Traxcess .007" Mini Guidewire is indicated for general intravascular use, including the neuro and peripheral vasculature. The guidewire can be steered to facilitate the selective placement of diagnostic or therapeutic catheters. This device is not indicated for use in coronary arteries.
Device Story
Traxcess .007" Mini Guidewire is a steerable, coiled guidewire used to navigate diagnostic or therapeutic catheters through neuro and peripheral vasculature. The device consists of a proximal stainless steel core wire and a distal nitinol core wire, with the distal section contained within a platinum/nickel coil. Both sections feature a hydrophilic coating for lubricity. The physician manually steers the guidewire to guide catheter placement. The device is supplied sterile for single use and includes a shaping mandrel, insertion tool, and torque device. By facilitating precise catheter navigation, the device assists clinicians in reaching target vascular locations for diagnostic or therapeutic procedures, potentially improving procedural outcomes.
Clinical Evidence
No clinical data. Substantial equivalence supported by performance bench testing and animal testing. Bench tests included dimensional inspection, tip shapeability, coating durability/lubricity, tensile strength, corrosion resistance, torque strength/response, fracture resistance, flexing, distal tip flexibility, particle testing, radiopacity, and simulated use. Animal testing confirmed trackability, friction, and handling equivalent to the predicate. Biocompatibility, packaging, and sterilization were validated via reference to the predicate device.
Indicated for general intravascular use, including neuro and peripheral vasculature, to facilitate selective placement of diagnostic or therapeutic catheters. Contraindicated for use in coronary arteries.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 30, 2016
MicroVention, Inc. Sapna Singh, MS, RAC Regulatory Affairs Project Manager 1311 Valencia Avenue Tustin, California 92780
Re: K161803
Trade/Device Name: Traxcess .007" Mini Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: MOF, DQX Dated: September 1, 2016 Received: September 2, 2016
Dear Ms. Singh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 ), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Carlos L. Pena -S
Carlos L. Peña, Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161803
Device Name Traxcess .007" Mini Guidewire
#### Indications for Use (Describe)
Traxcess .007" Mini Guidewire is indicated for general intravascular use, including the neuro and peripheral vasculature. The guidewire can be steered to facilitate the selective placement of diagnostic catheters. This device is not indicated for use in coronary arteries.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.6569 2.34315C14.2815 2.96774 14.2815 3.98364 13.6569 4.60823L4.60823 13.6569C3.98364 14.2815 2.96774 14.2815 2.34315 13.6569C1.71856 13.0323 1.71856 12.0164 2.34315 11.3918L11.3918 2.34315C12.0164 1.71856 13.0323 1.71856 13.6569 2.34315Z" fill="black"></path> <path d="M2.34315 2.34315C2.96774 1.71856 3.98364 1.71856 4.60823 2.34315L13.6569 11.3918C14.2815 12.0164 14.2815 13.0323 13.6569 13.6569C13.0323 14.2815 12.0164 14.2815 11.3918 13.6569L2.34315 4.60823C1.71856 3.98364 1.71856 2.96774 2.34315 2.34315Z" fill="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect height="15" stroke="black" width="15" x="0.5" y="0.5"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY [K161803]
This 510(k) summary for Traxcess .007" Mini Guidewire is submitted in accordance with the requirements of 21 CFR 807.87(h) and 807.92 and following the recommendations outlined in FDA Guidance, The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)], dated 28 July, 2014.
## SUBMITTER [807.92(a)(1)]
MicroVention, Inc. 1311 Valencia Avenue Tustin, California U.S.A.
| Telephone: | (714) 247-8162 |
|------------|----------------|
| Fax: | (714) 247-8014 |
| Contact Person: | Sapna Singh |
|-----------------|------------------------------|
| Email: | sapna.singh@microvention.com |
| Date Prepared: | September 30, 2016 |
## DEVICE [807.92(a)(2)]
| Name of Device: | Traxcess .007" Mini Guidewire |
|---------------------------|-----------------------------------------------------------------------------------------------|
| Common or Usual Name: | Traxcess Guidewire |
| Classification Name: | Catheter Guidewire |
| Product Code: | MOF, DQX |
| Regulatory Class: | Class II |
| Submission Type: | Special 510(K) |
| Regulation Number: | 21 CFR 870.1330 |
| Reviewing Product Branch: | Division of Neurological and Physical Medicine Devices<br>(Office of Device Evaluation, CDRH) |
## PREDICATE DEVICE [807.92(a)(3)]
Traxcess® 14 SELECT Guidewire (K153053)
## DEVICE DESCRIPTION [807.92(a)(4)]
Traxcess .007" Mini Guidewire is a coiled wire that is designed to fit inside a percutaneous microcatheter for the purpose of directing the catheter through a blood vessel. It consists of a
{4}------------------------------------------------
proximal coated Stainless Steel core wire, and a distal coated Nitinol core wire. The distal core wire is tapered at the distal tip and is contained within a Platinum/Nickel coil. The Platinum/Nickel coil is 6 cm in length. The distal 1.4 cm of the guidewire is shapeable by the physician.
Traxcess .007" Mini Guidewire distal and proximal sections are coated with hydrophilic coating. The purpose of the hydrophilic coating is to provide lubricity when the MicroVention guidewire is passed through microcatheters. A shaping mandrel, insertion tool, and torque device are also included with the device
# INDICATIONS FOR USE [807.92(a)(5)]
Traxcess .007" Mini Guidewire is indicated for general intravascular use, including the neuro and peripheral vasculature. The guidewire can be steered to facilitate the selective placement of diagnostic or therapeutic catheters. This device is not indicated for use in coronary arteries.
# CONTRAINDICATION
The device is not indicated to be used in coronary arteries.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS [807.92(a)(6)]
Traxcess .007" Mini Guidewire has the following similarities to the predicate device, Traxcess® 14 SELECT Guidewire (K153053):
- 1. Same intended use
- 2. Same operating principle
- 3. Incorporate the same basic guidewire design
- 4. Incorporate the same guidewire construction material
- 5. Are packaged and sterilized using the same materials and processes
The small distal tip of the guidewire and the application of hydrophilic coating on the distal and proximal sections of the guidewire does not change the indications for use of the Traxcess guidewires and is not a change to the fundamental scientific technology. The performance data below shows the device will perform as well as the previously marketed device.
The Table I states the comparison between Traxcess® 14 SELECT Guidewire (Predicate Device, K153053) and Traxcess .007" Mini Guidewire (Subject Device, K161803).
## Table I: Predicate Device vs Subiect Device Comparison Table
| | Traxcess 14 SELECT<br>Guidewire (Predicate Device,<br>K153053) | Traxcess .007" Mini<br>Guidewire (Subject Device,<br>K161803) |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | | |
| | | |
| Indications for Use Statement | Traxcess Guidewire is intended for<br>general intravascular use, including<br>the neuro and peripheral<br>vasculature. The guidewire can be<br>steered to facilitate the selective<br>placement of diagnostic or<br>therapeutic catheters. This device<br>is not intended for use in<br>coronary arteries. | Traxcess .007" Mini Guidewire is<br>indicated for general intravascular<br>use, including the neuro and<br>peripheral vasculature. The<br>guidewire can be steered to<br>facilitate the selective placement of<br>diagnostic or therapeutic catheters.<br>This device is not indicated for<br>use in coronary arteries. |
| Performance | | |
| Function | The steerable guidewire is used<br>to facilitate the selective<br>placement of diagnostic or<br>therapeutic catheters. | Same |
| Anatomical Location | General intravascular use,<br>including the neuro and peripheral<br>vasculature. | Same |
| Design | | |
| Overall Length | 200 cm | 210 cm |
| Diameter | Proximal = 0.014"<br>Distal = 0.012" | Proximal = Same<br>Distal = 0.007" |
| Core wire configuration | 60 cm Nitinol welded to Stainless<br>Steel | Same |
| Coil Length | 40 cm | 6 cm |
| Coil Configuration | 3cm Platinum Nickel alloy and<br>37cm Stainless Steel | 6cm Platinum Nickel alloy |
| Distal Shaft Length (Shapeable<br>Length) | 1.4 cm | Same |
| Docking Wire Compatibility | Yes | No |
| Material | | |
| Material | Core wire (proximal): Stainless<br>steel<br>Core wire (distal): Nickel titanium<br>(Nitinol) alloy<br>Coil: Platinum nickel alloy and<br>Stainless steel | Core wire (proximal): Same<br>Core wire (distal): Same<br>Coil: Platinum nickel alloy. No<br>Stainless steel |
| Coating Material | Coil and distal stainless steel<br>section: Hydrophilic Coating<br>[SLIP-COAT by Argon Medical] | Coil and distal stainless steel<br>section: Same<br>No PTFE Coating |
| | Proximal Stainless Steel Section:<br>PTFE | |
| Hydrophilic Coating Length | 98 cm | 158 cm |
| Other Attributes | | |
| Method of supply | Sterile and single use | Same |
| Sterilization method | Ethylene oxide gas | Same |
| Accessories | Shaping mandrel, Torque device,<br>and Insertion tool | Same |
| Package configuration | Placed into a Dispenser hoop,<br>Tyvek pouch, and Carton box | Same |
{5}------------------------------------------------
{6}------------------------------------------------
## PERFORMANCE DATA [807.92(b)]
Performance Bench Testing and Animal Testing: Results of the performance bench testing and animal testing (Table II) indicate that Traxcess .007" Mini Guidewire (subject device, K161803) meets established performance requirements, and is substantially equivalent for its intended use.
| Performance Bench Testing | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tests | Acceptance Criteria | Conclusion |
| <b>Dimensional Inspection (Visual):</b><br>The dimensional attributes of the test samples were inspected. | Test article should meet specified dimensional requirements for:<br>OD (Distal and Proximal) Overall Length Length of distal Pt/Ni coil section Length of hydrophilic coated section | Device met established dimensional specification. |
| <b>Tip Shapeability:</b><br>The test evaluates the shapeability and retention before and after simulated application of intraprocedural stresses. | Test article should be greater than or equal to existing tip shapeability specification. | Device met established tip shapeability specification. |
| Performance Bench Testing | | |
| Tests | Acceptance Criteria | Conclusion |
| Durability/Lubricity of<br>Hydrophilic Coating:<br>The durability/lubricity of the<br>coated test samples are<br>inspected by passing the sample<br>through two silicone clamp pads.<br>Machine starts pulling the<br>coated sample up through the<br>clamping pads, recording the<br>coating lubricity friction (in<br>grams) until the sample is<br>completely through. | Test article should meet existing<br>durability/lubricity of hydrophilic<br>coating specification. | Device met established<br>durability/lubricity of<br>hydrophilic coating<br>specification. |
| Tensile Strength:<br>The tensile strength of the distal<br>tip and proximal section (nitinol<br>and stainless steel weld) of the<br>guidewire was measured to make<br>sure it is sufficiently strong to<br>withstand normal tensile loading<br>for its intended use. | Test article should be greater than or<br>equal to existing tensile strength<br>specification for distal tip and<br>proximal joint section. | Device met established<br>distal tip and proximal joint<br>tensile strength<br>specification. |
| Corrosion Resistance:<br>The corrosion resistance of the<br>guidewire is tested to make sure<br>if the guidewire is corrosion<br>resistance. | Test article should be corrosion<br>resistant. | Device met established<br>corrosion resistance. |
| Surface Contamination and<br>Defects:<br>The surface contamination and<br>defect results of the guidewire is<br>tested to make sure it is free<br>from contamination and defects. | Test article when examined at<br>magnification, should meet existing<br>surface contamination and defects<br>specification. | Device met established<br>surface contamination and<br>defects specification. |
| Torque Strength:<br>The Torque Strength test was<br>performed to count the number of<br>turns to guidewire failure. | Test article should be greater than or<br>equal to existing torque strength<br>specification. | Device met established<br>torque strength<br>specification. |
| Torqueability:<br>The Torqueability test was<br>performed to measure the<br>difference in input angle (turn at<br>the proximal end at a set amount)<br>vs. the output angle and measure<br>how much the distal end turns. | Test article should be equal to, or<br>better than predicate device. | Subject device torque<br>response better than<br>predicate device. |
| Performance Bench Testing | | |
| Tests | Acceptance Criteria | Conclusion |
| Fracture resistance:<br>The Fracture Resistance Test<br>was performed to test for fracture<br>on the guidewire after winding<br>the guidewire around a cylindrical<br>former, then unwound and<br>examined for fracture of the<br>guidewire and the coating as well. | Test article should not show signs of<br>fracture. There should be no coating<br>flaking off the guidewire. | Device met established<br>fracture resistance<br>specification. |
| Flexing test:<br>The Flexing test was performed<br>to test the guidewire (distal and<br>proximal sections) under<br>repeated reverse bending and<br>straightening (flexing) and<br>examined for defects or damage. | Test article should not show signs of<br>defect, fracture or other damage.<br>There should be no coating flaking off<br>the guidewire. | Device met established<br>flexing test specification. |
| Distal Tip flexibility:<br>The distal tip flexibility testing<br>was performed to demonstrate<br>the force required to deflect the<br>distal tip of the guidewire. | Test article should be less than<br>existing distal tip specification to<br>deflect the distal tip of guidewire. | Device met established<br>distal tip flexibility<br>specification. |
| Particle Testing:<br>The particle testing analysis was<br>performed to quantify particulate<br>matter in injections of the<br>guidewire after<br>advancement/retraction<br>procedures. | Test article should meet established<br>particle testing specification. | Device met established<br>particle test specification. |
| Radiopacity:<br>The guidewire is placed under<br>fluoroscopy and digital images<br>are visually assessed for device<br>visibility. | Test article should be visible under<br>fluoroscopy. | Device met established<br>radiopacity specification. |
| In-Vitro Simulated Use<br>Testing:<br>Samples underwent simulated<br>use testing that included<br>introduction into and movement<br>within the catheter, tracking<br>guidewire/microcatheter system<br>in the model, system maximum<br>distal reach, overall performance<br>and any particles detected. | Test article should meet rating of 3 or<br>greater when tested with compatible<br>microcatheters. | Device met established<br>simulated use testing<br>specification. |
| Performance Bench Testing | | |
| Tests | Acceptance Criteria | Conclusion |
| Trackability, friction and handling<br>of the guidewire to the target<br>location | Equal to or better than the predicate<br>device. | Both subject device and<br>predicate device were able<br>to reach desired location<br>in the animal study with no<br>issues. No friction during<br>insertion.<br>Subject device trackability,<br>friction and handling<br>equivalent to predicate<br>device. |
| Radiopacity | Must be visible under fluoroscopy. | Both subject device and<br>predicate device were<br>visible under fluoroscopy. |
| Compatibility with Microcatheter | Must be compatible with<br>microcatheters. | Both subject device and<br>predicate device were<br>compatible with<br>microcatheters. |
| Overall Assessment | Equal to or better than the predicate<br>device. | Performance of both<br>subject device and<br>predicate device was<br>satisfactory.<br>Subject device overall<br>assessment is equivalent<br>to predicate device. |
### Table II: Performance Bench Testing and Animal Testing Summary
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
Biocompatibility: The biocompatibility studies were not repeated on the subject device, Traxcess .007" Mini Guidewire since it is made from the same material, same manufacturing processes, same sterility assurance level and same packaging configuration as those utilized in the fabrication of predicate device, Traxcess® 14 SELECT Guidewire (K153053). The results from biocompatibility testing conducted on predicate device showed that the acceptance criteria were met. Table III summarizes the biocompatibility testing conducted on Traxcess® 14 SELECT Guidewire.
### Table III: Biocompatibility Test Summary
| Biocompatibility Test<br>(ISO Standard) | Acceptance Criteria | Conclusion |
|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity - L929 MEM Elution<br>Test<br>(ISO 10993-5) | Test article meets the<br>requirements of the test if it does<br>not show greater than a mild<br>reactivity (Grade 2). | Test article exhibited a<br>biological reactivity grade of 0<br>(on a scale of 0 to 4).<br>(Non-cytotoxic). |
| Sensitization/Irritation - Kligman<br>Maximization Test<br>(ISO10993-10) | Test article meets the<br>requirements of the test if it does<br>not show a positive response in<br>atleast 10% of the test animals. | Test article exhibited 0%<br>sensitization.<br>(Non-sensitizer). |
| Biocompatibility Test<br>(ISO Standard) | Acceptance Criteria | Conclusion |
| Sensitization/Irritation -<br>Intracutaneous Injection Test<br>(ISO 10993-10) | Test article meets the<br>requirements of the test if it does<br>not produce irritation after<br>intracutaneous injection in New<br>Zealand White rabbits. | Test article did not show a<br>significantly greater biological<br>reaction than sites injected with<br>the control article. The<br>difference of the overall mean<br>score between the test article<br>and the control article was 0.0.<br>(Non-irritant). |
| Hemocompatibility - Hemolysis -<br>Direct and Indirect<br>(ISO 10993-4) | Test article meets the<br>requirements of the test if the<br>hemolytic index above the<br>negative control article is <5%. | Hemolysis index was above the<br>negative control of 0.77% via<br>direct contact method and<br>0.23% via indirect contact<br>method.<br>(Non-hemolytic). |
| Hemocompatibility - Unactivated<br>Partial Thromboplastin Time<br>(UPTT) Assay - Direct Contact<br>(ISO 10993-4) | Test article meets the<br>requirements of the test if no<br>statistical decrease is found<br>between UPTT of the plasma<br>exposed to the test article and<br>that of plasma exposed to<br>negative control or untreated<br>control. | No statistical decrease is found<br>between UPTT of the plasma<br>exposed to the test article and<br>that of plasma exposed to<br>negative control or untreated<br>control.<br>(Not considered to have an<br>effect on coagulation of human<br>plasma). |
| Hemocompatibility - C3A and<br>SC5B-9 Complement Activation<br>Test - Direct Contact<br>(ISO 10993-4) | Test article meets the<br>requirements of the test if the<br>concentration of C3A and SC5B-<br>9 in plasma exposed to test<br>article does not statistically<br>increase than the plasma<br>exposed to negative and<br>untreated controls. | The concentration of C3A and<br>SC5B-9 in plasma exposed to<br>test articles were not<br>statistically increased than the<br>plasma exposed to negative<br>and untreated controls.<br>(Not considered to activate the<br>complement system in human<br>plasma). |
| Hemocompatibility - In Vitro<br>Hemocompatibility Test - Direct<br>Contact<br>(ISO 10993-4) | Test article meets the<br>requirements of the test if no<br>statistical decrease (or<br>increase/decrease for hematocrit<br>and Mean corpuscular values) is<br>found between blood exposed to<br>test article and blood exposed to<br>negative control or untreated<br>control. | Test article did not have an<br>effect on the WBCs, Platelet<br>concentration and other<br>hematological parameters in<br>comparison to negative control<br>and untreated control.<br>(No effect on selected<br>hematological parameters). |
| Hemocompatibility - Dog<br>Thrombogenicity<br>(ISO 10993-4) | Test article meets the<br>requirements of the test if there<br>is minimal thrombosis for test<br>article (Grade 0-2). | Minimal thrombosis (Grade 0-1)<br>for test article and control sites.<br>(No significant thrombosis). |
| Systemic toxicity - Systemic<br>Injection Test<br>(ISO 10993-11) | Test article meets the<br>requirements of the test if it does<br>not induce a significantly greater<br>biological reaction than the<br>animal treated with the control<br>articles when injected into albino<br>mice | Test article did not induce a<br>significantly greater biological<br>reaction than the control<br>extracts when injected into<br>albino mice.<br>(No toxic effects). |
| Biocompatibility Test<br>(ISO Standard) | Acceptance Criteria | Conclusion |
| Systemic toxicity - Rabbit<br>Pyrogen Test<br>(ISO 10993-11) | Test article meets the<br>requirements of the test for the<br>absence of pyrogens, if no rabbit<br>shows an individual temperature<br>rise of 0.5°C or more above the<br>baseline temperature. | Temperature increases for the<br>test animals were all 0.0° C<br>from baseline.<br>(Non-pyrogenic). |
{10}------------------------------------------------
{11}------------------------------------------------
Packaging: The packaging validation, T=3 years accelerated aging was performed on predicate device, Traxcess® 14 SELECT Guidewire (K153053) and included testing devices via visual inspection, simulated use testing, sterile pouch seal strength testing, dye penetration testing and shipper box testing. The results from packaging testing conducted on predicate device showed that the acceptance criteria were met. The testing was not repeated on the subject device, Traxcess .007" Mini Guidewire since there is no change in the packaging material, packaging configuration, and method of supply or sterilization. The packaging still provides the same protection and sterile barrier requirements.
Sterilization: A product adoption study was performed on predicate device, Traxces® 14 SELECT Guidewire (K153053). The device is sterilized using 100% Ethylene Oxide (EtO) gas, per Cycle 11 to a sterility assurance level of 10 t in the same manner as our existing Headway 17 Microcatheter (K083343). The bacterial endotoxin testing was conducted on predicate device and found to be less than 0.011 EU/ml and met the acceptance criteria of no more than 2.15 EU/device. The sterilization testing was not repeated on subject device, Traxcess .007" Mini Guidewire since it is sterilized using 100% Ethylene Oxide (EtO) gas in the same manner as predicate device, Traxcess® 14 SELECT Guidewire (K153053). There are no changes in materials or other design attributes made to the subject device.
Shelf Life: The shelf life testing, T=3 years accelerated aging was performed on predicate device, Traxcess® 14 SELECT Guidewire (K153053) and included testing devices via simulated use, dimensional inspection, durability/lubricity of hydrophilic and PTFE coating, tensile strength, surface defects and contamination, torque response and particle testing. The results from shelf life testing conducted on predicate device showed that the acceptance criteria were met. The shelf life testing was not repeated on the subject device, Traxcess .007" Mini Guidewire since no new materials are added to the subject device. There will be no impact on device shelf life as material degradation rate is the same.
The conclusions drawn from the performance bench testing, biocompatibility, packaging, sterilization and shelf-life testing demonstrate compliance with the established testing acceptance criteria. Therefore, we can conclude that proposed device is substantially equivalent to legally marketed predicate device, Traxcess 14 SELECT Guidewire (K153053).
## CONCLUSIONS
Based on the 510(k) summary and information provided herein, we conclude the subject device, Traxcess .007" Mini Guidewire, is substantially equivalent in its intended use, design, guidewire
{12}------------------------------------------------
material, performance, and the underlying fundamental scientific technology used, to the predicate device, Traxcess® 14 SELECT Guidewire (K153053).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.