← Product Code [QJP](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/QJP) · K213390

# Benchmark BMX96 Access System (K213390)

_Penumbra, Inc. · QJP · Aug 18, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94cardiovascular-diagnostic-devices/QJP/K213390

## Device Facts

- **Applicant:** Penumbra, Inc.
- **Product Code:** [QJP](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/QJP.md)
- **Decision Date:** Aug 18, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Benchmark BMX96 Access System is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

## Device Story

Three-component system: Benchmark BMX96 Delivery Catheter, Neuron 6F Select Catheter, and Dilator; used for percutaneous access to peripheral, coronary, and neuro vasculature. Delivery catheter: single lumen, laser-cut hypotube-reinforced, variable stiffness, radiopaque distal marker, Luer hub. Neuron 6F Select: single lumen, braid-reinforced, variable stiffness, four tip shapes (SIM, H1, BER, SIM-V). Dilator: single lumen, tapered distal end, facilitates atraumatic skin-to-vessel entry. Operated by physicians in clinical settings to navigate vasculature and deliver interventional devices. System provides pathway for interventional tools; enables minimally invasive procedures; benefits patient by facilitating access to target anatomy.

## Clinical Evidence

Bench testing only. No animal or clinical studies conducted. Bench-top performance testing included particulate testing, coating integrity testing, and simulated use testing; all met acceptance criteria. Biocompatibility assessment based on identical materials and sterilization methods to previously cleared products per ISO 10993-1.

## Technological Characteristics

Three-component catheter system. Materials: laser-cut hypotube-reinforced (delivery catheter), braid-reinforced (Neuron 6F). Features: variable stiffness, radiopaque distal markers, Luer hubs. Compatible with 0.038 in. guidewires. Sterilization method: identical to predicate. No software or electronic components.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Benchmark BMX96 System ([K201271](/device/K201271.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 18, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Penumbra, Inc. Buu Buu Ly Regulatory Affairs Specialist III One Penumbra Place Alameda, California 94502

Re: K213390

Trade/Device Name: Benchmark BMX96 Access System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP, DQY Dated: July 13, 2022 Received: July 15, 2022

Dear Buu Buu Ly:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K213390

Device Name Benchmark BMX96 Access System

Indications for Use (Describe)

The Benchmark BMX96 Access System is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Penumbra, a healthcare company. The word "Penumbra" is written in a bold, sans-serif font in a dark red color. To the right of the word is a stylized "P" inside of a red circle, with a thin white line connecting the "a" in Penumbra to the "P" in the circle. The logo is simple and modern, and the red color gives it a sense of energy and urgency.

### 510(k) Summary

(as required by 21 CFR 807.92)

Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra, Inc. is providing the summary of Substantial Equivalence for the subject Benchmark BMX96 Access System

#### 1.1 Submitter

Penumbra. Inc. One Penumbra Place Alameda, CA 94502 USA

Contact person: Buu Buu Ly Regulatory Affairs Specialist III Tel: (510) 995-2332 Fax: (510) 217-6414 E-mail: bly(@penumbrainc.com

Date of preparation: June 01, 2022

### Subject Device 1.2

Benchmark BMX96 Access System

Regulatory Class: II Classification Panel: Neurology Classification Name: Catheter, Percutaneous, Neurovasculature Regulation Number: 21 CFR 870.1250 Product Code: QJP, DQY

### Predicate Device 1.3

Benchmark BMX96 System (K201271)

### 1.4 Device Description

The Benchmark BMX96 Access System is a three-component system comprised of the Benchmark BMX96 Delivery Catheter, Neuron 6F Select Catheter, and a Dilator. The Benchmark BMX96 Delivery Catheter can be used individually with a 0.038 in. [0.97 mm] guidewire or together with the Neuron 6F Select Catheter to access the desired anatomy.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The word "Penumbra" is written in a bold, red font. To the right of the name is a red circle with a white "P" inside, connected to the word "Penumbra" by a thin gray line.

## Benchmark BMX96 Delivery Catheter

The Benchmark BMX96 Delivery Catheter is a single lumen, laser-cut hypotubereinforced, variable stiffness catheter with a radiopaque marker band on the distal end and a Luer hub on the proximal end. The Benchmark BMX96 Delivery Catheter is compatible with introducer sheaths appropriately sized for the outer of Benchmark BMX96.

## Neuron 6F Select Catheter

The Neuron 6F Select Catheter is a single lumen, braid-reinforced, variable stiffness catheter with a radiopaque distal end and a Luer hub on the proximal end. The Neuron 6F Select Catheter is available in four tip shapes (SIM, H1, BER, or SIM-V). The Neuron 6F Select Catheter is compatible with the Benchmark BMX96 Delivery Catheter.

## Dilator

The Dilator is a single lumen, radiopaque catheter with a tapered distal end and a Luer hub on the proximal end. The Dilator is compatible with the Benchmark BMX96 Delivery Catheter. The Dilator facilitates the percutaneous entry of the Benchmark BMX96 Delivery Catheter by forming an atraumatic transition from the skin through the subcutaneous tissue to the vessel.

#### 1.5 Indications For Use

The Benchmark BMX96 Access System is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

### Comparison of Technological Characteristics with the Predicate Device 1.6

The subject and predicate devices have identical technological characteristics.

The only difference between the subject and predicate devices are additional instructions added to the instructions for use (IFU) for the subject device related to radial access use.

#### Performance Data 1.7

The following performance data were provided in support of the substantial equivalence determination:

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Penumbra logo. The word "Penumbra" is written in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. A thin white line extends from the right side of the word "Penumbra" to the left side of the "P".

- . Bench-top Performance.
The subject device met all established requirements.

### 1.7.1 Bench-top Performance

The following bench-top performance tests were performed on the subject device and all have met acceptance criteria:

- Particulate Testing and Coating Integrity Testing. .
- Simulated Use Testing. ●

### Biocompatibility 1.7.2

The subject Benchmark BMX96 Access System is categorized as a limited exposure (< 24 hours), externally communicating device with circulating blood contact in accordance with ISO 10993-1. The design and manufacturing of the subject Benchmark BMX96 Access System use identical materials, similar processing, and identical sterilization methods as products that Penumbra has already successfully conducted biocompatibility testing for per ISO 10993-1. Therefore, no additional biocompatibility testing is required.

### 1.7.3 Performance Data – Animal, Clinical

No animal or clinical studies were conducted because bench testing was determined sufficient for verification and validation purposes.

### 1.8 Conclusions

The subject Benchmark BMX96 Access System is substantially equivalent to the predicate device Benchmark BMX96 System. The subject device has the same indications for use as the predicate device. The device testing described in this 510(k) Summary demonstrates that the subject device is substantially equivalent to the predicate device in regard to operating principle, design concept, fundamental technology, and device performance. The changes to the instructions for use do not raise new questions of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94cardiovascular-diagnostic-devices/QJP/K213390](https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94cardiovascular-diagnostic-devices/QJP/K213390)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
