The AXS Vecta Aspiration Catheter, as part of the AXS Vecta Aspiration System, is indicated in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - MI and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.
Device Story
AXS Vecta Aspiration System is a neurovascular intervention tool for acute ischemic stroke. System components include an aspiration catheter, Scout/Peel-Away introducers, hemostasis valve, universal aspiration tubing, Medela Dominant Flex pump, and universal liner set. Operated by physicians in clinical settings, the device uses vacuum aspiration to remove thrombi from intracranial vessels. The pump provides adjustable vacuum pressure to the catheter tip, facilitating clot retrieval. The system is designed for revascularization within 8 hours of symptom onset. Clinical benefit includes restoring blood flow in large vessel occlusions. The device is a modification of the predicate system, specifically integrating the Medela Dominant Flex pump and associated tubing. Performance is verified through benchtop simulated use and usability testing, confirming that the system maintains required vacuum pressure and lumen patency during operation.
Clinical Evidence
No clinical study was conducted. Substantial equivalence is supported by bench testing (simulated use, usability, vacuum drop, lumen patency) and a literature review of clinical outcomes for technologically similar devices used for direct aspiration.
Technological Characteristics
System includes aspiration catheter, introducers, hemostasis valve, aspiration tubing, and Medela Dominant Flex pump. Operates via vacuum aspiration with adjustable vacuum pressure (max 28 in Hg). Components are single-use (catheter, tubing, liners) or reusable (pump). Connectivity is mechanical/pneumatic. Sterilization method not specified.
Indications for Use
Indicated for revascularization in patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (internal carotid, middle cerebral M1/M2, basilar, and vertebral arteries) within 8 hours of symptom onset. Applicable to patients ineligible for or failed IV t-PA therapy.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, featuring an abstract symbol. To the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
April 24, 2019
Stryker Neurovascular Shazia Hakim Senior Staff Regulatory Affairs Specialist 47900 Bayside Parkway Fremont, California 94538
Re: K190212
Trade/Device Name: AXS Vecta Aspiration System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: March 22, 2019 Received: March 25, 2019
Dear Shazia Hakim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Xiaolin Zheng him in the many of the subject of the subject of the station of the subject of the subject of the station of the station of the subject of the subjection of the subjection o -5 J Date: 2019.04.24 14:13:14 -04'00'
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190212
Device Name AXS Vecta Aspiration System
#### Indications for Use (Describe)
The AXS Vecta Aspiration Catheter, as part of the AXS Vecta Aspiration System, is indicated in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - MI and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
|---------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
## Submitter's Name and Address
Stryker Neurovascular 47900 Bayside Parkway Fremont, California 94538 Telephone: 1-510-413-2636 Facsimile: 1-510-413-2588
## Contact Information
Shazia Hakim Staff Regulatory Affairs Specialist Stryker Neurovascular 47900 Bayside Parkway Fremont, CA 94538 510-413-2636 (phone) 510-413-2588 (fax) Shazia.Hakim@stryker.com
#### Secondary Contact
Angelica Beckmann Director of Regulatory Affairs Stryker Neurovascular 47900 Bayside Parkway Fremont, CA 94538 650-224-9823 (phone) angelica.beckmann@stryker.com
#### Date Prepared
January 31, 2019
## Device Trade or Proprietary Name
AXS Vecta Aspiration System
{4}------------------------------------------------
### Device Common or Classification Name:
Catheter, Percutaneous, 21 CFR 870.1250, Class II
## Product Code:
NRY (Catheter)
ldentification of the Legally Marketed Devices to which Equivalence is Being Claimed
| Name of Predicate Device | Name of Manufacturer | 510(k) Number |
|-------------------------------|----------------------|---------------|
| Zenith Flex Aspiration System | InNeuroCo, Inc | K181354 |
| Name of Reference Device | Name of Manufacturer | 510(k) Number |
### Device Description
The Stryker AXS Vecta Aspiration System includes an aspiration catheter and its accessories, including the Scout Introducer, the Peel-Away Introducers, the Hemostasis Valve, the AXS Universal Aspiration Tubing, the Medela Dominant Flex Pump, and the AXS Universal Liner Set.
### Indications for Use
The AXS Vecta Aspiration Catheter, as part of the AXS Vecta Aspiration System, is indicated in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who failed IV t-PA therapy are candidates for treatment.
{5}------------------------------------------------
# Comparison to Predicate Device
| | Predicate Device<br>Stryker AXS Vecta<br>Aspiration System<br>(formerly<br>InNeuroCo Zenith<br>Flex System) | Reference Device<br>Stryker AXS<br>Universal<br>Aspiration System | Subject Device<br>AXS Vecta<br>Aspiration System<br>(with the Medela<br>Dominant Flex<br>Pump) |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| 510(k) Number | K181354 | K173841 | K190212 |
| Classification | Class II | Class II | Same |
| Product Code | NRY | NRY | Same |
| Review Panel | Neurology | Neurology | Same |
| Indications For<br>Use | The Zenith Flex<br>System, including the<br>Zenith Flex Catheter,<br>Aspiration Tubing<br>Set, and VC-701 Cliq<br>Aspirator Pump, is<br>indicated in the<br>revascularization of<br>patients with acute<br>ischemic stroke<br>secondary to<br>intracranial large<br>vessel occlusive<br>disease (within the<br>internal carotid,<br>middle cerebral – M1<br>and M2 segments,<br>basilar, and vertebral<br>arteries) within 8<br>hours of symptom<br>onset. Patients who<br>are ineligible for<br>intravenous tissue<br>plasminogen<br>activator (IV t-PA) or<br>who failed IV t-PA<br>therapy are<br>candidates for<br>treatment. | The AXS Catalyst<br>Distal Access<br>Catheter as part of<br>the AXS Universal<br>Aspiration System is<br>indicated for use in<br>the<br>revascularization of<br>patients with acute<br>ischemic stroke<br>secondary to<br>intracranial large<br>vessel occlusive<br>disease (in the<br>internal carotid,<br>middle cerebral - M1<br>and M2 segments,<br>basilar, and vertebral<br>arteries) within 8<br>hours of symptom<br>onset.<br>Patients who are<br>ineligible for<br>intravenous tissue<br>plasminogen<br>activator (IV t-PA) or<br>who failed IV t-PA<br>are candidates<br>for treatment. | Same |
{6}------------------------------------------------
| | Predicate Device<br>Stryker AXS Vecta<br>Aspiration System<br>(formerly<br>InNeuroCo Zenith<br>Flex System) | Reference Device<br>Stryker AXS<br>Universal<br>Aspiration System | Subject Device<br>AXS Vecta<br>Aspiration System<br>(with the Medela<br>Dominant Flex<br>Pump) | | | |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------|-----------------------------------------------|-------------|----------------------------|
| Patient Contacting<br>Components | AXS Vecta Aspiration<br>Catheter and its<br>Accessories | AXS Catalyst Distal<br>Access Catheter | Same as AXS Vecta<br>Aspiration System | | | |
| Pump | Cliq VC-701<br>Aspirator | Medela Dominant<br>Flex Pump | Same as AXS<br>Universal Aspiration<br>System | | | |
| Aspiration Tubing | Aspiration Tubing Set | AXS Universal<br>Aspiration Tubing | Same as AXS<br>Universal Aspiration<br>System | | | |
| Replacement<br>Components | Canister Set | AXS Universal Liner<br>Set | Same as AXS<br>Universal Aspiration<br>System | | | |
| Aspiration Method | Pump | Pump | Same | | | |
| Single Use<br>Components | Catheter and its<br>accessories,<br>Aspiration Tubing,<br>Canister Sets | Catheter and its<br>accessories, Aspiration<br>Tubing, Liner Set | Same as AXS<br>Universal Aspiration<br>System | | | |
| Reusable<br>Components | Pump | Pump | Same | | | |
| Maximum<br>Aspiration<br>Pressure | 27 in Hg | 28 in Hg | Same as AXS<br>Universal Aspiration<br>System | | | |
| Calculated Force<br>at Tip | Catheter ID | Calculated<br>Force at tip | Catheter ID | Calculated<br>Force at tip | Catheter ID | Calculated<br>Force at tip |
| | 0.071 in | 0.024 kgf | 0.060 in | 0.018 kgf | 0.071 in | 0.025 kgf |
| | 0.074 in | 0.026 kgf | | | 0.074 in | 0.027 kgf |
| Pressure<br>Regulator Method | Adjustable vacuum<br>pressure dial | Same | | Same | | |
| Flow Rate | Non-adjustable Flow<br>Rate | Three Selectable Air<br>Flow Rates: 40, 50,<br>60 mL | | Same as AXS<br>Universal Aspiration<br>System | | |
{7}------------------------------------------------
### Summary of Non-Clinical Data
The subject device has the same materials, packaging, manufacturing process, and sterilization process as the predicate device. Therefore, testing was only conducted for specifications that were affected by changing the pump and the tubing.
#### Animal Testing
Animal testing previously conducted for the Zenith Flex System was used to support this change. The GLP animal testing performed with the System included a safety evaluation and an efficacy evaluation at a maximum pressure rating of -27 inHq. However, a calculation of the force at the tip demonstrated that the pressure differences and forces exhibited on the product with the different pumps are minimal. Therefore, this is a modification of an interchangeable component. Since both components have already received clearance for this indication and because the bench top data supports that the catheter still works within the pressure ranges previously established as safe, the animal testing previously conducted can be used to support this change as well.
### AXS Vecta Performance Testing
To demonstrate substantial equivalence between the subject AXS Vecta Aspiration System and the predicate Zenith Flex System and the reference Stryker AXS Universal Aspiration System, performance testing was conducted. The tests were performed using standard test methods and pre-determined acceptance criteria and all samples passed. Therefore, this test data supports the argument that the AXS Vecta Aspiration System has similar performance characteristics as the predicate device and the reference device. All the testing conducted to demonstrate substantial equivalence are presented in the following table.
| Test | Test Method<br>Summary | Acceptance Criteria | Conclusions |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Simulated Use - Bench<br>(AXS Vecta Aspiration<br>System) | AXS Vecta Aspiration<br>System underwent<br>simulated use testing by<br>a physician in a<br>benchtop model | Test samples must meet<br>predetermined user needs | AXS Vecta Aspiration System<br>test samples met the<br>acceptance criteria for<br>Simulated Use<br>- Bench to demonstrate that<br>the AXS Vecta System is<br>substantially equivalent to the<br>predicate device. |
{8}------------------------------------------------
| Test | Test Method<br>Summary | Acceptance Criteria | Conclusions |
|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Simulated Use Testing<br>– Usability<br>(AXS Vecta Aspiration<br>System) | AXS Vecta Aspiration<br>System underwent<br>simulated use testing by a<br>physician in a benchtop<br>model | Test samples must meet<br>predetermined user needs | AXS Vecta Aspiration System<br>test samples met the<br>acceptance criteria for<br>Simulated Use<br>- Usability to demonstrate that<br>the AXS Vecta Aspiration<br>System is substantially<br>equivalent to the predicate<br>device. |
| Vacuum Drop / Suction<br>Connector Secure<br>Attachment<br>(Aspiration Tubing Set) | Vacuum pressure<br>measured at source and<br>tip to evaluate pressure<br>difference | Test sample results must<br>meet or exceed existing<br>pressure specifications. | AXS Vecta Aspiration System<br>test samples met the<br>acceptance criteria for<br>Vacuum Drop to demonstrate<br>that the AXS Vecta Aspiration<br>System is substantially<br>equivalent to the predicate<br>device. |
| Lumen Patency<br>(Aspiration Tubing Set) | Samples were evaluated<br>for lumen collapse during<br>aspiration | Test sample results must<br>meet or exceed existing<br>lumen patency<br>specifications. | AXS Vecta Aspiration System<br>test samples met the<br>acceptance criteria for Lumen<br>Patency to demonstrate that<br>the AXS Vecta Aspiration<br>System is substantially<br>equivalent to the predicate<br>device. |
## Performance Data - Clinical
No clinical study was conducted as bench and animal testing was determined sufficient for verification and validation purposes. A review was conducted considering published clinical study articles that featured the Reference Device and other devices with similar dimensions used for direct aspiration. The literature review was used to support the proposed indications for use under the NRY product code by using clinical outcomes from devices that are considered technologically similar. This documentation was reviewed within the scope of this change and determined to be applicable.
## Summary of Substantial Equivalence
The performance characteristics and the test results demonstrate that the AXS Vecta Aspiration System meets the acceptance criteria to determine that the AXS Vecta Aspiration System is substantially equivalent to the predicate device. Furthermore, the intended use, the operating principles, and the design are all equivalent and support the conclusion that all devices are technologically similar.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.