K231954 · Scientia Vascular, Inc. · MOF · Aug 1, 2023 · Cardiovascular
Device Facts
Record ID
K231954
Device Name
Aristotle 18 Guidewire; Aristotle 24 Guidewire
Applicant
Scientia Vascular, Inc.
Product Code
MOF · Cardiovascular
Decision Date
Aug 1, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
The Aristotle 18 Guidewire and Aristotle 24 Guidewire are intended for use by a physician to help introduce and position catheters or other interventional devices within the neuro and peripheral vasculature.
Device Story
Steerable guidewires with shapeable tips; used to access neuro and peripheral vasculatures. Input: manual manipulation by physician. Transformation: guidewire provides trackability and support for catheter placement. Output: physical positioning of interventional devices. Used in clinical settings (e.g., neuro/peripheral interventions). Accessories include introducer, torque device, and shaping mandrel. Benefits: facilitates navigation through complex vascular anatomy. No AI/ML or software components.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility and performance supported by data from previously cleared predicate and reference devices (K183608, K192783, K222437, K173235) per ISO 10993-1:2018 and ISO 14971 risk assessment.
Technological Characteristics
Materials: Stainless steel core, Nitinol distal tip, platinum marker coil, hydrophilic distal coating, PTFE proximal coating. Dimensions: 0.018" or 0.024" O.D., 200-300 cm length. Sterilization: 100% Ethylene Oxide (EO) to SAL 10^-6. Mechanical device; no software or energy source.
Indications for Use
Indicated for general vascular use within neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. Not intended for use in coronary vasculature.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
August 1, 2023
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. The FDA logo is on the right, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are placed side by side.
Scientia Vascular, Inc. Bailey Johannsen Regulatory Affairs Specialist 3487 West 2100 South, Suite 100 West Valley City, Utah 84119
Re: K231954
Trade/Device Name: Aristotle 18 Guidewire: Aristotle 24 Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: MOF, DQX Dated: June 30, 2023 Received: July 3, 2023
Dear Bailey Johannsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K231954
Device Name Aristotle 18 Guidewire: Aristotle 24 Guidewire
Indications for Use (Describe)
The Aristotle 18 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional device is not intended for use in the coronary vasculature.
The Aristotle 24 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.
| Type of Use (Select one or both, as applicable) | | | | |
|-------------------------------------------------|---------------------------------------------|--|--|--|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Scientia. The logo consists of a stylized, teal-colored symbol resembling an infinity loop or a stylized "S" shape. Below the symbol, the word "SCIENTIA" is written in a futuristic, sans-serif font, with the letters having rounded edges and a slightly blocky appearance.
# K231954 510(k) SUMMARY
| 510(k) Sponsor: | Scientia Vascular, Inc. |
|--------------------|-----------------------------------------|
| | 3487 West 2100 South, Suite 100 |
| | West Valley City, UT 84119 |
| | Tel (888) 385-9016 |
| Contact Person(s): | Bailey Johannsen |
| | Regulatory Affairs Specialist |
| | Tel: (888) 385-9016 |
| | E-mail: regulatory@scientiavascular.com |
| Date Prepared: | July 28, 2023 |
{4}------------------------------------------------
#### Table 1. Subject Device Information
| Aristotle 18 Guidewire | |
|--------------------------|-----------------------------------------|
| Trade Name: | Aristotle 18 Guidewire |
| Common Name: | Guidewire |
| Classification Name | Catheter Guide Wire per 21 CFR 870.1330 |
| Primary Product Code: | MOF |
| Subsequent Product Code: | DQX |
| Predicate Device: | Aristotle 18 Guidewire (K183608) |
| Reference Device: | Aristotle Colossus Guidewire (K222437) |
| Reference Device: | Aristotle 14 Guidewire (K173235) |
| Aristotle 24 Guidewire | |
| Trade Name: | Aristotle 24 Guidewire |
| Common Name: | Guidewire |
| Classification Name | Catheter Guide Wire per 21 CFR 870.1330 |
| Primary Product Code: | MOF |
| Secondary Product Code | DQX |
| Predicate Device: | Aristotle 24 Guidewire (K192783) |
| Reference Device: | Aristotle Colossus Guidewire (K222437) |
{5}------------------------------------------------
#### DEVICE DESCRIPTION
The Scientia Vascular's Aristotle 18 Guidewire and Aristotle 24 Guidewire are steerable guidewires with a shapeable tip to aid in accessing the neuro and peripheral vasculatures. The guidewires are supplied sterile, for single use only, in the following diameters, stiffness profiles, and lengths:
| Aristotle 18 Guidewire | |
|------------------------|-------------------------|
| Diameter | 0.018" |
| Stiffness Profiles | Soft, Standard, Support |
| Length | 200 cm, 300 cm |
| Aristotle 24 Guidewire | |
| Diameter | 0.024" |
| Stiffness Profiles | Soft, Standard, Support |
| Length | 200 cm, 300 cm |
The distal portion of each guidewire's tip includes a radiopaque platinum wire marker coil to facilitate fluoroscopic visualization. All guidewires have a hydrophilic polymer coating on the distal portion and a polytetrafluoroethylene (PTFE) coating on the proximal portion to reduce friction during manipulation in vessels.
The guidewires are provided with an accessory kit consisting of an introducer (to aid with the insertion of the guidewire into a catheter hub and/or a hemostasis valve), a torque device (to attach to the proximal portion of the guidewire to facilitate gripping and manipulation of the guidewire during use), and a shaping mandrel (to aid in shaping the flexible tip of the guidewire). These accessory devices are included to facilitate use of the guidewires and are not intended to contact the patient's body.
{6}------------------------------------------------
#### INDICATIONS FOR USE
#### Aristotle 18 Guidewire
The Aristotle 18 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.
#### Aristotle 24 Guidewire
The Aristotle 24 Guidewire is intended for general vascular use within the neuro and peripheral vasculatures to introduce and position catheters and other interventional devices. The guidewire is not intended for use in the coronary vasculature.
The indications for use of the subject Aristotle 18 Guidewire and Aristotle 24 Guidewire are the same as in prior, predicate submissions for the Aristotle 18 Guidewire (K183608) and Aristotle 24 Guidewire (K192783), respectively.
{7}------------------------------------------------
Scientia Vascular, Inc. Special 510(k)
#### INTENDED USE
The Aristotle 18 Guidewire and Aristotle 24 Guidewire are intended for use by a physician to help introduce and position catheters or other interventional devices within the neuro and peripheral vasculature.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The subject devices have the following similarities to the previously cleared predicate devices:
- The same indications for use, ●
- The same intended use,
- The same operating principle,
- The same basic guidewire design,
- The same materials, and ●
- The same packaging materials and sterilization processes. ●
The tables (Table 2 and Table 3) below provide additional details on the technological characteristics of the subject devices compared to their respective predicate and reference devices.
{8}------------------------------------------------
| Characteristic | Subject Device<br>Aristotle 18<br>Guidewire | Predicate<br>Aristotle 18<br>Guidewire (K183608) | Reference Device<br>Aristotle Colossus<br>Guidewire (K222437) | Reference Device<br>Aristotle 14<br>Guidewire (K173235) |
|-------------------------------------|---------------------------------------------------------|--------------------------------------------------|---------------------------------------------------------------|---------------------------------------------------------|
| Anatomical<br>Location | Same | Neuro and peripheral<br>vasculature | Neuro and peripheral<br>vasculature | Neuro and peripheral<br>vasculature |
| Dimensions | O.D.: 0.018" (0.46mm)<br>Length: 200cm,<br>300cm | O.D.: 0.018" (0.46mm)<br>Length: 200cm | O.D.: 0.035" (0.88mm)<br>Length: 150cm –<br>300cm | O.D.: 0.014" (0.36mm)<br>Length: 200cm,<br>300cm |
| Core Wire | Same | Stainless Steel | Stainless Steel | Stainless Steel |
| Distal Tip | Same | Shapeable<br>Length: 35cm<br>Material: Nitinol | Shapeable<br>Length: 35cm<br>Material: Nitinol | Shapeable<br>Length: 35cm<br>Material: Nitinol |
| Stiffness Profiles | Support, standard, soft | Support, standard, soft | Standard | Soft, standard |
| Coatings | Same | Distal End: Hydrophilic<br>Proximal End: PTFE | Distal End: Hydrophilic<br>Proximal End: PTFE | Distal End: Hydrophilic<br>Proximal End: PTFE |
| Radiopaque Marker | Same | 1 radiopaque marker at<br>distal tip | 1 radiopaque marker at<br>distal tip | 1 radiopaque marker at<br>distal tip |
| Centering Coil | Same | Stainless steel | Stainless steel | Stainless steel |
| Centering Washer | Does not contain a<br>washer | Does not contain a<br>washer | Stainless steel | Does not contain a<br>washer |
| Bushing | Does not contain<br>bushing | Does not contain<br>bushing | Nitinol Bushing | Does not contain<br>bushing |
| Shaping Mandrel<br>(Accessory) | Same | Provided with each<br>guidewire | Provided with each<br>guidewire | Provided with each<br>guidewire |
| Guidewire Introducer<br>(Accessory) | Same | Provided with each<br>guidewire | Provided with each<br>guidewire | Provided with each<br>guidewire |
| Torque Device<br>(Accessory) | Same | Provided with each<br>guidewire | Provided with each<br>guidewire | Provided with each<br>guidewire |
| Sterilization Method | Same | 100% Ethylene Oxide<br>(EO) | 100% EO | 100% EO |
| Shelf Life | 36 Months | 36 Months | 12 Months | 12 Months |
| Characteristic | Subject Device<br>Aristotle 24<br>Guidewire | Predicate<br>Aristotle 24<br>Guidewire (K192783) | Reference Device<br>Aristotle Colossus<br>Guidewire (K222437) | Reference Device<br>Aristotle 14<br>Guidewire (K173235) |
| Anatomical<br>Location | Same | Neuro and peripheral<br>vasculature | Neuro and peripheral<br>vasculature | Neuro and peripheral<br>vasculature |
| Dimensions | <b>O.D.:</b> 0.024" (0.61mm)<br>Length: 200cm,<br>300cm | <b>O.D.:</b> 0.024" (0.61mm)<br>Length: 200cm | <b>O.D.:</b> 0.035" (0.88mm)<br>Length: 150cm –<br>300cm | <b>O.D.:</b> 0.014" (0.36mm)<br>Length: 200cm,<br>300cm |
| Core Wire | Same | Stainless Steel | Stainless Steel | Stainless Steel |
| Distal Tip | Same | Shapeable<br>Length: 35cm<br>Material: Nitinol | Shapeable<br>Length: 35cm<br>Material: Nitinol | Shapeable<br>Length: 35cm<br>Material: Nitinol |
| Stiffness Profiles | Support, standard, soft | Support, standard, soft | Standard | Soft, standard |
| Coatings | Same | Distal End: Hydrophilic<br>Proximal End: PTFE | Distal End: Hydrophilic<br>Proximal End: PTFE | Distal End: Hydrophilic<br>Proximal End: PTFE |
| Radiopaque Marker | Same | 1 radiopaque marker at<br>distal tip | 1 radiopaque marker at<br>distal tip | 1 radiopaque marker at<br>distal tip |
| Centering Coil | Same | Stainless steel | Stainless steel | Stainless steel |
| Centering Washer | Does not contain a<br>washer | Does not contain a<br>washer | Stainless steel | Does not contain a<br>washer |
| Bushing | Nitinol bushing | Nitinol bushing | Nitinol bushing | Does not contain<br>bushing |
| Shaping Mandrel<br>(Accessory) | Same | Provided with each<br>guidewire | Provided with each<br>guidewire | Provided with each<br>guidewire |
| Guidewire Introducer<br>(Accessory) | Same | Provided with each<br>guidewire | Provided with each<br>guidewire | Provided with each<br>guidewire |
| Torque Device<br>(Accessory) | Same | Provided with each<br>guidewire | Provided with each<br>guidewire | Provided with each<br>guidewire |
| Sterilization Method | Same | 100% EO | 100% EO | 100% EO |
| Shelf Life | 36 Months | 36 Months | 12 Months | 12 Months |
Table 2. Aristotle 18 Guidewire Technological Characteristics Comparison (Differences are Bolded)
{9}------------------------------------------------
Table 3. Aristotle 24 Guidewire Technological Characteristics Comparison (Differences are Bolded)
{10}------------------------------------------------
The Aristotle 18 Guidewire and Aristotle 24 Guidewire have been modified to include additional models of longer length with added 100 cm of proximally PTFE coated guidewire for a total guidewire length of 300 cm. The Aristotle 24 Guidewires have been previously cleared with a total guidewire length of 200 cm in K183608 and K192783, respectively. This additional length does not impact any features or dimensions at the distal end of the guidewire and does not raise new questions of safety and effectiveness.
#### NON-CLINICAL PERFORMANCE TESTS
#### Biocompatibility
Biocompatibility testing was conducted for the reference devices (K222347, K173235). These results support the biocompatibility of the subject devices because they use the same materials, processes, components, and dimensions within the range of the previously cleared predicate and reference devices. The evaluation supports that the subject device with its accessories meets biological safety requirements per ISO 10993-1:2018 for externally communicating medical devices with circulating blood contact for a limited duration (less than 24 hours).
#### Sterilization
The subject devices have been adopted into a validated sterilization cycle that uses 100% EO to achieve a sterilization assurance level (SAL) of 10-6.
#### Bench Performance Testing
The bench performance testing conducted for the reference devices (K222347, K173235) and the predicate 200 cm long models of the subject guidewires cleared under K183608 and K192783, support the performance of the subject devices (300 cm) because they use the same materials, processes, components, and dimensions within the range of the previously cleared predicate and reference devices.
Following the risk assessments conducted in accordance with ISO 14971, no additional testing was deemed necessary to support the substantial equivalence of the subject device models.
#### Animal Testing
No animal testing was deemed necessary to support the substantial equivalence of the subject devices.
#### Clinical Testing
No clinical testing was deemed necessary to support the substantial equivalence of the subject devices.
{11}------------------------------------------------
### CONCLUSION
The Aristotle 18 Guidewire and Aristotle 24 Guidewire have the same intended use and indications for use statement as their respective predicate devices. The identified technological difference does not raise new questions of safety or effectiveness regarding the use of the subject devices. Risk evaluation was completed for the subject devices. The risk evaluation and testing of the predicate and reference devices demonstrate that the subject devices are substantially equivalent to their respective predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.