass |
| Air Leakage | The device was tested for air leakage into the hub per<br>ISO 10555-1. | Pass |
| Dynamic Burst | Tested to verify the device can withstand internal liquid<br>pressure under dynamic flow conditions with the distal<br>end open. | Pass |
| Torque Strength | The device was evaluated for torque strength by<br>measuring the number of catheter rotations until failure<br>after preconditioning in a simulated use model and<br>compared to the predicate. | Pass |
| Hub Validation<br>Testing | The device shall meet the established acceptance criteria<br>per ISO 80369-7. | Pass |
| Particulate Matter | The catheter underwent simulated use testing and<br>particulate testing was conducted including a reference<br>device for comparison. | Pass |
| Tip Buckling | The maximum force to cause catheter tip buckling while<br>constrained at varying distances was measured. | The tip stiffness was<br>comparable to the<br>predicate and other cleared<br>catheters. |
| Corrosion | Corrosion tested per ISO 10555-1. | No evidence of corrosion<br>and met requirements per<br>ISO 10555-1. |
| Static Burst | The distal tip of the catheter was blocked, and fluid was<br>injected into the lumen at increasing pressure until the<br>catheter burst per ISO 10555-1 and the static burst<br>pressure was compared with the maximum pressure<br>generated with manual syringe injection. | Pass |
| Test | Test Method Summary | Results |
| Coating Integrity | The coating integrity was inspected before and after preconditioning through a simulated use model. | No evidence of surface damage or coating defects. |
| Saline and<br>Contrast Exposure | After the device was used to deliver saline and contrast media, the device was inspected for damage, and dimensional attributes were measured. | No visual evidence of damage or dimensional changes. |
| Radiopacity<br>(Visibility) | The device was tested to demonstrate acceptable radiopacity. | Marker radiopacity is comparable to the predicate. |
| Design Validation<br>/Usability | The subject and predicate devices were prepared in accordance with their respective instructions for use and tested for device usability in a clinically relevant anatomical model. | Device preparation, introduction, trackability, and retrieval were comparable to the predicate. |
{6}------------------------------------------------
## Performance Data - Bench:
The following performance bench testing was conducted to assess the performance of the Next
Generation Access Catheter:
{7}------------------------------------------------
### Sterilization and Shelf-Life:
The Next Generation Access Catheters are sterilized using an ethylene oxide sterilization cycle that was verified to a sterility assurance level of 1 x 10° in accordance with ISO 11135. Accelerated aging testing for the Next Generation Access Catheter based on ASTM F1980 has established that the subject device and packaging remain functional for the labeled expiration date. Aging studies for packaging integrity, seal strength, and device functionality were performed and met the acceptance criteria.
### Performance Data - Animal:
No animal study was conducted because bench performance testing was deemed sufficient for verification and validation and to support substantial equivalence.
### Performance Data - Clinical:
No clinical study was conducted because bench performance testing was deemed sufficient for verification and validation and to support substantial equivalence.
### Conclusion:
The evidence presented in this 510(k) submission demonstrates substantial equivalence between the subject device and the predicate device. The subject and predicate devices have intended use and indications for use. The differences in technological characteristics do not raise new questions of safety and effectiveness. Nonclinical bench testing demonstrates the Next Generation Access Catheter meets the device specifications and perform as intended.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.