← Product Code [MOF](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MOF) · K191714

# ASAHI Neurovascular Guide Wire CHIKAI X 014 soft (K191714)

_Asahi Intecc Co., Ltd. · MOF · Jul 26, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MOF/K191714

## Device Facts

- **Applicant:** Asahi Intecc Co., Ltd.
- **Product Code:** [MOF](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MOF.md)
- **Decision Date:** Jul 26, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

This guide wire is intended to be used in the neuro vasculature to facilitate the placement and exchange of therapeutic devices such as cerebral catheters during intravascular therapy. This guide wire is intended for use only in the neuro vasculature.

## Device Story

Steerable guide wire for neurovascular procedures; facilitates placement/exchange of therapeutic devices (e.g., cerebral catheters). Constructed from stainless steel core wire with platinum-nickel and stainless steel coils; distal end features radiopaque tip for visibility; hydrophilic coating on distal portion. Used in clinical settings by physicians during intravascular therapy. Device provides mechanical support and guidance for catheter navigation through neurovascular anatomy. Benefits include improved access to target sites within cerebral arteries, enabling therapeutic interventions.

## Clinical Evidence

Bench testing only. Evaluated per FDA Coronary and Cerebrovascular Guidance (1995). Tests included tensile strength, torque strength, torqueability, tip flexibility, coating adherence, catheter compatibility, and simulated clinical use in models. All test articles met established acceptance criteria based on predicate device specifications.

## Technological Characteristics

Steerable guide wire, 0.014" diameter, 200cm length. Materials: stainless steel core, platinum-nickel/stainless steel coil assembly. Hydrophilic coating on distal section. Sterilization: Ethylene Oxide (SAL 10^-6). Shelf life: 1 year. Mechanical operation.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- ASAHI Neurovascular Guide Wire CHIKAI ([K110584](/device/K110584.md))

## Reference Devices

- ASAHI Neurovascular Guide Wire CHIKAI black ([K141751](/device/K141751.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the rest of the name. The words "U.S. FOOD & DRUG" are stacked on top of the word "ADMINISTRATION".

July 26, 2019

ASAHI INTECC CO., LTD % Cynthia Valenzuela Director, Regulatory Affairs ASAHI INTECC USA, Inc. 3002 Dow Avenue, Suite 212 Tustin, California 92780

Re: K191714

Trade/Device Name: ASAHI Neurovascular Guide Wire (CHIKAI X 014 soft) Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: MOF Dated: June 26, 2019 Received: June 26, 2019

Dear Cynthia Valenzuela:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Xiaolin Zheng, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K191714

Device Name

ASAHI Neurovascular Guide Wire (CHIKAI X 014 soft)

Indications for Use (Describe)

This guide wire is intended to be used in the neuro vasculature to facilitate the placement and exchange of therapeutic devices such as cerebral catheters during intravascular therapy. This guide wire is intended for use only in the neuro vasculature.

| Type of Use (Select one or both, as applicable)                                                                                                        |  |
|--------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <span style="font-family: Arial, sans-serif;"> <input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)         </span> |  |
| <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)         </span>                    |  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary [as required by 21CFR § 807.92(c)]

Image /page/3/Picture/1 description: The image shows the logo and contact information for ASAHI INTECC CO., LTD. The logo is in red and black. Below the logo is the text "Global Headquarters and R&D Center", followed by the address "3-100 Akatsuki-cho, Seto-shi, Aichi 489-0071 Japan". The phone number is +81-561-48-5551, the fax number is +81-561-48-5552, and the website is http://www.asahi-intecc.co.jp/.

ASAHI Neurovascular Guide Wire CHIKAI X 014 soft

| DATE PREPARED:         | 27JUN2019                                                                                                                                                                                                                                                        |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT:             | ASAHI INTECC CO., LTD<br>3-100 Akatsuki-cho, Seto<br>Aichi 489-0071, Japan                                                                                                                                                                                       |
| PRIMARY CONTACT:       | Mrs. Cynthia Valenzuela<br>Director, Regulatory Affairs<br>ASAHI INTECC USA, INC.<br>3002 Dow Avenue, Suite 212<br>Tustin, California 92780<br>Phone: (714) 442 0575<br>Fax: (949) 377 3255<br>Cell Phone: (949) 413 0071<br>Email: cynthiav@asahi-intecc-us.com |
| ALTERNATE CONTACT:     | Mr. Yoshi Terai<br>President, CEO<br>ASAHI INTECC USA, INC.<br>Tustin, CA 92780 USA<br>Phone: (949) 756 8901<br>Fax: (949) 377 3255<br>Email: yoshi@asahi-intec-us.com                                                                                           |
| TRADE NAME:            | ASAHI Neurovascular Guide Wire (CHIKAI X<br>014 soft)                                                                                                                                                                                                            |
| DEVICE CLASSIFICATION: | Class II, 21CFR § 870.1330                                                                                                                                                                                                                                       |
| CLASSIFICATION NAME:   | Wire, Guide, Catheter, Neurovascular                                                                                                                                                                                                                             |
| PRODUCT CODE:          | MOF - Catheter Guide Wire                                                                                                                                                                                                                                        |
| PREDICATE DEVICE(S):   | Primary Predicate:<br>ASAHI Neurovascular Guide Wire CHIKAI<br>(K110584)<br>Reference Device:<br>ASAHI Neurovascular Guide Wire CHIKAI<br>black(K141751)                                                                                                         |

{4}------------------------------------------------

## Intended Use/Indications for Use

This quide wire is intended to be used in the neuro vasculature to facilitate the placement and exchange of therapeutic devices such as cerebral catheters during intravascular therapy. This guide wire is intended for use only in the neuro vasculature.

# Description:

The ASAHI Neurovascular Guide Wire CHIKAI X 014 soft (hereafter "CHIKAI X 014 soft") is a steerable quide wire with a maximum diameter of 0.014 inches (0.36mm) and available in 200cm length. The extension wire is connected to the end of the guide wire outside the body for 200cm wire. The guide wire is constructed from a stainless steel core wire with a platinum-nickel and stainless steel coils. The coil assembly consists of an inner coil and an outer coil, and the coil assembly is soldered to the core wire. The distal end of the guide wire has a radiopaque tip to achieve visibility, and is available in a straight configuration and can be made to bend easily with the vessel curve. A hydrophilic coating is applied to the distal portion of the guide wire. The coil assembly construction is similar to the 510(K) cleared ASAHI Neurovascular Guide Wire CHIKAI with K110584.

## Comparison with Predicate Device(s):

| Predicate Device:                               | Reference Device:                                        |
|-------------------------------------------------|----------------------------------------------------------|
| ASAHI Neurovascular Guide Wire CHIKAI (K110584) | ASAHI Neurovascular Guide Wire CHIKAI black<br>(K141751) |

Comparisons of the CHIKAI X 014 soft and predicate devices show that the technological characteristics of the CHIKALX 014 soft such as components, design, materials, sterilization method, shelf life and operating principle are identical or similar to the currently marketed predicate device. A tabular comparison of the specific technological characteristics between the predicate device and subject device is provided below.

| Name of Device               | ASAHI Neurovascular Guide<br>Wire CHIKΑΙ Χ 014 soft                                                                                                                                                                                                                 | ASAHI Neurovascular Guidewire<br>CHIKAI  |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| 510(K)                       | K191714                                                                                                                                                                                                                                                             | K110584                                  |
| Intended Use and Indications | The guide wire is intended to be used in the neuro vasculature to facilitate<br>the placement and exchange of therapeutic devices such as cerebral<br>catheters during intravascular therapy. This guide wire is intended for use<br>only in the neuro vasculature. |                                          |
| Sterilization                | Ethylene Oxide to SAL 10-6                                                                                                                                                                                                                                          |                                          |
| Shelf Life                   | 1 Year                                                                                                                                                                                                                                                              | 3 Years                                  |
| Target Body Location         | Neurovascular                                                                                                                                                                                                                                                       |                                          |
| Outer Coil Material          | Stainless-Steel                                                                                                                                                                                                                                                     | Stainless-Steel<br>Platinum-Nickel Alloy |
| Core Wire Material           | Stainless-Steel                                                                                                                                                                                                                                                     | Stainless-Steel                          |
| Inner Coil Material          | Platinum-Nickel Alloy                                                                                                                                                                                                                                               | Stainless-Steel                          |
| Distal Tip Shape             | Straight                                                                                                                                                                                                                                                            | Straight                                 |
| Overall Length               | 200cm                                                                                                                                                                                                                                                               | 200cm, 300cm                             |

{5}------------------------------------------------

| Distal Section Coating length | 1700mm | 1700mm |
|-------------------------------|--------|--------|
| Outside Diameter of Wire      | 0.36mm | 0.36mm |

Non Clinical Testing / Performance Data:

The substantial equivalence of the CHIKAI X 014 soft line extension was evaluated in bench testing that followed the recommendations in the FDA guidance document; Coronary and Cerebrovascular Guidance, January 1995. Only those tests impacted by the device modifications were repeated. The table below provides a summary of the bench test methods, results and conclusions. Acceptance criteria for each of the tests were determined by prior comparative testing with predicate devices, ASAHI's established guide wire specifications, and clinical experience.

| Test                   | Test Method Summary                                                                                                                                                                                                                          | Results / Conclusions                                                                                                                                                                                           |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tensile Strength       | To determine maximum allowable<br>tensile load between connections, guide<br>wire is fixed in the Tensile Testing<br>Machine and pulled until failure.                                                                                       | All test articles met established<br>acceptance criteria. Acceptance<br>criteria determined by evaluation of<br>predicate devices and ASAHI's<br>established tensile strength<br>specifications.                |
| Torque Strength        | To determine torque strength, distal end<br>is inserted and advanced through<br>simulated model. Distal tip is held<br>stationary while proximal end is rotated<br>until failure.                                                            | All test articles met established<br>acceptance criteria. Acceptance<br>criteria determined by evaluation of<br>predicate devices and ASAHI's<br>established torque strength<br>specifications.                 |
| Torqueability          | To determine torque response,<br>guidewire is inserted through catheter<br>and into   Rotational Response model.<br>Proximal end is rotated from 0° to<br>720°. Torque response at distal end is<br>measured at each 90° angle.              | All test articles met the acceptance<br>criteria. Acceptance criteria<br>determined by evaluation of<br>predicate devices and ASAHI's<br>established torqueability<br>specifications.                           |
| Tip Flexibility        | To determine flexibility of the distal end,<br>the force to deflect the guide wire 45°<br>and 90° at 5, 10, and 20mm from distal<br>tip is measure by a force analyzer<br>attached to a load cell.                                           | All test articles met established Tip<br>Flexibility acceptance criteria.<br>Acceptance criteria determined by<br>evaluation of predicate devices and<br>ASAHI's established tip flexibility<br>specifications. |
| Coating Adherence      | Integrity of coated outer coil and core<br>wire is determined before, and after,<br>pretreatment and manipulation in<br>excess of that expected in clinical use.                                                                             | Test results confirmed the integrity of<br>the coating was maintained during<br>simulated clinical use in all test<br>articles.                                                                                 |
| Catheter Compatibility | Catheter compatibility is evaluated by<br>measuring the force to withdraw the<br>guide wire that has been inserted<br>through the test catheter.                                                                                             | All test articles met the acceptance<br>criteria. Resistance to catheter<br>withdrawal is similar or better than<br>predicate.                                                                                  |
| Bench Testing          | To simulate clinical use, guidewire is<br>inserted through guide catheter placed<br>in simulated model and advanced to<br>target area. Microcatheter is inserted<br>over guidewire and advanced to target<br>cerebral artery multiple times. | Test results on all test articles<br>confirmed guidewire performance.<br>Guidewire reached target area and<br>interventional catheter was<br>successfully advanced<br>over quide<br>wire to target site.        |

{6}------------------------------------------------

The in vitro bench tests demonstrated that the CHIKAI X 014 soft met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrate that the device functions as intended and has a safety and effectiveness profile that is similar to the predicate device.

#### BIOCOMPATIBILITY:

There have been no changes in materials of the Subject device as compared to the Predicate device. For this reason, no additional biocompatibility information is required for this Special 510(K).

### CONCLUSION:

The CHIKAI X 014 soft has identical intended use and the same or similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate and reference device. Performance data demonstrates that the device functions as intended.

Therefore, the CHIKAI X 014 soft is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MOF/K191714](https://fda.innolitics.com/submissions/NE/subpart-b%E2%80%94cardiovascular-diagnostic-devices/MOF/K191714)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
