MODIFIED TREVO PROVUE RETRIEVER

K133464 · Concentric Medical, Inc. · NRY · Jan 17, 2014 · Cardiovascular

Device Facts

Record IDK133464
Device NameMODIFIED TREVO PROVUE RETRIEVER
ApplicantConcentric Medical, Inc.
Product CodeNRY · Cardiovascular
Decision DateJan 17, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Device Story

Modified Trevo ProVue Retriever is a neurovascular thrombus retrieval device. Consists of flexible, tapered nitinol core wire with distal shaped section; radiopaque platinum/tungsten coils allow fluoroscopic visualization. Hydrophilic coating reduces friction. Used in neurovasculature to remove thrombi in ischemic stroke patients. Operated by physicians under fluoroscopic guidance. Device delivered via microcatheter; torque device and insertion tool facilitate navigation. Restores blood flow by mechanical clot retrieval. Benefits include rapid revascularization in acute stroke. Proximal end compatible with guide wire extensions for catheter exchange. Single-use, sterile device.

Clinical Evidence

Bench testing only. Simulated use testing performed in a silicone neurovascular model replicating human arterial tortuosity, diameter, and ICA siphon. Model utilized 37°C pressurized water bath (2-2.5 psi) to simulate circulation. Testing assessed deliverability, deployment, and reloadability using simulated thrombus. All samples met acceptance criteria.

Technological Characteristics

Materials: Nitinol core/shaped section, platinum/tungsten radiopaque coils, 304 stainless steel proximal coil, gold/tin solder, sodium hyaluronate hydrophilic coating. Dimensions: 180 cm length, 4 mm shaped section diameter. Energy: Mechanical (manual operation). Connectivity: None. Sterilization: 100% EtO. Form factor: Tapered core wire with distal shaped section.

Indications for Use

Indicated for patients experiencing ischemic stroke within 8 hours of symptom onset, including those ineligible for or failing IV t-PA therapy, to restore neurovascular blood flow by thrombus removal.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K133464 JAN 17 2014 510(k) Summary | Trade Name: | Modified Trevo ProVue Retriever | | |--------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------| | Common Name: | Catheter, Thrombus Retriever | | | Classification Name: | Thrombus Retriever, 21CFR 870.1250 Class II | | | Product Code: | NRY | | | Submitter: | Concentric Medical, Inc.<br>301 E. Evelyn Avenue<br>Mountain View, CA 94041<br>Tel 510-413-2681<br>Fax 510-413-2558<br>Facility Registration #2954917 | | | Contact: | Sarah Meyer<br>Senior Regulatory Affairs Specialist | | | Date Prepared: | December 13, 2013 | | | Predicate Device: | Concentric Trevo ProVue Retriever (K122478) | | | Feature | Cleared Trevo ProVue Retriever<br>(K122478) | Modified Trevo<br>ProVue Retriever | | Indications for Use | The Trevo ProVue Retriever is intended to restore<br>blood flow in the neurovasculature by removing<br>thrombus in patients experiencing ischemic stroke<br>within 8 hours of symptom onset. Patients who are<br>ineligible for intravenous tissue plasminogen<br>activator (IV t-PA) or who fail IV t-PA therapy are<br>candidates for treatment. | Same as cleared<br>device | | Device Description | The Retriever consists of a flexible, tapered core<br>wire with a shaped section at the distal end. A<br>platinum coil allows fluoroscopic visualization. In<br>addition, the shaped section is also radiopaque. The<br>Retriever has a hydrophilic coating to reduce friction.<br>The Retriever has a shaft marker to indicate<br>proximity of Retriever tip relative to Microcatheter tip.<br>A torque device and insertion tool are provided with<br>the Retriever. | Same as cleared<br>device | | Target Population | Patients with symptoms of an ischemic stroke | Same as cleared<br>device | | Anatomical Sites | Neurovasculature | Same as cleared<br>device | | Accessory Devices<br>Provided<br>(not in direct contact<br>with patient) | Insertion tool and torque device provided in product<br>package | Same as cleared<br>device | | Regulation Number | 21CFR 870.1250 | Same as cleared<br>device | | Regulation Name | Catheter, Thrombus Retriever | Same as cleared<br>device | | Regulatory Class | II | Same as cleared<br>device | | Product Code | NRY | Same as cleared<br>device | | Core Wire Material | Nitinol (nickel titanium alloy) | Same as cleared<br>device | | Distal Shaped<br>Section Material | Nitinol | Same as cleared<br>device | | Feature | Cleared Trevo ProVue Retriever<br>(K122478) | Modified Trevo<br>ProVue Retriever | | Coil Material Distal<br>to Distal Shaped<br>Section | Platinum/Tungsten | Same as cleared<br>device | | Shaped Section<br>Radiopaque Wire | Platinum/Tungsten | Same as cleared<br>device | | Coil Material<br>Proximal to<br>Shaped Section | 304 Stainless Steel | Same as cleared<br>device | | Solder | Gold/Tin | Same as cleared<br>device | | Hydrophilic Coating | Sodium hyaluronate mixture | Same as cleared<br>device | | Nominal Design Attributes | | | | Overall Length | 180 cm | Same as cleared<br>device | | Total Shaped<br>Section Length<br>(nominal) | 37 mm | Same as cleared<br>device | | Active Shaped<br>Section<br>Length (nominal) | 20 mm | Same as cleared<br>device | | Distal Tip Length<br>(nominal) | 4 mm | Same as cleared<br>device | | Proximal Core Wire<br>Diameter | 0.0180" | Same as cleared<br>device | | Shaped Section<br>Diameter (nominal) | 4 mm | Same as cleared<br>device | | Distal Taper<br>Length (nominal) | 10 mm | Same as cleared<br>device | | Packaging | | | | Materials and<br>Configuration | Polyethylene Hoop, polycarbonate mounting card,<br>Tyvek/Film Pouch, HDPE Tubing Clips, Chipboard<br>carton | Same as cleared<br>device | | Sterilization<br>Method | 100% EtO | Same as cleared<br>device | | How Supplied | Sterile/Single Use | Same as cleared<br>device | #### Device Description The Modified Trevo ProVue Retriever consists of a flexible, tapered core with a shaped section at the distal end. It is designed to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke. A radiopaque coil at the distal end allows fluoroscopic visualization. The shaped section is also radiopaque. Retriever dimensions are indicated on the product label. The Retriever has a hydrophilic coating to reduce friction during use. A torque device and an insertion tool are provided with the Retriever. The proximal end of the device is compatible with the Abbott guide wire extension to facilitate removal or exchange of a catheter while maintaining the Retriever position in the vessel. #### Indications for Use The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. {1}------------------------------------------------ #### Technological Characteristics and Product Feature Comparison The Modified Trevo ProVue Retriever with a modified platinum wire weave is substantially equivalent to the predicate device in terms of design, materials used, and function. A comparison of the subject device with predicate device is summarized in the below table. #### Comparison of Cleared Trevo ProVue Retriever to the Modified Trevo ProVue Retriever {2}------------------------------------------------ #### Risk Assessment Risk assessment of the modifications has been conducted in accordance with EN ISO 14971:2012. Concentric Medical, Inc. has determined the modifications to the predicate device raise no new questions of safety or effectiveness. {3}------------------------------------------------ Results of testing have demonstrated the Modified Trevo ProVue Retriever is substantially equivalent to the predicate device. Furthermore, the modifications did not result in any new failure modes nor were there any changes to existing failure modes. #### Testing Summary The results of testing conducted on the Modified Trevo ProVue Retriever demonstrate that it performs as designed, is suitable for its intended use and is substantially equivalent to the predicate device. Specifically, the following tests were performed on the proposed device: | Test | Test Method Summary | Results | |-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------| | Simulated<br>Use --<br>Deliverability,<br>Deployment,<br>and<br>Reloadability | Simulated use testing used a<br>silicone neurovascular model<br>cast from actual human<br>neurovascular arteries. This<br>bench testing model replicates<br>the tortuosity, diameter and<br>location of the arteries in the<br>neurovasculature including the<br>internal carotid artery (ICA)<br>siphon. The model ends at the<br>mid carotid arteries and proximal<br>support is provided by a guide<br>catheter. The model<br>incorporates a re-circulating<br>water bath at 37°C pressurized<br>between 2 - 2.5 psi (100 - 126<br>mm Hg) to simulate the human<br>arterial circulation. All testing<br>follows the procedural<br>instructions outlined in the<br>Instructions for Use. Simulated<br>thrombus is used to assess the<br>devices ability to retrieve clot | All samples passed the<br>acceptance criteria.<br><br>Device continues to meet<br>same design requirements<br>as predicate device<br>K122478. | In summary, the results of simulated testing for deliverability, deployment, and reloadability that were conducted on the Modified Trevo ProVue Retriever demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicate device. {4}------------------------------------------------ #### Summary of Substantial Equivalence The Modified Trevo ProVue Retriever is substantially equivalent to the predicate device with regard to device design, materials, intended use, and patient population. The conclusions drawn from risk assessments and the testing conducted using the Modified Trevo ProVue Retriever demonstrate that the device performs as designed, is suitable for its intended use and is substantially equivalent to the legally marketed predicate device. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the emblem. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 17, 2014 Concentric Medical % Ms. Sarah Meyer Senior Regulatory Affairs Specialist 301 East Evelyn Ave. Mountain View, CA 94041 Re: K133464 > Trade/Device Name: Trevo ProVue Retriever Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: December 18, 2013 Received: December 19, 2013 Dear Ms. Meyer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {6}------------------------------------------------ Page 2 - Ms. Sarah Meyer forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## Joyce M. Whang -S for Carlos L. Peña, Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ### Indications for Use 510(k) Number (if known): K133464 Device Name: Trevo ProVue Retriever Indications For Use: The Trevo Retriever is intended to restore blood flow in the neurovasculature by removing thrombus in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. > Prescription Use _ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use __ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health (CDRH) # Joyce M. Whang -S Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...