Route 92 Medical Full Length 054 Reperfusion System and Aspiration Tubing Set
Applicant
Route 92 Medical, Inc.
Product Code
NRY · Cardiovascular
Decision Date
Apr 23, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Route 92 Medical Full Length 054 Reperfusion System and Aspiration Tubing Set is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.
Device Story
System comprises 054 Aspiration Catheter, Tenzing 5 Delivery Catheter, and Aspiration Tubing Set. Used by neurovascular interventional physicians in hospital settings for mechanical thrombectomy. Catheters are single-lumen, coil-reinforced, variable stiffness, and hydrophilically coated. System connects to a hospital-grade vacuum pump via the tubing set to provide constant suction (-25 to -27.5 inHg). Physician navigates catheters to the occlusion site under fluoroscopic guidance; aspiration removes thrombus. Output is restored blood flow in the occluded vessel. Benefits include rapid revascularization for acute ischemic stroke patients.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including simulated use, lumen patency, dimensional verification, tensile strength, kink/torsion resistance, tip flexibility, leakage, burst pressure, coating integrity, and particulate evaluation. All tests met pre-determined acceptance criteria.
Technological Characteristics
Single-lumen, coil-reinforced, variable stiffness catheters with hydrophilic coating. Aspiration tubing made of high-pressure material with on/off switch. Sterilized via ethylene oxide (ISO 11135). Luer fittings per ISO 80369-7:2021. Leakage/burst testing per ISO 10555-1:2013.
Indications for Use
Indicated for patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (internal carotid, M1/M2 middle cerebral, basilar, and vertebral arteries) within 8 hours of symptom onset. Candidates include those ineligible for or failed IV t-PA therapy.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Penumbra Reperfusion Catheter 054 and Penumbra Separator 054 (K090752)
Reference Devices
Route 92 Medical Full Length 054 Access System (K231583)
Route 92 Medical Full Length 070 Reperfusion System and Aspiration Tubing Set (K223530)
Submission Summary (Full Text)
{0}------------------------------------------------
April 23, 2024
Image /page/0/Picture/1 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Route 92 Medical, Inc. Kirsten Valley Chief Operating Officer 155 Bovet Road, Suite 100 San Mateo, California 94402
### Re: K233329
Trade/Device Name: Route 92 Medical Full Length 054 Reperfusion System and Aspiration Tubing Set Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter
Regulatory Class: Class II Product Code: NRY Dated: March 24, 2024 Received: March 25, 2024
Dear Kirsten Valley:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K233329
Device Name
Route 92 Medical Full Length 054 Reperfusion System and Aspiration Tubing Set
Indications for Use (Describe)
The Route 92 Medical Full Length 054 Reperfusion System and Aspiration Tubing Set is indicated for use in the revascularization of patients with acute secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K233329 510(K) SUMMARY
| Sponsor: | Route 92 Medical, Inc.<br>155 Bovet Road, Suite 100<br>San Mateo, CA 94402<br>Phone: 650-581-1179 |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Kirsten Valley |
| Date Prepared: | April 21, 2024 |
| Device Name: | Route 92 Medical Full Length 054 Reperfusion System and<br>Aspiration Tubing Set |
| Common Name: | Percutaneous Catheter |
| Classification Name: | Catheter, Thrombus Retriever (Product Code NRY, 21<br>CFR 870.1250) |
| Device Classification | Class II |
| Predicate Device: | Penumbra Reperfusion Catheter 054 and Penumbra<br>Separator 054, K090752 |
| Reference Devices: | Route 92 Medical Full Length 054 Access System,<br>K231583<br>Route 92 Medical Full Length 070 Reperfusion System and<br>Aspiration Tubing Set, K223530 |
#### Device Description
The 054 Aspiration Catheter and Tenzing 5 Delivery Catheter comprise the Route 92 Medical Full Length 054 Reperfusion System. The Route 92 Medical Full Length 054 Reperfusion System is used with the Route 92 Medical Aspiration Tubing Set.
The 054 Aspiration Catheter is a single-lumen, coil-reinforced variable stiffness catheter. The Tenzing 5 Delivery Catheter is a single-lumen variable stiffness catheters are hydrophilically coated. The devices are provided sterile and non-pyrogenic and are intended for single use only.
The Route 92 Medical Aspiration Tubing Set is made from high-pressure tubing. An on/off switch is provided to control fluid flow. A suction connector allows connection to an aspiration pump canister and a Luer fitting allows connection to the aspiration catheter.
For the aspiration source, the Aspiration Catheter is used in conjunction with an aspiration pump with prespecified performance parameters that is connected using the Route 92 Medical Aspiration Tubing Set, along with a legally marketed canister and accessories kit.
{4}------------------------------------------------
#### Indications for Use
The Route 92 Medical Full Length 054 Reperfusion System and Aspiration Tubing Set is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.
### Comparison to the Predicate Device
The method of action, design, and materials of the Route 92 Medical Full Length 054 Reperfusion System and Aspiration Tubing Set are substantially equivalent to the predicate device as shown in the following table.
| Attribute | Predicate Device<br>Penumbra Reperfusion Catheter 054<br>and Penumbra Separator 054<br>(K090752) | Subject Device<br>Route 92 Medical Full Length 054<br>Reperfusion System and Aspiration<br>Tubing Set (K233329) |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Penumbra System is intended for<br>use in the revascularization of patients<br>with acute ischemic stroke secondary<br>to intracranial large vessel occlusive<br>disease (within the internal carotid,<br>middle cerebral – M1 and M2<br>segments, basilar, and vertebral<br>arteries) within 8 hours of symptom<br>onset. | The Route 92 Medical Full Length<br>054 Reperfusion System and<br>Aspiration Tubing Set is indicated for<br>use in the revascularization of patients<br>with acute ischemic stroke secondary<br>to intracranial large vessel occlusive<br>disease (within the internal carotid,<br>middle cerebral - M1 and M2<br>segments, basilar, and vertebral<br>arteries) within 8 hours of symptom<br>onset. Patients who are ineligible for<br>intravenous tissue plasminogen<br>activator (IV t-PA) or who failed IV t-<br>PA therapy are candidates for<br>treatment. |
| Device<br>Description | The Penumbra System consists of<br>three devices that work as a system to<br>remove thrombus including the<br>Penumbra Reperfusion Catheter,<br>Penumbra Separator and Aspiration<br>Tubing. The Penumbra System is<br>used with the Penumbra Aspiration<br>Pump. | The Route 92 Medical Full Length<br>054 Reperfusion System includes the<br>054 Aspiration Catheter and Tenzing<br>5 Delivery Catheter and is used with<br>the Route 92 Medical Aspiration<br>Tubing Set. The devices are sterile,<br>non-pyrogenic, and intended for<br>single use only. |
| User | Physicians trained in neurovascular<br>interventional techniques | Same |
| Attribute | Predicate Device<br>Penumbra Reperfusion Catheter 054<br>and Penumbra Separator 054<br>(K090752) | Subject Device<br>Route 92 Medical Full Length 054<br>Reperfusion System and Aspiration<br>Tubing Set (K233329) |
| Anatomical<br>Sites | Neurovasculature only | Same |
| Materials | Polymers and metals commonly used<br>in the manufacture of medical devices | Same |
| Sterilization | Ethylene oxide | Same |
| Shelf-life | 3 years | 6 months |
| Aspiration Catheter | | |
| Inner<br>Diameter<br>(nominal) | 0.054" | 0.054" |
| Outer<br>Diameter<br>(nominal) | 0.064" | 0.066" |
| Length | 135 cm | 125 cm and 148 cm |
| Delivery Catheter | | |
| Inner<br>Diameter | Not applicable | 0.019" proximal<br>0.015" distal |
| Outer<br>Diameter | Not applicable | 0.048" |
| Length | Not applicable | 167 cm |
| Compatibility:<br>Pump for<br>Aspiration | Penumbra Pump MAX | Commercially available vacuum<br>pump designed for use in hospital<br>settings that is compatible with the<br>Route 92 Medical Reperfusion<br>System and provides a constant<br>vacuum of -25 inHg to -27.5 inHg. |
| Aspiration Tubing | | |
| Inner<br>Diameter | 0.71"-0.110" | 0.110" |
| Outer<br>Diameter | Not available | 0.188" |
| Length | 112" | 112" |
| Attribute | Predicate Device<br>Penumbra Reperfusion Catheter 054<br>and Penumbra Separator 054<br>(K090752) | Subject Device<br>Route 92 Medical Full Length 054<br>Reperfusion System and Aspiration<br>Tubing Set (K233329) |
| Materials | Biocompatible materials, commonly<br>utilized for interventional devices | Same |
{5}------------------------------------------------
{6}------------------------------------------------
### Non-Clinical Testing
### Shelf Life and Sterility
The Route 92 Medical Full Length 054 Reperfusion System and Aspiration Tubing are sterilized using an ethylene oxide sterilization cycle that was verified to a sterility assurance level of 1 x 10° in accordance with ISO 11135. Aging studies established that the subject device and packaging remain functional for the labeled expiration date.
### Biocompatibility Testing
The patient-contacting materials of the Route 92 Medical Full Length 054 Reperfusion System and Aspiration Tubing Set were unchanged compared to the reference device K223530; therefore, no additional biocompatibility testing was required.
#### Performance Testing
The successful completion of the performance testing listed in the following tables demonstrates that the Route 92 Medical Full Length 054 Reperfusion System is suitable for its intended use.
| Full Length 054 Reperfusion System | | | |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|--|
| Test | Test Method | Results | |
| Simulated Use<br>Testing | Deliverability, clot retrieval, device integrity, kink<br>and aspiration resistance, lumen patency, durability<br>and compatibility with accessory devices were<br>evaluated in a neurovascular model. | PASS<br>All samples met the pre-<br>determined acceptance criteria | |
| Lumen Patency<br>Testing | Test specimens were tested for lumen patency under<br>vacuum. | PASS<br>All samples met the pre-<br>determined acceptance criteria | |
| Dimensional<br>Verification | Device dimensions were measured to confirm<br>conformance to the specifications. | PASS<br>All samples met the pre-<br>determined acceptance criteria | |
| Luer Integrity | Tested per ISO 80369-7:2021. | PASS<br>All samples met the pre-<br>determined acceptance criteria | |
| Full Length 054 Reperfusion System | | | |
| Test | Test Method | Results | |
| Tensile Strength | The tensile strength of the catheter sections and bonds<br>was tested. | PASS<br>All samples met the pre-<br>determined acceptance criteria | |
| Kink Resistance | Test specimen segments were formed into a defined<br>bend diameter to evaluate kink resistance. | PASS<br>All samples met the pre-<br>determined acceptance criteria | |
| Torsion<br>Resistance | The test specimens were rotated to evaluate integrity<br>after rotation. | PASS<br>All samples met the pre-<br>determined acceptance criteria | |
| Tip Flexibility | Test specimens were tested for tip flexibility. | PASS<br>All samples met the pre-<br>determined acceptance criteria | |
| Air Leakage | Tested per ISO 10555-1:2013 Annex D. | PASS<br>All samples met the pre-<br>determined acceptance criteria | |
| Liquid Leakage | Tested per ISO 10555-1:2013 Annex C. | PASS<br>All samples met the pre-<br>determined acceptance criteria | |
| Static Burst | Tested per ISO 10555-1:2013 Annex F. | PASS<br>All samples met the pre-<br>determined acceptance criteria | |
| Dynamic Burst | Mechanical integrity was assessed under pressure. | PASS<br>All samples met the pre-<br>determined acceptance criteria | |
| Hydrophilic<br>Coating Integrity | The integrity of the hydrophilic coating was evaluated<br>after multiple insertion and withdrawal cycles. | PASS<br>All samples met the pre-<br>determined acceptance criteria | |
| Aspiration<br>Pressure | The vacuum pressure at the tip of the catheter was<br>assessed during aspiration. | PASS<br>All samples met the pre-<br>determined acceptance criteria | |
| Particulate | The size and quantity of particulates generated during<br>simulated use were evaluated. | PASS<br>All samples met the pre-<br>determined acceptance criteria | |
{7}------------------------------------------------
## Animal Studies:
No animal study was deemed necessary to demonstrate substantial equivalence to the predicate device.
{8}------------------------------------------------
### Clinical Studies:
No clinical study was deemed necessary to demonstrate substantial equivalence to the predicate device. Conclusion
The Route 92 Medical Full Length 054 Reperfusion System and Aspiration Tubing Set has the same intended use, similar technological characteristics, and same method of action as the predicate device. Differences between the subject and predicate devices do not raise new questions of safety and effectiveness of the device. The successful completion of performance testing demonstrates that the Route 92 Medical Full Length 054 Reperfusion System and Aspiration Tubing Set is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.